  • Hello! I am an AI chatbot trained to assist you with the Roche MODULAR ANALYTICS E-MODULE User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Roche Diagnostics Replacement Page
Operator’s Manual · Version 4.1.1 D-35
MODULAR ANALYTICS EVO SWA 17 Data alarms
Data alarms D module and P module
SD!
Screen G*
*displayed on Calibration > Status > Calib Trace
Alarm SD limit exceeded
Description During nonlinear or multipoint linear calibration, the SD value was larger than the
SD limit programmed on Utility > Application.
Cause The standard solution is not placed at a correct position.
An improper SD limit value is specified.
Remedy
1
Check calibrator positions on Calibration > Calib.
2
Touch Utility > Application > Calib to check the SD Limit.
3
Check preparation and expiration dates of calibrators and reagents. Recalibrate
the affected test.
4
Check the standard concentrations on Calibration > Install. For a calibration with
automatic standard dilution; check whether the ratio between concentration,
sample, diluent volume and diluted sample is correct on Utility > Application >
Others.
5
Resume operation. If alarm recurs, call technical support.
Sens
Screen Y*
* displayed on Calibration > Status > Calib Trace
Alarm Sensitivity error
Description Sensitivity is checked for linear (2 to 6 points), nonlinear, or isozyme-P calibration.
This alarm is indicated if the sensitivity value obtained in a calibration falls out of the
sensitivity limits specified on Utility > Application > Calib.
A sensitivity value is calculated from the measured absorbances ( ) and given
concentration values ( ) of the blank calibrator ( ) and the calibrator :
Revision: CAPA 270021730
Equation revised.
, where
= Std 2 for 2 Point calibrations and span calibrator for multipoint calibrations
Cause o The standard solution or reagent is not placed at a proper position.
o The reagent has been prepared improperly or has deteriorated.
An improper sensitivity limit is specified.
The result of the calibration is updated.
Abs
Conc
S
1
S
N
Abs S
N
Abs S
1

Conc S
N
Conc S
1

-------------------------------------------------------- -
S
N
For span calibration, the previous S1 Abs (linear) or previous mean absorbance (nonlinear)
of standard solution (1) is used for the sensitivity check.
/