The algorithms are designed to work for VENTANA PD-L1
(SP263) Assay. The test results are only as good as the quality
and accuracy of the IHC slide that is imaged, and the subsequent
image that is analyzed.
The pathologist must validate the VENTANA PD-L1 (SP263)
Assay staining run by manual microscopic examination of the
PD-L1 control slides to verify that the expected results have been
obtained before images from patient slides are acquired by the
uPath RUO enterprise software.
The pathologist must follow the manufacturer’s
recommendations for the VENTANA PD-L1 (SP263) Assay
including using all the positive and negative quality control
materials for each staining run.
If the control slides are not acceptable with manual
microscopic examination, the patient tissues need to be re-
stained with acceptable results. (See the VENTANA PD-L1
(SP263) Assay method sheet for details about quality control
recommendations.)
The pathologist must follow the recommendations for VENTANA
PD-L1 (SP263) Assay interpretation as recommended in the
VENTANA PD-L1 (SP263) Assay method sheet (available at www.
ventana.com).
If the images captured have different staining (nuclear,
cytoplasm, etc.), incorrect results will be generated.
uPath PD-L1 (SP263) image analysis for NSCLC will reject nuclei
that are elongated regardless of the overall shape of the cell.
For this reason, tumors containing large numbers of cells with
elongated nuclei may need to be evaluated manually.
Although it is required for macrophages to be excluded from
the region of analysis, it is not always possible to capture all
macrophages. Thus, the algorithm result is influenced by
remaining macrophages in the region of analysis. This is critical
when a patient score is close to the cutoff (either 1% or 50%).
In addition, performance of the Roche uPath RUO enterprise
software for uPath PD-L1 (SP263) image analysis for NSCLC with
the following types of lung cancers has not been evaluated: small
cell carcinoma, fetal adenocarcinoma, enteric adenocarcinoma,
squamous cell carcinoma in situ (SCIS), combined small cell
carcinoma, large cell neuroendocrine carcinoma, typical
carcinoid tumor, atypical carcinoid tumor, pleomorphic
carcinoma, spindle cell carcinoma, giant cell carcinoma,
carcinosarcoma, pulmonary blastoma, lymphoepithelioma-
like carcinoma, NUT carcinoma, mucoepidermoid carcinoma,
adenoid cystic carcinoma, and epithelial-myoepithelial
carcinoma.
This device has not been tested, or its safety and effectiveness
validated, when used with a personal computer (PC) from home.
According to the 1988 Clinical Laboratory Improvement
Amendments (CLIA ‘88), each laboratory that introduces
an FDA cleared system must demonstrate that it can obtain
performance specifications comparable to those established by
the manufacturer.
Limitations
uPath PD-L1 (SP263) image analysis for Non-Small Cell Lung Cancer Algorithm Guide 5