For initial set up, we strongly recommend that you also refer to the relevant manual!
05/2020 – 4008215801 – QS OPTITEMP TT 12 C R01 en
Programmable 2-wire transmitter
For Pt100 and Pt1000 input
Installation, assembly, start-up and maintenance may only be performed by appropriately trained personnel.
Check the nameplate for operating conditions.
Instruments must not be connected to power supply before reading instructions described in the manual.
The responsibility as to the suitability, intended use of this device rests solely with the operator.
For complete documentation (manuals, supplementary manuals, data sheets and certificates) please refer to
www.krohne.com/Downloads.
1 Installation
•The TT 12 C is designed to fit inside connection heads type DIN B or larger.
•A convenient way to install the transmitter is to use the mounting kits for in-head and DIN rail
mounting. In order to minimize measuring errors make sure the connecting screws are tightened
enough, 0.5-0.6 Nm.
•Connect input, output and power supply according to “Electrical connection”.
•The transmitter is configured from a mobile device via NFC® communication, by using the app
OPTICHECK Temperature Mobile. The app is avaiable for free download on App Store and
Google Play™.
Transmitter Tmin. Tmax.
[°C] [ °F] [°C] [ °F]
TT 12 C -40 -40 +85 +185
Input Pt100 3W, 0 to 100 °C (equivalent to 212°F)
Output 4 to 20 mA
Filtering level Response time 0.9 s
Sensor error indication Upscale
.
Installation procedure: Step 1
1. Place the transmitter on the clip.
2. Press down the transmitter until it is attached to the clip.
Screw the transmitter on to the clip.
Installation procedure: Step 2
1. Hook one end of the clip into the rail.
2. Push the other end down until it snaps onto the rail.
3. Release by pushing the hook and at the same time lift the
clip out of the rail.
How to mount on a DIN-rail adapter