Installation
L
20
30 F
B
1 2
►Grease inner contact faces of the thread with suitable paste.
The paste must be suitable and approved for the application and compatible
with the elastomers used.
Recommendation: Klüber paste UH1 84-201 with USDA-H1 approval for the
food industry.
►Slip the thread onto the temperature probe and set the probe to the requested
installation depth (L).
►Tighten the locking nut (B) and the clamping device (F) with a spanner / torque
wrench. Tightening torque: 25 Nm.
The tightening torque is important to guarantee the pressure resistance of
the clamp connection up to 50 bar.
For mounting the Aseptoflex adapter → installation instructions of the adapter.
When removing the sensor, the clamping device (F) should be blocked by
means of a spanner to avoid loosening of the clamping device.