ABB 4234 600 User manual

Type
User manual

This manual is also suitable for

ABB MEASUREMENT & ANALYTICS | INSTRUCTION MANUAL | IM/4234600 REV. E
4234 600 and 4234 601
Katharometer power supply units
The constant-current supply for all
your katharometer applications
Measurement made easy
Introduction
This instruction manual describes the installation
and operation of the 4234 600 and 4234 601 power
supply units (PSUs), designed to provide a stabilized
(optional 500, 400, 350, 250 or 180 mA) DC supply
for operation of a single katharometer unit and an
auxiliary 10 V DC supply for katharometer
temperature control.
For more information
The Data Sheet for the 4234 600 and 4234 601 PSUs
is available for free download from:
www.abb.com/analytical
(see link and reference number below)
or by scanning this code:
Search for or click on
Data Sheet

Katharometer power supply units

Katharometer power
supply units
2 4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
Electrical safety
This equipment complies with the requirements of CEI/IEC


instrument is used in a manner NOT specified by the Company,
the protection provided by the instrument may be impaired.
Symbols
One or more of the following symbols may appear on the
equipment labeling:
Warning – Refer to the manual for instructions
Caution – Risk of electric shock
Protective earth (ground) terminal
Earth (ground) terminal
Direct current supply only
Alternating current supply only
Both direct and alternating current supply
The equipment is protected through double insulation
Information in this manual is intended only to assist our
customers in the efficient operation of our equipment. Use of
this manual for any other purpose is specifically prohibited and
its contents are not to be reproduced in full or part without
prior approval of the Technical Publications Department.
Health and safety
To ensure that our products are safe and without risk to health,
the following points must be noted:
The relevant sections of these instructions must be read
carefully before proceeding.
Warning labels on containers and packages must be
observed.
Installation, operation, maintenance and servicing must
only be carried out by suitably trained personnel and in
accordance with the information given.
Normal safety precautions must be taken to avoid the
possibility of an accident occurring when operating in
conditions of high pressure and/or temperature.
Safety advice concerning the use of the equipment described in
this manual or any relevant hazard data sheets (where
applicable) may be obtained from the Company address on the
back cover, together with servicing and spares information.
4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
3
Contents
1 Document symbols ......................... 4
2 Description ................................ 4
3 Specification ................................5
4 Mechanical installation.......................6
5 Electrical installation ........................7
Interconnection diagrams and cable requirements ...7
Katharometer connections ....................7
Temperature control circuit ...................7
Accessing the PSU terminals ....................... 8
Electrical connections ............................. 8
6 Servicing....................................9
Fuse replacement ................................. 9
4 4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
1 Document symbols
Symbols that appear in this document are explained below:
WARNING – BODILY INJURY
This symbol in conjunction with the signal word
WARNING’ indicates a potential electrical hazard.
Failure to observe this safety information will result in
death or severe injury.
CAUTION – MINOR INJURIES
This symbol in conjunction with the signal word
‘CAUTION’ indicates a potentially dangerous
situation. Failure to observe this safety information
may result in minor or moderate injury. The symbol
may also be used for property damage warnings.
IMPORTANT (NOTE)
This symbol indicates operator tips, particularly
useful information or important information about
the product or its further uses. The signal word
‘IMPORTANT (NOTE)’ does not indicate a dangerous
or harmful situation.
2 Description
The PSU is available in two separate versions defined in
Specification on page 5.
The unit is housed in a metal case fitted with mounting lugs.
Cable gland entries are provided at opposite ends of the case
for supply voltage input and output cables respectively.
A printed circuit board assembly (see Figure 1) contains the
circuit components. The supply voltage input is connected to
terminal block (TB1) located on the pcb, adjacent to the fuses
F2 and F3. The constant current output terminal block (TB2) is
located on the gland assembly at the opposite end of the
assembly. The auxiliary 10 V output terminal block (AUX 10V) is
located on the PCB near to TB1. The pcb is mounted on a steel
chassis and secured by six screws. The cover is perforated to
allow ventilation for cooling.
IMPORTANT (NOTE)
The PSU has NO serviceable parts.
Two identical fuses (L & N) are mounted within the unit on the
mains input. A third fuse is connected in the DC line circuit of
the transformer. A fourth fuse is connected in the auxiliary
10 V DC line. The L & N fuses are a high breaking capacity type
and must be replaced only with identical types. See
Specification on page 5 for rating.
TB1
T1
LNE
F2 F3
6
8
10
230 V
115 V
TB2
F1
B
A
C
X
D
E
M5 earth stud
Output
current
links
Cable gland for
DC output to
katherometer
Cable gland for
10 V output to
katherometer
Cable gland for
AC mains input
Mains input
terminal block
DC output
terminal block
Fuses
∅ 5 to 9 mm
cables
∅ 5 to 9 mm
cable
F4
AUX 10V
+
Red
Black
+
WARNING. HAZARDOUS VOLTAGES
There are no serviceable parts in
this unit.
If faulty, return to the manufacturer
or seek the services of a qualified
engineer.
Ensure the mains supply is
switched off and disconnected
before removing the cover.
Figure 1 Location of internal components
4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
5
3 Specification
4234-600 and 4234-601
Mains supply
 
 
Permissible supply
Voltage variation ±15 %
 
Power consumption
Katharometer 3.5 W max.
Katharometer with temperature control 13 W max.
Fuse ratings
L & N 250mA (T) (20 x 5 mm) 250 VA,
1500A h.b.c. cartridge type
DC line 400mA (F) (250V AC)
Auxiliary 10V DC line 1A (F) (250V AC)
DC output
Optional, stabilized 500, 400, 350, 250 or 180 mA
Load


Output
Output
180, 250, 350, 400 or 500 mA DC (constant current)
Output regulation
Within ±0.8 % for:
a) Load variation ±15 %
b) Supply variation ±6 %
c) Ambient temperature ±10 °C (50 °F)
Ripple

Output stability
Within ±0.7 % of initial setting over a period of one month
with load resistance, supply voltage and ambient
temperature at nominal stated values
Auxiliary output
10 V DC nominal at 1 A for katharometer temperature control
Environmental
Protection
IP30 (NEMA 1)
Ambient temperature range
20 to 55 °C (–4 to 131 °F)
Mounting centers
140 x 110 mm (5.5 x 4.33 in)
Overall dimensions (height, width, depth)
160 x 170 x 111 mm (6.3 x 6.7 x 4.4 in.)
Weight
2.12 kg (4.7 lb) approx.
F
6 4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
4 Mechanical installation
The PSU must be mounted in a clean, dry, well ventilated
position (on a wall or panel) in a safe area by means of the four
fixing lugs. Figure 2 gives dimensions and fixing details for the
unit.
Before commencing installation, ensure that the unit supplied
is suitable for the supply voltage available – see Specification
on page 5 and Figure 3.
Dimensions in mm (in)
140 (5.5)
between centers
All holes 7.0 diameter
20
(0.8)
20
(0.8)
60 (2.4)
dim
Ensure a clearance of at least 100 mm (4 in)
at both ends of the unit for access to the
cable glands and to minimize cable bends.
Mains cable
entry
Output to temperature control
circuit in katharometer only
Output to
katharometer
An adaptor kit (part number 4234 519) is available for mounting the
4234 600/601 PSU in place of the 004235 000/01/02 PSU.
All dimensions are nominal
mm (in) unless otherwise stated.
30
(1.2)
30
(1.2)
30
(1.2)
110 (4.3)
between centers
170 (6.7)
83 (3.3)
3 (0.1)
111 (4.4)
120 (4.7)
160 (6.3)
Figure 2 PSU dimensions
ABB Limited
Oldends Lane, Stonehouse,
Glos. England GL10 3TA
Code 4234/600XX
Serial No
Voltage 230 V AC
Wa tts
Year Manuf..
Manufactured by:
N4006
Figure 3 Typical identification labels and locations
4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
7
5 Electrical installation
Interconnection diagrams and cable
requirements
See Figure 4 for interconnections for typical gas analysis
systems.
Katharometer connections
Cable connecting the constant current DC supply from the PSU
to the katharometer should be selected so that the maximum

length between the katharometer and its associated PSU.
Note that the cable gland accepts a cable between 5 and 9 mm
in diameter.
Temperature control circuit
The cable connecting the auxiliary 10 V DC supply to the
katharometer temperature control circuit must have cable
conductance capable of passing 1 amp without undue voltage
drop or excessive heating.
Note also that the cable gland will accept a cable between 5 and
9 mm in diameter.
or
1–+2
C501
4
1+
8 –
3
2+
+
Power supply unit
4234600/601
Katharometer types
6515
6517
6518
6520
6522
6524
7 +
+
mA o/p
10 V o/p
For temperature-controlled
katharometer block type
6518, 6520, 6524
Mains in
4234 600 | 230 V AC
or
4234 601 | 115 V AC
4689 display
TBB
+12
TBB
10–
or
28–+29
C351
or
1–+2
Datum
L150
Commander series
Figure 4 Typical interconnecting diagram showing 4234 600/601 PSU, katharometer and displays
8 4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
…5 Electrical installation
Accessing the PSU terminals
WARNING. HAZARDOUS VOLTAGES
If the PSU is cubicle mounted or has its cover removed
for test purposes, AC mains supply voltage is exposed
at certain locations within the assembly when
supplies are connected. Thus, there is a risk of
electrocution.
All normal precautions must be taken to avoid the risk
of accidental electrical shock during installation and
maintenance of the equipment. Ensure that the
supply cable is disconnected at its source before
touching any electrical connections.
CAUTION – PROPERTY DAMAGE
Ensure that the correct version of the unit has been
supplied for the supply voltage available. A 115 V unit
cannot be used with a 230 V supply or vice versa.
Check that the current output meets the
katharometer input requirements (see katharometer
data label) – see Figure 7. If in doubt contact ABB
Limited.
Remove the power supply unit cover to obtain access to the
separate labeled terminal blocks – see Figure 5.
To gain access, remove the four
screws and lift off the cover
Figure 5 PSU cover removal
Electrical connections
The mains terminal block, TB1, is situated next to the
transformer, T1, and fuses, F2 and F3.
Connect the AC mains input cable as in Figure 6.
IMPORTANT (NOTE)
Earth bonding to be made via M5 earth stud.
Connect the DC output and 10 V cables as in Figure 7.
TB1
T1
F2 F3
6
8
10
230 V
115 V
L N E
115 V AC
or
230 V AC
supply
See CAUTION opposite
Tighten cable gland
M5 earth stud
Figure 6 AC mains supply connections
To katharometer
(see manual for
termination details)
To katharometer temperature
control circuit
(see manual for
termination details)
Tighten cable gland
TB2
+
F1
B
A
C
X
D
E
DC output
terminal block
Cable clamp
Links
A to X
B to X
C to X
D to X
E to X
Output current (mA)
500
400
350
250
180
Current output link
AUX 10V
+
Red
Black
Figure 7 Outputs connections
4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
9
6 Servicing
Fuse replacement
There are no serviceable components in the PSU; the only
servicing that can be performed is fuse replacement – see
Figure 8. The fuses must be replaced only with those stated in
the specification. If a fault occurs, return the PSU to the nearest
ABB Service Centre or to the factory for repair.
TB1
T1
L N E
F2 F3
6
8
10
230 V
115 V
TB2
F1
B
A
C
X
D
E
F4
AUX 10V
+
+
Switch off and disconnect the
mains power supply
Remove the four securing screws
and lift off the cover
Replace fuse(s)
Refit cover, connect to mains power supply and switch on.
1
2
3
4
IMPORTANT (NOTE)
Suitable fuses may be obtained from ABB using the
following part numbers:
F1 0231555 (400 mA(F) cartridge).
F2 and F3 0231577 (250 mA (T) 1500 A hbc cartridge)
F4 0231599 1 A(F) 250 V
Figure 8 Replacing the fuses
10 4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
Notes
4234 600 AND 4234 601 | KATHAROMETER POWER SUPPLY UNITS | IM/4234600 REV. E
11
IM/4234600 REV. E 10.2018
ABB Limited
Measurement & Analytics
Oldends Lane, Stonehouse
Gloucestershire, GL10 3TA
UK
Tel: +44 (0)1453 826661
Fax: +44 (0)1453 829671
ABB Inc.
Measurement & Analytics
125 E. County Line Road
Warminster, PA 18974
USA
Tel: +1 215 674 6000
Fax: +1 215 674 7183
abb.com/measurement
We reserve the right to make technical changes or modify the contents of this document
without prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.
We reserve all rights in this document and in the subject matter and illustrations contained
therein. Any reproduction, disclosure to third parties or utilization of its contents – in
whole or in parts – is forbidden without prior written consent of ABB.
Copyright © ABB 2018
All rights reserved
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12

ABB 4234 600 User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI