SICK TRANSIC151LP Operating instructions

Category
Measuring, testing & control
Type
Operating instructions
MMM
I
MMMOPERATING INSTRUCTIONS
OPERATING INSTRUCTIONS
TRANSIC151LP
Oxygen Measuring Device
Installation, Operation, Maintenance
Title Page
2
8015383/YSK2/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
Described Product
Product name: TRANSIC151LP
Document ID
Title: Operating Instructions
Part No.: 8015383
Version: V1-2
Release: 2016-07
Manufacturer
SICK AG
Erwin-Sick-Str. 1 · D-79183 Waldkirch · Germany
Phone: +49 7641 469-0
Fax: +49 7641 469-1149
Place of Manufacture
SICK AG
Nimburger Str. 11 · D-79276 Reute · Germany
Original Documents
The English edition 8015383 of this document is an original document from SICK AG.
SICK AG assumes no liability for the correctness of an unauthorized translation.
In case of doubt, contact SICK AG or your local representative.
Legal Information
Subject to change without notice.
© SICK AG. All rights reserved.
3
8015383/YSK2/V1-2/2016-07| SICK OPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
Warning Symbols
Warning Levels / Signal Words
DANGER
Risk or hazardous situation which will result in severe personal injury or death.
WARNING
Risk or hazardous situation which could result in severe personal injury or death.
CAUTION
Hazard or unsafe practice which could result in less severe or minor injuries.
NOTICE
Hazard which could result in property damage.
Information Symbols
Hazard (general)
Please consult the documentation.
Hazard through corrosive substances
Hazard in potentially explosive atmospheres
Hazard by explosive substances/mixtures
Hazards by noxious substances
Hazard by toxic substances
Hazard by laser radiation
Hazards through high temperature or hot
surfaces
Hazard for the environment/nature/organic
life
Information on product condition with
regard to protection against explosions
(general)
Important technical information for this
product
Important information on electric or
electronic functions
Nice to know
4
8015383/YSK2/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
Supplementary information
Link to information at another place
CONTENTS
5
8015383/YYJ7/V1-2/2016-07 | SICK OPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
1 Important Information ............................................................................10
1.1 Main operating information ..........................................................................10
1.1.1 Installation location ......................................................................10
1.2 Intended use .................................................................................................11
1.2.1 Purpose of the TRANSIC151LP ....................................................11
1.3 Responsibility of user....................................................................................12
2 Product Description.................................................................................13
2.1 Product identification....................................................................................13
2.2 Functional principle/measuring principle ....................................................14
2.2.1 TRANSIC151LP probe design.......................................................15
2.3 TRANSIC151LP variants ...............................................................................16
2.3.1 Variant for in-situ-measurement ..................................................16
2.3.2 Variant for extractive measurement ............................................17
2.3.3 Variant for ambient gas measurement........................................18
2.4 Explosion protection according to ATEX and IECEx .....................................19
3 Installation ................................................................................................21
3.1 Project planning ............................................................................................21
3.1.1 Chemical tolerance .......................................................................21
3.1.2 Temperature conditions ...............................................................21
3.1.3 Strong light sources near the oxygen measuring probe .............22
3.1.4 Pressure ........................................................................................22
3.1.5 Installation angle .........................................................................22
3.2 Information on installation in potentially explosive atmospheres ..............23
3.3 Installation.....................................................................................................24
3.3.1 Safety information: Installation....................................................24
3.4 Installation options .......................................................................................24
3.4.1 Process conditions for installation options .................................24
3.4.2 TRANSIC151LP installation - in-situ with flange..........................25
3.4.3 TRANSIC151LP installation - extractive.......................................27
3.4.4 TRANSIC151LP installation - ambient gas measurements .......30
3.5 Connections...................................................................................................31
3.5.1 Connecting the TRANSIC151LP transmitter................................32
3.5.2 Connecting the TSA151 voltage supply unit ...............................33
3.5.3 Connecting the analog and digital outputs..................................34
3.5.4 Connecting the power supply to the TSA151 voltage supply
unit ................................................................................................34
3.5.5 Gas connection (optional) ............................................................35
4 Operation ..................................................................................................36
4.1 Interfaces ......................................................................................................36
4.1.1 Control via keypad ........................................................................36
4.1.2 Maintenance interface .................................................................37
4.1.3 Analog output................................................................................37
4.1.4 NAMUR digital output ...................................................................37
CONTENTS
6
8015383/YYJ7/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
4.2 Settings using the keypad ............................................................................ 38
4.2.1 Short description: Settings using the keypad ............................. 38
4.2.2 Safety information on using the password: ................................ 38
4.3 Menu navigation without password ............................................................. 39
4.3.1 Oxygen statistics (O
2
) ................................................................... 39
4.3.2 Temperature statistics (T) ............................................................ 39
4.3.3 Calibration gas, actual value (CAL.C) .......................................... 39
4.3.4 Signal intensity(SIL)...................................................................... 40
4.3.5 Errors not deleted (ERR) .............................................................. 40
4.3.6 Entering the password (PAS) ....................................................... 40
4.4 Menu navigation with password .................................................................. 41
4.4.1 Process pressure (APP)................................................................ 41
4.4.2 H
2
0 content in the process gas (H2O) ........................................ 41
4.4.3 C0
2
content in process gas (CO2) ............................................... 41
4.4.4 One-point adjustment (CAL1) ...................................................... 41
4.4.5 Two-point adjustment (CAL2)....................................................... 41
4.4.6 Analog output (AOU) ..................................................................... 42
4.4.7 Resetting to the factory calibration (FAC) ................................... 42
4.4.8 Scaling the analog output (ASCL) ................................................ 42
4.4.9 Digital output (ALA) ...................................................................... 43
4.4.10 Resetting the TRANSIC151LP (rESE)........................................... 43
4.5 Maintenance interface ................................................................................. 43
5 Setting Ambient Parameters .................................................................44
5.1 Compensation of ambient parameters........................................................ 44
5.1.1 Pressure compensation ............................................................... 45
5.1.2 Background gas effects ............................................................... 45
6 Adjustment ............................................................................................... 48
6.1 Hardware preparations................................................................................. 48
6.1.1 Setting up the gas supply ............................................................ 49
6.1.2 Information on calibration gases................................................. 51
6.2 Calibration..................................................................................................... 51
6.2.1 Using bottled gas.......................................................................... 52
6.3 Adjustment.................................................................................................... 53
6.3.1 Adjustment options ...................................................................... 53
6.3.2 One-point adjustment using the keypad ..................................... 53
6.3.3 Two-point adjustment using the keypad ..................................... 54
CONTENTS
7
8015383/YYJ7/V1-2/2016-07 | SICK OPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
6.4 Adjustment for TRANSIC151LP ambient gas measurement.......................55
6.4.1 Setting up the gas supply .............................................................55
6.4.2 Calibration.....................................................................................56
6.4.2.1 Using ambient air ......................................................56
6.4.3 Using bottled gas ..........................................................................56
6.4.3.1 Calibration gases ......................................................57
6.4.4 Adjustment....................................................................................57
6.4.5 Adjustment options.......................................................................57
6.4.6 One-point adjustment using the keypad......................................58
6.5 Restoring factory calibration ........................................................................58
7 Maintenance.............................................................................................59
7.1 Field maintenance ........................................................................................59
7.1.1 Assembly and disassembly ..........................................................59
7.1.1.1 Safety information: Maintenance work ....................59
7.1.2 Draining condensation .................................................................60
7.1.3 Cleaning the optical components ................................................60
7.1.4 Checking the temperature probe .................................................62
7.1.5 TRANSIC151LP filters...................................................................62
7.1.6 Cleaning the filter .........................................................................63
7.2 Spare parts and accessories ........................................................................64
8 Troubleshooting .......................................................................................65
