WIKA TR31 Operating instructions

Type
Operating instructions
DE
EN
Operating instructions
Betriebsanleitung
Model TR31-3 Model TR31-K
OEM miniature resistance thermometer
Models TR31-3, TR31-K
OEM-Miniatur-Widerstandsthermometer
Typen TR31-3, TR31-K
70018194
2
14096788.07 03/2018 EN/DE
WIKA operating instructions models TR31-3, TR31-K
DE
EN
Operating instructions models TR31-3, TR31-K
Page 3 - 34
Betriebsanleitung Typen TR31-3, TR31-K
Seite 35 - 66
Further languages can be found at www.wika.com.
© 07/2014 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA
®
is a registered trademark in various countries.
WIKA
®
ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
3
Contents
Contents
Declarations of conformity can be found online at www.wika.com.
1. General information 4
2. Safety 6
3. Specications 10
4. Design and function 14
5. Transport, packaging and storage 17
6. Commissioning, operation 18
7. Conguration 24
8. CongurationsoftwareWIKAsoft-TT 25
9. Connecting PU-548 programming unit 28
10. Maintenance and cleaning 30
11. Faults 30
12. Dismounting, return and disposal 32
Appendix:CSAcontroldrawing 34
EN
14096788.07 03/2018 EN/DE
4 WIKA operating instructions models TR31-3, TR31-K
1. General information
The resistance thermometer described in the operating instructions
has been designed and manufactured using state-of-the-art
technology. All components are subject to stringent quality and
environmental criteria during production. Our management systems
are certied to ISO 9001 and ISO 14001.
These operating instructions contain important information on
handling the instrument. Working safely requires that all safety
instructions and work instructions are observed.
Observe the relevant local accident prevention regulations and
general safety regulations for the instrument's range of use.
The operating instructions are part of the product and must be kept
in the immediate vicinity of the instrument and readily accessible to
skilled personnel at any time.
Skilled personnel must have carefully read and understood the
operating instructions prior to beginning any work.
The manufacturer's liability is void in the case of any damage caused
by using the product contrary to its intended use, non-compliance
with these operating instructions, assignment of insuciently qualied
skilled personnel or unauthorised modications to the instrument.
The general terms and conditions contained in the sales
documentation shall apply.
Subject to technical modications.
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: TE 60.31
- Application consultant: Tel.: +49 9372 132-0
Fax: +49 9372 132-406
1. General information
EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
5
Explanationofsymbols
WARNING!
... indicates a potentially dangerous situation that can result
in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result
in light injuries or damage to equipment or the environment,
if not avoided.
Information
... points out useful tips, recommendations and information
for ecient and trouble-free operation.
DANGER!
... identies hazards caused by electrical power. Should the
safety instructions not be observed, there is a risk of serious
or fatal injury.
WARNING!
... indicates a potentially dangerous situation that can result
in burns, caused by hot surfaces or liquids, if not avoided.
Abbreviations
2-wire The lead resistance is recorded as an error in the measurement.
3-wire With a cable length of 30 m or longer, measuring errors can
occur.
4-wire The lead resistance can be neglected.
1. General information
EN
14096788.07 03/2018 EN/DE
6 WIKA operating instructions models TR31-3, TR31-K
2. Safety
WARNING!
Before installation, commissioning and operation, ensure
that the appropriate resistance thermometer has been
selected in terms of measuring range, design, specic
measuring conditions and appropriate wetted parts' materials
(corrosion).
Non-observance can result in serious injury and/or damage
to the equipment.
Depending on the type of application, the electrical
connection must be protected against mechanical damage.
Further important safety instructions can be found in the
individual chapters of these operating instructions.
2.1 Intendeduse
The model TR31 resistance thermometer is used as a general-
purpose thermometer for the measurement of temperatures from
-50 … +150 °C or -58 ... +302 °F (without neck tube) and -50 … +250 °C
or -58 ... +482 °F (with neck tube) in liquid and gaseous media. The
version with mineral-insulated sheathed cable and neck tube allows
to measure temperatures up to 300 °C or 572 °F. It can be used for
pressures up to 140 bar with 3 mm sensor diameters and up to 270 bar
with 6 mm sensor diameters, dependent on the instrument version.
The instrument has been designed and built solely for the intended use
described here, and may only be used accordingly.
