  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo Thornton 4000TOC Sensor THO_4000TOC_58 130 196_en_May 2014 Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Total Organic Carbon Measurement System
Operation Manual
4000TOC Sensor
M300 TOC Transmitter
4000TOC Sensor
M300 TOC Transmitter
58 130 196 Rev B
4000TOC Sensor 2
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4000TOC Sensor 3
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
Operation Manual
4000TOC Sensor
M300 TOC Transmitter
4000TOC Sensor 4
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4000TOC Sensor 5
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
Content
1 Introduction ______________________________________________________________________________________________ 9
2 Safety instructions________________________________________________________________________________________ 10
2.1 Denitionofequipmentanddocumentationsymbolsanddesignations __________________________________________ 10
2.2 Correctdisposaloftheunit ____________________________________________________________________________ 11
3 Unit overview____________________________________________________________________________________________ 12
3.1 Overview, 4000TOC Sensor ____________________________________________________________________________ 12
3.1.1 SensorDetails ________________________________________________________________________________ 12
3.1.2 HighPressureApplications ______________________________________________________________________ 13
3.1.3 PotentialCondensationApplications ______________________________________________________________ 13
3.1.4 HighTemperatureApplications ___________________________________________________________________ 13
3.2 Overview, M300 TOC _________________________________________________________________________________ 13
3.2.1 1/4DIN _____________________________________________________________________________________ 13
3.2.2 1/2DIN _____________________________________________________________________________________ 14
3.3 Control/NavigationKeys _______________________________________________________________________________ 14
3.3.1 MenuStructure _______________________________________________________________________________ 14
3.3.2 Navigationkeys _______________________________________________________________________________ 15
3.3.2.1 Navigatingthemenutree _______________________________________________________________ 15
3.3.2.2 Escape
_____________________________________________________________________________ 15
3.3.2.3 Enter _______________________________________________________________________________ 15
3.3.2.4 Menu
_______________________________________________________________________________ 15
3.3.2.5 Calibrationmode _____________________________________________________________________ 15
3.3.2.6 Infomode ___________________________________________________________________________ 15
3.3.3 Navigationofdataentryelds ____________________________________________________________________ 15
3.3.4 Entryofdatavalues,selectionofdataentryoptions ___________________________________________________ 15
3.3.5 NavigationwithuinDisplay _____________________________________________________________________ 16
3.3.6 ”Savechanges”dialog _________________________________________________________________________ 16
3.3.7 SecurityPasswords ____________________________________________________________________________ 16
3.4 Display
____________________________________________________________________________________________ 16
4 M300 TOC Installation instruction ___________________________________________________________________________ 17
4.1 UnpackingandinspectionoftheM300TOC _______________________________________________________________ 17
4.1.1 Panelcutoutdimensionalinformation–
1/4DINmodels _______________________________________________________________________________ 17
4.1.2 Installationprocedure–1/4DINmodels ____________________________________________________________ 18
4.1.3 Panelcutoutdimensionalinformation–
1/2DINmodels _______________________________________________________________________________ 19
4.1.4 Installationprocedure–1/2DINmodels ____________________________________________________________ 20
4.1.5 Assembly–1/2DINversion ______________________________________________________________________ 20
4.1.6 1/2DINversion–Dimensiondrawings _____________________________________________________________ 21
4.1.7 1/2DINversion–Pipemounting __________________________________________________________________ 21
4.1.8 1/4DINversion–Dimensiondrawings _____________________________________________________________ 22
4.2 Connectionofpowersupply ____________________________________________________________________________ 23
4.2.1 1/4DINhousing(panelmount) ___________________________________________________________________ 23
4.2.2 1/2DINhousing(wallmount) ____________________________________________________________________ 24
4.3 ConnectorPINdenition _______________________________________________________________________________ 24
4.3.1 TB1andTB2for1/2DINand1/4DINversions _______________________________________________________ 25
4.3.2 TB3for1/2DINand1/4DINversions ______________________________________________________________ 25
4.4 ConnectionofTOCSensor _____________________________________________________________________________ 25
5 4000TOC Installation Instructions ___________________________________________________________________________ 26
5.1 Unpackingandinspectionofthe4000TOC ________________________________________________________________ 26
5.2 Installationofthe4000TOC ____________________________________________________________________________ 26
5.3 SampleTubingConnections ____________________________________________________________________________ 27
5.4 SampleConditioningCoil ______________________________________________________________________________ 28
5.5 ACPowerConnection _________________________________________________________________________________ 29
5.6 SensorConnection ___________________________________________________________________________________ 30
6 Placing transmitter in, or out, of service _____________________________________________________________________ 31
6.1 Placingtransmitterinservice ___________________________________________________________________________ 31
6.2 Placingtransmitteroutofservice _______________________________________________________________________ 31
7 4000TOC Sensor Operation ________________________________________________________________________________ 32
7.1 InitialStartup _______________________________________________________________________________________ 32
7.2 SettingSampleFlowRate ______________________________________________________________________________ 32
7.3 StartTOCMeasurement _______________________________________________________________________________ 33
7.4 NormalOperation ____________________________________________________________________________________ 33
7.5 SystemSuitabilityTesting ______________________________________________________________________________ 34
4000TOC Sensor 6
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
7.6 Calibration _________________________________________________________________________________________ 34
7.6.1 TOCCalibrationMethods ________________________________________________________________________ 34
8 Quick Setup _____________________________________________________________________________________________ 36
9 Sensor Calibration _______________________________________________________________________________________ 37
9.1 EnterCalibrationMode ________________________________________________________________________________ 37
10 Conguration
____________________________________________________________________________________________ 38
10.1 EnterCongurationMode ______________________________________________________________________________ 38
10.2 Measurement
_______________________________________________________________________________________ 38
10.2.1 ChannelSetup ________________________________________________________________________________ 38
10.2.2 Conductivity/ResistivityTemperatureCompensation ___________________________________________________ 39
10.2.3 SetAveraging ________________________________________________________________________________ 39
10.3 TOCSetup __________________________________________________________________________________________ 40
10.3.1 TOCGeneralParameters ________________________________________________________________________ 40
10.3.2 LampControl _________________________________________________________________________________ 41
10.3.3 Autobalance
_________________________________________________________________________________ 41
10.4 AnalogOutputs ______________________________________________________________________________________ 42
10.5 ____________________________________________________________________________________________________
Setpoints __________________________________________________________________________________________ 43
10.6 Alarm
_____________________________________________________________________________________________ 46
10.7 Display
____________________________________________________________________________________________ 47
10.7.1 Measurement
_________________________________________________________________________________ 47
10.7.2 Resolution
___________________________________________________________________________________ 48
10.7.3 Backlight
____________________________________________________________________________________ 48
10.7.4 Name ______________________________________________________________________________________ 48
10.8 HoldAnalogOutputs _________________________________________________________________________________ 49
11 System _________________________________________________________________________________________________ 50
11.1 SetLanguage _______________________________________________________________________________________ 50
11.2 USB ______________________________________________________________________________________________ 50
11.3 Passwords
_________________________________________________________________________________________ 51
11.3.1 ChangingPasswords __________________________________________________________________________ 51
11.3.2 ConguringMenuAccessforOperator _____________________________________________________________ 51
11.4 Set/ClearLockout ____________________________________________________________________________________ 51
11.5 Reset _____________________________________________________________________________________________ 52
11.5.1 ResetSystem _________________________________________________________________________________ 52
11.5.2 ResetAnalogCalibration ________________________________________________________________________ 52
11.5.3 ResetSensorCalibrationDatatoFactorySettings _____________________________________________________ 53
11.5.4 ResetCalibrationDataofSensorElectronicstoFactorySettings __________________________________________ 53
12 4000TOC Sensor Service Requirements ______________________________________________________________________ 54
12.1 UVLampReplacement ________________________________________________________________________________ 54
12.2 HighCapacityInletFilterReplacement ____________________________________________________________________ 55
13 M300 TOC Service _______________________________________________________________________________________ 56
13.1 EnterServiceMenu ___________________________________________________________________________________ 56
13.2 Diagnostics
________________________________________________________________________________________ 56
13.2.1 Model/SoftwareRevision ________________________________________________________________________ 56
