Gilson Adjustment Bracket for 5qt Lab Mixer User manual

Type
User manual

This manual is also suitable for

INTRODUCTION
When ๎€Ÿtted with the MAA-30A Clearance Adjustment Bracket
and properly adjusted, the MA-52 Laboratory Mixer meets ASTM
C109, C227, C305, AASHTO T106, and T162.
INSTALLATION INSTRUCTIONS
A. Mounting and Adjusting the Bracket
1. Remove two front existing motor mounting bolts (see
Figure 1).
2. Attach bracket with long replacement mounting bolts
provided (see Figure 2).
3. For ASTM C305, adjust clearance between the paddle and
bowl to between 0.8 and 2.5mm.
4. For ASTM C227, adjust clearance between the paddle and
bowl to between 5mm and 6mm.
5. If required, notch yoke to obtain proper clearance as follows:
a. Remove yoke from mixer by removing four mounting
screws (see Figure 3).
b. Remove two yoke mounting screws (see Figure 4).
c. File or cut notch in yoke as indicated in Figure 5.
d. Reassemble by reversing directions a. through c.
6. Repeat step 3.
(Continued on back.)
Figure 1
Adjusting
Screws (2)
Motor Mounting Bolts (2)
Locking Nuts (2)
Clearance Adjustment Bracket for
MA-52 & MA-52X
OPERATING MANUAL
Rev: 07/2019
PHONE: 800-444-1508 P.O. Box 200, Lewis Center, Ohio 43035-0200
740-548-7298 E-mail: customerservice@gilsonco.com Product Web Page: www.globalgilson.com
FAX: 800-255-5314
740-548-5314
โ†’
2-1/2in
1/8in
โ†‘
โ†“
Figure 2
Figure 4
Figure 5
Figure 3
Adjusting Screws (2)
Yoke Mounting Screws (2)
Base Mounting Screws (4)
Motor Mounting Bolts (2)
Locking Nuts (2)
  • Page 1 1
  • Page 2 2

Gilson Adjustment Bracket for 5qt Lab Mixer User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI