cobas c111 system
Operator’s Manual
Version 3.0
cobas c111
Roche Diagnostics
2 Operator’s Manual · Version 3.0
Document information
Revision history
Edition notice The cobas c111 instrument is a continuous random-access analyzer intended for the
in vitro determination of clinical chemistry and electrolyte parameters in serum,
plasma, urine or whole blood (HbA1c). It is optimized for small throughput
workloads of approximately samples per day, utilizing photometric analysis and an
optional unit for ion selective electrodes (ISE).
This manual is for users of the cobas c 111 instrument.
Every effort has been made to ensure that all the information contained in this
manual is correct at the time of printing. However, Roche Diagnostics GmbH reserves
the right to make any changes necessary without notice as part of ongoing product
development.
Any customer modification to the instrument will render the warranty or service
agreement null and void.
Software updates are done by service representatives.
Intended use The cobas c111 instrument is a continuous random-access analyzer intended for the
in vitro determination of clinical chemistry and electrolyte parameters in serum,
plasma, urine or whole blood (HbA1c).
It is important that the operators read this manual thoroughly before using the
system.
Copyrights © 2009, Roche Diagnostics GmbH. All rights reserved.
Tradem ark s The following trademarks are acknowledged:
COBAS, COBAS C and LIFE NEEDS ANSWERS are trademarks of Roche.
All other trademarks are the property of their respective owners.
Manual version Software version Revision date Changes
1.0 July 2006 First publication.
2.0 2.0 December 2007 Full mode added.
Improved calibration concept.
Additional maintenance and
troubleshooting information.
Additions, improvements, and
corrections.
3.0 3.0 June 2009 Inventory, processing sequence,
and ratio functions added.
Improvements and corrections.
Layout upgraded.
Roche Diagnostics
Operator’s Manual · Version 3.0 3
cobas c111
Instrument approvals The cobas c111 instrument meets the protection requirements laid down in IVD
Directive 98/79/EC and the European Standard EN 591. Furthermore, our
instruments are manufactured and tested according to the following international
standards:
o EN/IEC 61010-1 2
nd
Edition
o EN/IEC 61010-2-101 1
st
Edition
The Operators manual meets the European Standard EN 591.
Regulatory compliance is demonstrated by the following marks:
Contact addresses
Manufacturer
Distributor
Complies with European Union (EU) Directive 98/79/EC.
CUS
®
Issued by Underwriters Laboratories, Inc. (UL) for Canada and the
US.
Roche Diagnostics Ltd.
Forrenstrasse
CH-6343 Rotkreuz
Switzerland
Roche Diagnostics GmbH
Sandhofer Strasse 116
D-68305 Mannheim
Germany
cobas c111
Roche Diagnostics
4 Operator’s Manual · Version 3.0
Roche Diagnostics
Operator’s Manual · Version 3.0 5
cobas c111
Table of contents
Document information 2
Contact addresses 3
Table of contents 5
Preface 7
How to use this manual 7
Online Help system 7
Conventions used in this manual 8
System Description Part A
1Safety
Safety classification A–5
Safety information A–5
Data security A–10
License notices A–11
Legal liability A–11
Disposal recommendation A–12
Safety labels A–13
2 Introduction to the instrument
Overview A–17
User interface A20
Wizards A21
Daily operation A–22
Maintenance A–34
System status A–35
3 Hardware
Covers and panels A–39
LEDs A–41
Main components A–43
Hardware overview A–44
Technical specifications A–66
5Software
Introduction A–71
Screen layout A–72
Display items A–73
Workflows and wizards A–74
Working with the user interface A–75
Key screens A–84
Color interpretation for LEDs A–122
Buttons A–124
Operation Part B
6 Daily operation
Introduction B–5
Starting the shift B–10
Preparing the system B–12
Analyzing samples B–34
Validating sample results B–51
Performing calibrations B–59
Performing QC B–68
Finishing the shift B–77
Logging off B–88
Switching off the system B–88
