11M800 Water EIP / PN
5 Schematic Diagram
pH
ISM Ch1 M1-M4
AI1 - AI4
AI5 - AI8
AI9 - AI10
AI11 - AI12
Transmitter and sensors status
Maintenance
ISM Ch2 M1-M4
Analog input groupDI groupAO groupDO group
Flow Ch1 M1-M2
Flow Ch1 M1-M2
Cond
TOC
Flow1 Ch1
Flow2 Ch2
ISM Chx
(x could be 1 or 2)
O
2
M800
M800
Profinet, EtherNet/IP
Note: Flow sensors are reserved for future use.
6 Function Block Denition
6.1 Block Mapping and Default Values
Channel Slot Number Slot(In PLC) Block Slot Name Bytes Data Type
AI group
ISM Ch1 1 1.2 Analog Input 1 ISM Ch1-M1 Value and status 4 Float32
2 1.3 Analog Input 2 ISM Ch1-M2 Value and status 4 Float32
3 1.4 Analog Input 3 ISM Ch1-M3 Value and status 4 Float32
4 1.5 Analog Input 4 ISM Ch1-M4 Value and status 4 Float32
ISM Ch2 5 1.6 Analog Input 5 ISM Ch2-M1 Value and status 4 Float32
6 1.7 Analog Input 6 ISM Ch2-M2 Value and status 4 Float32
7 1.8 Analog Input 7 ISM Ch2-M3 Value and status 4 Float32
8 1.9 Analog Input 8 ISM Ch2-M4 Value and status 4 Float32
Flow Ch1 9 1.10 Analog Input 9 Flow1-M1 Value 4 Float32
10 1.11 Analog Input 10 Flow1-M2 Value 4 Float32
Flow Ch2 11 1.12 Analog Input 11 Flow2-M1 Value 4 Float32
12 1.13 Analog Input 12 Flow2-M2 Value 4 Float32
DI group
System 13 1.14 Discrete Input 1 HW and setpoint status 4 Dword
14 1.15 Discrete Input 2 Measurement Status 4 Dword
ISM Ch1 15 1.16 Discrete Input 3 ISM Ch1 Status 4 Dword
ISM Ch2 16 1.17 Discrete Input 4 ISM Ch2 Status 4 Dword
AO group
ISM Ch1 17 1.18 Analog Output 1 1st Proc. Pressure(mbar) 4 Float32
ISM Ch2 18 1.19 Analog Output 2 2nd Proc. Pressure(mbar) 4 Float32
DO group 19 1.20 Discrete Output 1 Hold and LED Control 4 Dword
6.2 Discrete Output Function Blocks
Bit Meaning
Ch1
0
System Hold Control
On – 1
Off = 0
1
000 – Invalid
001 – LED on
010 – LED off
Only valid when LED mode is Auto and only for
1–ch PN version
2
3
4
5
6
7
Ch2
8 System Hold, same as Bit0
9
same as Ch1 ODO LED10
11
12
13
14
15