The technical specications contained in these operating instructions
must be observed. Improper handling or operation of the instrument
outside of its technical specications requires the instrument to be taken
out of service immediately and inspected by an authorised WIKA service
engineer.
2. Safety
EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
7
If the instrument is transported from a cold into a warm environment, the
formation of condensation may result in instrument malfunction. Before
putting it back into operation, wait for the instrument temperature and the
room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on
operation contrary to the intended use.
2.2 Personnelqualication
WARNING!
Riskofinjuryshouldqualicationbeinsucient!
Improper handling can result in considerable injury and
damage to equipment.
The activities described in these operating instructions
may only be carried out by skilled personnel who have the
qualications described below.
Keep unqualied personnel away from hazardous areas.
Skilled electrical personnel
Skilled electrical personnel are understood to be personnel who, based
on their technical training, know-how and experience as well as their
knowledge of country-specic regulations, current standards and directives,
are capable of carrying out work on electrical systems and independently
recognising and avoiding potential hazards. The skilled electrical personnel
have been specically trained for the work environment they are working
in and know the relevant standards and regulations. The skilled electrical
personnel must comply with current legal accident prevention regulations.
Special operating conditions require further appropriate knowledge, e.g.
of aggressive media.
2. Safety
EN
14096788.07 03/2018 EN/DE
8 WIKA operating instructions models TR31-3, TR31-K
2.3 Special hazards
WARNING!
For hazardous media such as oxygen, acetylene,
ammable or toxic gases or liquids, and refrigeration plants,
compressors, etc., in addition to all standard regulations,
the appropriate existing codes or regulations must also be
followed.
WARNING!
Protection from electrostatic discharge (ESD) required!
The proper use of grounded work surfaces and personal
wrist straps is required when working with exposed circuitry
(printed circuit boards), in order to prevent static discharge
from damaging sensitive electronic components.
To ensure safe working on the instrument, the operating
company must ensure
that suitable rst-aid equipment is available and aid is
provided whenever required.
that the operating personnel are regularly instructed in all
topics regarding work safety, rst aid and environmental
protection and know the operating instructions and in
particular, the safety instructions contained therein.
DANGER!
Danger of death caused by electric current
Upon contact with live parts, there is a direct danger of death.
Electrical instruments may only be installed and
connected by skilled electrical personnel.
Operation using a defective power supply unit (e.g. short
circuit from the mains voltage to the output voltage) can
result in life-threatening voltages at the instrument!
2. Safety
EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
9
WARNING!
Residual media in dismounted instruments can result in a
risk to persons, the environment and equipment.
Take sucient precautionary measures.
Do not use this instrument in safety or emergency stop
devices. Incorrect use of the instrument can result in injury.
Should a failure occur, aggressive media with extremely high
temperature and under high pressure or vacuum may be
present at the instrument.
2.4 Labelling, safety marks
Productlabel(examples)
2. Safety
Model
Date of manufacture (Year-Month)
Approval logos
Information on version (measuring element, output signal, measuring range...)
Thermometer with transmitter and 4 ... 20 mA output signal
Thermometer with direct sensor output with Pt100 and Pt1000
Serial number, TAG no.
Before mounting and commissioning the instrument,
ensure you read the operating instructions!
EN
14096788.07 03/2018 EN/DE
10 WIKA operating instructions models TR31-3, TR31-K
3. Specications
Thermometerwithtransmitterandoutputsignal4...20mA
(modelTR31-x-x-TT)
Temperature range Without neck tube -30 ... +150 °C (-22 ... +302 °F)
With neck tube -30 ... +250 °C (-22 ... +482 °F)
1) 2)
Measuring element Pt1000
Connection method 2-wire
Tolerancevalueofthe
measuring element
Class A
(per IEC 60751)
Measuringdeviationofthe
transmitter
±0.25 K
(per IEC 60770)
Totalmeasuringdeviationin
accordancewithIEC60770
Measuring deviation of the measuring element +
the transmitter
Measuring span Minimum 20 K, maximum 300 K
Basicconguration Measuring range 0 ... 150 °C (-32 ... +302 °F),
other measuring ranges are adjustable
Analogueoutput 4 ... 20 mA, 2-wire
Linearisation Linear to temperature per IEC 60751
Linearisation error ±0.1 %
3)
Switch-on delay, electrical Max. 4 s (time before the rst measured value)
Warming-up period After approx. 4 minutes, the instrument will
function to the specications (accuracy) given in
the data sheet.
