Precision Power Analyzer
Communication Interface
WT5000
i
IM WT5000-17EN
Thank you for purchasing the WT5000 Precision Power Analyzer.
This Communication Interface User’s Manual explains the following interface features and commands.
Ethernet interface
USB interface
GP-IB interface
To ensure correct use, please read this manual thoroughly before operation.
Keep this manual in a safe place for quick reference in the event a question arises.
List of Manuals
The following manuals, including this one, are provided as manuals for this instrument. Please read all
the manuals.
Manual Title Manual No. Description
WT5000
Precision Power Analyzer
Features Guide
IM WT5000-01EN The supplied CD contains the PDF file of this
manual. This manual explains all the instrument’s
features other than the communication interface
features.
WT5000
Precision Power Analyzer
User’s Manual
IM WT5000-02EN The supplied CD contains the PDF file of this
manual. The manual explains how to operate this
instrument.
WT5000
Precision Power Analyzer
Getting Started Guide
IM WT5000-03EN The manual explains the handling precautions and
basic operations of this instrument.
WT5000
Precision Power Analyzer
Communication Interface User’s Manual
IM WT5000-17EN This manual. The supplied CD contains the PDF
file of this manual. This manual explains the
instrument’s communication interface features and
instructions on how to use them.
WT5000
Precision Power Analyzer
IM WT5000-92Z1 Document for China
The “EN” and “Z1” in the manual numbers are the language codes.
Contact information of Yokogawa offices worldwide is provided on the following sheet.
Document No. Description
PIM 113-01Z2
List of worldwide contacts
Notes
The contents of this manual are subject to change without prior notice as a result of continuing
improvements to the instrument’s performance and functionality. The figures given in this manual
may differ from those that actually appear on your screen.
Every effort has been made in the preparation of this manual to ensure the accuracy of its
contents. However, should you have any questions or find any errors, please contact your nearest
YOKOGAWA dealer.
Copying or reproducing all or any part of the contents of this manual without the permission of
YOKOGAWA is strictly prohibited.
The TCP/IP software of this product and the documents concerning it have been developed/created
by YOKOGAWA based on the BSD Networking Software, Release 1 that has been licensed from
the Regents of the University of California.
1st Edition: September 2018 (YMI)
All Rights Reserved, Copyright © 2018 Yokogawa Test & Measurement Corporation
ii IM WT5000-17EN
Trademarks
Microsoft, Internet Explorer, MS-DOS, Windows, Windows 7, Windows 8.1, and Windows 10 are
either registered trademarks or trademarks of Microsoft Corporation in the United States and/or
other countries.
Adobe and Acrobat are either registered trademarks or trademarks of Adobe Systems Incorporated.
In this manual, the ® and TM symbols do not accompany their respective registered trademark or
trademark names.
Other company and product names are registered trademarks or trademarks of their respective
holders.
About the Ethernet Interface and USB Interface
To use the Ethernet communication features, your PC must have the following:
Communication library (TMCTL)
To use the USB communication features, your PC must have the following:
Communication library (TMCTL)
USB device driver for connecting this instrument to the PC
You can download the library and driver from the following web page.
https://tmi.yokogawa.com/library/
Revisions
1st Edition: September 2018
iii
IM WT5000-17EN
How to Use This Manual
Structure of the Manual
This manual contains seven chapters and an appendix.
Chapter 1 Ethernet Interface
Describes the features and specifications of the Ethernet interface.
Chapter 2 USB Interface
Describes the features and specifications of the USB interface.
Chapter 3 GP-IB Interface
Describes the GP-IB interface features and specifications.
Chapter 4 Programming Overview
Describes command syntax and other programming information.
Chapter 5 Commands
Describes every command individually.
Chapter 6 Status Reports
Describes the status byte, various registers, and queues.
Chapter 7 Commands Compatible with Legacy Models
Describes the WT3000E, WT3000, WT1800E, WT1800, or WT1600 Compatible
Command type.
Appendix
Describes error messages and provides other information.
Index
iv IM WT5000-17EN
Symbols and Notation Used in This Manual
Notes and Cautions
The notes and cautions in this manual are categorized using the following symbols.
WARNING Calls attention to actions or conditions that could cause serious or fatal injury to
the user, and precautions that can be taken to prevent such occurrences.
CAUTION Calls attention to actions or conditions that could cause light injury to the user or
damage to the instrument or user’s data, and precautions that can be taken to
prevent such occurrences.
French
AVERTISSEMENT Attire l’attention sur des gestes ou des conditions susceptibles
de provoquer des blessures graves (voire mortelles), et sur les
précautions de sécurité pouvant prévenir de tels accidents.
ATTENTION Attire l’attention sur des gestes ou des conditions susceptibles de
provoquer des blessures légères ou d’endommager l’instrument ou les
données de l’utilisateur, et sur les précautions de sécurité susceptibles
de prévenir de tels accidents.
Note Calls attention to information that is important for proper operation of the
instrument.
Character Notations
Menu Names and Panel Key Names in Bold Characters
Indicate controls such as menu commands, tabs, and buttons that appear on the screen and front
panel keys.
Unit
k: Denotes 1000. Example: 100 kHz
K: Denotes 1024. Example: 720 KB (file size)
Metasyntax
The following table contains the symbols that are used in the syntax discussed mainly in chapters 4
and 5. These symbols are referred to as BNF (Backus-Naur Form) symbols. For details on how to
write data using these symbols, see pages 4-6 and 4-7.
Symbol Description Example Example of Input
<> A defined value ELEMent<x> <x> = 1 to 7 ELEMENT2
{}
|
Select an option in { }
Exclusive OR
SQFormula {TYPE1|TYPE2|TYPE3} SQFORMULA TYPE1
[] Can be omitted NUMeric[:NORMal]:VALue? NUMERIC:VALUE?
v
IM WT5000-17EN
1
2
3
4
5
6
7
App
Index
Contents
List of Manuals ................................................................................................................................... i
How to Use This Manual .................................................................................................................. iii
Symbols and Notation Used in This Manual .................................................................................... iv
Chapter 1 Ethernet Interface
1.1 Component Names and Functions ................................................................................... 1-1
1.2 Ethernet Interface Features and Specifications ............................................................... 1-2
1.3 Connecting to the Ethernet Interface ................................................................................ 1-3
1.4 Configuring the Instrument Ethernet Settings .................................................................. 1-4
Chapter 2 USB Interface
2.1 Component Names and Functions ................................................................................... 2-1
2.2 USB Interface Features and Specifications ...................................................................... 2-2
2.3 Connecting to the USB Interface ...................................................................................... 2-3
2.4 Configuring the Instrument USB Settings ......................................................................... 2-4
Chapter 3 GP-IB Interface
3.1 Component Names and Functions ................................................................................... 3-1
3.2 GP-IB Interface Features and Specifications ................................................................... 3-2
3.3 Connecting to the GP-IB Interface ................................................................................... 3-4
3.4 Configuring the Instrument GP-IB Settings ...................................................................... 3-6
3.5 Responses to Interface Messages ................................................................................... 3-7
Chapter 4 Programming Overview
4.1 Messages ......................................................................................................................... 4-1
4.2 Commands ....................................................................................................................... 4-3
4.3 Responses ....................................................................................................................... 4-5
4.4 Data .................................................................................................................................. 4-6
4.5 Synchronization with the Controller .................................................................................. 4-8
Chapter 5 Commands
5.1 List of Commands ............................................................................................................ 5-1
5.2 AOUTput Group ............................................................................................................. 5-15
5.3 AUX Group ..................................................................................................................... 5-16
5.4 COMMunicate Group ..................................................................................................... 5-18
5.5 CURSor Group ............................................................................................................... 5-20
5.6 DISPlay Group ............................................................................................................... 5-23
Function Option List (Settings That Can Be Used for <Function>) .............................. 5-36
5.7 FILE Group ..................................................................................................................... 5-40
5.8 HARMonics Group ......................................................................................................... 5-43
5.9 HOLD Group .................................................................................................................. 5-44
5.10 IMAGe Group ................................................................................................................. 5-45
5.11 INPut Group ................................................................................................................... 5-47
5.12 INTEGrate Group ........................................................................................................... 5-62
5.13 MEASure Group ............................................................................................................. 5-65
5.14 MOTor Group .................................................................................................................. 5-70
5.15 NUMeric Group .............................................................................................................. 5-76
Numeric Data Formats ................................................................................................. 5-81
List of Numeric Data Output Items That Are Preset ..................................................... 5-82
vi IM WT5000-17EN
Contents
5.16 STATus Group ................................................................................................................ 5-85
5.17 STORe Group................................................................................................................. 5-86
5.18 SYSTem Group .............................................................................................................. 5-90
5.19 UPDate Group ................................................................................................................ 5-93
5.20 WAVeform Group ............................................................................................................ 5-94
5.21 Common Command Group ............................................................................................ 5-96
Chapter 6 Status Reports
6.1 About Status Reports ....................................................................................................... 6-1
6.2 Status Byte ....................................................................................................................... 6-3
6.3 Standard Event Register .................................................................................................. 6-4
6.4 Extended Event Register .................................................................................................. 6-5
6.5 Output and Error Queues ................................................................................................. 6-6
Chapter 7 Commands Compatible with Legacy Models
7.1 Command Type Compatible with Legacy Models ...............................................................7-1
Appendix
Appendix 1 Error Messages .................................................................................................. App-1
Appendix 2 About the IEEE 488.2-1992 Standard ................................................................App-4
Index
1-1
IM WT5000-17EN
Ethernet Interface
Chapter 1 Ethernet Interface
1
2
3
4
5
6
7
App
Index
1.1 Component Names and Functions
Front Panel
Setup icon
Remote control settings
Time out period (see sections 1.2 and 1.4)
TCP/IP (see section 13.2, in the Uses manual, IM WT5000-02EN)
Command type (see chapter 7)
MENU key in SETUP (see section 1.4)
SET key
(see section 1.4)
CURSOR keys
(see section 1.4)
ESC key (see section 1.4)
REMOTE LED
Off: Local mode
Solid: Remote mode
UTILITY key
Switching from remote to local mode
(see section 1.2)
This key is disabled when local lockout has
been activated by a controller (e.g., PC).
Showing the utility settings overview screen
(see sections 1.4)
Rear Panel
Ethernet port
This port is for connecting this instrument
to a controller using an Ethernet cable.
See section 1.3.
1-2 IM WT5000-17EN
1.2 Ethernet Interface Features and Specifications
Ethernet Interface Features
Reception Feature
You can use the reception feature to specify the same settings that you specify by using the touch
panel or the front panel keys.
Receive output requests for measured and computed data, panel setup parameters, and error
codes.
Transmission Feature
The instrument can (1) transmit measured and computed data, (2) transmit panel setup parameters
and the status byte, and (3) error codes when errors occur.
Ethernet Interface Specifications
Electrical and mechanical specifications Complies with IEEE802.3, Auto-MDIX
Simultaneous connections 1
Communication protocol TCP/IP (VXI-11)
Connector RJ-45
Switching between Remote and Local Modes
Switching from Local to Remote Mode
The instrument switches to remote mode when it is in local mode and it receives a :COMMunicate:
REMote ON command from the PC.
In Remote Mode
The REMOTE LED turns on.
Touch panel operations and front panel key operations except the UTILITY key are disabled.
The local mode settings are retained even when this instrument switches to remote mode.
Switching from Remote to Local Mode
When the instrument is in remote mode and you press UTILITY, the instrument switches to local mode.
However, this does not work if the instrument has received a :COMMunicate:LOCKout ON command
from the PC. The instrument switches to local mode when it receives a :COMMunicate:REMote OFF
command from the PC, regardless of the local lockout state.
In Local Mode
The REMOTE LED turns off.
Touch panel operations and front panel key operations are enabled.
Settings entered in remote mode are retained even when this instrument switches to local mode.
Note
You cannot use the Ethernet interface simultaneously with other interfaces (GP-IB and USB interfaces).
Setting the Timeout Value
If the instrument is not accessed within a given period of time (specified by the timeout value), it will
disconnect from the network. The timeout value can be set to Infinite or in the range of 1 s to 3600 s.
The default value is Infinite.
For instructions on how to set the timeout value, see section 1.4, “Configuring the instrument Ethernet
Settings.”
1-3
IM WT5000-17EN
Ethernet Interface
1
2
3
4
5
6
7
App
Index
1.3 Connecting to the Ethernet Interface
Connection Procedure
Connect a UTP (Unshielded Twisted-Pair) or STP (Shielded Twisted-Pair) cable that is connected to a
hub or other network device to the Ethernet port on the rear panel of this instrument.
Hub or router that supports 1000BASE-T
UTP or STP cable
WT5000
Controller
(PC or work station)
Ethernet port
RJ-45 modular jack
Notes about Connections
To connect the instrument to a PC, be sure to connect through a hub or router. Proper operation is
not guaranteed for a one-to-one connection.
Use a network cable that supports the data rate of your network.
Note
For details on how to connect the instrument to a network, see section 13.1, “Connecting to a Network” in the
User’s Manual, IM WT5000-02EN.
1-4 IM WT5000-17EN
1.4 Configuring the Instrument Ethernet Settings
This section explains the settings listed below. You must configure these settings when controlling the
instrument remotely through an ethernet interface.
Network connection timeout setting
Utility Settings Overview (Utility)
You can display the remote control setting screen from the utility settings overview screen.
The utility settings overview screen can be opened mainly using the following two methods.
* For information about the utility settings overview screen, see section 1.4 in the User’s Manual, IM WT5000-
02EN.
Procedure Using the Setup Menu
1. Tap the Setup icon ( ), or press MENU under SETUP.
2. Tap the Utility tab. The utility settings overview screen appears.
Pressing ESC closes the list.
Utility tub
Note
You can also display the utility settings overview screen by moving the cursor on the Utility tab using the
arrow keys and then pressing SET.
You can also display the list from the navigation screen that appears immediately after power-on.
Display the list (Setup menu)
Navigation screen
Procedure Using the UTILITY Key
1. Press UTILITY on the front panel. The utility settings overview screen appears.
Pressing ESC closes the list.
1-5
IM WT5000-17EN
Ethernet Interface
1
2
3
4
5
6
7
App
Index
Remote Control Settings (Remote Control)
1. Use either method below to open the utility settings overview screen.
Procedure using the Setup menu: See the previous page.
Procedure using the UTILITY key: See the previous page.
2. Tap the Remote Control button.
The remote control setting screen (Network (VXI-11), GP-IB, USB (USB-TMC)) appears.
3. Tap Time Out. Use the displayed input box to set the timeout value.
Remote Control button
Set the timeout period
(Infinite, 1 to 3600 sec)
IP address
For the setup procedure, see section 13.2 in
the User’s Manual IM WT5000-02EN.
Set the command type
(see chapter 7).
Note
Only use one communication interface: Network, GP-IB, or USB. If you send commands simultaneously from
more than one communication interface, this instrument will not execute the commands properly.
TCP/IP Settings
To use the Ethernet interface, you must specify the following TCP/IP settings.
IP address
Subnet mask
Default gateway
For instructions on how to specify these settings, see section 13.2, “Configuring TCP/IP Settings” in
the User’s Manual, IM WT5000-02EN.
1.4 Configuring the Instrument Ethernet Settings
2-1
IM WT5000-17EN
USB Interface
1
2
3
4
5
6
7
App
Index
2.1 Component Names and Functions
Front Panel
Setup icon
Remote control settings
Viewing the serial number that is used in
USB TMC communications (see section 2.4)
Command type (see chapter 7) MENU key in SETUP (see section 1.4)
SET key
(see section 1.4)
CURSOR keys
(see section 1.4)
ESC key (see section 1.4)
REMOTE LED
Off: Local mode
Solid: Remote mode
UTILITY key
Switching from remote to local mode
(see section 2.2)
This key is disabled when local lockout has
been activated by a controller (e.g., PC).
Showing the utility settings overview screen
(see sections 1.4 and 2.4)
Rear Panel
USB port
This port is for connecting this instrument to a
controller (such as a PC) using a USB cable.
See section 2.3.
Chapter 2 USB Interface
2-2 IM WT5000-17EN
2.2 USB Interface Features and Specifications
USB Interface Features
Reception Feature
You can use the reception feature to specify the same settings that you specify by using the touch
panel or the front panel keys.
Receive output requests for measured and computed data, panel setup parameters, and error
codes.
Transmission Feature
The instrument can (1) transmit measured and computed data, (2) transmit panel setup parameters
and the status byte, and (3) error codes when errors occur.
USB Interface Specifications
Electrical and mechanical specifications Complies with USB Rev. 3.0
Connector USB 3.0 Type B connector (receptacle)
Number of ports 1
Power supply Self-powered
System requirements A PC with a USB port, running Windows 7, Windows 8.1, or
Windows 10.
A separate device driver is required to enable the connection
with the PC.
Switching between Remote and Local Modes
Switching from Local to Remote Mode
The instrument switches to remote mode when it is in local mode and it receives a :COMMunicate:
REMote ON command from the PC.
In Remote Mode
The REMOTE LED turns on.
Touch panel operations and front panel key operations except the UTILITY key are disabled.
The local mode settings are retained even when this instrument switches to remote mode.
Switching from Remote to Local Mode
When the instrument is in remote mode and you press UTILITY, the instrument switches to local mode.
However, this does not work if this instrument has received a :COMMunicate:LOCKout ON command
from the PC. The instrument switches to local mode when it receives a :COMMunicate:REMote OFF
command from the PC, regardless of the local lockout state.
In Local Mode
The REMOTE LED turns off.
Touch panel operations and front panel key operations are enabled.
Settings entered in remote mode are retained even when the instrument switches to local mode.
Note
You cannot use the USB interface simultaneously with other interfaces (GP-IB and Ethernet interfaces).
2-3
IM WT5000-17EN
USB Interface
1
2
3
4
5
6
7
App
Index
2.3 Connecting to the USB Interface
Notes about Connections
Be sure to insert the USB cable connector firmly into the USB port.
If you are connecting multiple devices by using a USB hub, connect this instrument to the USB hub
port that is closest to the port that the controller is connected to.
Do not connect or remove USB cables from the time when this instrument is turned on until
operation becomes available (approximately 20 to 30 seconds). Doing so may damage this
instrument.
2-4 IM WT5000-17EN
2.4 Configuring the Instrument USB Settings
This section explains the settings listed below. You must configure these settings when controlling the
instrument remotely through a USB interface.
Viewing the serial number that is used in USB TMC communications
Remote Control Settings (Remote Control)
1. Use either method below to open the utility settings overview screen.
Procedure using the Setup menu: See section 1.4.
Procedure using the UTILITY key: See section 1.4.
2. Tap the Remote Control button.
The remote control setting screen (Network (VXI-11), GP-IB, USB (USB-TMC)) appears.
Check the serial number.
(see chapter 7).
Note
Only use one communication interface: Network, GP-IB, or USB. If you send commands simultaneously
from more than one communication interface, this instrument will not execute the commands properly.
Install the YOKOGAWA USB driver on your PC. For information about how to obtain the YOKOGAWA
USB driver, contact your nearest YOKOGAWA dealer. You can also access the YOKOGAWA USB driver
download web page and download the driver.
https://tmi.yokogawa.com/library/
3-1
IM WT5000-17EN
GP-IB Interface
1
2
3
4
5
6
7
App
Index
3.1 Component Names and Functions
Front Panel
UTILITY key
Switching from remote to local mode
(see sections 3.2 and 3.5)
This key is disabled when local lockout has
been activated by a controller (e.g., PC).
Showing the utility settings overview screen
(see sections 1.4 and 3.4)
MENU key in SETUP (see section 1.4)
SET key
(see section 1.4)
CURSOR keys
(see section 1.4)
ESC key (see section 1.4)
REMOTE LED
Off: Local mode
Solid: Remote mode
Setup icon
Remote control settings
GP-IB address (see section 3.4)
Command type (see chapter 7)
Rear Panel
GP-IB port
This port is for connecting this instrument to a
controller (such as a PC) using a GP-IB cable.
See section 3.3.
Chapter 3 GP-IB Interface
3-2 IM WT5000-17EN
3.2 GP-IB Interface Features and Specifications
GP-IB Interface Features
Listener Capability
You can use the GP-IB interface to specify the same settings that you can using the instrument’s touch
panel or keys. You cannot use it, however, to turn the instrument on and off or change communication
settings.
Output requests for measured and computed data, panel setup parameters, and error codes can be
received.
Commands such as status report commands can be received.
Talker Capability
The instrument can (1) transmit measured and computed data, (2) transmit panel setup parameters
and the status byte, and (3) error codes when errors occur.
Note
Talk-only, listen-only, and controller capabilities are not available on the instrument.
GP-IB Interface Specifications
Supported Devices National Instruments Corporation
PCI-GPIB or PCI-GPIB+ • GPIB-USB-HS
PCIe-GPIB or PCIe-GPIB+ • GPIB-USB-HS+
PCMCIA-GPIB or PCMCIA-GPIB+
Driver NI-488.2M Version 1.60 or later
Electrical and mechanical
specifications
Complies with IEEE St’d 488-1978
Functional specifications See the following table.
Protocol Complies with IEEE St’d 488.2-1992
Code ISO (ASCII) code
Mode Addressable mode
Address settings You can set the address in the range of 0 to 30 from the utility settings overview
screen > the remote control setting screen.
Clearing remote mode Clear remote mode by pressing UTILITY key.
This key is disabled when local lockout has been activated by a controller.
Remote mode can be cleared by receiving the following interface message from
the controller.
- Set REN (Remote Enable) to false.
- GTL(Go To Local)
Functional Specifications
Function Subset Name Description
Source handshaking SH1 Full source handshaking capability
Acceptor handshaking AH1 Full acceptor handshaking capability
Talker T6 Basic talker capability, serial polling, and untalk on MLA (My Listen
Address). No talk-only capability.
Listener L4 Basic listener capability, unlisten on MTA (My Talk Address), and
no listen-only capability
Service request SR1 Full service request capability
Remote local RL1 Full remote/local capability
Parallel polling PP0 No parallel poll capability
Device clear DC1 Full device clear capability
Device trigger DT1 Device trigger capability
Controller C0 No controller capability
Electric characteristics E1 Open collector
3-3
IM WT5000-17EN
GP-IB Interface
1
2
3
4
5
6
7
App
Index
Switching between Remote and Local Modes
Switching from Local to Remote Mode
The instrument switches to remote mode when it is in local mode and receives an interface message*
that sets REN (Remote Enable) to true from the controller.
In Remot Mode
The REMOTE LED turns on.
Touch panel operations and front panel key operations except the UTILITY key are disabled.
The local mode settings are retained even when this instrument switches to remote mode.
Switching from Remote to Local Mode
When the instrument is in remote mode and you press UTILITY, the instrument switches to local
mode. However, if the instrument has received an LLO (Local Lockout) interface message* from the
controller, the instrument is in local lockout mode and, the UTILITY key is also disabled.
If the instrument receives an interface message* that sets REN (Remote Enable) to false from the
controller, local lockout is cleared, and all devices on the bus are set to local mode.
When the instrument is in remote mode and receives a GTL (Go To Local) interface message* from
the controller, the device designated as a listener switches to local mode.
In Local Mode
The REMOTE LED turns off.
Touch panel operations and front panel key operations are enabled.
Settings entered in remote mode are retained even when this instrument switches to local mode.
Note
You cannot use the GP-IB interface simultaneously with other interfaces (USB and Ethernet interfaces).
* For information about the interface messages, see section 3.5.
3.2 GP-IB Interface Features and Specifications
3-4 IM WT5000-17EN
3.3 Connecting to the GP-IB Interface
GP-IB Cable
The instrument is equipped with an IEEE St’d 488-1978 24-pin GP-IB connector. Use GP-IB cables
that comply with IEEE St’d 488-1978.
Connection Procedure
Connect the cable as shown below.
Notes about Connections
Firmly tighten the screws on the GP-IB cable connector.
On the PC end, use a GP-IB board (or card) made by National Instruments. For details, see section
3.2.
The instrument may not operate properly if the instrument is connected to the PC through converters
(such as a GP-IB to USB converter). For more details, contact your nearest YOKOGAWA dealer.
Several cables can be used to connect multiple devices. However, no more than 15 devices,
including the controller, can be connected on a single bus.
When connecting multiple devices, you must assign a unique address to each device.
Use cables that are 2 m or shorter in length to connect devices.
Make sure the total length of all cables does not exceed 20 m.
When devices are communicating, have at least two-thirds of the devices on the bus turned on.
To connect multiple devices, wire them in a daisy-chain or star configuration as shown below. You
can also mix these configurations. Loop configuration is not allowed.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146

YOKOGAWA WT5000 User manual

Category
Measuring, testing & control
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI