Samsung SONOACE X4 User manual

Type
User manual

Samsung SONOACE X4 is a high-performance ultrasound system that provides exceptional image quality and advanced features for a wide range of clinical applications. With its compact design and intuitive user interface, the SONOACE X4 is easy to use and can be tailored to meet the specific needs of your practice.

Some of the key features and capabilities of the SONOACE X4 include:

  • Exceptional image quality: The SONOACE X4 utilizes advanced imaging technologies to provide crystal-clear images with excellent resolution and contrast. This allows you to visualize anatomical structures and pathologies with greater detail and accuracy.

Samsung SONOACE X4 is a high-performance ultrasound system that provides exceptional image quality and advanced features for a wide range of clinical applications. With its compact design and intuitive user interface, the SONOACE X4 is easy to use and can be tailored to meet the specific needs of your practice.

Some of the key features and capabilities of the SONOACE X4 include:

  • Exceptional image quality: The SONOACE X4 utilizes advanced imaging technologies to provide crystal-clear images with excellent resolution and contrast. This allows you to visualize anatomical structures and pathologies with greater detail and accuracy.
Chapter 7
Maintenance
System Maintenance......................................................................................................2
Installation Requirements ......................................................................................................2
Cleaning and disinfections.....................................................................................................3
Fuse Replacement.................................................................................................................4
Cleaning The Air Filters..........................................................................................................5
Monitor Filter Replacement....................................................................................................6
Accuracy Check.....................................................................................................................7
Administration of Information..........................................................................................8
User Setting Back up.............................................................................................................8
Patient information Back-up...................................................................................................8
Software.................................................................................................................................8
7 - 2 SONOACE X4 Operation Manual
System Maintenance
NOTE
The user must ensure that the safety inspections are performed every 2
years according to the requirements of safety standard EN 60601-1.
Only trained persons are allowed to perform the safety inspections
mentioned above.
Installation Requirements
When installing:
Optimal conditions for the system are temperatures of 10° ~ 35° C and humidity of
30% ~ 75%.
Avoid humidity.
Avoid direct sunlight.
Avoid places with extreme temperature variations.
Avoid heat sources.
Avoid dusty and unventilated areas.
Avoid places where the system is likely to be exposed to vibration or impacts.
Avoid places where the system is likely to be exposed to chemical substances or gases.
CAUTION
Placing the system near generators, X-Ray machines, or broadcast cables
may result in screen noise and abnormal visual images. Using the power
source with other electric devices may also induce noise.
Chapter 7. Maintenance 7 - 3
Cleaning and disinfections
WARNING
Always use protective eyewear and gloves when cleaning and disinfecting
the equipment.
Cleaning
1. Turn off the system and disconnect the system power cord from the wall outlet.
2. Use a soft cloth lightly dampened in a mild soap or detergent solution to clean exterior
surfaces on the system.
Disinfection
CAUTION
Use only recommended disinfectants on system surfaces.
A disinfectant qualified by the FDA 510(k) process is recommended. The following
disinfectants are recommended because of both their biological effectiveness (as qualified
through the FDA 510(k) process) and their chemical compatibility with MEDISON ultrasound
products.
Solutions Country
Type Active ingredient FDA 510(k)
Cidex USA Liquid Gluteraldehyde K934434
Cidex Plus USA Liquid Gluteraldehyde K923744
1. Turn off the system and disconnect the system power cord from the wall outlet.
2. Mix the disinfection solution compatible with your system according to label instructions
for solution strength.
3. Wipe the system surfaces with the disinfectant solution, following the disinfectant label
instructions for wipe durations, solution strength, and disinfectant contact duration.
4. Air dry or towel dry with a sterile cloth according to the instructions on the disinfectant
label.
7 - 4 SONOACE X4 Operation Manual
Fuse Replacement
The power protection fuse protects the product from excess current. If the power monitoring
protection circuit detects excess current, it shuts off the current to the equipment in order to
prevent overheating and to restrict the ultrasound power output.
If the fuse blows, replace it as shown below.
[Figure 7.1 Fuse Replacement]
1. Disconnect the power cord from the system. (See
)
2. Take the fuse out by pressing the middle of fuse. (See
)
3. Insert the fuse holder into the appliance inlet.
4. After installing the new fuse, connect the power cord to the system.
Fuse information is shown in the following table.
DANGER
To avoid risk of electric shock, always disconnect the plug from the system
prior to fuse replacement.
Input Ratings Fuse Ratings Company Order No.
100-120VAC
8A/250V Orisel 55T210000
200-240VAC
5A/250V Orisel 55T210000
Chapter 7. Maintenance 7 - 5
Cleaning The Air Filters
The air filters minimize the indraft of dust. Clean the air filter to ensure that a clogged filter does
not cause the system to overheat and reduce the noise and the system performance.
It is recommended the air filters be cleaned once every three months.
[Figure 7.2 The location of the air filters]
1. Pull the filter under the front of the console
to away from the product.
2. Shake the filter to remove the dust and
wash in a mild soapy solution.
3. Rinse and air dry or dry with a cloth.
4. Slide the filter back into the product.
CAUTION
Be sure to lock the brakes on the front wheels before cleaning the air filters
to avoid injury by any unexpected movement of the product.
NOTE
Allow the wet filter to dry thoroughly before installing. The wet filter can
cause the malfunction.
7 - 6 SONOACE X4 Operation Manual
Monitor Filter Replacement
Monitor filter not only makes the eyes less tired but also allow clear viewing of images even in
dark places. Monitor filter also protects the monitor surface, preventing scratches and other
similar damages.
Follow the instructions below to replace monitor filter.
[Figure 7.3 Monitor filter replacement]
1. Press the two latches on the upper part of the monitor.
2. Hold the upper part of the monitor filter and pull it out carefully. Be careful not to drop
the monitor filter when detaching it.
3. Push the new monitor filter in, aligning the lower part of the filter to the slot on the
bottom of the monitor.
4. Press the latches again and lightly press the upper part of the filter.
NOTE
You can easily press the latches and detach the filter by using a coin or a
thin object.
Chapter 7. Maintenance 7 - 7
Accuracy Check
The product’s maintenance status may affect the measurements obtained using the product.
The product should be maintained in an optimal state to ensure reliable measurements.
To ensure optimal operation of the product, perform an accuracy check every year. The equati-
ons and table related to measurement accuracy are included in “Chapter 5 Measurements” in
this manual.
7 - 8 SONOACE X4 Operation Manual
Administration of Information
CAUTION
You may lose information files on user settings or patients, because of shock on
the product or internal error. Thus, back-up on a regular basis.
User Setting Back up
- Always keep a backup copy of all information related to the user settings in case of data loss.
- Clients cannot back-up the user settings of the product. Contact the MEDISON Customer
Service Department to attain support for back-up.
- However, clients may back up the user setting on GA Table used in obstetrics diagnosis. For
further information please refer to “Chapter 3. Settings”.
Patient information Back-up
The SonoView program can be used for backing up patients’ basic information and scanned
images. The user can choose to save the data, and the data is also saved in the system by
default. If the system needs to be reinstalled due to product failure, etc., the MEDISON
customer support staff will restore the patients’ basic information and scanned images that are
saved in the system. For more information on this, see “Chapter 6 Image Managements.”
Software
The product software may be updated to enhance performance. The user cannot make any changes to
the software. Please contact the MEDISON customer service representative for help in software changes.
CAUTION
Minor software updates may be carried out without the prior notice from the
manufacturer.
Should errors occur in the operating system (Linux), and should you desire to upgrade the
operating system, please follow the instructions of the operating system manufacturer.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8

Samsung SONOACE X4 User manual

Type
User manual

Samsung SONOACE X4 is a high-performance ultrasound system that provides exceptional image quality and advanced features for a wide range of clinical applications. With its compact design and intuitive user interface, the SONOACE X4 is easy to use and can be tailored to meet the specific needs of your practice.