Current signal for fault
signal
Congurable in accordance with NAMUR NE43
downscale ≤ 3.6 mA upscale ≥ 21.0 mA
Sensor short-circuit Not congurable, in accordance with
NAMUR NE43 downscale ≤ 3.6 mA
Sensor current < 0.3 mA (self-heating can be neglected)
Load R
A
R
A
≤ (U
B
- 10 V) / 23 mA with R
A
in Ω and U
B
in V
Eectofload ±0.05 % / 100 Ω
Power supply U
B
DC 10 ... 30 V
Max.permissibleresidual
ripple
10 % generated by U
B
< 3 % ripple of the output
current
Power supply input Protected against reverse polarity
Powersupplyeect ±0.025 % / V (depends on the power supply)
3.Specications
EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
11
Thermometerwithtransmitterandoutputsignal4...20mA
(modelTR31-x-x-TT)
Inuenceoftheambient
temperature
0,1 % of the span / 10 K T
a
Electromagnetic
compatibility(EMC)
5)
EN 61326 emission (group 1, class B) and
interference immunity (industrial application)
4)
,
conguration at 20 % of the full measuring range
Temperature units Congurable °C, °F, K
Infodata TAG No., descriptor and user message can be
stored in transmitter
Congurationand
calibration data
Permanently stored
Electrical connection
M12 x 1 circular connector (4-pin)
Directly connected cable
Readings in % refer to the measuring span
For a correct determination of the overall measuring deviation, both sensor and transmitter
measuring deviations have to be considered.
1)
The temperature transmitter should therefore be protected from temperatures over 85 °C (185 °F).
2) Version with mineral-insulated sheathed cable can be used up to 300 °C (572 °F).
3) ±0.2 % for measuring ranges with a lower limit less than 0 °C (32 °F)
4) Use resistance thermometers with shielded cable and ground the shield at at least one end
of the lead if the lines are longer than 30 m or leave the building. Operate the instrument in a
grounded state.
5) During transient interferences (e.g. burst, surge, ESD) take into account an increased measuring
deviation of up to 2 %.
3.Specications
EN
14096788.07 03/2018 EN/DE
12 WIKA operating instructions models TR31-3, TR31-K
Thermometer with direct sensor output with Pt100
(modelTR31-x-x-Px)andPt1000(modelTR31-x-x-Sx)
Temperature range
Class A
Class B
Without neck tube -30 ... +150 °C (-22 ... +302 °F)
With neck tube -30 ... +250 °C (-22 ... +482 °F)
6)
Without neck tube -50 ... +150 °C (-58 ... +302 °F)
With neck tube -50 ... +250 °C (-58 ... +482 °F)
6)
Temperature at the
connector or at the directly
connected cable
Max. 85 °C (185 °F)
Measuring element
Pt100 (measuring current: 0.1 ... 1.0 mA)
Pt1000 (measuring current: 0.1 ... 0.3 mA)
Connection method
2-wire
3-wire
4-wire
Tolerancevalueofthemeasu-
ringelementperIEC60751
Class A
Class B at 2-wire
Electrical connection
M12 x 1 circular connector (4-pin)
Directly connected cable
6) Version with mineral-insulated sheathed cable can be used up to 300 °C (572 °F).
For detailed specications for Pt sensors, see Technical information IN 00.17 at
www.wika.com.
Case
Material Stainless steel
Ingressprotection
Case with connected connector
7) 8)
or directly connected cable
Coupler connector, not connected
IP67 and IP69 per EN/IEC 60529,
IP69K per ISO 20653
IP67 per EN/IEC 60529
Weight in kg Ca. 0.2 ... 0.7
(depending on version)
Dimensions See “Dimensions in mm”
7) The stated ingress protection only applies when plugged in using mating connectors that have
the appropriate ingress protection.
