Thermo Fisher Scientific Waters LC User guide

Type
User guide
XCALI-97192 Revision B July 2007
Waters LC
LC Devices
Getting Connected
© 2007 Thermo Fisher Scientific Inc. All rights reserved.
Waters® is a registered trademark and ACQUITY UPLCis a trademark of Waters Company.
Xcaliburis a trademark of Thermo Fisher Scientific Inc.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Revision A for LC Devices 2.02, Revision B for LC Devices 2.1.0. The instructions for
connecting the Waters 2795 Separation Module are backward compatible to LC Devices 2.0.2.
For Research Use Only. Not regulated for medical or veterinary diagnostic use by U.S. Federal Drug
Administration or other competent authorities.
Thermo Scientific Waters LC Devices Getting Connected iii
C
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
About This Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Safety and Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vi
Chapter 1 Connecting the Waters 2795 Separations Module . . . . . . . . . . . . . . . . . . . . . . . . . .1
Required Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Connecting the Contact Closure Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Connecting the Serial Communication Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Setting Up the Separations Module for Remote Control . . . . . . . . . . . . . . . . . . . 5
Checking the Firmware Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Setting the Operating Mode to Remote Control . . . . . . . . . . . . . . . . . . . . . . . 6
Configuring the Waters 2795 Separations Module . . . . . . . . . . . . . . . . . . . . . . . 7
Chapter 2 Connecting a Waters Acquity LC System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Connecting the Waters Acquity LC to the Data System Computer . . . . . . . . . . 11
Installing and Configuring a Network Interface Card . . . . . . . . . . . . . . . . . . 11
Connecting the Ethernet Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Connecting the Contact Closure Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Configuring the Waters Acquity Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Troubleshooting Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Starting the Waters Server Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Upgrading the Firmware of the Waters Acquity Devices . . . . . . . . . . . . . . . . 28
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Contents
Thermo Scientific Waters LC Devices Getting Connected v
P
Preface
About This Guide
This guide describes how to connect a Waters LC system to a Thermo Scientific mass
spectrometer and the data system computer.
Related Documentation
The Xcalibur LC Devices instrument control software provides Help.
Safety and Special Notices
Make sure you follow the precautionary statements presented in this guide. The safety and
other special notices appear in boxes.
Safety and special notices include the following:
CAUTION Highlights hazards to humans, property, or the environment. Each
CAUTION notice is accompanied by an appropriate CAUTION symbol.
CAUTION Highlights electric shock related hazards to human beings. Each electric shock
notice is accompanied by the international High Voltage symbol.
IMPORTANT Highlights information necessary to prevent damage to software, loss of
data, or invalid test results; or might contain information that is critical for optimal
performance of the system.
Note Highlights information of general interest.
Tip Highlights helpful information that can make a task easier.
Preface
vi Waters LC Devices Getting Connected Thermo Scientific
Contacting Us
There are several ways to contact Thermo Fisher Scientific.
YTo contact Technical Support
Find software updates and utilities to download at www.mssupport.thermo.com.
YTo contact Customer Service for ordering information
YTo suggest changes to documentation or to Help
Fill out a reader survey online at www.thermo.com/lcms-techpubs.
Send an e-mail message to the Technical Publications Editor at
techpubs.finnigan-lcms@thermofisher.com.
Phone 800-685-9535
Fax 561-688-8736
E-mail TechSupport.C+MS@thermofisher.com
Knowledge base www.thermokb.com
Phone 800-532-4752
Fax 561-688-8731
Web site www.thermo.com/finnigan
Thermo Scientific Waters LC Devices Getting Connected 1
1
Connecting the Waters 2795 Separations Module
Required Parts
To connect the Waters 2795 Separations Module to a Thermo Scientific mass spectrometer,
order the Waters Interface kit from Thermo Fisher Scientific. This kit contains the cables
listed in Tab le 1. The serial communication cable connects the Waters 2795 Separations
Module to the data system computer. The contact closure (trigger) cable connects the Inject
Out pins of the separations module to the Start In pins on the I/O panel of the mass
spectrometer.
Connecting the Contact Closure Cable
If you want the separations module to trigger the start of data acquisition from a Thermo
Scientific mass spectrometer during a run, connect the Inject Out pins of the Waters 2795
Separations Module to the Start In pins on the I/O panel of the mass spectrometer.
Contents
Required Parts
Connecting the Contact Closure Cable
Connecting the Serial Communication Cable
Setting Up the Separations Module for Remote Control
Configuring the Waters 2795 Separations Module
Table 1. Xcalibur kit used with the Waters 2795 Separations Module
Part Number Description of Kit
OPTON-21710 Waters Interface Kit
Serial communication cable with two DB9 connectors
2-wire contact closure cable, with an 8-pin connector and
a 2-pin connector
1 Connecting the Waters 2795 Separations Module
Connecting the Contact Closure Cable
2Waters LC Devices Getting Connected Thermo Scientific
YTo hardwire the Waters 2795 Separations Module to the MS detector
1. Connect the 8-pin connector of the 2-wire, contact closure cable (supplied in the Waters
Interface kit) to the contact closure terminal (labeled B) on the back panel of the Waters
2795 Separations Module.
2. Connect the 2-pin connector of the 2-wire, contact closure cable to the User I/O of the
MS detector. Follow the wiring scheme listed in Tabl e 2 and shown in Figure 1.
Table 2. Wiring the MS detector and the Waters 2795 Separations Module for contact closure
Waters 2795
Contact closure terminal B MS detector
Pin 1 Start In +
Pin 2 Start In -
1 Connecting the Waters 2795 Separations Module
Connecting the Contact Closure Cable
Thermo Scientific Waters LC Devices Getting Connected 3
Figure 1. Contact closure connection between the Waters 2795 Separations Module and the MS detector
Surveyor MSQ
(back panel)
+-
Not drawn to scale
LCQ Fleet, LXQ, or LTQ
MS Detector
(right-side panel)
START IN
START IN
TSQ Series MS Detector
(back panel)
(right-side
panel)
Start In +
Start In -
START IN
(5V Max)
(5V Max)
+
LCQ Series
(back panel)
(left-side
panel)
Peripheral
Control
BA
RS232
Back panel of Waters 2795
BA
White Wire
Brown Wire
White wire
Brown wire
(right-side
panel)
(right-side panel)
(left-side
panel)
(back panel)
(back panel)
(back panel)
Back panel of Waters 2795
Not drawn to scale
1 Connecting the Waters 2795 Separations Module
Connecting the Serial Communication Cable
4Waters LC Devices Getting Connected Thermo Scientific
Connecting the Serial Communication Cable
As Figure 2 shows, the data system computer communicates with the Waters 2795
Separations Module through an RS232 serial communication cable. This cable is provided in
the Waters Interface kit.
YTo connect the serial communication cable
1. Connect one end of the serial communication cable to the RS232 port on the back panel
of the Waters 2795 Separations Module. See Figure 2.
2. Connect the other end of the serial communication cable to an RS232 port on the data
system computer.
Figure 2. Serial communication connection
RS232
BA
Computer
Waters 2795 Separations Module
(back panel)
Serial communication cable
(with two DB9 connectors)
1 Connecting the Waters 2795 Separations Module
Setting Up the Separations Module for Remote Control
Thermo Scientific Waters LC Devices Getting Connected 5
Setting Up the Separations Module for Remote Control
The Xcalibur based instrument control software for the Waters 2795 Separations Module is
compatible with firmware version 2.04. To control the Waters 2795 Separations Module, set
the separations module to operate by remote control through an RS232 connection to the
computer where Xcalibur resides.
To set up the separations module for remote control, follow these procedures:
Checking the Firmware Version
Setting the Operating Mode to Remote Control
Checking the Firmware Version
YTo check the firmware version for the Waters Separations Module
Turn on the instrument and watch the front panel LCD display as the instrument cycles
through the system startup.
At the end of the system initialization process the Main LCD display appears. See Figure 3.
The firmware version is listed in the lower-right corner of the display.
Figure 3. Main LCD display
Main
<Default> Idle
Separations Module
Press Menu/
Status for monitoring and direct control
User:
Revision 2.04
Develop
Methods Run
Samples Diag Config Log
01/02/2007 09:00:00 pm
© 2001 Waters Corporation
Tester
Waters 2795
**
1 Connecting the Waters 2795 Separations Module
Setting Up the Separations Module for Remote Control
6Waters LC Devices Getting Connected Thermo Scientific
Setting the Operating Mode to Remote Control
YTo set up the Waters 2795 Separations Module for remote control
1. From the Main page on the LCD display of the Waters 2795 Separations Module, press
the Config screen key. See Figure 4.
Figure 4. Front panel LCD display and keypad for the Waters 2795 Separations Module
The Configuration page appears. See Figure 5.
Figure 5. Configuration page
2. Press the < > arrow key to highlight the System box, and then press the Enter key on the
front panel keypad.
?
123
456
789
0.
Exit Menu/
Status
Stop
Flow
Enter
Clear
Display
Contrast
Arrow Keys
Config Key
Diag Config Log
Main
<Default> Idle
Separations Module
Press Menu/
Status for monitoring and direct control
User:
Revision 2.04
Develop
Methods Run
Samples Diag Config Log
01/02/2007 09:00:00 pm
© 2001 Waters Corporation
Teste r
Waters 2795
**
Arrow keys
Config key
Configuration
<Default> Idle
Tester
Pressure Units:
500
Syringe Size: Loop Size:
System
Remote Control via RS-232
Printer Date Format: Alarm Sound:
User
Names
Detec-
tors Options
Set
Clock
Events
In
More
psi
MDY
Baud Rate:
1200
None Continuous
50
µL µL
IEEE 488 Address:
10
o
1 Connecting the Waters 2795 Separations Module
Configuring the Waters 2795 Separations Module
Thermo Scientific Waters LC Devices Getting Connected 7
The list of system options appears as shown in Figure 6.
Figure 6. Configuration page with the list of system options
3. Select Controlled via RS232 by pressing the < > and < > arrow keys, and then press
the Enter key on the front panel keypad to save the change.
4. Press the Exit key on the front panel keypad to return to the Main page.
Configuring the Waters 2795 Separations Module
To establish communication between the data system and the separations module, specify the
communication port to which you have connected the separations module. To make sure that
the separations module draws sample from the correct location, specify the plates you are
using.
YTo configure the Waters 2795 Separations Module
1. From the Windows XP taskbar, choose Start > All Programs > Xcalibur > Instrument
Configuration.
The Instrument Configuration dialog box appears. See Figure 7.
Configuration
<Default> Idle
Tester
Pressure Units:
500
Syringe Size: Loop Size:
System
Remote Control via RS-232
Printer Date Format: Alarm Sound:
User
Names
Detec-
tors Options
Set
Clock
Events
In
More
psi
MDY
Baud Rate:
1200
None Continuous
50
µL µL
IEEE 488 Address:
10
o
No Interaction
System Controller
Controlled by Millennium 32
Controlled by MassLynx
Controlled by Millennium 2.XX
Operate gradient by event in
Controlled via RS232
1 Connecting the Waters 2795 Separations Module
Configuring the Waters 2795 Separations Module
8Waters LC Devices Getting Connected Thermo Scientific
Figure 7. Instrument Configuration dialog box
2. Under Available Devices, double-click .
A copy of the Waters 2795 button appears under Configured Devices.
3. Under Configured Devices, double-click .
The Waters Alliance 2795 Configuration dialog box appears. See Figure 8.
1 Connecting the Waters 2795 Separations Module
Configuring the Waters 2795 Separations Module
Thermo Scientific Waters LC Devices Getting Connected 9
Figure 8. Alliance 2795 Configuration dialog box
4. In the Port list, confirm that the selected COM port is the RS232 port on the data system
computer to which you physically connected the separations module.
For information on connecting the serial communication cable, see “Connecting the
Serial Communication Cableon page 3.
5. In the Carrier Plate boxes, specify the tray type that you are using and the sequential
reference counting order for each plate.
6. Click Autoconfigure, and then click OK to save the settings and close the dialog box.
The separations module is ready for remote control from the Xcalibur data system.
Port list
Thermo Scientific Waters LC Devices Getting Connected 11
2
Connecting a Waters Acquity LC System
Connecting the Waters Acquity LC to the Data System Computer
The Waters Acquity LC stack communicates with the data system computer through an
Ethernet connection. To make this connection, follow these procedures:
1. Installing and Configuring a Network Interface Card
2. Connecting the Ethernet Cables
Installing and Configuring a Network Interface Card
The Waters Acquity LC stack requires a dedicated network interface card.
YTo install an additional network interface card in the data system computer
Follow the instructions in the manuals provided by the manufacturer of the computer.
YTo set the IP address for the dedicated network interface card
1. To open the Internet Protocol (TCP/IP) Properties dialog box, do the following:
a. From the computer desktop, choose Start > Control Panel > Network Connections
> Local Area Connection.
The General page of the Local Area Connection Status dialog box appears.
Contents
Connecting the Waters Acquity LC to the Data System Computer
Connecting the Contact Closure Cable
Configuring the Waters Acquity Devices
Troubleshooting Communications
CAUTION To prevent personal injury caused by an electric shock, always unplug your
computer from the electrical outlet before removing the cover.
2 Connecting a Waters Acquity LC System
Connecting the Waters Acquity LC to the Data System Computer
12 Waters LC Devices Getting Connected Thermo Scientific
b. Click Properties.
The Local Area Connection Properties dialog box appears.
c. In the This connection uses the following items list, select Internet Protocol TCP/IP
and click Properties.
The Internet Protocol (TCP/IP) Properties dialog box appears. See Figure 9.
Figure 9. Internet Protocol (TCP/IP) Properties dialog box with the IP address for the
dedicated Ethernet card
2. Select the Use the following IP address option.
3. In the IP address box, type 192.168.0.1.
4. In the Subnet mask box, type 255.255.255.0.
5. Click OK to accept the new IP address and subnet mask, and exit the Internet Protocol
(TCP/IP) Properties dialog box.
6. Click OK to exit the Local Area Connection dialog box.
2 Connecting a Waters Acquity LC System
Connecting the Waters Acquity LC to the Data System Computer
Thermo Scientific Waters LC Devices Getting Connected 13
Connecting the Ethernet Cables
The Waters Acquity LC system comes with an Ethernet switch. Connect each device of the
Waters Acquity LC stack to the Ethernet switch, and connect the Ethernet switch to the data
system computer with category 5, shielded Ethernet cables.
YTo connect the Waters Acquity LC to the data system computer
1. Connect one end of a category 5, shielded Ethernet cable to the Ethernet switch supplied
with the Waters Acquity LC system.
2. Connect the other end of the category 5, shielded Ethernet cable to the Ethernet port of
the network interface card dedicated to the Waters Acquity LC system. See Figure 10.
Figure 10. Ethernet connections
CAUTION To ensure compliance with safety and EMC regulations, use the Ethernet
cables supplied in the Ethernet Communication kit or category 5, shielded Ethernet
cables no longer than 3 m (10 ft) in length.
Ethernet
Ethernet
Ethernet
Ethernet
III
Ethernet switch
Dedicated
network
interface card
2 Connecting a Waters Acquity LC System
Connecting the Contact Closure Cable
14 Waters LC Devices Getting Connected Thermo Scientific
Connecting the Contact Closure Cable
The LC system sends a ready signal after the Waters Acquity Sample Manager (autosampler)
makes an injection. If you want the Thermo Scientific mass spectrometer to start data
acquisition after the autosampler starts an injection, you must hardwire the output terminals
of the autosampler to the start terminals on the I/O panel of the mass spectrometer.
YTo connect the contact closure cable
1. Connect one end of the two-wire, contact closure cable in the Waters accessory kit to
pin 1 and pin 2 of the contact closure terminal (labeled I) on the back of the Waters
Acquity Sample Manager.
2. Connect the other end of the two-wire contact closure cable to the User I/O of the mass
spectrometer. Follow the wiring scheme listed in Table 3 and shown in Figure 11.
Table 3. Connecting the contact closure cable
Waters
Contact closure terminal I MS detector
Pin 1 Start In +
Pin 2 Start In -
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36

Thermo Fisher Scientific Waters LC User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI