ROBOTIQ PALLETIZING SOLUTION User guide

Type
User guide
Robotiq Palletizing Solution - Troubleshooting guide
1. How to consult this troubleshooting guide
We tried to make this guide as convenient and intuitive as possible. Here is an example of how to read this table.
1
2. Copilot and Polyscope update
The following recommandations take into account that the latest version of the Copilot URcap and the Polyscope operating system are used.
Visit our support page (https://robotiq.com/support) to download the latest URcaps from Robotiq.
3. Air Supply requirements
The air supply must be dried and filtered according to ISO 8573-1 class 3.4.3. Air pressure must be between 3 and 7 bar with an optimal
pressure at 7 bar. (6.5 bar for Airpick serial number ≤ V-01351)
4. Table of contents
1. How to consult this troubleshooting guide 1
2. Copilot and Polyscope update 2
3. Air Supply requirements 2
5. Troubleshooting table 3
APPENDIX A 9
APPENDIX B 10
APPENDIX C 11
© 2022 Robotiq inc. 2
5. Troubleshooting table
Problem / issue
Cause(s)
Solution(s)
Collision
Collision with the pallet
sensor on one side, but
not on the other.
[AX series] Linear axis is not aligned with the
base.
[Manual] Execute the centering procedure.
Ch. 3, “Mechanical and Electrical Installation”, “Centering the Solution” section.
[PE series] Pedestal is not aligned with the base.
[Manual] Execute the centering procedure.
Ch. 3, “Mechanical and Electrical Installation”, “Centering the Solution” section.
Collision with a box
already placed on the
pallet.
Collision between the gripped box and the box
on the pallet.
A small Z offset accumulates over many layers.
[Box] Increase the box height parameter (C) to compensate for box variability.
Use the average box height : Measure the height of a full stack of boxes and divide
it by the number of layers.
Box faces are rounded.
[Box] Increase the box width and/or length parameter (A and/or B).
Wrong pallet state.
[Pallet viewer] Restart the program and select the correct pallet state. Enter the
amount of boxes already present on the pallet.
Pallet height is set too low.
[Pallet] Measure the pallet height and set the parameter correctly.
Collision occurred during
palletizing with End Of
Arm Tooling (EOAT) .
TCP configuration is incorrect.
[Software][Installation] Enter correct TCP orientation and position from the manual.
[Info] When using a gripper other than AirPick, refer to the supplier's manual,
measure it yourself or use Polyscope's TCP wizard.
TCP was set without considering the use of an
extension bracket.
Gripper is larger than the box (X and/or Y
dimensions).
[Info] Collision model will consider the gripper
has the same X and Y dimensions as the box.
[Support] Contact Robotiq's Coach team.
[Info] [Manual] Consult Ch. 3, “Other Grippers” section, for more details.
[PE series] Pedestal height does not match the
height configured in the installation tab.
[Online configurator][Installation] Adjust the actual pedestal height and/or the
parameter in the installation tab according to the configurator's results.
(
Installation
->
URCaps
->
Copilot
->
Configuration
)
© 2022 Robotiq inc. 3
Problem / issue
Cause(s)
Solution(s)
Error message
"Discontinuous path /
joint solution impossible
detected, please change
your starting position."
Box pick position is too close to the robot’s base.
[Online configurator] Simulate with your actual installation until it works. Move the
conveyor to move the pick position.
Adjust your real installation correspondingly.
Conveyor position does not allow trajectory.
Robot is not able to generate a vertical trajectory
to avoid the conveyor.
[Info] Conventional conveyor height is between
700 and 900mm.
[Online configurator] Adjust conveyor position in your simulation until getting
satisfying results.
[Conveyor] Set the conveyor according to the simulation's results.
Joint has reached or is close to its limits.
[Pick position] [Move] Move the tool to the pick position.
Move each joint near the center of their range (°0). Visualize the range using the
UR
Move
tab. Teach the new pick position. [Info] While running the program, the
software will reorient Wrist #3.
OR
Set the robot back to
Home position
(arm straight in the air) and restart the
program.
TCP orientation at pick position is wrong.
[Software][Installation] Enter the correct TCP orientation and position from the
manual.
[Info] When using a gripper other than AirPick, refer to the supplier's manual,
measure it yourself or use Polyscope's TCP wizard.
Specific point not reachable on the pattern (on
both pallets).
[Online configurator] Adjust pattern and label orientation until every point is
reachable.
[Software] [Pattern] Apply the changes required to have the simulation running.