Using the barcode scanner B–89
6 Special operations
Deleting sample orders B–93
Deleting sample results B–94
Calibration B–95
Deleting QC results B–96
Lot handling B–98
Exporting data B–104
Importing data B–111
Preparing a new disk B–116
Assigning tests to test tabs B–119
Deleting bottle sets from the Inventory list B–120
Refilling printer paper B–122
Removing condensation water
from the reagent cooler B–124
Replacing the probe B–125
Connecting and disconnecting the
external fluid containers B–127
Adjusting the touchscreen B–130
Cleaning the touchscreen B–131
7 Configuration
Introduction B–135
Applications B–137
Configuration B–157
Maintenance Part C
8 General maintenance
Overview C–5
Maintenance actions C–8
Roche Diagnostics
6 Operator’s Manual · Version 3.0
cobas c111
Troubleshooting Part D
9 Messages and alarms
About messages D–5
Message screen D–5
Acoustic signals D–6
Alarm monitor D–6
List of alarm messages D–10
10 Result flags
About flags D–25
Safety D–27
List of flags D–28
11 Troubleshooting
Introduction D–41
Dealing with exceptional situations D–42
Reacting to messages D–44
Detailed procedures D–46
ISE Part E
12 ISE description
Overview E–5
Hardware E–9
Basic operation E–12
Technical specifications E–13
13 ISE operation
Daily operation E–17
Replacing ISE fluid bottles E–30
Replacing electrodes E–32
Cleaning the ISE tower off the instrument E–36
14 ISE maintenance
Introduction E–41
ISE maintenance actions E–42
15 ISE troubleshooting
Introduction E–69
Safety E–70
List of ISE flags E–71
Reacting to error messages E–80
Glossary and Index Part F
Glossary F–3
Index F–11
Revisions Part G
18 Revisions
Roche Diagnostics
Operator’s Manual · Version 3.0 7
cobas c111
Preface
The cobas c111 instrument is a continuous random-access analyzer intended for the
in vitro determination of clinical chemistry and electrolyte parameters in serum,
plasma, urine or whole blood (HbA1c). It is optimized for small throughput
workloads of approximately samples per day, utilizing photometric analysis and an
optional unit for ion selective electrodes (ISE).
This manual describes the cobas c111 features and general operational concepts, and
it provides operating, maintenance, and emergency procedures.
How to use this manual
To help you find information quickly, there is a table of contents at the beginning of
the manual and each chapter. In addition, a complete index can be found at the end.
Online Help system
The cobas c111 instrument has a context-sensitive online Help feature to aid in its
operating. “Context-sensitive” means that wherever you are located within the
cobas c111 software, choosing Help ( ) displays Help text relating to that area of
the software. The online Help offers a quick and convenient way of finding
information, such as explanations of screens and dialog boxes and on how to perform
particular tasks.
o Keep this Operator’s Manual in a safe place to ensure that it is not damaged and
remains available for use.
o This Operator’s Manual should be easily accessible at all times.
cobas c111
Roche Diagnostics
8 Operator’s Manual · Version 3.0
Conventions used in this manual
Visual cues are used to help locate and interpret information in this manual quickly.
This section explains the formatting conventions used in this manual.
Symbols The following symbols are used:
Buttons When used for identification purposes, a generic form of the buttons is used, without
color or navigation indicators.
Screenshots The screen representations shown in this publication are for illustrative purposes
only. The screens do not necessarily show valid data.
Symbol Used for
a Start of procedure
o List item
e
Cross-reference
h Call-up (software navigation path)
Color of display item on the screen
Tip
Safety alert
Electrical and electronic equipment marked with this symbol are
covered by the European directive WEEE.
The symbol denotes that the equipment must not be disposed of in
the municipal waste system.