8.1 Function errors ..............................................................................................65
8.1.1 Self-test .........................................................................................65
8.1.2 Error control and error categories................................................65
8.1.3 TRANSIC151LP behavior when errors occur ...............................65
8.1.4 Error display ..................................................................................66
8.1.5 Error Table.....................................................................................66
8.2 LED display in the TSA151 ...........................................................................67
9 Shutdown ..................................................................................................68
9.1 Safety information: Shutting down...............................................................68
9.2 Preparations for shutdown ...........................................................................68
9.3 Switching the TRANSIC151LP off .................................................................68
9.4 Protecting a shutdown TRANSIC151LP........................................................68
9.5 Disposal .........................................................................................................68
9.6 Shipping the TRANSIC151LP to SICK...........................................................68
10 Specifications...........................................................................................69
10.1 Compliances..................................................................................................69
10.1.1 Electrical protection......................................................................69
10.2 Ex approvals ..................................................................................................69
CONTENTS
8
8015383/YYJ7/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
10.3 Technical Data .............................................................................................. 70
10.3.1 Dimensions and drill pattern ....................................................... 70
10.3.2 Measured value recording ........................................................... 76
10.3.3 Ambient conditions ...................................................................... 76
10.3.3.1 Interfaces .................................................................. 76
10.3.4 TRANSIC151LP electrical connection ......................................... 76
10.3.4.1 (Terminals) TRANSIC151LP connection data.......... 77
10.3.5 TRANSIC151LP safety-related technical characteristic data
(IECEx/ATEX)................................................................................. 77
10.3.5.1 TRANSIC151LP voltage supply input ....................... 77
10.3.5.2 TRANSIC151LP analog output ................................. 77
10.3.5.3 TRANSIC151LP switching output (NAMUR) ............. 77
10.3.6 TSA151 voltage supply input ....................................................... 78
10.3.6.1 (Terminals) TSA151 connection data ...................... 78
10.3.7 TSA151 safety-related technical characteristic data
(IECEx/ATEX)................................................................................. 78
10.3.7.1 TSA151 intrinsically safe output voltage................. 78
10.3.8 Connection lines........................................................................... 79
10.3.8.1 Maximum line length from TSA151 to
TRANSIC151LP ......................................................... 79
10.3.8.2 Maximum line length from AO and DO on TSA151
to the isolation amplifiers ........................................ 79
10.3.8.3 Maximum line length 24 V PELV power supply
unit to TSA151 .......................................................... 79
10.3.9 Dimensions and mechanics......................................................... 80
10.3.9.1 Pressure suitability ................................................... 80
10.3.9.2 Options and accessories .......................................... 80
11 Annex......................................................................................................... 81
11.1 Type code ...................................................................................................... 81
11.2 Humidity Conversion Table........................................................................... 82
11.3 Influence of background gases on oxygen measurement .......................... 83
11.4 Password....................................................................................................... 84
CONTENTS
9
8015383/YYJ7/V1-2/2016-07 | SICK OPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
10
8015383/YYJ7/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
1 IMPORTANT INFORMATION
1 Important Information
1.1 Main operating information
Observe the following safety precautions:
1.1.1 Installation location
The TRANSIC151LP can be used indoors and outdoors.
Altitude: Up to 2000 m above sea level
Max. air humidity: 100% r.h., non-condensing
CAUTION: TRANSIC151LP is a protection class 1 laser product.
Normal handling and operation of the TRANSIC151LP is eye-safe, because laser
radiation is collimated and maintained inside the probe, as shown schematically on
page 15.
!
Do not place objects with reflecting surfaces (tools, etc.) directly onto the probe
when the TRANSIC151LP is in operation to prevent the laser beam being reflected
out of the probe.
NOTICE: Protection against electrostatic discharges
SICK products are adequately protected against ESD when used as intended.
Observe the generally valid ESD regulations to avoid damaging the TRANSIC151LP
through electrostatic discharges when touching parts within the enclosure.
WARNING: Risk of fire through strong oxidation processes
High O
2
concentrations have strong oxidizing properties. These increase the risk of
combustion and can trigger violent reactions with inflammable substances.
Before installation, check the TRANSIC151LP is suitable for your application with
regard to all ambient conditions.
11
8015383/YYJ7/V1-2/2016-07 | SICK OPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
IMPORTANT INFORMATION 1
1.2 Intended use
1.2.1 Purpose of the TRANSIC151LP
The TRANSIC151LP is a stationary oxygen transmitter and serves continuous
measurement of oxygen in the industrial sector.
There are 3 variants of the TRANSIC151LP:
1. In-situ measurement
2. Extractive measurement
3. Measurement in ambient air
The TRANSIC151LP is certified for a process-side operating range of 800 mbar abs. to
1400 mbar abs. Usage with deviating pressures can void the Ex certification.
The TRANSIC151LP must be operated within the specifications described in the Section
Technical Data (see “Technical Data”, page 70). The Ex certification becomes void when
the TRANSIC151LP is operated outside the specifications.
The TRANSIC151LP has not been evaluated regarding the safety function as laid out in
ATEX 2014/34/EC, Annex II, Section 1.5.
Operation in potentially explosive atmospheres
The TRANSIC151LP is suitable for the following zones
In accordance with ATEX (EN60079-10) and
In accordance with IECEx (IEC60079-10)
Gas:
Measuring probe: Category 1G EPL Ga (Zone 0)
Transmitter / voltage supply unit: Category 2G EPL Gb (Zone 1)
Dust:
Transmitter/voltage supply unit/measuring probe: Category 2D EPL Db (Zone 21)
The TRANSIC151LP may be used in zones in which inflammable or explosive gases in
groups I, IIA and IIB or inflammable dust in groups IIIA, IIIB and IIIC are present.
Temperature class:
Gas: T4 (max. surface temperature 135°C)
Dust: T85°C
Approvals
Transmitter:
–Gas:
outside process: II 2G Ex ib IIB T4 Gb
within process: II 1G Ex op is IIB T4 Ga
–Dust:
Outside process: II 2D Ex ib tb IIIC T85°C Db
within process: II 2D Ex ib tb IIIC T85°C Db
Voltage supply unit TSA151:
–Gas:
II 2G Ex e mb [ib] IIB T4 Gb
–Dust:
II 2D Ex tb [ib] IIIC Db
12
8015383/YYJ7/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
1 IMPORTANT INFORMATION
1.3 Responsibility of user
Designated users
All operators of the TRANSIC151LP should be specifically trained on this device,
knowledgable of relevant regulations, and able to assess potential hazards related to its
operation.
Correct use
Use the device only as described in these Operating Instructions.
The manufacturer bears no responsibility for any other use.
Carry out the specified maintenance work.
Do not remove, add or change any components in or on the device unless such changes
are officially allowed and specified by the manufacturer. Failure to observe these
precautions could result in
voiding the manufacturer's warranty
causing the TRANSIC151LP to could become dangerous
Special operating conditions
TSA151 Power Supply: Protective Extra Low Voltage (PELV) power supply is required. The
24V inlet of the TSA151 is specified with a fault voltage of Um=60V.
The partition wall thickness of the temperature probe is 0.2<t<1 mm. In order to
maintain the zone separation, the temperature probe may not be exposed to corrosive
atmosphere that can endanger zone separation.
The service interface with the SICK service interface cable (Part No. 2066710) may only
be used outside the Ex area.
Special local conditions
Follow all local laws, regulations and company policies applicable at the respective
installation location.
Retention of documents
These Operating Instructions:
Must be available for reference.
Must be conveyed to new owners.
Keep passwords in a separate, safe place and secure against unauthorized use.
This Manual presumes that the TRANSIC151LP has been delivered as specified
during project planning (i.e., based on the SICK application questionnaire) and the
relevant delivery state of the TRANSIC151LP.
Contact SICK Customer Service if you are not sure whether the TRANSIC151LP
corresponds to the state defined during project planning or to the delivered
system documentation. Please contact SICK Customer Service.
