ABB 4234 600 User manual

Type
User manual

This manual is also suitable for

ABB 4234 600 is a Power Supply Unit designed to provide a stabilized DC supply for operation of a single katharometer unit. It also provides an auxiliary 10V DC supply for katharometer temperature control, making it suitable for various gas analysis systems. The device ensures precise and stable output voltage, with regulation within ±0.8% for load, supply, and temperature variations. Its compact design and easy installation process make it convenient to integrate into existing setups.

ABB 4234 600 is a Power Supply Unit designed to provide a stabilized DC supply for operation of a single katharometer unit. It also provides an auxiliary 10V DC supply for katharometer temperature control, making it suitable for various gas analysis systems. The device ensures precise and stable output voltage, with regulation within ±0.8% for load, supply, and temperature variations. Its compact design and easy installation process make it convenient to integrate into existing setups.

4234 600 Series
Power Supply Unit
Instruction Manual
Models 4234 600 & 4234 601
ABB LIMITED
The Company
ABB Ltd. is an established world force in the design and manufacture of instrumentation for
industrial process control, flow measurement, gas and liquid analysis and environmental
applications.
As a part of ABB, a world leader in process automation technology, we offer customers
application expertise, service and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and
unrivalled service and support.
The quality, accuracy and performance of the Company’s products result from over 100 years
experience, combined with a continuous program of innovative design and development to
incorporate the latest technology.
The NAMAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants
operated by the Company, and is indicative of ABB’s dedication to quality
and accuracy.
Use of Instructions
Warning.
An instruction that draws attention to the risk of injury or
death.
Caution.
An instruction that draws attention to the risk of damage to
the product, process or surroundings.
Note.
Clarification of an instruction or additional information.
Information.
Further reference for more detailed information or
technical details.
Although
Warning hazards are related to personal injury, and Caution hazards are associated with equipment or property damage,
it must be understood that operation of damaged equipment could, under certain operational conditions, result in degraded
process system performance leading to personal injury or death. Therefore, comply fully with all
Warning and Caution notices.
Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual
for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of
Marketing Communications Department, ABB Ltd.
Health and Safety
To ensure that our products are safe and without risk to health, the following points must be noted:
1. The relevant sections of these instructions must be read carefully before proceeding.
2. Warning labels on containers and packages must be observed.
3. Installation, operation, maintenance and servicing must only be carried out by suitably trained personnel and in accordance with the
information given.
4. Normal safety precautions must be taken to avoid the possibility of an accident occurring when operating in conditions of high pressure
and/or temperature.
5. Chemicals must be stored away from heat, protected from temperature extremes and powders kept dry. Normal safe handling procedures
must be used.
6. When disposing of chemicals ensure that no two chemicals are mixed.
Safety advice concerning the use of the equipment described in this manual or any relevant hazard data sheets (where applicable) may be
obtained from the Company address on the back cover, together with servicing and spares information.
BS EN ISO 9001
Cert. No. Q5907
R
E
G
I
S
T
E
R
E
D
EN 29001 (ISO 9001)
Lenno, Italy – Cert. No. 9/90A
0255
Stonehouse, U.K.
1
1.1 General
This instruction manual describes the installation and
operation of the 4234 600 Series Power Supply Unit (PSU),
designed to provide a stabilised (optional 500, 400, 350, 250
or 180 mA) d.c. supply for operation of a single katharometer
unit and an auxiliary 10 V d.c. supply for katharometer
temperature control.
1.2 Description – Fig. 1.1
The PSU is available in two separate versions defined in
Section 2.
The unit is housed in a metal case fitted with mounting lugs.
Cable gland entries are provided at opposite ends of the case
for supply voltage input and output cables respectively.
A printed circuit board assembly (see Fig. 1.1) contains the
circuit components. The supply voltage input is connected to
terminal block (TB1) located on the pcb, adjacent to the fuses
F2 and F3. The constant current output terminal block (TB2) is
located on the gland assembly at the opposite end of the
assembly. The auxiliary 10 V output terminal block (AUX 10V)
is located on the PCB near to TB1. The pcb is mounted on a
steel chassis and secured by six screws. The cover is
perforated to allow ventilation for cooling.
Note. The PSU has NO serviceable parts.
Two identical fuses (L & N) are mounted within the unit on the
mains input. A third fuse is connected in the d.c. line circuit of
the transformer. A fourth fuse is connected in the auxiliary
10 V d.c. line. The L & N fuses are a high breaking capacity
type and must only be replaced with identical types. See
Section 2 for rating.
1 INTRODUCTION
Fig. 1.1 Location of Internal Component
TB1
T1
L N E
F2 F3
Warning. Hazardous Voltages
There are no servicable parts in this unit.
Please return to the manufacturer if faulty,
or seek the services of a qualified engineer.
Ensure that the mains supply is switched
off and disconnected before removing the
cover for any reason.
6
8
10
230 V
115 V
TB2
+
F1
B
A
C
X
D
E
M5 Earth Stud
Cable Gland for
a.c. Mains Input
Output
Current
Links
Cable Gland for
d.c. Output to
Katharometer
Mains Input
Terminal Block
d.c. Output
Terminal Block
Fuses
5 to 9 mm
Cables
5 to
9 mm
Cable
F4
AUX 10V
+
Cable Gland for
10 V Output to
Katharometer
CONTENTS
1 INTRODUCTION ................................................. 1
1.1 General ...................................................... 1
1.2 Description.................................................1
2 SPECIFICATION ................................................. 2
3 MECHANICAL INSTALLATION .......................... 3
4 ELECTRICAL INSTALLATION ........................... 4
4.1 Interconnection Diagrams and
Cable Requirements ..................................4
4.1.1 Katharometer Connections............4
4.1.2 Temperature Control Circuit........... 4
4.2 Accessing the PSU Terminals....................5
4.3 Electrical Connections ...............................5
5 SERVICING ......................................................... 6
5.1 Fuse Replacement.....................................6
2
Supply:
Model 4234600............... 230 V a.c., 50/60 Hz
Model 4234601............... 115 Va.c., 50/60 Hz
Power
Consumption
Katharometer.................. 3.5 W maximum
Katharometer with
temperature control..... 13 W maximum
Fuse Ratings:
L & N ................................ 250 mA (20 x 5 mm) 250 VA,
1500A h.b.c. cartridge type
d.c. line ............................ 400 mA (250 V a.c.)
Aux. 10 V d.c line ............ 1 A (250 V) F.B. (20 x 5 mm)
D.C. Output:...................... optional, stabilised, 500, 400,
350, 250 or 180 mA
Load: ................................. one katharometer (12.5 max.)
plus interconnecting cable (2
max)
Ambient Temperature
Range:............................... 20 to +55°C
2 SPECIFICATION
Permissible Supply
Voltage Variation: ........... ±15%
Frequency Range........... 46 to 64 Hz
Output
Regulation: ....................... Within ±0.8% for
a)Load variation ±15%
b)Supply variation ±6%
c) Ambient temperature ±10°C
Ripple:............................... Less than 17.5 mV across 10
load peak to peak
Stability:............................ Within ±0.7% of initial setting
over a period of one month with
load resistance, supply voltage
and ambient temperature at
nominal stated values.
Auxiliary Output............... 10 V d.c. nominal at 1 A for
katharometer temperature
control.
Dimensions: ..................... see Fig. 3.1
Fixing: ............................... see Fig. 3.1
Weight:.............................. 2.12 kg approximately
3
Fig. 3.1 PSU Dimensions
Fig. 3.2 Typical Identification Labels and locations
3 MECHANICAL INSTALLATION
ABB Limited
Oldends Lane, Stonehouse,
Glos. England GL10 3TA
Code 4234/600XX
Serial No
Voltage 230 Vac
Watts
Year Manuf.
Manufactured by:
N4006
The PSU must be mounted in a clean, dry, well ventilated position (on a wall or panel) in a safe area by means of the four fixing lugs.
Fig. 3.1 gives dimensions and fixing details for the unit.
Before commencing installation, check that the unit supplied is suitable for the supply voltage available see Section 2 and the data
plate on the unit cover, Fig. 3.2.
140 CTRS
110 CTRS
170
111
120
All holes 7.0 diameter
160
83
20 20
3030
30
60 DIM
Ensure a clearance of at least 100 mm at both
ends of the unit for acces to the cable glands
and to minimise cable bends.
Mains Cable
Entry
Output to Temperature Control
Circuit in Katharometer only
Output to
Katharometer
3.0
An adaptor kit (part number 4234 519) is available for mounting the
4234 600/601 PSU in place of the 004235 000/01/02 PSU.
All dimensions are nominal
mm unless otherwise stated.
4
4 ELECTRICAL INSTALLATION
or
1+2
C501
4
1+
8
3
2+
-
+
Power Supply Unit
4234600/601
Katharometer Types
6515
6517
6518
6520
6522
6524
7 +
-
+
mA o/p
10V o/p
For temperature controlled
katharometer block type
6518, 6520, 6524
Mains in
4234 600 230 V a.c.
or
4234 601 115 V a.c.
4689 Display
TBB
+12
TBB
10
or
28+29
C351
or
1+2
Datum
L150
Commander Series
Fig. 4.1 Typical Interconnecting Diagram Showing 4234 600/601 PSU, Katharometer and Displays
4.1 Interconnection Diagrams and Cable
Requirements
See Fig. 4.1 for interconnections for typical gas analysis
systems.
4.1.1 Katharometer Connections
Cable connecting the constant current d.c. supply from the PSU
to the katharometer should be selected so that the maximum
cable resistance is less than 2 ohms. This limits the maximum
cable length between the katharometer and its associated PSU.
Note that the cable gland will accept a cable between 5 to 9 mm
in diameter.
4.1.2 Temperature Control Circuit
The cable connecting the auxiliary 10 V d.c. supply to the
katharometer temperature control circuit must have cable
conductance capable of passing 1 amp without undue voltage
drop or excessive heating.
Note also that the cable gland will accept a cable between 5 to
9 mm in diameter.
5
To gain access, remove the four
screws and lift off the cover.
Fig. 4.2 PSU Cover Removal
TB1
T1
F2 F3
6
8
10
230 V
115 V
L N E
115 V a.c.
or
230 V a.c.
Supply
See Caution opposite
Tighten Cable Gland
M5 Earth Stud
Fig. 4.4 Connecting Outputs
Fig. 4.3 Connecting to a.c. Supply
4 ELECTRICAL INSTALLATION
To Katharometer
(see manual for
termination details)
Tighten Cable Gland
TB2
+
F1
B
A
C
X
D
E
d.c. Output
Terminal Block
Cable Clamp
Output Current (mA) Links
500 A to X
400 B to X
350 C to X
250 D to X
180 E to X
Current Output Link
To Katharometer
temperature control
circuit
(see manual for
termination details)
AUX 10V
+
4.2 Accessing the PSU Terminals
Warning. If the PSU is cubicle mounted or has its
cover removed for test purposes, a.c. mains supply
voltage is exposed at certain locations within the
assembly when supplies are connected. Thus, there is a
risk of electrocution.
All normal precautions must be taken to avoid the risk of
accidental electrical shock during installation and
maintenance of the equipment. Ensure that the supply
cable is disconnected at its source before touching any
electrical connections.
Caution. Ensure that the correct version of the unit
has been supplied for the supply voltage available. A 115 V
unit cannot be used with a 230 V supply or vice versa.
Check that the current output meets the katharometer
input requirements (see katharometer data label) see
Fig. 4.4. If in doubt contact ABB Limited.
Remove the power supply unit cover to obtain access to the
separate labelled terminal blocks see Fig. 4.2.
4.3 Electrical Connections
The mains terminal block, TB1, is situated next to the transformer,
T1, and fuses, F2 and F3.
Connect the a.c. mains input cable as in Fig. 4.3.
Note. Earth bonding to be made via M5 Earth Stud.
Connect the d.c. output and 10 V cables as in Fig. 4.4.
6
5 SERVICING
5.1 Fuse Replacement
There are no serviceable components in these PSU's other than to replace the fuses see Fig. 5.1. These fuses must be replaced
only with those stated in the specification. If a fault occurs the PSU should be returned to the nearest ABB Service Centre, or
returned to the factory for repair.
Fig. 5.1 Changing Fuses
TB1
T1
L N E
F2 F3
6
8
10
230 V
115 V
TB2
F1
B
A
C
X
D
E
F4
AUX 10V
+
+
Switch off and disconnect
the mains supply
1
2
Remove the four screws and lift
off the cover.
Replace fuse(s)
3
Note. Suitable fuses may be obtained
from ABB with the following part numbers:
F1..............0231555 (400 mA cartridge)
F2 & F3....0231577 (250 mA 1500 A
hbc cartridge)
F4.............0231599 1 A (F.B.) 250 V
4
Replace cover, connect to mains power source
and switch on.
7
NOTES
8
…NOTES
PRODUCTS & CUSTOMER SUPPORT
Products
Automation Systems
for the following industries:
Chemical & Pharmaceutical
Food & Beverage
Manufacturing
Metals and Minerals
Oil, Gas & Petrochemical
Pulp and Paper
Drives and Motors
AC and DC Drives, AC and DC Machines, AC motors to 1kV
Drive systems
Force Measurement
Servo Drives
Controllers & Recorders
Single and Multi-loop Controllers
Circular Chart , Strip Chart and Paperless Recorders
Paperless Recorders
Process Indicators
Flexible Automation
Industrial Robots and Robot Systems
Flow Measurement
Electromagnetic Magnetic Flowmeters
Mass Flow Meters
Turbine Flowmeters
Wedge Flow Elements
Marine Systems & Turbochargers
Electrical Systems
Marine Equipment
Offshore Retrofit and Referbishment
Process Analytics
Process Gas Analysis
Systems Integration
Transmitters
Pressure
Temperature
Level
Interface Modules
Valves, Actuators and Positioners
Control Valves
Actuators
Positioners
Water, Gas & Industrial Analytics Instrumentation
pH, conductivity, and dissolved oxygen transmitters and
sensors
ammonia, nitrate, phosphate, silica, sodium, chloride,
fluoride, dissolved oxygen and hydrazine analyzers.
Zirconia oxygen analyzers, katharometers, hydrogen purity
and purge-gas monitors, thermal conductivity.
Customer Support
ABB Ltd. provides a comprehensive after sales service via a
Worldwide Service Organization. Contact one of the following
offices for details on your nearest Service and Repair Centre.
United Kingdom
ABB Ltd.
Tel: +44 (0)1453 826 661
Fax: +44 (0)1453 827 856
United States of America
ABB Inc.
Tel: +1 (0) 755 883 4366
Fax: +1 (0) 755 883 4373
Client Warranty
Prior to installation, the equipment referred to in this manual
must be stored in a clean, dry environment, in accordance with
the Company's published specification. Periodic checks must be
made on the equipment's condition.
In the event of a failure under warranty, the following
documentation must be provided as substantiation:
1. A listing evidencing process operation and alarm logs at time
of failure.
2. Copies of operating and maintenance records relating to the
alleged faulty unit.
IM/4234600 Issue 1
The Company’s policy is one of continuous product
improvement and the right is reserved to modify the
information contained herein without notice.
Printed in UK (01.02)
© ABB 2002
ABB Ltd.
Oldends Lane, Stonehouse
Gloucestershire, GL10 3TA
UK
Tel: +44 (0)1453 826 661
Fax:+44 (0)1453 827 856
ABB Inc.
2175 Lockheed Way
Carson City, NV 89706
USA
Tel: +1 (0) 775 883 4366
Fax:+1 (0) 775 883 4373
ABB has Sales & Customer Support
expertise in over 100 countries worldwide
www.abb.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12

ABB 4234 600 User manual

Type
User manual
This manual is also suitable for

ABB 4234 600 is a Power Supply Unit designed to provide a stabilized DC supply for operation of a single katharometer unit. It also provides an auxiliary 10V DC supply for katharometer temperature control, making it suitable for various gas analysis systems. The device ensures precise and stable output voltage, with regulation within ±0.8% for load, supply, and temperature variations. Its compact design and easy installation process make it convenient to integrate into existing setups.