Mettler Toledo 10 & 50 Constant Conductivity/Resistivity Sensors Thornton Instruments Operating instructions

Type
Operating instructions
THORNTON
Leading Pure Water Analytics
TYPICAL
NOT
INSTALLATIONS
READ THIS FIRST
RECOMMENDED
RECOMMENDED
MOUNTING
INLET
84299F2
OUTLET
This is the ideal
mounting position
84299F1
OUTLET
INLET
The solution flow should be directed
into the end of the sensor and allowed
to pass out the circulation holes.
The sensor should be mounted so the
cavity does not trap air or solids.
Mettler-Toledo Thornton does not
recommend mounting the sensor in a
vertical pipe with the flow direction
down, because air bubbles or air
pockets will affect the sensor
measurement.
INLET
OUTLET
84299F5
INLET
OUTLET
AIR
OUTLET
OUTLET
AIR
84299F3
INLET
SOLIDS
INLET
OUTLET
84299F4
OM84299-B
INSTALLATION INSTRUCTIONS FOR
10 AND 50 CONSTANT
CONDUCTIVITY SENSORS
SENSOR MAINTENANCE
Clean 10 and 50 constant sensors only chemically with dilute acid or other suitable cleaning agent. Do NOT clean the
inside of the cavities with a small brush or pipe cleaner because this would result in permanent damage to the internal
electrodes and cause erroneous measurements.
WARNING: USE STANDARD SAFETY PRECAUTIONS IF CLEANING WITH ACID.
Mettler-Toledo Thornton, Inc. Toll-Free: 800-510-PURE
36 Middlesex Turnpike Fax: 781-271-0214
Bedford, MA 01730 [email protected]
(781) 301-8600
www.thorntoninc.com
OM84299 Rev. B 11/03