13.2.2 DigitalInput __________________________________________________________________________________ 57
13.2.3 Display
_____________________________________________________________________________________ 57
13.2.4 Keypad
_____________________________________________________________________________________ 57
13.2.5 Memory
_____________________________________________________________________________________ 57
13.2.6 SetRelay ____________________________________________________________________________________ 58
13.2.7 ReadRelays _________________________________________________________________________________ 58
13.2.8 SetAnalogOutputs ____________________________________________________________________________ 58
13.2.9 ReadAnalogOutputs ___________________________________________________________________________ 58
13.3 Calibrate
___________________________________________________________________________________________ 59
13.3.1 VerifyConductivityElectronics ____________________________________________________________________ 59
13.3.2 CalibrateAnalog ______________________________________________________________________________ 59
13.3.3 CalibrateUnlock ______________________________________________________________________________ 60
13.4 TechService ________________________________________________________________________________________ 60
14 Info ____________________________________________________________________________________________________ 61
14.1 InfoMenu __________________________________________________________________________________________ 61
14.2 Messages
__________________________________________________________________________________________ 61
14.3 CalibrationData _____________________________________________________________________________________ 61
14.4 Model/SoftwareRevision ______________________________________________________________________________ 62
14.5 TOCSensorInfo _____________________________________________________________________________________ 62
15 Maintenance ____________________________________________________________________________________________ 63
15.1 ForTechnicalSupport _________________________________________________________________________________ 63
15.2 FrontPanelCleaning _________________________________________________________________________________ 63
4000TOC Sensor 7
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
16 Troubleshooting the M300 TOC Transmitter ___________________________________________________________________ 64
16.1 ChangingtheFuse ___________________________________________________________________________________ 64
16.2 Warning-andAlarmindicationonthedisplay _____________________________________________________________ 65
16.2.1 Warningindication ____________________________________________________________________________ 65
16.2.2 Alarmindication ______________________________________________________________________________ 65
17 Troubleshooting the 4000TOC Sensor ________________________________________________________________________ 66
17.1 BasicTroublshooting _________________________________________________________________________________ 66
17.2 FaultandErrorMessages ______________________________________________________________________________ 66
18 Accessories and Spare Parts _______________________________________________________________________________ 68
18.1 M300 TOC Transmitter ________________________________________________________________________________ 68
18.2 4000TOC Sensor ____________________________________________________________________________________ 68
18.2.1 4000TOCSensorSpareParts ____________________________________________________________________ 68
18.2.2 4000TOCSensorConsumableItems ______________________________________________________________ 69
4000TOC Sensor 8
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4000TOC Sensor 9
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
1 Introduction
Statementofintendeduse–The4000TOCsensorisaTotalOrganicCarbonsensordesignedto
measuretheconcentrationoforganicsubstancesinpureandultrapurewaterapplications.The
M300TOCisasinglechanneltransmitterwithalargefourlinebacklitLiquidCrystalDisplay
whichconveysmeasuringandsetupinformationforthe4000TOCsensor.
Whenusedincombination,the4000TOCsensorandM300TOCtransmitter,formaTOCmea-
surementsystemwhichprovidesthedesignflexibilityforfullintegrationintowatersystems.The
separatecomponentsallowforclosepositioningofthesensortothesamplepointforfastestre-
sponse,whileallowingintegrationofthetransmitterintothecontrolsystem.
Thismanualappliestothe4000TOCsensor,andM300TOCtransmitter:
4000TOC
Designation Orderno.
4000TOCSensor,110VAC,50/60Hz 58036021
4000TOCSensor,220VAC,50/60Hz 58036022
M300 TOC
Designation Orderno.
M300 TOC 1/4DIN 58 006 302
M300 TOC 1/2DIN 58 006 312
Throughoutthismanual,theterms‘M300’and‘M300TOC’areused.Withinthecontextofthis
document,bothtermsshouldbetakentorefertotheM300TOCtransmitter.
Theprintscreenimagesinthismanualhaveageneralexplainingcharacterandcandifferfrom
therealdisplayinyourtransmitter.
Thisdescriptioncorrespondstothefirmwarerelease,version1.0oftheM300TOCtransmitter.
Changesaretakingplaceconstantly,withoutpriornotification
4000TOC Sensor 10
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
2 Safety instructions
Thismanualincludessafetyinformationwiththefollowingdesignationsandformats.
2.1 Denitionofequipmentanddocumentationsymbols
and designations
a
WARNING:POTENTIALFORPERSONALINJURY.
a
CAUTION:possibleinstrumentdamageormalfunction.
h
NOTE:Importantoperatinginformation.
a
On the transmitter or in this manual text indicates:Cautionand/orotherpossiblehazardin-
cludingriskofelectricshock(refertoaccompanyingdocuments).
Thefollowingisalistofgeneralsafetyinstructionsandwarnings.Failuretoadheretothesein-
structionscanresultindamagetotheequipmentand/orpersonalinjurytotheoperator.
TheM300Transmittershouldbeinstalledandoperatedonlybypersonnelfamiliarwith
thetransmitterandwhoarequalifiedforsuchwork.
TheM300Transmittermustonlybeoperatedunderthespecifiedoperatingconditions
(seesection19).
RepairoftheM300Transmittermustbeperformedbyauthorized,trainedpersonnelonly.
Withtheexceptionofroutinemaintenance,cleaningproceduresorfusereplacement,as
describedinthismanual,theM300Transmittermustnotbetamperedwithoralteredin
anymanner.
Mettler-Toledoacceptsnoresponsibilityfordamagecausedbyunauthorizedmodifica-
tions
tothetransmitter.
Followallwarnings,cautions,andinstructionsindicatedonandsuppliedwiththisprod-
uct.
Installequipmentasspecifiedinthisinstructionmanual.Followappropriatelocaland
nationalcodes.
Protectivecoversmustbeinplaceatalltimesduringnormaloperation.
Ifthisequipmentisusedinamannernotspecifiedbythemanufacturer,theprotection
providedbyitagainsthazardsmaybeimpaired.
WARNINGS:
Installationofcableconnectionsandservicingofthisproductrequireaccesstoshock
hazardvoltagelevels.
Mainpowerandrelaycontactswiredtoseparatepowersourcemustbedisconnectedbe-
foreservicing.
Switchorcircuitbreakershallbeincloseproximitytotheequipmentandwithineasy
reachoftheOPERATOR;itshallbemarkedasthedisconnectingdevicefortheequip-
ment.
Mainpowermustemployaswitchorcircuitbreakerasthedisconnectingdeviceforthe
equipment.
a
4000TOC Sensor 11
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
ElectricalinstallationmustbeinaccordancewiththeNationalElectricalCodeand/orany
otherapplicablenationalorlocalcodes.
Safetyandperformancerequirethatthisinstrumentbyconnectedandproperlygrounded
throughathree-wirepowersource.
h
NOTE! RELAY CONTROL ACTION:theM300transmitterrelayswillalwaysde-energizeonlossof
power,equivalenttonormalstate,regardlessofrelaystatesettingforpoweredoperation.Config-
ureanycontrolsystemusingtheserelayswithfail-safelogicaccordingly.
h
NOTE! PROCESS UPSETS:Becauseprocessandsafetyconditionsmaydependonconsistent
operationofthistransmitter,provideappropriatemeanstomaintainoperationduringsensor
cleaning,replacementorsensororinstrumentcalibration.
h
NOTE: Thisisa4-wire-productwithanactive4–20mAanalogoutput.
PleasedonotsupplytoPin1–Pin6ofTB2.
2.2 Correct disposal of the unit
Whenthetransmitterisfinallyremovedfromservice,observealllocalenvironmentalregulations
forproperdisposal.
4000TOC Sensor 12
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
3 Unit overview
M300modelsareavailableinbotha1/4DINand1/2DINcasesize.The1/4DINisapanel-
mountonlydesignandthe1/2DINmodelsprovideanintegralIP65housingforwall-,orpipe-
mount.
3.1 Overview, 4000TOC Sensor
3.1.1 Sensor Details
The4000TOCSensorisaTotalOrganicCarbonsensordesignedforusewiththeM300TOC
transmitter.TheM300TOCwillautomaticallyidentifythe4000TOCSensorwhenitisconnect-
ed,andallfactorycalibrationdataisautomaticallyreadandusedforoperationandindication.
The4000TOCSensorandM300TOCtransmitterareconfiguredwithfactorydefaultsettingsto
minimizemeasurementsetuptime.Thesedefaultsettingsallowforsensoroperationwithmini-
malkeystrokes.The4000TOCSensorhasmorefeaturesthanothersensorsandmayrequire
additionalconfigurationdependantupontheoperationalsetup.Oncetheseparametersareset,
theTOCsensorwillfunctionautomatically,andwillrequireuserinterfaceonlyforperiodicser-
viceandmaintenance.Measurementsavailablefromthe4000TOCSensorincludeTOC,Con-
ductivity/Resistivity(compensatedanduncompensated)andTemperature.
The4000TOCSensorisdesignedwithfourlocalLEDindicatorsandaUVLampcontrolkeyon
thefrontpanel,asshownintheillustrationbelow.TheLEDsaredesignedtoprovidelocalindi-
cationofsensorstatus.TheoperationoftheLEDsissynchronizedwiththeM300TOCdisplay
statusmessagesthatappearinthe6-digitmeasurementnamelocation.