[Info] When possible, allow the gripper to pick the box at 90° and/or 180°
(
Palletize
r node ->
Settings
).
Specific point not reachable on the pattern (on
only one side)
[Software] [Pattern] Center your pattern on the pallet in the
Palletizer
node.
OR
Create a
Palletizer
node for each side, allowing to bring the pattern closer to the
base for the left and right sides.
© 2022 Robotiq inc. 4
[PE series] Incorrect pedestal height for a specific
box and pattern.
[Online configurator] Simulate your use case online, note the pedestal height
selected in the KPI window.
[Software][Installation] Set the pedestal height according to the online configurator
results.
(
Installation
->
URCaps
->
Copilot
->
Configuration
)
[Installation] Set the actual pedestal height accordingly.
None of the above has worked.
[Support] Contact Robotiq's Coach team.
[AX series] "Please use
the Safeguard reset
before starting the linear
axis"
System has been restarted or a new installation
file has been loaded.
Press the safeguard reset button.
Cables were installed or removed in the
controller while the unit was powered on.
Power-off then power-on (power cycle) the linear axis controller.
Linear axis security relay latched. Dual channel
security signal mismatched for more than 0.5s.
Wires from the linear axis are not properly
connected into the UR controller.
[Manual] [I/O] Connect the wires according to the "Safety connections" section.
"Safety configuration" and/or "System
Emergency Stopped" are not properly
configured.
[Installation] [I/O] Verify that "Safeguard reset" and "System Emergency Stopped"
functions are properly configured.
(
Installation
->
Safety
->
I/O
)
Safeguard function is not used and not bypassed.
[I/O] If the safeguard function is not used, bypass this function by adding a jumper
between pin 1-2 and 3-4 of the provided connector.
[Manual] See "Safety connections" section.
An OSSD safeguard device is used directly.
A safety relay to convert the signal from OSSD to standard signal is required.
[Info] OSSD signals are not compatible with the AX series controller.
[Info] We recommend the PSR-MS60-2NO-24DC-SC by Phoenix Contact.
Emergency output from the UR controller is set to
"OSSD signal".
(e-Series only)
Verify that the OSSD checkbox for the "System Emergency Stopped" output is
unchecked. (Installation -> Safety -> I\O)
[Info] OSSD signals are not compatible with the AX series controller.
[AX series] "Unable to
find the linear axis. Make
sure it is connected and
try again."
Ethernet connection between the UR controller
and linear axis controller is faulty.
Verify connection and power cycle (power-off then power-on) both controllers.
USB to Ethernet adapter is not used.
[Manual] Refer to the installation steps.
© 2022 Robotiq inc. 5
Problem / issue
Cause(s)
Solution(s)
"Collision detected"
during program
execution
Pallet viewer state does not match reality.
Box number parameter was set higher than what
is really on the pallet.
[Pallet viewer] Restart the program and select the right step using the “
+
” and “
-
buttons.
No feasible trajectory was found, the program
cannot avoid a collision.
[Online configurator] Use the simulation to confirm the pattern is possible with the
actual installation. Consider changing the following elements :
- TCP offset (length of the extension bracket);
- Label orientation;
- Pattern.
[Installation] Apply the changes to the actual installation, if required.
[Software] When possible, allow the gripper to pick the box at 90° and/or 180°.
(
Palletize
r node ->
Settings
)
[Pick position] When possible, change the gripper orientation when teaching the
pick position.
(
Palletizer
node ->
Box
)
Virtual collision between the box and the virtual
planes around the pallet.
Increase the distance from these planes to the pallet (
Palletize
r node ->
Settings
).
[Info] A minimal distance of 50 mm is recommended.
"RTDE interface cannot
initialize. Make sure that
Ethernet/IP adapter and
Profinet are disabled, and
that no other URCaps use
the RTDE interface."
Register conflict.
[Software] Upgrade to the latest version of Copilot and Polyscope.
[Support] Contact Robotiq’s technical support if the problem persists.
© 2022 Robotiq inc. 6
Installation issues
[AX series] Unable to
scan the Robotiq
controller.
Ethernet connection between the UR controller
and the linear axis controller is faulty.
Connect the Ethernet adapter.
Shut down and restart both controllers.
[Support] Contact Robotiq’s technical support if the problem persists.
Unable to touch the base
with the robot’s wrist
when performing the
centering procedure.
The robot’s first joint, also known as the base
joint, is near its limit (close to +/- 360 degrees).
Center the base joint (close to 0°) using the
Move
tab.
[Manual] Perform the centering procedure.
Ch. 3, “Mechanical and Electrical Installation”, “Centering the Solution” section.
Protective stop
Protective stop occurs
after a couple of layers.
Collision between the gripped box and the box
on the pallet.
A small Z offset accumulates over many layers.
[Software][Box] Increase the box height parameter (C) to compensate for box
variability.
Use the average box height : Measure the height of a full stack of boxes and divide
it by the number of layers.
Protective stop occurs
when placing a box on
the first layer.
Box height is set too low.
[Software][Box] Adjust the box height parameter (C) to compensate for box
variability.
Use the average box height : Measure the height of a full stack of boxes and divide
it by the number of layers.
Pallet height is set too low.
[Software][Pallet] Measure your pallet and set the pallet height in the
Palletizer
node.
Wrong TCP.
[Software][Installation] Change your TCP Z value to match your gripper.
[Manual] Refer yourself to the documentation (Ch. 6, “Center of Mass, Tool Center
Point (TCP)” section),
or use Polyscope’s TCP wizard.
Protective stop occurs
while palletizing.
Cable management system hinders the robot’s
movements.
Install the cable management system according to the procedure.
[See appendix B]
[Manual] Additional info in Ch. 3, “Mechanical and Electrical Installation” section.
© 2022 Robotiq inc. 7
Problem / issue
Cause(s)
Solution(s)
Unexpected behavior
Palletizer releases boxes
from too high.
A small Z offset accumulates over many layers.
[Box] Decrease the box height parameter (C).
Use the average box height : Measure the height of a full stack of boxes and divide
it by the number of layers.
Wrong TCP.
[Software][Installation] Validate the TCP is located over the box at the pick position.
Pallet height is set too low.
[Pallet] Measure the pallet height and set the parameter correctly.
[PE series] The pedestal height does not match
the height configured in the installation tab.
[Installation] Adjust either the actual pedestal height or the parameter in the
installation tab accordingly.
(
Installation
tab ->
URCaps
->
Copilot
->
Configuration
)
Boxes are not always
placed straight on the
pallet.
Box pick position is not stable.
[Conveyor] [Pick position] Adjust the conveyor stop rig to generate a precise and
repeatable box pick position.
A stable box position ensures a reliable box positioning on the pallet.
Box pick position is incorrect.
[Box][Pick position] While teaching the pick position, make sure that the vacuum
gripper is squared and centered with the box.
TCP is not defined correctly.
[Installation] Adjust the orientation of the TCP to match the orientation of the
gripper.
[AX series] Linear axis is not aligned with the
base.
[Manual] Perform the centering procedure.
Ch. 3, “Mechanical and Electrical Installation”, “Centering the Solution” section.
[PE series] Pedestal is not aligned with the base.
[Manual] Perform the centering procedure.
Ch. 3, “Mechanical and Electrical Installation”, “Centering the Solution” section.
Robot is waiting over the
pick box position
indefinitely.
Box presence input is not properly configured.
[Box sensor] When using a custom box sensor, the sensor should be set to "active
high" = Normally Open.
Box presence input is not well defined.
[Software] [Box sensor] [I/O] Set the Box presence input to the one the box sensor
is connected to.
(
Palletizer
node ->
Settings
)
Palletizer stays idle after
the pallet has been
changed.
[PE series] "Pallet confirmed" safety option is
enabled.
Press the "Pallet confirmed" button once the pallet has been changed.
[AX series] Active drive
axis position is wrong.
Active Drive menu shows a negative “Actual
position” for the linear axis.
[Support] Contact Robotiq’s technical support.
A program will be provided to fix the issue.
© 2022 Robotiq inc. 8
APPENDIX A
Air Supply setup recommendations
The air supply must be dried and filtered according to ISO 8573-1 class 3.4.3.
Air supply pressure must be between 3 and 7 bar with an optimal pressure at 7 bar. (6.5 bar for Airpick serial number ≤ V-01351)
© 2022 Robotiq inc. 9
APPENDIX B
Cable management system (valid for AX and PE series)
Notes :
-Cut the air tube at the right length (815 mm / 32 in) past the plastic strain relief.
-For more details about the cable management system, consult the AX or PE series manual (Ch. 3, “Mechanical and Electrical
Installation” section).
© 2022 Robotiq inc. 10
APPENDIX C
Safety connections
© 2022 Robotiq inc. 11
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11

ROBOTIQ PALLETIZING SOLUTION User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI