Dial Caliper Best Practices
• Makesuretothoroughlycleanandwipedry
the jaws before each use.
• Beforeusingthecalipers,check,andif
necessary, adjust the dial needle to zero with
the jaws closed together.
• Topreventexcessivestressontheprecision
parts of the caliper, always use the
thumbwheel to move the jaws and dial.
• Whenmakingmeasurements,alwayshold
the caliper perpendicular to the measured
surface.
• Asthejawsopen,therackisexposed.
Always keep the rack and its channel free
from dirt and debris; otherwise, accuracy will
be lost and internal damage may occur.
• Donotdroporbendthecalipers.
• Cleanandwipedrytheentirecaliperafter
each use and before placing it in the case.
Reading Calipers
The calipers illustrated in Figure 2 are set at
2.147".
Figure 2. Calipers set at 2.147".
© 2001 Grizzly Industrial, Inc.351201622
Dial Calipers can measure objects with an accuracy of 0.001". Some rules of use should be observed to
maximize precision.
• Wipe the jaws clean before use.
• Check the position of zero on the dial with the jaws closed and adjust if needed by loosening the dial lock
and rotating the bezel.
• Always use the thumb wheel to close or open the jaws and never apply too much pressure.
• Hold the caliper perpendicular to the measured surface.
• As the jaws open, the rack is exposed. Do not allow dust and debris to settle here or accuracy will be
lost and internal damage may occur.
• Do not drop the calipers.
In the illustration above, the calipers have been set at 2.147". The following describes how this measure-
ment was read.
The jaw index has passed the 2 Inch Mark, A. 2.000"
The jaw index has passed the 1 on the Revolution Scale, B. .100"
The needle has passed the 40 on the Dial Numbers. .040"
There are 7 White Lines after the 40 mark, C. .007"
The total measurement = 2.147"
Figure 2. Measuring 2.147".
0 1 2
1 2 3 4 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 2 3 4 5 6 7 8 9
83
0
10
20
30
40
50
60
70
80
90
The following describes how this measurement is
read:
1. The jaw index is past the "2" inch mark.
A = 2.000"
2. The jaw index is past the "1" on the
revolution scale.
B = 0.100"
3. The dial needle is aligned with "47" on the
dial.
C = 0.047"
4. The total measurement is the sum of A, B,
and C.
2.000" + 0.100" + 0.047" = 2.147"
Measuring Objects
The calipers can be used to measure both
internal and external dimensions of objects,
as well as depth dimensions. The examples
in Figures 3–5 illustrate each type of
measurement.
0 1 2
1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
3 4 5 6
1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
10
20
30
40
50
60
70
80
90
0
10
20
30
40
50
60
70
80
90
Figure 3. Taking an outside measurement.
0 1 2
1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
3 4 5 6
1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
10
20
30
40
50
60
70
80
90
0
10
20
30
40
50
60
70
80
90
Figure 4. Taking an inside measurement.
0 1 2
1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
3 4 5
1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9
1 2 3
10
20
30
40
50
60
70
80
90
0
10
20
30
40
50
60
70
80
90
Figure 5. Taking a depth measurement.