AMSCO®600 Medium
Leak Isolation Kit Procedure
Instructions for Use
1 — Leak Isolation Kit Setup
2
11050355_A AMSCO®600 Medium Instructions for Use
Leak Isolation Kit Setup
Introduction
Installation
Setup Leak Function on Heise Gauge
Introduction
The purpose of the leak isolation kit is to help find a leak in a sterilizer by isolating the chamber drain, steam to
chamber, and transducer piping. The leak isolation tool uses the sterilizer’s vacuum source to pull a vacuum on the leak
isolation test kit’s piping, the chamber drain piping, the steam to chamber piping, and the transducer piping. Using the
ball valves on the tool, the piping sections can be isolated to help determine the source of the leak.
NOTE: The leak isolation tool does not pinpoint the exact component that is causing the leak; it just shows which piping
assembly has a leak. If the leak is not in the chamber drain, steam to chamber, and transducer piping, then it will not
show up while using this tool.
Figure 1-1 shows the leak isolation kit connected to a sterilizer.
Figure 1-1. Leak Isolation Kit Connected to Sterilizer
1 — Leak Isolation Kit Setup
3
Instructions for Use AMSCO®600 Medium 11050355_A
Installation
WARNING
PERSONAL INJURY
Wear proper PPE and follow LOTO procedure for the steam supply and all safety procedures.
1. Open door and insert safety pins.
2. Turn off the steam and electric to the sterilizer and use LOTO. Allow chamber to cool.
3. Install the correct Rubber Port Plug and washers onto the drain plug adapter shaft. Make sure the lock nut is
tightened all the way to the end of the threads, but do not over tighten.
4. Install the correct Rubber Port Plug and washers onto the Transducer plug adapter shaft. Make sure the lock nut
is tightened all the way to the end of the threads, but do not over tighten.
Figure 1-2. Transducer Plugs
5. Remove the chamber drain strainer.
6. Verify that the chamber drain port is clean and free of debris. Clean as necessary.
7. Install the drain plug adapter in the chamber drain port. Make sure the rubber sealing part of the plug is inside
the drain nipple.
IMPORTANT: Do not push plug adapter to the bottom of the drain.
8. Tighten the plug adapter until it is secure.
IMPORTANT: Do not overtighten!
9. Verify that the transducer port is clean and free of debris. Clean as necessary.
10. Install the transducer plug adapter in the transducer port. Make sure the rubber sealing part of the plug is inside
the transducer port nipple.
IMPORTANT: Do not push plug adapter to the bottom of the transducer pipe.
11. Tighten the plug adapter until it is secure.
IMPORTANT: Do not overtighten!
1 — Leak Isolation Kit Setup
4
11050355_A AMSCO®600 Medium Instructions for Use
Figure 1-3. Install Transducer Plug Adapter
IMPORTANT: The rubber plug must be inside the nipple of the drain and transducer port. Do NOT install the rubber plug
above or below the nipple.
12. Connect the Heise gauge to the Heise gauge port in the box (port without the valve).
Figure 1-4. Connect Heise Gauge
13. Connect tubing from the ports in the box to the drain plug adapter and the transducer plug adapter. One ball
valve on the box will be empty.
14. Inspect the tubing ends to ensure it’s not damaged. If necessary, cut off damaged end with tubing cutter.
Setup Leak Function on Heise Gauge
PTE-1 Heise Gauge
1. Make sure there is a 0-60 PSI ABS module with a ¼” tubing push connect installed on left side port of the Heise
gauge.
1 — Leak Isolation Kit Setup
5
Instructions for Use AMSCO®600 Medium 11050355_A
.
2. Power up the Heise gauge.
3. Press the TARE button until the pressure displays 0.00 or 0.000 PSI.
4. Press the SETUP button.
5. Press the Right Arrow button 10 times, or until LeakRate is flashing.
6. Press ENT.
7. Press Right Arrow button to select Decay.
8. Press ENT.
9. Typa 60.
10. Press ENT.
PTE-2 Heise Gauge
1. Make sure there is a 0-60 PSI ABS module with a ¼” tubing push connect installed on left side port of the Heise
gauge.
.
2. Power up the Heise gauge.
3. Press the TARE button until the pressure displays 0.00 or 0.000 PSI.
4. Press the ENTER button. Screen will display 0.00 or 0.000 PSI Tare.
2 — Leak Isolation Procedure
6
11050355_A AMSCO®600 Medium Instructions for Use
Leak Isolation Procedure
Test Piping in the BOX and Heise Gauge Piping
Test Drain and Transducer Sterilizer Piping and Test Fixture
Test Drain Piping
Test Transducer Piping
Test Steam Piping
Analyzing the Results Obtained from Previous Tests
Test Piping in the BOX and Heise Gauge Piping
1. Open all the ball valves on the test ports in the box except the one with no tubing connected.
2. Power up the sterilizer and go to the I/O screen in service mode.
3. Turn on VP1 vacuum pump and S3 valve. Pull a vacuum on the test fixture and piping to the limit of system.
4. Close the ball valves on the test ports in the box.
5. Turn off S3 valve and VP1 vacuum pump. Be sure to close S3 before turning off VP1.
6.
PTE-1 Heise Gauge: Press the LEAK button on the Heise gauge until TIMING is displayed to leak test the
piping in the box and the piping to Heise. A 60 second timer will start counting down on the Heise gauge. At
the end of the 60 seconds, the Heise will display a decay value. If the decay value is less than .2, continue
with the next step; if the decay is greater than .2, fix the leak in the test fixture prior to going to the next
step.
PTE-2 Heise Gauge: To run a Decay leak test:
a. Press the MENU button.
b. Select Modes and press the ENTER button.
c. Select Decay leak test and press the ENTER button.
d. Select 60 seconds and press the ENTER button.
A 60 second timer will start counting down on the Heise gauge. At the end of the 60 seconds, the Heise will
display a decay value. If the decay value is less than .2, continue with the next step; if the decay is greater
than .2, fix the leak in the test fixture prior to going to the next step.
Test Drain and Transducer Sterilizer Piping and Test Fixture
1. Open all the ball valves on the test ports in the box except the one with no tubing connected.
2. Turn on VP1 vacuum pump and S3 valve. Pull a vacuum on the test fixture and piping to the limit of system.
3. Turn off S3 valve and VP1 vacuum pump. Be sure to close S3 before turning off VP1.
4.
PTE-1 Heise Gauge: Press the LEAK button on the Heise gauge until TIMING is displayed to leak test the drain
piping, steam piping, and transducer piping. A 60 second timer will start counting down on the Heise gauge.
At the end of the 60 seconds, the Heise will display a decay value. Make a note of the total decay.
2 — Leak Isolation Procedure
7
Instructions for Use AMSCO®600 Medium 11050355_A
PTE-2 Heise Gauge: Press the ZERO button while the results from the previous decay leak test are displayed
to run another decay leak test. A 60 second timer will start counting down on the Heise gauge. At the end of
the 60 seconds, the Heise will display a decay value. Make a note of the total decay.
Test Drain Piping
1. Turn on VP1 vacuum pump and S3 valve. Pull a vacuum on the test fixture and piping to the limit of system.
2. Close the ball valve on the test ports in the box going to the transducer port.
3. Turn off S3 valve and VP1 vacuum pump. Be sure to close S3 before turning off VP1.
4.
PTE-1 Heise Gauge: Press the LEAK button on the Heise gauge until TIMING is displayed to leak test the drain
piping, steam piping, and transducer piping. A 60 second timer will start counting down on the Heise gauge.
At the end of the 60 seconds, the Heise will display a decay value. Make a note of the total decay.
PTE-2 Heise Gauge: Press the ZERO button while the results from the previous decay leak test are displayed
to run another decay leak test. A 60 second timer will start counting down on the Heise gauge. At the end of
the 60 seconds, the Heise will display a decay value. Make a note of the total decay.
Test Transducer Piping
1. Open all the ball valves on the test ports in the box except the one with no tubing connected.
2. Turn on VP1 vacuum pump and S3 valve. Pull a vacuum on the test fixture and piping to the limit of system.
3. Close the ball valve on the test ports in the box going to the drain port.
4. Turn off S3 valve and VP1 vacuum pump. Be sure to close S3 before turning off VP1.
5.
PTE-1 Heise Gauge: Press the LEAK button on the Heise gauge until TIMING is displayed to leak test the drain
piping, steam piping, and transducer piping. A 60 second timer will start counting down on the Heise gauge.
At the end of the 60 seconds, the Heise will display a decay value. Make a note of the total decay.
PTE-2 Heise Gauge: Press the ZERO button while the results from the previous decay leak test are displayed
to run another decay leak test. A 60 second timer will start counting down on the Heise gauge. At the end of
the 60 seconds, the Heise will display a decay value. Make a note of the total decay.
Test Steam Piping
1. Move the Plug Adapter located in the Transducer Port to the Steam Port.
2. Open all the ball valves on the test ports in the box except the one with no tubing connected.
3. Turn on VP1 vacuum pump and S3 valve. Pull a vacuum on the test fixture and piping to the limit of system.
4. Close the ball valve on the test ports in the box going to the drain port.
5. Turn off S3 valve and VP1 vacuum pump. Be sure to close S3 before turning off VP1.
6.
PTE-1 Heise Gauge: Press the LEAK button on the Heise gauge until TIMING is displayed to leak test the drain
piping, steam piping, and transducer piping. A 60 second timer will start counting down on the Heise gauge.
At the end of the 60 seconds, the Heise will display a decay value. Make a note of the total decay.
PTE-2 Heise Gauge: Press the ZERO button while the results from the previous decay leak test are displayed
to run another decay leak test. A 60 second timer will start counting down on the Heise gauge. At the end of
the 60 seconds, the Heise will display a decay value. Make a note of the total decay.
2 — Leak Isolation Procedure
8
11050355_A AMSCO®600 Medium Instructions for Use
Analyzing the Results Obtained from Previous Tests
1. Compare the decays from the drain port, the transducer port, and the steam piping. If done properly, the piping
section with the largest decay is most likely the source of the leak.
2. Repair leak and repeat the test to verify that the leak is fixed.
3. After the leak has been repaired, air break the system, disconnect and remove all testing components, and
return the sterilizer normal operating condition.
NOTE: When the plug adapters are not in use, loosen the wingnut to take the pressure off the rubber plugs and to
prevent the rubber plugs from getting deformed.