Installation information METTLER TOLEDO Spider and Spider S terminals
2
Contents
1. Important information before you start............................................................... 3
1.1 Introducing the Spider and Spider S terminals .................................................................... 3
1.2 What you should know about these instructions ................................................................ 3
1.3 Safety has priority........................................................................................................... 3
2. Setting up a weighing system ........................................................................... 4
2.1 Determining configuration data ........................................................................................ 4
2.2 Measurement ranges of the Spider and Spider S terminals................................................... 6
3. Attaching the weighing platform to the terminal ................................................. 7
3.1 Opening the terminal ...................................................................................................... 7
3.2 Attaching the weighing cell .............................................................................................. 8
3.3 Closing the terminal...................................................................................................... 10
4. Configuring the terminal ................................................................................. 11
4.1 Recall and operation of the service mode ........................................................................ 11
4.2 Layout of the service mode ............................................................................................ 12
4.3 Selecting basic unit ...................................................................................................... 13
4.4 Requesting ramp .......................................................................................................... 13
4.5 Requesting or changing the serial number....................................................................... 13
4.6 Entering parameters of the weighing cell ......................................................................... 14
4.7 Selecting the weighing range ......................................................................................... 15
4.8 Selecting resolution ...................................................................................................... 15
4.9 Setting geo adjustment value, specify preload and performing calibration ........................... 16
4.10 Correcting linearity error ................................................................................................ 17
4.11 Printing the service mode data ....................................................................................... 19
4.12 Exiting the service mode ............................................................................................... 19
4.13 Deleting the data in the EAROM ...................................................................................... 20
5. Troubleshooting and malfunction correction ..................................................... 21
6. Technical data................................................................................................ 24
6.1 Requirements regarding the weighing cell ....................................................................... 24
6.2 Specifications of the terminal ......................................................................................... 25
6.3 Notes on certified scales ............................................................................................... 25
7. Table of the geo adjustment values ................................................................. 26
8. Notes on digital Spider weighing platforms for ID terminals ............................. 27