8) Not tested at UL
3.Specications
EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
13
Ambientconditions
Ambienttemperaturerange
M12 x 1 circular connector
Model TR31-3-x-TT
Models TR31-3-x-Px, TR31-3-x-Sx
Directly connected cable
(model TR31-K-x-xx)
-40 ... +85 °C (-40 ... +185 °F)
-50 ... +85 °C (-58 ... +185 °F)
-20 ... +80 °C (-4 ... +176 °F)
Storage temperature range
M12 x 1, 4-pin circular connector
(model TR31-3-x-xx)
Directly connected cable
(model TR31-K-x-xx)
-40 ... +85 °C (-40 ... +185 °F)
-20 ... +80 °C (-4 ... +176 °F)
ClimateclassperIEC60654-1
M12 x 1 circular connector
Model TR31-3-x-TT
Models TR31-3-x-Px, TR31-3-x-Sx
Directly connected cable
(model TR31-K-x-xx)
Cx (-40 ... +85 °C or -40 ... +185 °F,
5 ... 95 % r. h.)
Cx (-50 ... +85 °C or -58 ... +185 °F,
5 ... 95 % r. h.)
Cx (-20 ... +80 °C or -4 ... +176 °F,
5 ... 95 % r. h.)
Maximumpermissiblehumidityper
IEC60068-2-30var.2
Relative humidity 100 %, condensation
allowed
Maximumoperatingpressure
9)10)
140 bar with 3 mm sensor diameter
270 bar with 6 mm sensor diameter
VibrationresistanceperIEC60751 10 ... 2,000 Hz, 20 g
9)
Shock resistance
perIEC60068-2-27
50 g, 6 ms, 3 axis, 3 faces, 3 times for
each face
Salt fog IEC 60068-2-11
9) Dependent on the instrument version
10)
Reduced operating pressure when using a compression tting:
Stainless steel
PTFE
max. 100 bar
max. 8 bar
3.Specications
EN
14096788.07 03/2018 EN/DE
14 WIKA operating instructions models TR31-3, TR31-K
Patents/property rights
M12x1adaptertoangular
connectorDINEN175301-803
No. 001370985
Conditionsforoutdooruse(forULapprovalonly)
The instrument is suitable for applications with pollution degree 3.
The power supply must be suitable for operation above 2,000 m
should the temperature transmitter be used at this altitude.
The instrument shall be installed in locations sheltered from the
weather.
The instrument shall be installed “sun/UV radiation protected”.
For further specications see WIKA data sheet TE 60.31 and the order
documentation.
4. Design and function
4.1 Description
The model TR31 resistance thermometer consists of a thermowell with
a xed process connection and is screwed directly into the process.
It is designed to be impact and vibration resistant and all electrical
components are protected against humidity (IP67 or IP69K). The
vibration resistance conforms to IEC 60751 (20 g, dependent on the
instrument version). The impact resistance of all versions meets the
requirements of IEC 60751. The electrical connection is made via an
M12 x 1 circular connector or via the directly connected cable.
An adapter for electrical connection with angular connector per
DIN EN 175301-803 is optionally available for the M12 x 1 circular
connector version.
3.Specications/4.Designandfunction
EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
15
X
X
X
X
Thread Height process connection X
G 1/2 11
G 3/8 11
G 1/4 10
M12 11
M20 11
1) For process temperature > 150 °C (302 °F) a neck length N (M
H
) of 70mm is required,
otherwise N (M
H
) selectable (55, 65 or 70 mm).
140069565.02
without
neck tube
without process
connection
with compression
tting
with
neck tube
Legend:
A (U
1
) Insertion length
N (M
H
) Neck length
X Height process connection
with direct cable outlet
connecting cable
14097792.01
4.2 Dimensions in mm
Process connection with parallel thread (or without process connection)
4. Design and function
Ød Sensor diameter
W Length of directly connected cable
L Length of bare wires
EN
14096788.07 03/2018 EN/DE
16 WIKA operating instructions models TR31-3, TR31-K
X
X
X
Process connection with tapered thread
4. Design and function
with direct cable outlet
connecting cable
14097792.01
140069565.02
without
neck tube
with compression
tting
with
neck tube
4.3 Scopeofdelivery
Cross-check scope of delivery with delivery note.
Thread Height process connection X
1/4NPT 15
1/2NPT 19
1) For process temperature > 150 °C (302 °F) a neck length N (M
H
) of 70mm is required,
otherwise N (M
H
) selectable (55, 65 or 70 mm).
Legend:
A (U
2
) Insertion length
N (M
H
) Neck length
X Height process connection
Ød Sensor diameter
W Length of directly connected cable
L Length of bare wires
EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
17
5. Transport, packaging and storage
5.1 Transport
Check the instrument for any damage that may have been caused by
transport. Obvious damage must be reported immediately and damaged
instruments must not be used.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport
(e.g. change in installation site, sending for repair).
5.3 Storage
Permissibleconditionsattheplaceofstorage:
Storage temperature:
- M12 x 1, 4-pin circular connector: -40 ... +85 °C (-40 ... +185 °F)
- Directly connected cable: -20 ... +80 °C (-4 ... +176 °F)
Humidity: 5 ... 95 % r. h.
Avoidexposuretothefollowingfactors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down hard)
Soot, vapour, dust and corrosive gases
Potentially explosive environments, ammable atmospheres
Store the instrument in its original packaging in a location that fulls the
conditions listed above. If the original packaging is not available, pack
and store the instrument as described below:
1. Wrap the instrument in an antistatic plastic lm.
2. Place the instrument along with shock-absorbent material in the
packaging.
3. If stored for a prolonged period of time (more than 30 days), place a
bag containing a desiccant inside the packaging.
WARNING!
Before storing the instrument (following operation), remove
any residual media. This is of particular importance if
the medium is hazardous to health, e.g. caustic, toxic,
carcinogenic, radioactive, etc.
5. Transport, packaging and storage
EN
14096788.07 03/2018 EN/DE
18 WIKA operating instructions models TR31-3, TR31-K
6. Commissioning, operation
WARNING!
Avoid putting any mechanical loading on the electrical
connections and on the enclosures. Connections must only
be opened once the instrument has been depressurised and
has cooled down.
Maximumpermissibletemperatures:
At case with transmitter: 85 °C (185 °F)
With directly connected cable: -20 ... +80 °C (-4 ... +176 °F)
Class A:
Without neck tube -30 ... +150 °C (-22 ... +302 °F)
With neck tube -30 ... +250 °C (-22 ... +482 °F)
1)
Class B:
Without neck tube -50 ... +150 °C (-58 ... +302 °F)
With neck tube -50 ... +250 °C (-58 ... +482 °F)
1)
1) Version with mineral-insulated sheathed cable can be used up to 300 °C (572 °F).
6.1 Mounting
These resistance thermometers are designed for screw-tting directly
into the process. The insertion length, along with the ow velocity and
viscosity of the process media, may reduce the max. loading on the
thermowell.
The housing must be grounded against electromagnetic elds and
electrostatic discharge. It is not necessary to connect the housing
separately to the equipotential bonding system, provided that it has a
xed and secure contact to the metallic vessel, its components or pipes,
and that these are connected to the equipotential bonding system.
When there is a non-metallic contact with the vessel, or with its
structural components or piping, the instrument must be provided with
equipotential bonding.
6. Commissioning, operation
EN
WIKA operating instructions models TR31-3, TR31-K
14096788.07 03/2018 EN/DE
19
Installation on pipes
a on elbows
b in smaller pipe, inclined
c perpendicular to ow direction
Insertion length A
WARNING!
Neither repairs nor structural modications are permitted,
and any would void the guarantee.
Installationexamples
For information on tapped holes, refer to DIN 3852 or for NPT threads to
ANSI B 1.20.
6.1.1Tighteningtorquesforcompressionttings
Sealing Turns Max.pressureinbar
Stainless steel ferrule 1 ¼ ... 1 ½ 100
Stainless steel compression ring 1 ¼ ... 1 ½ 100
PTFE ferrule 1 ¼ ... 1 ½ 8
6.1.2 Tightening torque for M12 mating connector or M12 adapter
Choose a tightening torque of 0.6 Nm.
6. Commissioning, operation
EN
14096788.07 03/2018 EN/DE
20 WIKA operating instructions models TR31-3, TR31-K
6.2 Electrical connection
The electrical connection is made via a circular connector.
Output signal Pt100 and Pt1000 (standard)
M12 x 1 circular connector (4-pin) Directly connected cable
6. Commissioning, operation
2-wire
red
white
white
white
white
red
red
red
red
3-wire
4-wire
mA
4 ... 20 mA
M12x1connector Angularconnector
Accessories:M12x1PtadaptertoDINEN175301-803angularconnector
Alternative pin assignments possible.
For further information see order documentation.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68

WIKA TR31 Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI

in other languages