Additional statement added:
LaserXL analysis software was designed to meet ISO 230-2 1997. The
current standard (ISO 230-2 2006) altered the analysis of the data by
introducing the determination and reporting of a value for
measurement uncertainty, which is not reported in LaserXL analysis
software. There has been no change to the data capture process or to
the previous analysis calculations from ISO 230-2 1997.
It is up to the user to define and complete a range of tests for the
uncertainty parameters specified in ISO 230-2 2006. Once a total
measurement uncertainty value has been calculated, the new
Renishaw analysis software (XCal-View) provides the user the space to
manually enter it into reporting.
ISO 230-2 2006 clarification
Section – Software,
Analysis,
Analysis standards