The Ion Chef™ System features a simple interface for loading chips, cleaning the
instrument, and performing system maintenance and conguration tasks.
2
1
3 4 5
1Set up Run button – Set up the Ion Chef™
template preparation and chip-loading
routine. Choose Step by Step to have the
instrument lead you stepwise through
installation of reagents and consumables,
or choose Quick Start to proceed if you
have already installed the consumables.
2Open Door button.
3Notifications button – View notifications
about instrument status during and
between runs.
4Quick Start button – Proceed directly to
the Quick Start instrument setup mode.
User verifies the loading of a new pipette
tip cartridge and an empty pipette tip rack
to hold waste tips generated during the
run, before proceeding to Deck Scan.
5Settings button – Advance to a screen
providing the following options:
•Notifications: view notifications about
instrument status during and between
runs
•Instrument Settings: view current
settings and network configuration,
set instrument name, adjust time zone
•Service tools: access screens for
service-related maintenance and
instrument diagnostics
•Torrent Server: add and manage
Torrent Server connections
•Clean Ion Chef: proceed directly to the
instrument cleaning routine
•Check for updates: check availability
of system software updates
Precautions
IMPORTANT! A primary source of contamination is spurious DNA fragments from
previous sample processing steps. Do not introduce amplied DNA into the library
preparation laboratory or work area.
IMPORTANT! To avoid possible damage to the chip due to electrostatic discharge,
ground yourself before picking up a chip or placing a chip on a surface such as a lab
bench. For example, touch the metal trim on the chip compartment before inserting or
removing a chip from the chip clamp.
About the Ion
Chef™ System
touchscreen
interface
Avoid nucleic acid
contamination
Avoid chip damage
Chapter 1 Product information
Precautions
1
14
Ion 520
™
& Ion 530
™
Kit − Chef User Guide