Function Color Operation
Fault Red
FlashesduringFaultcondition,sensoroperationstopped
Error Amber IlluminatesduringErrorcondition,sensorremainsoperational
SensorStatus Green
IlluminateswheneverACPowerisonandtheM300TOCis
connectedtothesensor
UVLampOn Green IlluminateswhenevertheUVLampison
TheUVLampcontrolkeyisdesignedtoprovidealocalOn/OffcontrolfortheUVlamptoallow
forquicklampchange-outandassisttheoperatorduringtroubleshooting,ifnecessary.
NOTE:IftheUVLampKeyispressedtoturntheUVLampoff,theUVLampcannotbeturnedon
fromtheM300TOC,itcanonlybeturnedonbypressingthesensorkeyagain.
h
4000TOC Sensor 13
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
Thefrontcoverofthesensorishingedontheleftside.Locatedatthetopandbottomrightcor-
nersofthesensorenclosurearetwotriangle-shapeddoorfasteners.Theinstallationkitprovid-
edwiththeSensorincludesthespecialtoolneededtoloosenthesefastenersandopenthefront
doortotheenclosure.Periodicaccesswillberequiredtoperformroutineserviceandmainte-
nance.AdditionalfrontcovertoolsareavailablefromMettler-ToledoThornton,Inc.See
Section18forpartnumbersanddescriptions.
3.1.2 High Pressure Applications
Forinstallationswithprocesspressureabove85psig(5.9bar),theHighPressureRegulator
(p/n58091552)isrequired(seesection5.3).Installationinstructionsaresuppliedwiththe
HighPressureRegulator.
3.1.3 Potential Condensation Applications
Certainambientandprocesstemperaturesmaycausecondensationtooccurinsidethe
4000TOCSensor.TheMettler-ToledoThorntonSampleConditioningCoilprovidedwiththe
sensorinstallationkitisdesignedtoreducecondensationconditionsbyallowingthesample
temperaturetopartiallyequalizewiththeambienttemperature.
3.1.4 High Temperature Applications
Forwatertemperaturesabove70°C(158°F),theSampleConditioningCoil(p/n58091511)
isrequired.Inhighpressureapplications,theHighPressureRegulator(p/n58091552)must
beinstalleddownstreamoftheSampleConditioningCoil.
3.2 Overview, M300 TOC
3.2.1 1/4DIN
4
5
6
7
8
4.01
[102]
4.01
[102]
EnterInfo
METTLER TOLEDO
ESC
Menu Cal
M300
1
2
3
1:HardPolycarbonatecase 5:USBInterfacePort
2:FiveTactile-FeedbackNavigationKeys 6:RelayOutputTerminals
3:Four-lineLCDDisplay 7:AnalogOutput/DigitalInputTerminals
4:PowerSupplyTerminals 8:SensorInputTerminals
4000TOC Sensor 14
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
3.2.2 1/2DIN
5.90
[150]
5.90
[150]
EnterInfo
CalMenu
M300
1
2
3
ESC
4
5
6
7
8
1:HardPolycarbonatecase 5:USBInterfacePort
2:FiveTactile-FeedbackNavigationKeys 6:RelayOutputTerminals
3:Four-lineLCDDisplay 7:AnalogOutput/DigitalInputTerminals
4:PowerSupplyTerminals 8:SensorInputTerminals
3.3 Control/Navigation Keys
3.3.1 Menu Structure
BelowisthestructureoftheM300menutree:
Measurement
Mode M300
CalMenu Info
Quick Setup Configure System Service Messages
TOC Sensor
Info
Measurement
Analog Outputs
Set Points
Alarm/Clean
Display
Set Language
USB
Passwords
Set/Clear
Lockout
Reset
Calibrate
Tech Service
Diagnostics
Hold Outputs
Calibration Data
Model/Software
Revision
Channel Setup
Analog Outputs
Set Points
Set Flow
Set Flow
UV Lamp
4000TOC Sensor 15
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
3.3.2 Navigation keys
Menu Cal
Info
Enter
ESC
3.3.2.1 Navigating the menu tree
EnterthedesiredmainMenubranchwiththe or keys.Usetheandkeystonavi-
gatethroughtheselectedMenubranch.
h
NOTE:Inordertobackuponemenupage,withoutescapingtothemeasurementmode,
movethecursorundertheUPArrowcharacter(u) atthebottomrightofthedisplayscreenand
press[Enter].
3.3.2.2 Escape
Presstheandkeysimultaneously(escape)toreturntotheMeasurementmode.
3.3.2.3 Enter
Usetheekeytoconfirmactionorselections.
3.3.2.4 Menu
PressthekeytoaccessthemainMenu.
3.3.2.5 Calibration mode
PressthekeytoenterCalibrationMode.
3.3.2.6 Info mode
PressthekeytoenterInfoMode
3.3.3 Navigationofdataentryelds
Usethekeytonavigateforwardorthe keytonavigatebackwardswithinthechangeable
dataentryfieldsofthedisplay.
3.3.4 Entry of data values, selection of data entry options
4000TOC Sensor 16
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
Usethekeytoincreaseorthekeytodecreaseadigit.Usethesamekeystonavigatewith-
inaselectionofvaluesoroptionsofadataentryfield.
h
NOTE:Somescreensrequireconfiguringmultiplevaluesviathesamedatafield(ex:configuring
multiplesetpoints).Besuretousethe or keytoreturntotheprimaryfieldandthe or
keytotogglebetweenallconfigurationoptionsbeforeenteringtothenextdisplayscreen.
3.3.5 Navigation with u in Display
Ifauisdisplayedonthebottomrighthandcornerofthedisplay,youcanusetheorthe
keytonavigatetoit.Ifyouclick[ENTER]youwillnavigatebackwardsthroughthemenu
(gobackonescreen).Thiscanbeaveryusefuloptiontomovebackupthemenutreewithout
havingtoexitintothemeasuringmodeandre-enterthemenu.
3.3.6 ”Save changes” dialog
Threeoptionsarepossibleforthe”Savechanges”dialog:Yes&Exit(Savechangesandexitto
measuringmode),”Yes&u”(Savechangesandgobackonescreen)and”No&Exit”(Don’t
savechangesandexittomeasuringmode).The”Yes&u”optionisveryusefulifyouwantto
continueconfiguringwithouthavingtore-enterthemenu.
3.3.7 Security Passwords
TheM300transmitterallowsasecuritylock-outofvariousmenus.Ifthesecuritylock-out
featureofthetransmitterhasbeenenabled,asecuritypasswordmustbeenteredtoallowac-
cesstothemenu.Seesection11.3formoreinformation.
3.4 Display
h
NOTE:IntheeventofanalarmorothererrorconditiontheM300Transmitterwilldisplayaflash-
inga intheupperrightcornerofthedisplay.Thissymbolwillremainuntiltheconditionthat
causedithasbeencleared.
h
NOTE:Duringcalibrations,clean,DigitalInwithAnalogOutput/Relay/USBinHoldstate,
aflashingHwillappearintheupperleftcornerofthedisplay.Thissymbolwillremainfor
20secondsuntilafterthecalibrationorcleaniscompleted.Thissymbolwillalsodisappear
whenDigitalInisdeactivated.
4000TOC Sensor 17
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4 M300 TOC Installation instruction
4.1 Unpacking and inspection of the M300 TOC
Inspecttheshippingcontainer.Ifitisdamaged,contacttheshipperimmediatelyforinstructions.
Donotdiscardthebox.
Ifthereisnoapparentdamage,unpackthecontainer.Besureallitemsshownonthepacking
listarepresent.
Ifitemsaremissing,notifyMettler-Toledoimmediately.
4.1.1 Panel cutout dimensional information –
1/4DIN models
1/4DINModeltransmittersaredesignedforpanel-mountinstallationonly.Eachtransmitteris
suppliedwithmountinghardwaretoprovidefastandsimpleinstallationtoaflatpanelorflaten-
closuredoor.ToinsureagoodsealandmaintainIPintegrityofinstallation,thepanelordoor
mustbeflatandhaveasmoothfinish.Hardwareconsistsof:
2snap-onMountingbrackets
1mountinggasketseal
Transmitterdimensionsandmountingareshowninthefiguresbelow.
3.62
+0.02
–0
[92
+0.5
]
–0
3.62
+0.02
–0
[92
+0.5
]
–0
PANEL CUT-OUT
4000TOC Sensor 18
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4.1.2 Installation procedure – 1/4DIN models
Makecutoutinpanel(seedimensionscutoutdrawing).
Besuresurfacesurroundingcutoutisclean,smoothandfreeofburrs.
Slidefacegasket(suppliedwithtransmitter)aroundtransmitterfromthebackoftheunit.
Placetransmitterintocutouthole.Besuretherearenogapsbetweenthetransmitter
andpanelsurface.
Placethetwomountingbracketsoneithersideofthetransmitterasshown.
Whileholdingtransmitterfirmlyintothecutouthole,pushthemountingbrackets
towardthebacksideofpanel.
Oncesecure,useascrewdrivertotightenthebracketsagainstthepanel.Inordertopro-
videIP65environmentalenclosurerating,thetwoclampsprovidedshallbesecurely
tightenedtocreateanadequatesealbetweenthepanelenclosureandM300frontface.
Facegasketwillcompressbetweentransmitterandpanel.
a
CAUTION:Donotovertightenbrackets.
4.01
[102]
4.33
[110]
3.55
[90]
4.01
[102]
3.55
[90]
4.01
[102]
4.96
[126]
1.77
[45]
3.54
[90]
4.33
[110]
MAX. 0.118
[3]
FLAT GASKET
MOUNTING SCREW
2 PLACES
MOUNTING
BRACKET
2 PLACES
PowerFuse
USB
1234567
1234567
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
4.92
[125]
4.01
[102]
4.33
[110]
3.55
[90]
4.01
[102]
3.55
[90]
4.01
[102]
4.96
[126]
1.77
[45]
3.54
[90]
4.33
[110]
MAX. 0.118
[3]
FLAT GASKET
MOUNTING SCREW
2 PLACES
MOUNTING
BRACKET
2 PLACES
PowerFuse
USB
1234567
1234567
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
4.92
[125]
4000TOC Sensor 19
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4.1.3 Panel cutout dimensional information –
1/2DIN models
1/2DINModeltransmittersaredesignedwithanintegralrearcoverforstand-alonewallmount
installation.
Theunitmayalsobewallmountedusingtheintegralrearcover.Seeinstallationinstructionsin
Section4.1.4.
Belowarecut-outdimensionsrequiredbythe1/2DINmodelswhenmountedwithinaflatpanel
oronaflatenclosuredoor.Thissurfacemustbeflatandsmooth.Texturedorroughsurfacesare
notrecommendedandmaylimittheeffectivenessofthegasketsealprovided.
5.39
+0.02
–0.00
[137
+0.5
]
–0.0
5.39
+0.02
–0.00
[137
+0.5
]
–0.0
PANEL CUT-OUT
Optionalhardwareaccessoriesareavailablethatallowforpanel-orpipe-mount.
RefertoSection18“AccessoriesandSpareParts”fororderinginformation.
4000TOC Sensor 20
© 05 / 14 Mettler-Toledo Thornton 4000TOC Sensor
Printed in USA
4.1.4 Installation procedure – 1/2DIN models
General:
Orientthetransmittersothatthecablegripsfacedownward.
Wiringroutedthroughthecablegripsshallbesuitableforuseinwetlocations.
InordertoprovideIP65enclosureratings,allcableglandsmustbeinplace.Eachcable
glandmustbefilledusingacable,orsuitableCableGlandHoleSeal.
For Wall Mount:
Removerearcoverfromfronthousing.
Startbyunscrewingthefourscrewslocatedonthefaceofthetransmitter,ineachcorner.
Thisallowsthefrontcovertoswingawayfromtherearhousing.
Removethehinge-pinbysqueezingthepinfromeachend.
Thisallowsthefronthousingtoberemovedfromtherearhousing.
Mountrearhousingtowall.SecuremountingkittotheM300accordingtothesupplied
instructions.Attachtowallusingappropriatemountinghardwareforwallsurface.Be
sureitislevelandsecurelyfastenedandtheinstallationadherestoanyandallclearance
dimensionsrequiredfortransmitterserviceandmaintenance.Orientthetransmitterso
thatthecablegripsarefacingdownward.
Replacethefronthousingtotherearhousing.Securelytightentherear-coverscrewsto
ensurethatIP65enclosureenvironmentalratingisprovided.Theunitisreadytobe
wired.
For Pipe Mount:
Useonlymanufacturer-suppliedcomponentsforpipe-mountingtheM300transmitterand
installperthesuppliedinstructions.Seesection18fororderinginformation.
4.1.5 Assembly – 1/2DIN version
1
2
3
1:3Pg13.5cableglands
2: 2plasticplugs
3: 4screws
/