SONOACE R7

Samsung SONOACE R7 User manual

  • Hello! I am an AI chatbot trained to assist you with the Samsung SONOACE R7 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Volume 2
Operation Manual
Version 3.01.00
English
Diagnostic Ultrasound System
MI68-02773A
5
Table of Contents
Table of Contents – Volume 2
Chapter6 Image Management
Reviewing Images (Cine / Loop) ............................................................................................. 6-3
Annotating Images ................................................................................................................... 6-6
Text .......................................................................................................................................................................................... 6-6
BodyMarker ......................................................................................................................................................................... 6-9
Indicator .............................................................................................................................................................................6-11
e-Motion Marker (Optional) .........................................................................................................................................6-13
Saving, Playing and Transferring Images ............................................................................ 6-16
Saving Images ..................................................................................................................................................................6-16
Playing Images ..................................................................................................................................................................6-18
Transferring Images .......................................................................................................................................................6-19
Printing and Recording Images ........................................................................................... 6-20
Printing Images .................................................................................................................................................................6-20
Recording Images ...........................................................................................................................................................6-20
SonoView ...................................................................................................................................6-21
Exam Mode ......................................................................................................................................................................... 6-21
Compare Mode .................................................................................................................................................................6-23
Managing Exam Images ................................................................................................................................................6-24
Chapter7 Utility
System Settings .......................................................................................................................... 7-3
General System Settings .................................................................................................................................................7-3
Display Settings .................................................................................................................................................................. 7-7
Annotate Settings ............................................................................................................................................................7-11
Peripherals Setup .............................................................................................................................................................7-15
System Information ........................................................................................................................................................7-18
DICOM Setup (Optional) ................................................................................................................................................7-19
Utilities Setup ....................................................................................................................................................................7-35
Option Setup .....................................................................................................................................................................7-38
Auto Calc .............................................................................................................................................................................7-39
6
Operation Manual
Measurement Setup ................................................................................................................7-40
General Measurement Setup ......................................................................................................................................7-40
Obstetrics Measurement Setup ..................................................................................................................................7-54
Cardiac Measurement Setup .......................................................................................................................................7-62
Vascular Measurement Setup .....................................................................................................................................7-63
Urology Measurement Setup ......................................................................................................................................7-64
Fetal Echo Measurement Setup..................................................................................................................................7-66
Utility .......................................................................................................................................... 7-67
ECG ........................................................................................................................................................................................7-69
Biopsy ...................................................................................................................................................................................7-71
Histogram ...........................................................................................................................................................................7-73
Post Curve ...........................................................................................................................................................................7-75
Setup ..................................................................................................................................................................................... 7-80
Measure Setup ..................................................................................................................................................................7-80
Storage Manager ..............................................................................................................................................................7-81
BodyMarker Edit ...............................................................................................................................................................7-83
e-Motion Marker ..............................................................................................................................................................7-85
Menu Edit ............................................................................................................................................................................7-86
Stress Echo (Optional) ............................................................................................................. 7-88
Setting Up a Protocol ......................................................................................................................................................7-89
Starting a Protocol and Capturing Images .............................................................................................................7-93
Reviewing an Image .......................................................................................................................................................7-97
Strain Image (Optional) ........................................................................................................7-101
Quantication ................................................................................................................................................................. 7-102
Strain Review .................................................................................................................................................................. 7-104
Auto EF .............................................................................................................................................................................. 7-109
7
Table of Contents
Chapter8 Maintenance
System Maintenance ................................................................................................................. 8-3
Installation Requirements............................................................................................................................................... 8-3
Cleaning and disinfections ............................................................................................................................................ 8-4
Fuse Replacement ............................................................................................................................................................. 8-6
Cleaning The Air Filters .................................................................................................................................................... 8-7
Accuracy Check ................................................................................................................................................................... 8-8
Administration of Information ................................................................................................ 8-9
User Setting Back up ......................................................................................................................................................... 8-9
Patient information Back-up ..........................................................................................................................................8-9
Software................................................................................................................................................................................. 8-9
Chapter9 Probes
Probes .......................................................................................................................................... 9-3
Ultrasound transmission Gel .......................................................................................................................................9-19
Sheaths.................................................................................................................................................................................9-20
Probe Precautions ............................................................................................................................................................9-21
Cleaning and Disinfecting the Probe .......................................................................................................................9-23
Biopsy .........................................................................................................................................9-35
Biopsy KIT Components ................................................................................................................................................9-35
Using the Biopsy Kit ........................................................................................................................................................9-36
Cleaning and Disinfecting Biopsy Kit .......................................................................................................................9-38
Assembling the Biopsy Kit ............................................................................................................................................9-40
** Reference Manual
Samsung Medison is providing an additional SONOACE R7 Reference Manual. GA tables and references
for each application are included in the Reference Manual.
Chapter 6
Reviewing Images (Cine / Loop) .....................6-3
Annotating Images .........................................6-6
Text ................................................................................................6-6
BodyMarker ..............................................................................6-9
Indicator ...................................................................................6-11
e-Motion Marker (Optional) ..............................................6-13
Saving, Playing and Transferring Images .. 6-16
Saving Images ........................................................................6-16
Playing Images .......................................................................6-18
Transferring Images .............................................................6-19
Printing and Recording Images .................. 6-20
Printing Images ......................................................................6-20
Recording Images ................................................................6-20
SonoView ....................................................... 6-21
Exam Mode ..............................................................................6-21
Compare Mode ......................................................................6-23
Managing Exam Images .....................................................6-24
Image Management
Chapter6 Image Management
6-3
Chapter 6 Image Management
Reviewing Images (Cine / Loop)
Images are automatically saved in the memory during scanning. The saved images can be useful when a
patient is diagnosed and reviewed.
The saved images can be in Cine or Loop depending on the diagnosis mode.
X Cine: Images that are saved in all modes other than M Mode and Spectral Mode.
X Loop: Images that are saved in M Mode and Spectral Mode.
[Figure 6.1 Reviewing Images]
Starting and Finishing Image Review
During scanning, press the
Freeze
button on the control panel. The scanning is stopped and the system
switches to the image review mode screen.
Press the
Freeze
button again to return to the scan mode.
6-4
Operation Manual
How to Review Images
Use the
Trackball
on the control panel. You can search for saved images in chronological order by rotating
the
Trackball
to the left or right. The number of images saved so far and the number of the image currently
under review are displayed in the user information area.
The Soft Menu is changed when images are reviewed.
Auto Run
Use the
Soft Menu
dial-button
[2]
. You can adjust the automatic playback speed for Cine or Loop by
rotating the dial-button. The allowable range for adjustment is 10-200% in increments/decrements of
10%.
Run / Stop
Use the
Soft Menu
dial-button
[2]
. You can stop or resume playback of an image by pressing the dial-
button.
Cine / Loop
Use the
Soft Menu
dial-button
[3]
. This option appears only when both Cine and Loop are available, as
in M Mode or Spectral Doppler Mode. You can select an image to play in Cine and Loop by pressing the
dial-button. The selection is displayed in the left side of the user information area.
First Pos
Use the
Soft Menu
dial-button
[4]
. You can specify the rst frame in the range in which Cine or Loop
will be saved. After specifying the position of the rst frame by rotating the dial-button or using the
Trackball
, press the dial-button to save it.
Last Pos
Use the
Soft Menu
dial-button
[5]
. You can specify the last frame in the range in which Cine or Loop
will be saved. After specifying the position of the last frame by rotating the dial-button or using the
Trackball
, press the dial-button to save it.
Trim First Pos
Use dial button
[4]
on the soft menu. Set the rst frame of the range for saving when saving Cine or Loop.
Set a location for the rst frame using the
Trackball
and press the dial button to save it.
6-5
Chapter 6 Image Management
Trim Last Pos
Use dial button
[5]
on the soft menu. Set the last frame of the range for saving when saving Cine or Loop.
Set a location for the last frame using the
Trackball
and press the dial button to save it.
Cine Save
Use the
Soft Menu
dial-button
[6]
. This has the same function as when the Clip Store button on the
control is pressed. You can save the specied image area by pressing the dial-button.
The saved image appears in the thumbnail list on the screen, and can be reloaded or replayed in scan
mode or SonoView. You can select Cine or Loop.
Reviewing Images in Multi-Image Mode
Only images in an active area can be reviewed. To review images in another area, change the active area by
using the
Dual
button on the control panel. Alternatively, after pressing
Pointer
on the control panel, place
the cursor in the area to activate and press the
Set
button on the control panel.
6-6
Operation Manual
Annotating Images
Text
You can enter comments or text over an image. This function can be useful when the diagnosis area is
dierentiated or displayed.
Starting Text Input Mode
Press the F3 button in the Alphanumeric Keyboard. The system will switch to text input mode.
NOTE: When Quick Text is enabled, the system switches to text input mode immediately after the
Alphanumeric Keyboard is pressed.
[Figure 6.2 Text Input Mode]
6-7
Chapter 6 Image Management
Typing Text
Use the Alphanumeric Keyboard. Use the
Document
menu or the
Soft Menu
dial-button
[3]
to enter text
automatically. You can move the cursor by using the
Trackball
or the arrow keys on the keyboard.
Font Size / Default size
Use the
Soft Menu
dial-button
[1]
. You can specify the font size, from 10 – 30, by rotating the dial-button.
Press the dial-button to return to the default size (Size 11).
Home Position
Use the
Soft Menu
dial-button
[2]
. Home Position represents the default position of the cursor in the
text mode.
Select Load or Set by rotating the dial-button. When ‘Load’ is selected, press the dial-button to return the
cursor to the default position. When ‘Set is selected, press the dial-button to the default position of the
cursor can be reset.
Text / Input
Use the
Soft Menu
dial-button
[3]
. You can select a predened text and enter it into an image.
Select a text to enter from Text 1 – 10 by rotating the dial-button.
Press the dial-button to enter the selected text in an image.
Current App. / Edit
Use the
Soft Menu
dial-button
[4]
. The current application is
displayed in Current App.
Tips!
Application
Applications displayed on the Soft Menu in Text Mode are used to search text that is to be entered
automatically. This has no relationship to the applications set in the Probe Selection screen.
Press the dial-button to display the Text Edit window. Edit Text 1 to 10 using the keyboard. Press OK to com-
plete editing. Press Cancel to cancel.
6-8
Operation Manual
[Figure 6.3 Text Edit]
Change App.
Use the
Soft Menu
dial-button
[5]
. Change the application by rotating the dial-button. As the application
changes, the details of the
Document
menu change too. The applications change in the following order:
Userset 3D OB Gynecology Cardiac Pediatric Cardiology Urology TCD Smallparts
Abdomen Vascular Pediatric MSK.
Autotext
This function allows you to enter text automatically by using an abbreviation. In this way, you can enter text
easily and quickly. When Autotext is enabled, The Autotext List appears on the screen.
[Figure 6.4 The Autotext List]
NOTE: Select Utility > Setup > Annotate > AutoText Setup > AutoText Edit.
Deleting Text
Press the
Clear
button on the control panel. All the text entered on the screen will be deleted.
Exiting Text Input Mode
Press
Exit
on the control panel.
6-9
Chapter 6 Image Management
BodyMarker
You can place a BodyMarker over an image. This function can be useful when the diagnosis area is dierentiated
or displayed.
Starting BodyMarker Input Mode
Press the
BodyMarker
button on the control panel. The system will switch to BodyMarker input mode and
BodyMarkers will appear in the user information area.
[Figure 6.5 BodyMarker]
6-10
Operation Manual
Entering BodyMarker
1.
BodyMarkers will appear on the Soft Menu.
X The types of BodyMarker shown in the Soft Menu vary with the diagnosis mode.
X A maximum of 6 BodyMarkers are displayed on the screen at any one time. When there are more
BodyMarkers than this, press the
BodyMarker
button repeatedly to move to other pages.
2.
Press the
Soft Menu
button for the BodyMarker that you want. The BodyMarker will appear in the image.
3.
Position the probe cursor on the BodyMarker using the
Trackball
.
4.
Use the
Soft Menu
dial-button
[1]
~
[5]
or the
Menu / Angle
dial-button on the control panel to adjust
the angle of the probe cursor.
X To change the angle degree, use the
Soft Menu
dial-button
[6] Angle Degree. Sele
ct 15 or 45.
5.
Press the
Set
button to nish.
NOTE:
X
Adding BodyMarker to a User Key
You can assign the BodyMarker key to a User Key. Select Setup > Utility > User Key Setup.
X
Changing BodyMarker Size
You can change the size of BodyMarker. At Setup > Annotate > BodyMarker group, select
from Small/Medium/Large.
Changing BodyMarker Position
1.
Press the
Change
button on the control panel.
2.
Move the BodyMarker to a desired position by using the
Trackball
.
3.
Press the
Change
button on the control panel again to conrm the new position.
Deleting BodyMarker
Press the
Clear
button on the control panel.
Exiting BodyMarker Input Mode
Press the
Exit
button on the control panel.
6-11
Chapter 6 Image Management
Indicator
You can place an Indicator over an image. This function can be useful when the diagnosis area is dierentiated
or displayed. You can place 50 Indicators on specic positions.
[Figure 6.6 Indicator Mode]
Starting Indicator Input Mode
Press the
Indicator
[F2]
in the keyboard. The system will switch to the Indicator input mode.
Entering Indicator
1.
Press the
Indicator
[F2]
in the keyboard and an Indicator will appear on the screen.
2.
Move the Indicator to a desired position by using the
Trackball
.
3.
Adjust the Indicator direction by using the
Menu/Angle
dial-button on the control panel.
4.
Press the
Set
button to nish. Press the
Exit
button to cancel.
6-12
Operation Manual
Deleting Indicator
Press the
Clear
button on the control panel and all Indicators displayed on the screen will be deleted.
Press the
Soft Menu
dial-button
[1]
Undo
to delete the last entered indicator only.
Indicator Input Mode Soft Menu
Size
Select the size of the Indicator. Select a size from 1 through 3 by rotating the Soft Menu dial-button
[1].
Undo
Press the Soft Menu dial-button [1] to delete the last indicator entered.
Type
Select the shape of the Indicator. Select a shape from 1 through 4 by rotating the Soft Menu dial-
button [2].
/