Roche Diagnostics
Operator’s Manual · Version 3.0 9
cobas c111
Abbreviations The following abbreviations are used:
Abbreviation Definition
C
Cfas Calibrator for automated systems
D
DIL Diluent
DM Data management
DRAM Dynamic random access memory
E
e.g. Exempli gratia – for example
EMC Electromagnetic compatibility
EN European standard
I
i.e. Id est – that is to say
IEC International Electrical Commission
ISE Ion selective electrode
L
LED Light-emitting diode
LIS Laboratory information system
LLD Liquid level detection
N
n/a Not applicable
Q
QC Quality control
R
REF Reference solution for ISE unit
ROM Read only memory
S
SD Standard deviation
SRAM Static random access memory
cobas c111
Roche Diagnostics
10 Operator’s Manual · Version 3.0
Units
Abbreviation Description
°C degree centigrade
µL microliter
µm micrometer
Aampère
cm centimeter
hhour
Hz hertz
LB pound (weight)
in inch
kg kilogram
kVA kilo volt-ampere
Lliter
mmeter
MB megabytes
min minute
mL milliliter
mm millimeter
nm nanometer
ssecond
Vvolt
VA v o l t - a m p è r e
V AC volt alternating current
V DC volt direct current
Wwatt
1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
2 Introduction to the instrument . . . . . . . . . . . . . . . . . . . . . . . . . A-15
3 Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-37
5 Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-69
System Description
A
Roche Diagnostics
Operator’s Manual · Version 3.0 A-3
cobas c111 1Safety
Table of c ontents
Safety
Protecting yourself and the environment
In this chapter, you will find information on the safe operation of the cobas c111
instrument.
Safety classification ......................................................................................................A-5
Safety information ......................................................................................................A-5
Transport ...............................................................................................................A-5
Electrical safety ......................................................................................................A-5
Optical safety .........................................................................................................A-6
Mechanical safety ..................................................................................................A-6
Instrument covers ..................................................................................................A-6
Operation and maintenance .................................................................................A-6
Biohazardous materials .........................................................................................A-6
Waste ......................................................................................................................A-7
Reagents and other working solutions .................................................................A-7
Installation .............................................................................................................A-7
Environmental conditions ....................................................................................A-7
Power interruption ................................................................................................A-8
Electromagnetic devices ........................................................................................A-8
Approved parts ......................................................................................................A-8
Third-party software .............................................................................................A-8
Operator qualification ...........................................................................................A-9
Operation over an extended period of time ......................................................... A-9
Cross contamination of sample ............................................................................A-9
Insoluble contaminants in sample ........................................................................A-9
Spillage ...................................................................................................................A-9
Data security ..............................................................................................................A-10
License notices ...........................................................................................................A-11
Legal liability ..............................................................................................................A-11
In this chapter
Chapter
1
Roche Diagnostics
A-4 Operator’s Manual · Version 3.0
1Safety cobas c111
Tab le o f cont e n t s
Disposal recommendation ........................................................................................A-12
Disposal label .......................................................................................................A-12
Disposal of external components .......................................................................A-12
Disposal of the instrument .................................................................................A-12
Constraint ............................................................................................................A-12
Safety labels ................................................................................................................A-13
Roche Diagnostics
Operator’s Manual · Version 3.0 A-5
cobas c111 1Safety
Safety classification
Safety classification
Before you attempt to use the cobas c111 instrument, you must be fully familiar with
the following symbols and their meanings:
Safety information
Before operating the cobas c111 instrument, it is essential that you both read and
understand the safety information listed below.
Read all Roche safety notices carefully and make sure you understand them.
Transport
Electrical safety
WARNING
Warning
Indicates a hazardous situation which, if not avoided, could result in death or serious
injury.
CAUTION
Caution
Indicates a hazardous situation which, if not avoided, may result in minor or moderate
injury.
NOTICE
Notice
Indicates a hazardous situation which, if not avoided, may result in property damage.
WARNING
Injury from heavy loads
You may injure your hands, fingers, or back when putting the analyzer in place. Carry the
analyzer according to the transport instructions.
WARNING
Electrical shock by electronic equipment
Do not attempt to work in any electronic compartment. Installation, service, and repair
must only be performed by authorized and qualified personnel.
Electrical safety
Connect the analyzer to grounded power outlets only (IEC protection class 1). All periph-
eral devices that are connected to the cobasc111 instrument must comply with safety
standard IEC 60950 for information technology equipment, or with IEC 61010-1, UL 61010-
1 for laboratory use instruments.
Roche Diagnostics
A-6 Operator’s Manual · Version 3.0
1Safety cobas c111
Safety information
Optical safety
Mechanical safety
Instrument covers
Operation and maintenance
Biohazardous materials
WARNING
Loss of sight
The intense light of the LEDs may severely damage you eyes. Do not stare into the LEDs.
Scanning equipment using LED technology is covered by the international standard IEC
60825-1 LED Safety: Class 1.
WARNING
Personal injury or damage to the analyzer due to contact with instrument
mechanism
Do not touch moving parts during instrument operation.
WARNING
Personal injury or damage to the analyzer due to contact with instrument
mechanism
Keep all covers closed, operate them as instructed on the screen.
WARNING
Personal injury or damage to the analyzer due to contact with instrument
mechanism
Do not touch any parts of the instrument other than those specified. During operation and
maintenance of the instrument, proceed according to the instructions.
WARNING
Infection by biohazardous materials
Contact with samples containing material of human origin may result in infection. All
materials and mechanical components associated with samples of human origin are
potentially biohazardous.
o Be sure to wear protective equipment. Take extra care when working with protective
gloves; these can easily be pierced or cut, which can lead to infection.
o If any biohazardous material is spilled, wipe it up immediately and apply disinfectant.
o If waste solution contacts your skin, wash it off immediately with water and apply a dis-
infectant. Consult a physician.
Roche Diagnostics
Operator’s Manual · Version 3.0 A-7
cobas c111 1Safety
Safety information
Waste
Reagents and other working solutions
Installation
Environmental conditions
e
For details on the required environmental conditions, see Environmental conditions on
page A-66.
WARNING
Infection by waste solution
Contact with waste solution may result in infection. All materials and mechanical compo-
nents associated with the waste systems are potentially biohazardous.
o Be sure to wear protective equipment. Take extra care when working with protective
gloves; these can easily be pierced or cut, which can lead to infection.
o If any biohazardous material is spilled, wipe it up immediately and apply disinfectant.
o If waste solution contacts your skin, wash it off immediately with water and apply a dis-
infectant. Consult a physician.
WARNING
Injury through reagents and other working solutions
Direct contact with reagents, cleaning solutions, or other working solutions may cause
personal injury.
When handling reagents, exercise the precautions required for handling laboratory
reagents, observe the cautions given in the package insert, and observe the information
given in the Safety Data Sheets available for Roche Diagnostics reagents and cleaning
solutions.
Skin inflammation caused by reagents
Direct contact with reagents may cause skin irritation, inflammation, or burns.
When handling reagents, be sure to wear protective equipment and observe the cautions
given in the package insert.
WARNING
Incorrect results or damage to the analyzer due to wrong installation
Follow the specified installation instructions carefully.
WARNING
Incorrect results or damage to the analyzer due to heat and humidity
Use the instrument indoor only.
Roche Diagnostics
A-8 Operator’s Manual · Version 3.0
1Safety cobas c111
Safety information
Power interruption
Electromagnetic devices
Approved parts
Third-party software
NOTICE
Data loss or damage to the system due to voltage drop
By a power failure or momentary voltage drop the operation unit or software of this system
may get damaged or data loss may occur. Use only uninterruptible power supply.
WARNING
Malfunction of instrument and incorrect results due to interfering electromagnetic
fields
Devices that emit electromagnetic waves may cause the instrument to malfunction. Do not
operate the following devices in the same room where the system is installed:
o Mobile phone
o Transceiver
o Cordless phone
o Other electrical devices that generate electromagnetic waves
NOTICE
Instructions for in vitro diagnostic (IVD) equipment for professional use
The IVD equipment complies with the emission and immunity requirements described in
the particular requirements for IVD medical equipment of the EN/IEC 61326-2-6 standard.
The electromagnetic environment should be evaluated prior to operation of the device.
Class B FCC rule compliance
This equipment has been tested and found to comply with the limits for Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reason-
able protection against harmful interferences when the equipment is operated in a resi-
dential area. However, this equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the present manual, may cause
harmful interference to radio communications.
The electromagnetic environment should be evaluated prior to operation of the device.
WARNING
Malfunction of instrument and incorrect results due to nonapproved parts
Use of nonapproved parts or devices may result in malfunction of the instrument and may
render the warranty null and void. Only use parts and devices approved by Roche Diag-
nostics.
WARNING
Malfunction of instrument and incorrect results due to third-party software
Installation of any third-party software that is not approved by Roche Diagnostics may
result in incorrect behavior of the system. Do not install any nonapproved software.
Roche Diagnostics
Operator’s Manual · Version 3.0 A-9
cobas c111 1Safety
Safety information
Operator qualification
Operation over an extended period of time
Cross contamination of sample
Insoluble contaminants in sample
Spillage
WARNING
Incorrect results or damage to the analyzer due to wrong operation
Operators are required to have a profound knowledge of relevant guidelines and norms as
well as the information and procedures contained in the Operator’s Manual.
o Do not carry out operation and maintenance unless you have been trained by Roche
Diagnostics.
o Carefully follow the procedures specified in the Operator’s Manual for the operation
and maintenance of the system.
o Leave maintenance that is not described in the Operator’s Manual to trained service
representatives.
o Follow standard laboratory practices, especially when working with biohazard material.
CAUTION
Fatigue due to long hours of operation
Looking at the monitor screen over an extended period of time may lead to fatigue of your
eyes or body. Take a rest for 10 to 15 minutes every hour to relax. Avoid spending more
than 6 hours per day looking at the monitor screen.
WARNING
Incorrect results due to carryover
Traces of analytes or reagents may be carried over one test to the next. Take adequate
measures (e.g. sample aliquoting) to safeguard additional testing and to avoid potentially
false results.
WARNING
Incorrect results and interruption of analysis due to contaminated samples
Insoluble contaminants in samples may cause clogging or pipetting volume shortage and
deterioration in measurement accuracy. When loading samples on the instrument, make
sure that samples contain no insoluble contaminants such as fibrin or dust.
NOTICE
Malfunction due to spilled liquid
Any liquid spilled on the instrument may result in malfunction of the instrument. If liquid
does spill on the instrument, wipe it up immediately and apply disinfectant.
Roche Diagnostics
A-10 Operator’s Manual · Version 3.0
1Safety cobas c111
Data security
Data security
e
For further information, contact your Roche service representative.
CAUTION
Unauthorized access and data loss due to malicious software and hacker attacks
Portable storage media can be infected with and transmit computer malware, which may
be used to gain unauthorized access to data or cause unwanted changes to software.
The cobas c111 is not protected against malicious software and hacker attacks.
The customers are responsible for IT security of their IT infrastructure and for protecting it
against malicious software and hacker attacks. Failure to do so may result in data loss or
render the cobas c111 unusable.
Roche recommends the following precautions:
o Allow connection to authorized external devices only.
o Ensure that all external devices are protected by appropriate security software.
o Ensure that access to all external devices is protected by appropriate security equip-
ment.
o Do not copy or install any software on the cobasc111 unless it is part of the system
software or you are instructed to do so by a Roche service representative.
o If additional software is required, contact your Roche service representative to ensure
validation of the software in question.
o Do not use the USB ports to connect other storage devices unless you are instructed to
do so by official user documentation or a Roche service representative.
o Exercise utmost care when using external storage devices such as USB flash drives,
CDs, or DVDs. Do not use them on public or home computers while connecting to the
cobasc111.
o Keep all external storage devices in a secure place and ensure that they can be
accessed by authorized persons only.
/