Invacare XPO2 is a portable oxygen concentrator that provides a convenient and reliable source of oxygen therapy for active individuals with respiratory conditions. It features a compact and lightweight design, making it easy to carry and transport. With its long battery life and pulse-dose delivery system, the XPO2 ensures uninterrupted oxygen delivery during various activities. Its user-friendly controls and simple operation make it suitable for patients of all ages. Additionally, the XPO2's low noise level and discreet design allow for comfortable and discreet use in various settings.

Invacare XPO2 is a portable oxygen concentrator that provides a convenient and reliable source of oxygen therapy for active individuals with respiratory conditions. It features a compact and lightweight design, making it easy to carry and transport. With its long battery life and pulse-dose delivery system, the XPO2 ensures uninterrupted oxygen delivery during various activities. Its user-friendly controls and simple operation make it suitable for patients of all ages. Additionally, the XPO2's low noise level and discreet design allow for comfortable and discreet use in various settings.

DEALER:Keepthismanual.TheproceduresinthismanualMUSTbeperformedbyaqualified
technician.
I EN I
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorseriousinjury.
Warningindicatesapotentiallyhazardoussituationwhich,
ifnotavoided,couldresultindeathorseriousinjury.
Cautionindicatesapotentiallyhazardoussituationwhich,
ifnotavoided,mayresultinpropertydamageorminor
injuryorboth.
Informationgivesusefultips,recommendationsand
informationforefficient,trouble-freeuse.
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.
ThelaunchdateofthisproductisstatedintheCE
declarationofconformity.
1.2KitOverview
ThiskitcontainscomponentstoreplacetheXPO
2
internalbattery.
Theproperinstallationofthiskitinvolvesperformingtheprocedures
listedbelow:
qAccesstheXPO
2
battery.
qReplacethebattery.
qReassembletheXPO
2
.
1.3KitInformation
PARTNUMBER
DESCRIPTION
n/a
InternalBatteryPack
n/aSeal/NoticeLabel
PARTNUMBER
DESCRIPTION
n/a
Ferrite
1154285BatteryReplacementService
Manual
1.4CustomerServiceandWarrantyInformation
Forwarrantyinformation,pleaserefertotheoriginalowner’s
manualwhichcamewiththisproduct,orcontactInvacareformore
information.
2Safety
2.1GeneralGuidelines
TheSafetysectioncontainsimportantinformationforthesafe
operationanduseofthisproduct
WARNING!
DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand
understandingtheseinstructionsandanyadditional
instructionalmaterialsuchasowner’smanuals,service
manualsorinstructionsheetssuppliedwiththis
productoroptionalequipment.Ifyouareunableto
understandthewarnings,cautionsorinstructions,
contactahealthcareprofessional,dealerortechnical
personnelbeforeattemptingtousethisequipment-
otherwise,injuryordamagemayoccur.
WARNING!
ACCESSORIES
Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare
accessories.Accessoriesdesignedbyother
manufacturershavenotbeentestedbyInvacareand
arenotrecommendedforusewithInvacareproducts.
Invacare®XPO
2
Portable
OxygenConcentrator
KitNo.1157695
ENBatteryReplacement
ServiceManual....................................1
FRRemplacementdelabatterie
/CPWGNFʼnGPVTGVKGP...............................3
DEBatterieaustausch
Servicehandbuch . . . . ..............................6
XPO2™PortableOxygenConcentrator
2.2Installation
WARNING!
Riskofinjury.
TheseproceduresMUSTbeperformedbyaqualified
technician.
WhenaccessingtheXPO
2
battery,rememberthat
batterypowerisstillbeingappliedtotheinternal(PCB)
printedcircuitboard.BatterypowerremainsonthePCB
aslongasthebatteryconnectionismadetothePCB.
CareMUSTbetakentominimizecontactwiththe
PCB.DONOTuseanykindoftool(screwdriver,
pliers,etc.)whenworkingontheinternalstructure.
Otherwise,injuryand/ordamagemayoccur.
CAUTION!
Riskofdamage.
DONOTdamageribboncablewhenremovingor
installingfrontcover.
DONOTuseMethylEthylKetone(MEK).Thiswill
damagetheplastic.
Checkallpartsforshippingdamage.Ifdamageisfound,DO
NOTuse.ContactCarrier/Invacareforfurtherinstructions.
ToensurethesafeandproperreplacementoftheXPO
2
battery,
theseinstructionsMUSTbefollowed.
3Service
3.1AccessingtheXPO
2
Battery
1.RemovethelabelAfromthetopoftheXPO
2
B(Figure1).
Thislabeladherestothefrontandrearsectionsofthe
XPO
2
.ThislabelMUSTberemovedtoseparatethetwo
sections.
2.PositiontheXPO
2
sothatthebackpanellabelCisfacingup
(Figure2).
3.UseaT15Torx®screwdrivertoremovethefourmounting
screwsDfromthebackoftheXPO
2
.
4.Whileholdingtheunittogether,rotatetheXPO
2
andpositionit
sothatthebackpanellabelisfacingdown.
TheribboncableEisconnectedtothefrontcoverF
andtheinternalprintedcircuitboard(PCB).Onlyopen
theXPO
2
asshowninFigure3andFigure4.
5.Slowly,openthefrontcoverF,ensuringnottodamagethe
ribboncableE(Figure3andFigure4).
3.2ReplacingtheXPO
2
Battery
CAUTION!
Beforehandlingtheinternalprintedcircuitboard
(PCB),ensureyouareproperlygroundedtoprevent
staticdamagetothecomponentsoftheboard.
AstaticcuffMUSTbewornandproperlygrounded
usinganalligatorclip.
Electricalconduitorwaterpipeisnormallysufficient
whenaknowngoodgroundisunavailable.
Careshouldbetakentoensurethatthealligatorclip
contactsabaremetalsurface.
1.Performthestepsin3.1AccessingtheXPO
2
Battery,page2.
2.RemovetheinletfilterHwithtubingIattachedfromtheholder
(Figure5).
3.DisconnectthebatterywiringharnessJfromtheinternal
printedcircuitboard(PCB)G.
4.RemovethebatteryKfromthethermalpadandbatterytray.
2
1154285B
Thethermalpadisbetweenthebatteryandthebottom
ofthetray.Itisasoft“dough-like”materialthatadheres
tothetray.Ifthermalpadisdamagedorpulledupduring
batteryremovalitMUSTbereplaced.
5.LocatetheblackwireLandtheredwireMonthenewbattery
pack(Figure6).
6.Movetheblackandredwiresawayfromtherestofthewires
(yellow,whiteandgreen).
TheredandblackwiresCANNOTbetwisted.See
Figure6.
7.OpentheferriteN.
8.Placetheferritearoundtheblackandredwires.
9.Closeandlocktheferritearoundtheblackandredwires.
10.Slidetheferritetoensureitisnotpinchingthewires.
11.PositionthenewbatteryonthethermalpadasshowninFigure5.
Wheninstallingthenewbattery,ensurethebatterylabel
facesup.
12.ConnectthebatterywiringharnessJtotheinternalprinted
circuitboard(PCB)G.
13.Re-assembletheXPO2.Referto3.3ReassemblingtheXPO
2
,
page3.
3.3ReassemblingtheXPO
2
1.Carefully,re-installthefrontcoverFontotheXPO
2
B(Figure
4).
2.TurntheXPO
2
Ontoensureproperoperation.
3.TurntheXPO
2
Off.
4.Whileholdingtheunittogether,positiontheXPO
2
sothatthe
backpanellabelCisfacingup.
5.InstallfourmountingscrewsDandsecurethetwosectionsof
theXPO
2
together(Figure2).Securelytighten.
6.RemovetheglueresiduefromthetopoftheXPO
2
,wherethe
labelwasremoved.Apply91%isopropylalcohol,andwipewitha
cleancloth.Makesurethesurfaceisdryandfreeofanyresidue.
Thiswillensureagoodsealforthenewlabel.
7.InstallreplacementlabelAonthetopoftheXPO
2
sothatit
adherestoboththefrontandrearsections(Figure1).
I FR I
1Généralités
1.1Symboles
Lesmentionsd’avertissementutiliséesdansleprésentmanuel
s’appliquentauxrisquesouauxpratiquesdangereusesquipourraient
provoquerdesblessuresoudesdommagesmatériels.Reportez-vous
auxinformationsci-dessouspourladéfinitiondesmentions
d’avertissement.
LamentionDangerindiqueunesituationdangereuse
imminentequi,siellen’estpasévitée,provoquerades
blessuresgravesoulamort.
LamentionAvertissementindiqueunesituation
potentiellementdangereusequi,siellen’estpasévitée,
pourraitprovoquerdesblessuresgravesoulamort.
LamentionAttentionindiqueunesituation
potentiellementdangereusequi,siellen’estpasévitée,
peutprovoquerdesdommagesmatérielset/oudes
blessureslégères.
1154285B3
UnitedKingdomInvacareLimitedPencoedTechnologyPark,Pencoed,BridgendCF35
5HZTel:(44)(0)1656776222Fax:(44)(0)1656776220
www.invacare.co.uk[email protected]
Ireland
InvacareIrelandLtdUnit5SeatownBusinessCampusSeatownRoad
SwordsCountyDublinIrelandTel:(353)18107084Fax:(353)
18107085www.invacare.ie[email protected]
France
InvacarePoirierSASRoutedeStRochF-37230FondettesTel:
(33)(0)247626466Fax:(33)(0)247421224www.invacare.fr
Switzerland
InvacareAGBenkenstrasse260CH-4108WitterswilTel.:(41)
(0)614877080Fax.:(41)(0)614877081www.invacare.ch
Deutschland
InvacareAquatecGmbHAlemannenstraße10D-88316IsnyTel:
(49)(0)75627000Fax:(49)(0)756270066www.invacare.de
España
InvacareSAc/Arenys/nPolígonIndustrialdeCelràE-17460
Celrà(Girona)Tel:(34)(0)972493200Fax:(34)(0)972493220
www.invacare.es[email protected]
Belgium&LuxemburgInvacarenvAutobaan22B-8210LoppemTel:(32)(0)50831010
Fax:(32)(0)50831011[email protected]www.invacare.be
EuropeanDistributorOrganisation
InvacareEuropeanDistributorOrganisationKleiststraße49D-32457
PortaWestfalicaTel:(49)(0)5731754540Fax:(49)(0)5731754
541[email protected]www.invacare.eu.com
Nederland
InvacareBVCelsiusstraat46NL-6716BZEdeTel:(31)(0)318
695757Fax:(31)(0)318695758[email protected]
www.invacare.nl[email protected]
Österreich
InvacareAustriaGmbHHerzogOdilostrasse101A-5310
MondseeTel:(43)623255350Fax:(43)623255354
[email protected]www.invacare.at
Portugal
InvacareLdaRuaEstradaVelha949P-4465-784LeçadoBalioTel:
(351)(0)225105946/47Fax:(351)(0)225105739www.invacare.pt
Danmark
InvacareA/SSdr.Ringvej37DK-2605BrøndbyTel:(45)(0)36
900000Fax:(45)(0)36900001[email protected]
www.invacare.dk
Norge
InvacareASGrensesvingen9Postboks6230EtterstadN-0603
OsloTel:(47)(0)22579500Fax:(47)(0)22579501
Sverige&Suomi
InvacareABFagerstagatan9S-16391SpångaTel:(46)(0)876170
90Fax:(46)(0)876181[email protected]www.invacare.se
ItaliaInvacareMeccSans.r.l.,ViadeiPini62,I-36016Thiene(VI)Tel:
(39)0445380059Fax:(39)0445380034[email protected]
www.invacare.it
©2013Invacare®Corporation.Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithoutprior
writtenpermissionfromInvacare.Trademarksareidentifiedby™and®.AlltrademarksareownedbyorlicensedtoInvacareCorporationor
itssubsidiariesunlessotherwisenoted.
TorxisaregisteredtrademarkofAcumentIntellectualProperties,LLC.
EURepresentative
InvacareDeutschlandGmbH
Kleiststraße49
D-32457PortaWestfalica
Tel:+49(0)57317540
Tel:+49(0)573175452191
Manufacturer
InvacareCorporation
2101E.LakeMaryBlvd.
Sanford,FL32773
USA
Tel:407–321–5630
InvacareCorporation
USA
OneInvacareWay
Elyria,OhioUSA
44036–2125
440–329–6000
800–333–6900
TechnicalServices
440–329–6593
800–832–4707
www.invacare.com
1154285-B2013-01-29
*1154285B*
MakingLife’sExperiencesPossible™
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12

Invacare XPO2 User manual

Type
User manual

Invacare XPO2 is a portable oxygen concentrator that provides a convenient and reliable source of oxygen therapy for active individuals with respiratory conditions. It features a compact and lightweight design, making it easy to carry and transport. With its long battery life and pulse-dose delivery system, the XPO2 ensures uninterrupted oxygen delivery during various activities. Its user-friendly controls and simple operation make it suitable for patients of all ages. Additionally, the XPO2's low noise level and discreet design allow for comfortable and discreet use in various settings.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI

in other languages