13
8015383/YYJ7/V1-2/2016-07 | SICK OPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
PRODUCT DESCRIPTION 2
2 Product Description
2.1 Product identification
Type plates
The type plate is located on the outer left hand side of the enclosure.
Fig. 1: Example - type plate for the TRANSIC151LP
Table 1: Type code:
Product name: TRANSIC151LP
Manufacturer:
SICK AG
Erwin-Sick-Str. 1 · D-79183 Waldkirch · Germany
CAUTION: Observe marking for special operating conditions
Marking “X” on the TRANSIC151LP type plate draws attention to the special operating
conditions to be complied with:
The partition wall thickness of the temperature probe is 0.2<t<1 mm. In order to
maintain the zone separation, the temperature probe may not be exposed to
corrosive atmosphere that can endanger zone separation.
Check the condition of the temperature probe during every regular maintenance.
The service interface with the SICK service interface cable (Part No. 2066710) may
only be used outside the Ex area.
Code TRANSIC151LP version
A With flange adapter <0.5 bar for process measurements
B For ambient gas measurement, wall fitting
C With sample gas cell for extractive measurement PN10
F With flange adapter for process measurement PN10
G With welding adapter for process measurement PN10
H With clamping flange 3"/DN65 for process measurement PN10
Complete explanation on type codes in Annex, see “Type code”, page 81.
14
8015383/YYJ7/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
2 PRODUCT DESCRIPTION
Fig. 2: Example - type plate for voltage supply unit TSA151
2.2 Functional principle/measuring principle
The TRANSIC151LP functions using light absorption of a tunable diode laser (Tunable
Diode Laser Absorption Spectroscopy TDLAS). The gas concentration is measured using
the damping of a laser beam sent from a tunable diode laser source in the gas sample. For
oxygen measuring, the laser beam wavelength is set to match one of the characteristic
absorption lines of oxygen in the wavelength range of around 760 nm in the near infrared
range (NIR) of the electromagnetic spectrum. During measurement, the diode laser beam
wavelength is continuously modulated to scan across one of the oxygen absorption lines.
This generates a periodic signal in a photodetector, the amplitude of which is proportional
to the amount of oxygen in the laser beam path.
CAUTION: Observe marking for special operating conditions
Special operating conditions: Marking “X” on the TSA151 type plate means the 24V
input of the TSA151 is specified with a fault voltage of Um=60V and therefore an
appropriate PELV power supply unit is mandatory.
Observe “Special local conditions”, page 12.
The dust load in the process does not falsify the measured O
2
value.
The TRANSIC151LP outputs a maintenance signal when the dust load is too high.
15
8015383/YYJ7/V1-2/2016-07 | SICK OPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
PRODUCT DESCRIPTION 2
2.2.1 TRANSIC151LP probe design
The sensor is designed as a probe for direct fitting at the measuring location. The diode
laser light source, and the photodetector that measures the light, are located in a
transmitter behind a protective window. The light is directed onto the photodetector using a
focusing mirror at the far end of the probe.
Fig. 3: Probe design and laser beam guidance within the probe
Further information on components having contact with the sample gas probe, see “Pres-
sure suitability”, page 80.
Light source
Mirror
Photodetector
16
8015383/YYJ7/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
2 PRODUCT DESCRIPTION
2.3 TRANSIC151LP variants
2.3.1 Variant for in-situ-measurement
Fig. 4: Flange mounted TRANSIC151LP-A/-F/-G/-H
Environment Measuring environment (process side)
Zone: 1/21
0
2
: 2 ... 21 vol%
T: -20 ... +60 °C
p: 0.8 ... 1.1 bar abs.
Zone: 0/21
0
2
: 0 ... 21% vol
T: -20 ... +80 °C
p: 0.8 ... 1.4 bar abs.
PS: 10 bar
V: 0.28 L
DN: 50 ... 65 mm (2" ... 3")
Non-ex zone
TSA151 voltage supply unit
Connection TSA151 with TRAN-
SIC151LP
Ex ib M20 x 1.5
3 x 2 x 0.5 mm
2
twisted pair, shielded
t: 7 ... 12 mm
L: 0.5 ... 5 m (longer lines not allowed)
Shield laid on both sides of the EMC
cable gland on the enclosure.
Analog/digital
output
Ex ib M20 x 1.5
2 x 2 x 0.5 mm
2
twisted pair,
shielded
t: 7 ... 12 mm
Voltage:
DC 24 V PELV
Um: 60V
Ex e M20 x 1.5
2 x 1.5 mm
2
2 x 1.0 mm
2
Depending on
line length.
t: 5 ... 9 mm
TRANSIC151LP
DC 24 V
PELV
WARNING: Ex certification becomes void for unallowed wiring
Comply with wiring requirements, see “Connections”, page 31.
17
8015383/YYJ7/V1-2/2016-07 | SICK OPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
PRODUCT DESCRIPTION 2
2.3.2 Variant for extractive measurement
Fig. 5: TRANSIC151LP-C with sample gas cell and wall bracket
Gas feed line from
process
Sample gas cell
Outlet (back to
process)
Environment
Zone: 1/21
0
2
: 2 ... 21% vol
T: -20 ... +60 °C
p: 0.8 ... 1.1 bar abs.
Measuring environment (in sample gas cell)
Zone: 0/21
0
2
: 0 ... 21% vol
T: -20 ... +80 °C
p: 0.8 ... 1.4 bar abs.
PS: 10 bar
V: 0.28 L
DN: 50 ... 65 mm (2" ... 3")
Non-ex zone
TSA151 voltage supply unit
Connection TSA151 with
TRANSIC151LP
Ex ib M20 x 1.5
3 x 2 x 0.5 mm
2
twisted pair,
shielded
t: 7 ... 12 mm
L: 0.5 ... 5 m (longer lines not
allowed)
Shield laid on both sides of the EMC
cable gland on the enclosure.
Analog/digital
output
Ex ib M20 x 1.5
2 x 2 x 0.5 mm
2
twisted pair,
shielded
t: 7 ... 12 mm
Voltage:
DC 24 V PELV
Um: 60V
Ex e M20 x 1.5
2 x 1.5 mm
2
2 x 1.0 mm
2
Depending on
line length.
t: 5 ... 9 mm
TRANSIC151LP
DC 24 V
PELV
WARNING: Ex certification becomes void for unallowed wiring
Comply with wiring requirements, see “Connections”, page 31.
18
8015383/YYJ7/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
2 PRODUCT DESCRIPTION
2.3.3 Variant for ambient gas measurement
Fig. 6: TRANSIC151LP-B for ambient gas measurement and wall bracket
When configuring ambient gas measurements, the complete TRANSIC151LP must be
exposed to the O
2
concentration to be measured.
The TRANSIC151LP variant for ambient measurements measures oxygen concentrations
from 2 ... 21 vol% O
2
.
The measuring function is lost for oxygen concentrations under 2 vol%.
Measuring environment
TRANSIC151LP complete in differing
O
2
concentrations
Zone: 1/21
0
2
: 2 ... 21 vol%
T: -20 ... +60 °C
p: 0.8 ... 1.1 bar abs.
Analog/digital
output
Ex ib M20 x 1.5
2 x 2 x 0.5 mm
2
twisted pair,
shielded
t: 7 ... 12 mm
Non-ex zone
TSA151 voltage supply unit
Connection TSA151 with
TRANSIC151LP
Ex ib M20 x 1.5
3 x 2 x 0.5 mm
2
twisted pair,
shielded
t: 7 ... 12 mm
L: 0.5 ... 5 m (longer lines not
allowed)
Shield laid on both sides of the EMC
cable gland on the enclosure.
Voltage:
DC 24 V PELV
Um: 60V
Ex e M20 x 1.5
2 x 1.5 mm
2
2 x 1.0 mm
2
Depending on
line length.
t: 5 ... 9 mm
TRANSIC151LP
DC 24 V
PELV
NOTICE: The O
2
concentration in the transmitter is part of the O
2
measurement
WARNING: Ex certification becomes void for unallowed wiring
Comply with wiring requirements, see “Connections”, page 31.
19
8015383/YYJ7/V1-2/2016-07 | SICK OPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
PRODUCT DESCRIPTION 2
2.4 Explosion protection according to ATEX and IECEx
The TRANSIC151LP is certified for a process-side operating range of 800 mbar abs. to
1400 mbar abs. Usage with deviating pressures voids the Ex certification.
The TRANSIC151LP must be operated within the specifications described in the Section
Technical Data, see “Technical Data”, page 70. Operating the TRANSIC151LP outside
the specifications voids the Ex certification.
The TRANSIC151LP has not been evaluated regarding the safety function as laid out in
ATEX 2014/34EC, Annex II, Section 1.5.
Operation in potentially explosive atmospheres
The TRANSIC151LP is suitable for the following zones
In accordance with ATEX (EN60079-10) and IECEx (IEC60079-10)
Gas:
Measuring probe: Category 1G EPL Ga (Zone 0)
Transmitter / voltage supply unit: Category 2G EPL Gb (Zone 1)
Dust:
Transmitter/voltage supply unit/measuring probe: Category 2D EPL Db (Zone 21)
The TRANSIC151LP may be used in zones in which inflammable or explosive gases in
groups I, IIA and IIB or inflammable dust in groups IIIA, IIIB and IIIC are present.
Temperature class: Gas: T4 (max. surface temperature 135°C); dust: T85°C
Transmitter:
Transmitter
–Gas
outside the process: II 2G Ex ib IIB T4 Gb
within the process: II 1G Ex op is IIB T4 Ga
–Dust
outside the process: II 2D Ex ib tb IIIC T85°C Db
within the process: II 2D Ex ib tb IIIC T85°C Db
TSA151 voltage supply unit
–Gas
II 2G Ex e mb [ib] IIB T4 Gb
–Dust
II 2D Ex tb [ib] IIIC Db
The Ex identification is on the type plate.
Location of Ex relevant subassemblies: see “Flange mounted TRANSIC151LP-A/-F/-G/-
H”, page 16 and/or “TRANSIC151LP-C with sample gas cell and wall bracket”, page 17
“TRANSIC151LP-B for ambient gas measurement and wall bracket”, page 18.
Do not remove, add or change any components in or on the device unless such changes
are officially allowed and specified by the manufacturer. Otherwise the approval for use
in potentially explosive atmospheres becomes void.
Adhere to the prescribed maintenance intervals, see “Safety information: Maintenance
work”, page 59.
The TRANSIC151LP may only be used in zones corresponding to the Ex identification
(see type plate) of the transmitter and the power supply unit. Check the requirements of
the installation location with the Ex certification of the device.
The TRANSIC151LP may only be installed by persons trained in the standards used
(e.g., EN/IEC 60079-14).
The TRANSIC151LP may not be modified. Any modification to the device voids the Ex
certification.
20
8015383/YYJ7/V1-2/2016-07 | SICKOPERATING INSTRUCTIONS | TRANSIC151LP
Subject to change without notice
2 PRODUCT DESCRIPTION
CAUTION: Observe marking for special operating conditions
Special operating conditions: Marking “X” on the TRANSIC151LP type plate means the
wall thickness of the temperature probe is 0.2<t<1 mm. In order to maintain the zone
separation, the temperature probe may not be exposed to corrosive atmosphere which
can endanger zone separation.
Check the condition of the temperature probe during every regular maintenance.
CAUTION: Observe marking for special operating conditions
Special operating conditions: Marking “X” on the TSA151 type plate means a PELV
power supply unit is mandatory for voltage supply to the TSA151.
See also Section “Special local conditions”, page 12.
TSA151 Power Supply: Protective Extra Low Voltage (PELV) power supply is required.
The 24V inlet of the TSA151 is specified with a fault voltage of Um=60V.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88

SICK TRANSIC151LP Operating instructions

Category
Measuring, testing & control
Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI