Installation of Transfer-Line Degasser 9000-1549
Figure 1
The unit can be vertical or horizontal
in orientation. Flow can enter at the
top or bottom of the unit.
A vacuum source (minimum 13.3 kPa
absolute) such as the Systec SST
pump (PN: 9000-1584 or 9000-1585)
is required.
● 1. Connect a 0.062 inch/1.57MM
ID vacuum tube to vacuum con-
nection. (See Figure 1).
● 2. Two 0.15 inch/3.8MM holes are
provided for mounting.
(See Figure 1).
● 3. Connect your in/outlet fluid line
into the 1/4-28 port on the debub-
bler chamber. (See Figure 2).
Maximum Operating Pressure:
200 kPa (30 psi) @ 25ºC
Degassing Performance:
<4 ppm dissolved O2 @ 5 mL/min
Compatible with aqueous salts, ac-
ids, and bases without detergents.
For other solutions containing organ-
ics contact IDEX H&S.
Important Note: Degassing tail is
flexible and can be maneuvered
around components allowing easy
retrofitting to installed instruments.
However, over twisting or bending of
the degassing tail can cause a kink in
the tube which can lead to decreased
performance or failure.
Technical Support
Call +1-707-588-2038
VACUUM CONNECTION
BARB FITTING, (0.062 IN/1.57MM)
ID TUBING FLUID IN/OUTLET
UNF 1/4-28 THREADED
FLAT-BOTTOM PORT
FLUID IN/OUTLET
UNF 1/4-28
THREADED FLAT-
BOTTOM PORT
DEGASSING
TAIL
Figure 2
MOUNTING
HOLES
0.15 INCH
3.8 MM Y-BLOCK
VACUUM
PORT
2322046A
03/28/2011