Steris Kerna Surgical Table Operating instructions

Type
Operating instructions
Customer manual
[ use & maintenance ]
OM 201-01EN 2023-06 REVA KERNA
General Surgical Table
KERNA
General Surgical Table - KERNA™
[2]
CONTENTS
1. QUALITY REQUIREMENTS .......................................................................................................................................... 2
2. SAFETY INSTRUCTIONS............................................................................................................................................. 3
3. SYMBOLS USED ......................................................................................................................................................... 5
4. OVERVIEW ................................................................................................................................................................. 7
5. SETTING UP THE TABLE ............................................................................................................................................ 10
6. CONNECT AND OPERATE THE CONTROL DEVICES .................................................................................................... 13
7. OVERRIDE SYSTEM ................................................................................................................................................... 15
8. WEIGHT CAPACITIES ................................................................................................................................................. 16
9. CLEANING / DISINFECTION ....................................................................................................................................... 18
10. MAINTENANCE PROGRAMME ................................................................................................................................. 18
11. KERNA™ SPARE PARTS LIST .................................................................................................................................. 19
12. TROUBLESHOOTING ................................................................................................................................................. 19
13. DISPOSAL OF WASTE EQUIPMENT .......................................................................................................................... 19
14. ELECTROMAGNETIC ENVIRONMENT ....................................................................................................................... 20
15. TFT HAND CONTROL (OPTIONAL) ............................................................................................................................ 21
16. IN SERVICE FORM ................................................................................................................................................... 31
17. IN SERVICE FORM ................................................................................................................................................... 33
Dear Customers,
STERIS thanks you for choosing the KERNA™ General Surgical Table, which helps you focus on your patient.
Please read this manual carefully to ensure that our products can offer you optimum efficiency and safety in use.
1. QUALITY REQUIREMENTS
The STERIS quality system is certified
ISO 13485*
for the design, manufacture, sales, installation and after-sales service of operating tables, platforms and transfer carts and their extensions and
accessories.
This table is designed in compliance with international regulations for medical electrical equipment :
IEC 60601-1 (EN 60601-1)* general requirements for basic safety and essential performance
IEC 60601-2-46 (EN 60601-2-46)* safety rules applicable to operating tables
IEC 60601-1-2 (EN 60601-1-2)* relating to electromagnetic compatibility
Intended use: KERNA™ General Surgical Table is medical device intended to support and position the patients during surgical procedures
carried out in an operating room.
KERNA™ General Surgical Table and associated extensions and accessories shall be operated in a controlled environment by a healthcare
professional familiar with appropriate techniques with such equipment.
Essential performance of KERNA™ General Surgical Table: supporting a patient without unwanted movement in a single-fault condition.
Any utilisation going beyond the aforementioned conditions is deemed to be not for intented purposes. Only the user is liable for any loss or
damage arising from this.
The KERNA™ General Surgical Table is Class 1 medical devices and are registered with ANSM (French National Agency for the Safety of Drugs
and Healthcare Products)
KERNA™ General Surgical Table is available in 4 models:
KERNA HV : Kerna Table high height velcro version
KERNA LV : Kerna Table low height velcro version
KERNA HSX : Kerna Table high height snapLINX™ version
KERNA LSX : Kerna Table low height snapLINX™ version
In this manual, these models will be regularly mentioned in order to provide the user with a better understanding of the similiarities and
differences between these models of operating tables.
Placing on the market
KERNA™ General Surgical Table was first placed on the market in 2020.
*in accordance with the most up to date version applicable
STERIS SAS
116 Avenue de MAGUDAS
33185 LE HAILLAN - FRANCE
General Surgical Table - KERNA
[3]
2. SAFETY INSTRUCTIONS
Any serious incident that occurs in relation to this medical device should be reported to STERIS and the competent authority of the member state in
which the user and/or patient is established.
Warning: The modification of the medical device without manufacturer authorisation is prohibited.
Pinch points
! During patient positioning, moving dierent sections of the tabletop and using extensions
and accessories can cause pinch points to appear which can injure the patient or the
surgical team. Users must therefore be aware of the risks and pay particular attention to
the following zones :
Precaution during the movements
! Before any movement activation, be sure that there is no risk of contact or collision with
an object, an equipment or a person.
Floor lock
! Even if the table is not locked to the floor, certain tabletop movements are still possible.
Lock the table to the floor before positioning the patient and starting the surgical
procedure.
! Do not try to move the locked table.
! Once the table is locked to the floor, ensure that it is stable. Move it if necessary to find
a flatter floor surface.
Moving the table with a patient on it
! When moving the table with a patient on it, the following conditions are required:
-The tabletop must be centered over the column or above the long part of the base.
-The patient should be in normal position.
! Never try to unlock the table and move it with the patient, if the table is not in the
transport position. The table may tip.
! If moving the table with a patient be careful to hold the patient in place.
! Before going anywhere, ensure that there are no obstacles along the way you wish to
take. The table is not intended to cross thresholds, to ascend or to descend steps.
Always move the table very carefully.
! Do not keep objects on the base as they may fall during motorised manoeuvres and cause
jamming when lowering the columns sliding covers.
! Consider a safe zone where you’ll not put anything on it.
Safe zone
Switching the table on
! Every time you turn the table on, check that the pre-selected patient orientation
corresponds to what you want to do. When you switch o the table, the patient
orientation pre-selection is memorised.
! If the hand control and the optional pedal are used at the same time, the hand control
overrides the pedal.
Switching the table off
! The STOP buttons on the main hand control and on the override system allow you to
stop all table movement. They deactivate the optional pedal and other external control
mechanisms.
! Note that the battery of the optionnal wireless remote control do not charge anymore when
the table is switched o by the STOP function, even if the table is connected to the Mains.
! The table stops automatically at the end of a preprogrammed cycle (3’ by default with
standard remote control in wired mode. 240’ in wireless mode with optional remote
control).
Handling the surgical table
! Avoid handling the surgical table roughly during the operation. Do not exert excessive
lateral loads on it, do not pull or push the surgical table with the accessories or with the
rails.
! Avoid handling the remote control and the override system roughly. Do not press the
keypad with too much force or using unsuitable objects. Avoid shock on the control
means of the table.
! Avoid handling the cables roughly when connecting or disconnecting the remote control,
the foot pedal or the power cord set.
! Do not sit on one of the two ends of the table, especially when the tabletop is not
centered, as this can cause the table to tilt.
! Do not stand on the base of the surgical table because it can damage the covers.
! When a power cord is connected to the surgical table, and therefore is lying on the floor,
pay attention to this power cord when moving around the surgical table and remove the
cable before moving the table.
Weight capacity
! Do not exceed a patient+accessory weight of 270 kg for the standard described
positions and respect the indications and warnings signalled.
! The use of certain extensions may decrease this patient+accessories weight. When
extensions are installed on the KERNA™ General Surgical Table, you should also
consult the user manuals of these extensions to know the possible weight restrictions
related to the use of these extensions.
! In positions where the tabletop will be significantly slide away from the column, ensure
that there is sucient stability for the type of surgical procedure requirements.
! For all other uses, please contact our network for advice. If you are working intensively
on the patient’s chest, we recommend that you bring the tabletop back to a centred
position on the column.
Positioning the patient
! The positioning of the patient remains under the responsibility of healthcare professional.
It is the responsibility of the operator to ensure that the movements made are under his
control and do not generate any instability or danger to the patient during the surgery
performed.
! Ensure the stability of the patient and the table when making combined position.
! Secure the patient properly (ex. using straps), especially when the table has a strongt tilt inclination.
! Long surgical procedures can cause compression points on patients and have post
operative consequences. Monitor patient during surgical procedures and use gel
cushions for extented procedures.
General Surgical Table - KERNA
[4]
Autonomy
! Two 12V sealed gelified-lead battery is integrated into the base and mean that
KERNA™ General Surgical Table is autonomous. Before using the table, check that the
battery is suciently charged.
Connections
! The sockets are equipped with locks: do not pull on the hand control or pedal cables to
disconnect them.
! Never try to connect a device not supplied by STERIS to the surgical table.
! Always use the power cord provided by STERIS.
! The operator should never touch the patient when the operator is plugging or
unplugging the remote control or the the pedal, or when the operator is handling the
connectors of these devices.
Potential equalisation conductor
! Connect together potential equalisation conductors to overcome the problem of diering
potentials across networks.
! Originally intended for cardiac areas, it protects the patient from electric shock during
procedures where even small potentials may cause harm.
! When a potential equalisation conductor is connected, the table MUST NOT be used
with a defibrillator
! When the use of the potential equalisation conductor allows to create an electromedical
system, then this system must meet the requirements of IEC 60601-1.
Antistatic link
! If antistatic connection is necessary, we recommend that you only use the pads that
have been specially developed for this product. Put the patient in direct contact with
them and use the table on an antistatic floor, or connect it to an equipotential cable
using the socket designed for this.
Electrical safety
Warning: To prevent any risk of electric shock, the
appliance should only be connected to an earthed
power supply system.
!
! The KERNA™ General Surgical Table must always be positioned so that the power cord
can be disconnected easily when connected to the table.
! The power cord with its main electrical plug is thus the device used to isolate the table
KERNA™ from the main electrical supply.
! It is preferable to use the battery instead of the main electrical supply if you have doubts
about the protective ground in the operating room.
! The use of high-frequency surgical equipment, cardiac defibrillators, and cardiac
defibrillator monitors in association with the KERNA™ General Surgical Table creates a
risk of burns, explosion, or electrical shock. The user must be aware of this.
! In case of any damage to the power supply or Mains cable or plug or IEC socket, do not
connect the equipment and immediatly contact STERIS technicians.
Accessories
! If accessories made by other manufacturers are used, it is the user’s responsibility to
check that they are compatible with KERNA™ General Surgical Table and that they do
not endanger the patient and the surgical team.
! Before connecting accessories to the surgical table, check the general state of them to
ensure that they are not degraded and that the connection system is not damaged.
! The end user shall especially ensure that the intended use and the conditions of use
(including the safe working load) of these accessories made by other manufacturers are
compatible with the claims of KERNA™ General Surgical Tables.
! Ensure that the extensions and accessories are correctly clamped to the surgical table
before each surgical procedure.
Electromagnetic compatibility
! If the table is used close to other electrical equipment, ensure that these comply with the
current regulations regarding electromagnetic compatibility.
! If high frequency equipment (such as a defibrillator) is being used, make sure that the
manufacturer’s instructions are rigorously followed.
Emergency mode
! When the override control panel is being used, the other control systems including the
main hand control are automatically switched off.
! Lock / Ensure the table to the floor before starting any movements with a patient.
Cleaning
! Ensure that the socket protection correctly protects the plugs so that fluids do not get in.
After cleaning procedures, ensure no residual water remains near the plug.
! Before cleaning the table, ensure that it is switched o. Put on the socket protection.
! Do not immerse the hand control and do not spray cleaning products directly onto the
sockets.
! Hypochlorite solutions can damage metal parts. After disinfection, ensure it is rinsed o
thoroughly with water and dry the surface. Prolonged exposure to hypochlorite solutions
may degrade mattress material. Avoid hypochlorite solutions and other liquids coming
into contact with any internal parts of the table.
Hand control storage
! We recommend hanging the hand control on a table rail when the hand control is not in
use, to prevent any accidental pressing of the buttons. Handle the hand control and the
associated cable with care.
Environmental conditions
Trans
port / storage : temperature 0°C-60°C, humidity 10-85% non-condensing, pressure 500-
1060 hPa, table on castors, battery fuse removed..
Use : temperature 10°C-40°C, humidity 10-75% non-condensing, pressure 800-1060hPa.
Local pressure on the floor: 17kg/cm2 (1667kPa) with a patient+accessories load of 270kg
Floor surface punching pressure : 1,3t/m2 (13kPa) with a patient+accessories load of 270kg
! Do not use the table in the presence of flammable anesthetic gas, or in areas subject to
explosive risks.
! KERNA™ General Surgical Table and associated extensions and accessories must not be
used in the vicinity of strong magnetic fields (MRI environment).
! KERNA™ General Surgical Table and associated extensions and accessories are not
intended to operate in an oxygen rich environment.
! KERNA™ General Surgical Table and associated extensions and accessories are
not intended for outdoor use, and shall not be exposed to high humidity for extented
periods.
X-Ray radiation
! KERNA™ General Surgical Table shall not be considered patient supporting system for
X-Ray equipment for interventional procedures. KERNA™ General Surgical Table is not
an associated equipment of X-Ray equipment in a radiological installation. The operating
table materials located in the X-Ray beam therefore do not have optimal radiolucent
performance and can operate parasite radiation that is potentially harmful to users.
Radiological protection devices should therefore be worn by users when using a X-Ray
equipment with KERNA™ General Surgical Table.
Maintenance
! The safety, reliability and proper operation of the surgical table and associated
extensions and accessories can only be ascertained and granted by STERIS. Repair and
adjustments should only be assigned to STERIS or approved technicians.
! Observe the maintenance schedule set out by STERIS and described in this document.
! Never try to adjust or repair the surgical table yourself.
! No interventions may be undertaken on KERNA™ General Surgical Table before all
electrical power is switched o and the supplies made safe.
! During maintenance and repair, only original parts must be used.
! No part of the surgical table shall be serviced or maintained while in use with the patient.
General Surgical Table - KERNA
[5]
3. SYMBOLS USED
Connection panel
Main hand control
Optional pedal
ACC
OUT
Optional motorised legs
Trend / Rev trend
Table up / Table down
Right tilt / Left tilt
Back up (seat down in reverse
orientation)
Back down (seat up in reverse
orientation)
Sliding towards the feet
Sliding towards the head
Selection
unlocked / locked
Table off button
ON
Start the table
(Option : start the table in bluetooth)
Emergency mode
(in combination with the
desired movement)
Override control
Main hand control (standard version)
Battery discharge status STOP
Trend Rev trend
Table up Table down
Left tilt Right tilt
Slide foot Slide head
Return-to-level
Patient orientation on the
table
Normal / Reverse
Selection
unlocked / locked
Reflex Flex
Back up
(leg up in reverse orientation)
Back down
(leg down in reverse
orientation)
General Surgical Table - KERNA
[6]
Warning:
consult the user’s manual
Risk of tipping:
Do not sit
Equipotential plug
WEEE Directive:
the product must be recycled
Name and address
of manufacturer
Direct Current
Alternating Current
Protection from electric shock
class 1, type B
IPX4 Protection against water projection
(for surgical table and its accessories)
IPX6 Protection against water projection
(optional pedal)
Table battery charge status
Date of manufacture
(YYYY-MM-DD)
REF Reference of the product
SN Serial number
MD Medical Device
Caution
Compliance with requirements of European legislation
1060
500
Atmosphere pressure limitation
85
10
Humidity limitation
65°C
0°C
Temperature limitation
Nstacking limit by number (1 for KERNA™)
Main electrical connection
Press and release any
pedal to start
This way up
Fragile, handle with care
Handle with care
UKCA Marked (UK Conformity Assessed)
Compliance with essential requirements of UK Medical Devices
Regulations 2002
REPUK
UK Authorized Representative
REP REPCH
Swiss Authorized Representative
Importer
Base and column labels + packaging labels
General Surgical Table - KERNA
[7]
4. OVERVIEW
1. Mobile base
2. Column
3. Tabletop
4. Removable head section
5. Half back section
6. Seat section
7. Removable leg sections
8. Power panel
9. Power cord
10. Battery fuse
11. Connection panel
12. Main hand control
13. Override control
14. Equipotential plug
15. Electrical characteristics label
16. Identification label
9
14
10
8
15
16
3
4
5
7
1
6
2
13
11
16 12
The model shown is the KERNA™ General Surgical Table (with standard hand control) HEAD05B and LEG20C in patient normal orientation
General Surgical Table - KERNA
[8]
Characteristics
The KERNA™ General Surgical Table is a multi-sectional, autonomous, electro-
hydraulic surgical table.
The table will be presented in the rest of the manual equipped with a double
articulation head section and split leg sections for convenience.
KERNA™ General Surgical Table offers exceptional radiological access without
the need for additional extensions or for the patient to be reversed.
The KERNA™ General Surgical Table is suitable for most surgical procedures
and has been designed to support patients+accessories weighing up to 300 kg.*
A standard remote control or an optional remote control can be both used with
the table.
Movements may also be carried out by using the optional pedal.
If using motorized legs (LEG10M or LEG01M), the optional remote control is
necessary.
* Veuillez vous référer au chapitre 8 Capacités de poids
An original integrated power system enables the table to run on battery or on
Mains depending on user preference.
The tabletop movements are motorised, with the exception of the headrest and
manual split leg sections.
KERNA™ General Surgical Table’s intelligent electronic E-Serve™ system
allows users to operate the table remotely, and also allows for diagnostic
and maintenance via computer performed by authorised and trained STERIS
personnel (errors, history).
The integrated ALS™ linear and angular position sensors avoid collisions
between the motorised sections and the table.
Finally, an integrated backup system (emergency panel) covering as many
breakdown issues as possible, allows the user to continue the “in process”
procedure and to remove the table from the room.
Electrical characteristics
Mains 100-240 VAC 50-60Hz
420VA max 1 phase
Protection Class 1 type B
Equipotential plug DIN42801, cord not supplied
Mains cable provided 5m (main sector)
Internal fuses
(accessible to maintenance)
2XT6,3A 250V 5x20 PC1,5kA (Mains)
(T=Slow acting Fuse ; PC=Breaking Capacity)
3xF15A 32V PC1000A (power output)
(F=Fast acting Fuse ; PC=Breaking Capacity)
Batteries 2x 12V 17Ah sealed lead
Batteries fuse 1xF16A 250V 5x20 PC750A
(F=Fast acting Fuse ; PC=Breaking Capacity)
(accessible to operator)
Batteries of table Operating Temperature Range : -15°C to +50°C
Battery of remote control Operating Temperature Range : 0°C to +45°C
Autonomy Up to 60 procedures
Maximum charge duration 10h
Charge /discharge indicator Base LED / hand control LED
Intermittent runningI 3 min/h during surgical prodedure
Resistance to liquids IPX4 (for surgical table and its accessories)
IPX6 (for optional pedal)
Antistatic path Using pads and antistatic feet
Applied parts Pads
Radiocommunication characteristics
(for optional remote control)
Frequency bands used 2,40Ghz - 2,4835 Ghz
Maximal radiofrequency power
transmitted on used frequency bands 10dBm
Mechanical characteristics
Table up/Table down 595mm/945mm (versions LV and LSX)
695mm/1045mm (versions HV and HSX)
Left tilt/Right tilt 20°/20°
Trend/Rev trend 30°/-30°
Back up / Back down 90°/90°
Slide head / Slide foot 0mm / 230mm (stroke 230mm)
Flex / Reflex -220°/ 120°
Tabletop length (excluding extensions) 840mm (2025mm with headrest and split legs)
Tabletop width 520mm (560mm including rails)
Off-load table weight 185kg
Maximum weight of patient + accessories 270kg
Maximum capacity up/down 300kg horizontal tabletop, centered
Floor clerance / wheel diam 16mm / 100mm
Main material used
Covers Stainless steel and ABS
Structures High performance coated steel and aluminium
Tabletops Phenolic laminate
Bellows PVC
Pads Polyurethane foam welded seams
General Surgical Table - KERNA
[9]
Flex / Reflex
Tilt
Rev trend / Trend
down / up
Sliding
BackBack
Tabletop dimensions (Example LEG20C / HEAD05B)
90°
90°
30°
20°
230
2055
530310330
840
885
560
520
220° 120°
Movement possibilities (seat and half back section)
30°
20°
0
595mm/945mm (versions LV and LSX)
695mm/1045mm (versions HV and HSX)
595mm
695mm
945mm
1045mm
General Surgical Table - KERNA
[10]
5. SETTING UP THE TABLE
Clic!
To remove the head section, you only need to press the 2 buttons located on each
side of the tabletop once because of the push spring integrated into the Hi-lock™
system.
Pull the pins out completely while supporting the weight of the head section.
Please refer to the head section user manual.
5.1 Attaching/removing snapLINX™ mattresses
Warning! INFECTION HAZARD
Ensure mattresses are free from cuts, splits and abrasions. Replace worn mattresses for new ones as soon as possible to eliminate possible biological
hazards.
Y
To fix the mattress to the section:
1. Ensure that the tabletop is clean and dry before installation.
2. Offer up the mattress to the section so that the snapLINX™ pins are lined up with attachments in the underside of the mattress.
3. Gently ease one mating attachment at a time onto and over the respective snapLINX™ pin.
4. The mattress should lie flat on the section when this has been achieved without evidence that the snapLINX™ pins are holding the mattress
off the section.
5. Check the fixation, trying to move the mattress laterally.
5.2 Mounting / demounting extensions
The KERNA™ General Surgical Table is equipped with Hi-lock™ locking system using side buttons. It is compact, easy to use and secure.
With Hi-lock™, an extension is automatically locked-in when it is inserted, and just by using the side buttons it is then removed.
Depending on the type of extension and the required safety level, it may be necessary to press the button twice in order to remove the extension. Do not force the
extension if it does not release after pressing the button once.
For convenience, it is preferable to install and remove the extensions when the table is in a horizontal position.
Example – installing / removing a head section
Insert simultaneously the two pins of the head section into the two holes in the back or seat section of the table. The head section should attach easily; you will
hear a click on each side when the automatic locking happens. Check that it is locked correctly by trying to pull the head section out.
Depending on your choice you will have either mattresses with Velcro attachment or mattresses with snapLINX™ attachments.
General Surgical Table - KERNA
[11]
Click!
Example – installing / removing the split leg sections
For each leg section, insert the pins into the corresponding holes (left or right) in the back or seat section of the table. When the pin is 3/4 inserted, make sure
that the anti-rotation pin is aligned with the casing on the table upright. Do not force it: the leg section should attach easily. You will hear a click on each side
when it becomes locked automatically. Check that it is locked correctly by attempting to pull out the leg section.
The leg sections are equipped with a double security system: to remove them, the side button on the relevant leg section must first be pressed in order to release
the anti-rotation pin.
There are then two ways to remove the leg section completely. The user must be ready to support the weight of the leg section being removed:
-Press the side button on the relevant leg section a second time and completely extract the pin while supporting the weight;(see steps below)
-Rotate the leg section while continuing to extract the pin and supporting the weight.
5.3 Inserting and removing the battery fuse:
For reasons of safety during transport, the battery fuse is packed separately.
1. Insert the fuse into the fuse holders provided separately.
2. Insert all of this into the fuse holders.
3. Push and turn using a screwdriver in order to lock the fuse holder.
4. Check by switching on the table with the main hand control without using the
Mains. The three colored LED (green, orange and red) will flash for a few seconds.
If the table is not used for a long period of time (>1 month), remove the fuse in order
to limit battery discharge and maintain battery functionality
For further instructions, please refer to the leg section user manual.
A B C
General Surgical Table - KERNA
[12]
Hand control onStandard On the base Procedures Battery status
30-60 sufficient
20-30 Mains connection recommended
1-20 Mains connection mandatory
0
Caution : The blinking of the red LED warns a problem, the battery must be recharged imperatively or the table will automatically shut down. If the problem
persists, please contact STERIS technical department.
5.5 Using the table on battery
&&
In standard version:
The battery status is displayed on the hand control and on the base
To extend the working life of the battery, we recommend the following:
! Charge battery systematically each day (also in no use conditions)
! Charge battery systematically before the indicator on the screen of the remote control is red
! Avoid the deep discharge to prevent degradation of battery life time
! Carry out a preventive maintenance (a control of battery each year)
5.4 Connecting the table to the main electrical supply and control
the battery charge level
&
&
1. Check that the Mains voltage matches the indications on the table.
2. Lift the hatch of the rear panel, and connect the power cord to the IEC connector.
3. Connect the power cord to a Mains socket.
The power supply allows the user to choose whether to operate the table on battery or mains power.
It has three coloured LED used to monitor the battery status when connected to Mains
&
&
1. If the battery is charging, the coloured LED light up sequentially orange and green.
2. If the battery is charged, the coloured LED light up steadily orange, and green:
This can also be checked on the remote control.
If the 3 coloured LED (red, orange, green) flash when the Mains is connected, this indicates a battery
charge error. In this case, please contact the STERIS technical department.
The charger built into the power supply automatically regulates battery charge, so that the table can be
left permanently connected to the Mains.
Bip!
Bip!
General Surgical Table - KERNA
[13]
6. CONNECT AND OPERATE THE CONTROL DEVICES
Connecting and activating the main hand control
Connection:
1. Lift the protection cover off the corresponding socket.
2. Align the red dot on the hand control cable connector towards the top, with those on the socket on the table side.
3. Plug the connector until it “clicks, indicating that it is correctly connected.
The wired main hand control haspriorityover the optional pedal. To activate the hand control, press any button except
the STOP button.
It switches off automatically after 3 minutes if it has not been used.
Disconnection:
1. Switch off the hand control by pressing STOP.
2. The connector is equipped with a lock: unlock the connector by pulling the ring backwards. Never just pull on the cable.
3. Replace the protection covers to protect the sockets.
Connecting and activating the optional pedal
Connection:
1. Lift the protection cover off the corresponding socket.
2. Align the red dot on the pedal cable connector with that on the socket on the table side.
3. Plug the connector until it “clicks, indicating that it is correctly connected.
The pedal enables users to adjust the 3 main table movements: up/ down, trend/rev trend, and right/left tilt. To activate
it, just press a pedal and then release on any movement. It switches off automatically after 3 minutes if it has not been used.
Disconnection:
1. Switch off the pedal by using the STOP button on the hand control.
2. The connector is equipped with a lock: unlock the connector by pulling the ring backwards. Never just pull on the cable.
3. Replace the protection covers to protect the sockets.
Connecting an optional motorized leg extension
(LEG01M and LEG10M) with optional TFT remote control only
For more information about the motorized legs, refer to the user manual related to these extensions.
Connecting the E-Serve™ maintenance software
The specific E-Serve™ cable connects to the same port as the main hand control. Therefore, when the E-Serve™
cable is connected, the main hand control cannot be connected.
Connection:
1. Lift the protection cover off the corresponding socket.
2. Align the red dot on the specific E-Serve™ connector, with that on the socket on the table side.
3. Plug the connector until it “clicks”, indicating that it is correctly connected.
Disconnection:
1. Switch off the table by pressing STOP button on the override system.
2. The connector is equipped with a lock: unlock the connector by pulling the ring backwards. Never just pull on the cable.
3. Replace the protection covers to protect the sockets.
General Surgical Table - KERNA
[14]
Tabletop movements / LED indicators
STOP
Return to level
Tilt right
Rev trend
Return to level LED
Trend
Tilt left
Battery discharge status
Floor lock LED
Back up
(leg up in reverse orientation)
Back down
(leg down in reverse orientation)
Reverse patient orientation
Normal patient orientation
Flex
Reflex
• Lock/unlock table to floor
• Blue LED “locked”
Blue LED fixed :the table is locked
Blue LED blinking : the table is unlocked
Slide foot Slide head
Warning:
Tactile: in the case of an anti-collision alert, the hand control vibrates (ALS)
Audible: beep in the case of a movement requiring special attention or at the end of a movement
6.1 OPERATING THE KERNA™ MAIN HAND CONTROL (standard version)
The main hand control includes:
1. A 5 m spiral cord with a connector at each end making it easy to replace.
2. A plastic housing equipped with a hook so that it can be attached to the table’s side rails.
3. A keyboard allowing direct access to the function of the table
To level the table:
The blue LED will permanently light up when the table is in horizontal position.
The return-to-level button enables the tabletop to be returned to a horizontal position.
During the return to level tabletop movements, even if it is automatically realised, always ensure patient safety.
1. To return the tabletop to level: (the level LED is switched off if the tabletop is not horizontal)
2. Press the 0 button; the blue LED will flash during the tabletop movements.
3. The blue LED will be permanently on when the tabletop is horizontal.
Switching on the table
Pressing any button (except the STOP button) turns the table on but does not start any movements. A welcome screen is then displayed for a few seconds. You
can then press the button for the required movement.
Switching o the table
When the STOP button is pressed, the table immediately stops.
The Auto-off function turns off the table after 3 minutes if it is no longer in use in order to avoid any unintentional movement and to conserve battery life.
Locking / unlocking to floor
The mobile base of the KERNA™ General Surgical Table allows it to be moved around the operating theatre so that it can be relocated or so that the floor can be cleaned.
Locking / unlocking to the floor is controled automatically from the hand control.
During locking / unlocking (which takes a few seconds) all of the buttons, except the STOP button, are disabled.
The Locking/unlocking button has to be pressed more than 1 second.
When the blue LED blinks the table is on its castors. When the blue LED is fixed the table is locked.
All movements are authorized even when the table is on its castors.
Before starting any surgical procedure, check that the table is locked to the floor and stable. The floor locking pads can compensate by up to 4mm if the floor is
not quite horizontal: if necessary, move the table to find a more stable location.
up
down
General Surgical Table - KERNA
[15]
7. OVERRIDE SYSTEM
If the table does not work correctly with the battery or the Mains, an override keypad located on the
column shall be used to replace the main hand control.
When the override system is in use:
-the main hand control is automatically switched off;
-the patient orientation described by the buttons is always in normal orientation;
-the sensors, and therefore the ALS™ system, are on standby.
You should therefore pay particular attention to the patient and look out for potential table section
conflicts, and contact the STERIS service.
Operation with the override keypad (powered mode)
-Simultaneously press the desired movement and the Powered Override Validation button. .
-The movement starts with a beep indicating that attention is required.
NB:
-Placing the table on the castors requires also the use of the button .
To unlock the table, keep pressing the buttons long enough until floor lock cylinders are
fully retracted.
To lock the table , keep pressing the buttons long enough until the floor lock cylinders
are fully extracted.
ON
Trend
(Rev trend if patient in rev
orientation)
Tilt left
(Right if patient in rev orientation)
Unlocking Move to castors
Tilt right
(Left if patient in rev orientation)
Rev trend
(Trend if patient in rev
orientation)
Back up
(Seat if patient in rev orientation)
Slide head
(Foot if patient in rev orientation)
Slide foot
(Head if patient in rev orientation)
Back down
(Seat if patient in rev orientation)
Table up
Table down
Table ON / Bluetooth
STOP / Stop the table
Floorlock
Override motorised Validation
Override system
General Surgical Table - KERNA
[16]
8. WEIGHT CAPACITIES
Man height according to current standard 190cm
Table shown with HEAD04B + BACK02B + LEG30C
Patient in reverse orientation
Reverse patient orientation state for the head on the long side of the base. All movements of the table can be performed. Only height adjustment can be permitted
for patient over 270kg.
Patient 270kg
Patient 270kg
Patient 165kg
Patient 165kg
Patient 300kg
Centered
Fully slided
Centered
Centered
Lithotomy
Centered
Centered
Centered
Fully slided
Fully slided
Lithotomy
Patient 270kg
Patient 270kg
Patient 300kg
Patient 270kg
For the weight capacities shown below, the centre of gravity (usually the umbilicus) must be kept within the blue and green and yellow highlighted areas.
The weight of additional accessories fitted to the table must be included in any estimation of applied load.
Due to the modular design of the KERNA™ General Surgical Table, the maximum permissible patient and accessory load will change depending on tabletop
configuration, patient position and accessories used. Please refer to the User Instructions of the Extension and/or Accessory to determine the maximum
permissible load.
?cm - Column - 48cm
Patient in normal orientation
Normal patient orientation state for the head on the short side of the base. All movements of the table can be performed. Only height adjustment can be permitted
for patient over 270kg.
Green : 300kg
Blue : 270kg
Green : 300kg
Blue : 270kg
Yellow : 165kg
35cm - 10cm - Column - 3cm - 55cm
General Surgical Table - KERNA
[17]
Normal patient orientation
The head support is placed on the half backrest side and the leg supports
on the seat side
- Maximum sliding on head side (intervention via the upper body)
- Maximum sliding on foot side (intervention via the lower body)
Reverse patient orientation
The head support is placed on the seat side and the leg supports on the half
backrest side.
- Maximum sliding on head side (intervention via the upper body)
- Maximum sliding on foot side (intervention via the lower body)
Radiological access
Use of the KERNA™ General Surgical Table with a C-arm have been of particular focus.
It is possible to reverse the orientation of the KERNA™ General Surgical Table in relation to the base by taking off the head and leg sections and attaching them to
opposite sides so that the table is optimally arranged for the type of surgery required. These two options give different kinds of radiological access.
NB: every time the table orientation is reversed, you will need to select the correct patient orientation on the main hand control so that the buttons continue to control the
correct movements.
Below table is presented with LEG20C and HEAD05B+BACK02B - Dimensions will vary depending on extensions mounted.
900mm 1060mm
980mm 980mm
1620mm
1210mm 750mm 670mm 1290mm
Below table is presented with CARBEXT + Carbon headrest
The CARBEXT extension is placed on the seat side
General Surgical Table - KERNA
[18]
9. CLEANING / DISINFECTION
General information
Before every cleaning session, we recommend to turn off the table and
disconnect it from the charger. Make sure that the socket covers are fitted
correctly.
Do not submerge the hand control.
Before you start, ensure that the active components of the products used are
compatible with the table materials.
You must follow the dilution and temperature instructions recommended by the
detergent-disinfectant manufacturer.
Wear the recommended personal protective equipment (gloves, goggles)
After cleaning procedures, ensure no residual water remains near the plug.
Cleaning between 2 procedures
Cleaning the hand control:
-Wipe it with a paper tissu (or non-woven equivalent) impregnated with a
detergent-disinfectant solution.
-Leave to dry.
Cleaning the table’s visible surfaces:
-Spray a detergent-disinfectant solution on all the visible surfaces including
the pads.
-Clean with a paper tissu. Spray the detergent-disinfectant on all the visible
surfaces.
-Leave to dry.
Full cleaning at the end of the day
The pads can be easily removed thanks to their Velcro bands or their
snapLINX™:
-Remove the pads.
-Spray a detergent-disinfectant directly onto the pads.
-Clean with a paper tissue. Spray the detergent-disinfectant on all the visible
surfaces.
-Leave to dry.
-If necessary brush the Velcro bands, the pads and the tabletop with a non-
metallic brush (e.g. with a clothes brush).
Cleaning the table:
-Raise the table to its maximum height in order to get optimal access to the
column.
-Put the tabletop in a horizontal position.
-Clean with a paper tissue. Spray the detergent-disinfectant on all the visible
surfaces.
-Leave to dry.
If there is a risk of streaming, spray the detergent-disinfectant directly onto the
paper tissu.
Slide rails are sensitive to corrosion and must stay lubricated. If a cleaning
has to be done in this area, remember to lubricate them afterwards with an oil
film..
Cleaning the castors:
-Lock the table so that the castors are no longer touching the floor.
-Spray the detergent-disinfectant on all visible surfaces.
-Leave to dry.
Recommended products / products to avoid
Recommended active ingredients:
You should preferably use products designed for above-floor surfaces.
STERIS recommends the use of cleaning and disinfectant chemicals whose
compounds are based on isopropyl alcohol (<45%) or based on a quaternary
ammonium.
The use of cleaning chemicals not recommended by STERIS is forbidden
because the compatibility and the effectiveness of the chemicals have not
been demonstrated.
If users are unsure of which products to use they should consult the hygienists
in their hospital or their supervisory authority.
Do not use products that are meant for cleaning floors.
Hypochlorite solutions can damage metal parts.
After disinfection, ensure it is rinsed off thoroughly with water and dry the
surface.
Prolonged exposure to hypochlorite solutions may degrade mattress material.
Avoid hypochlorite solutions and other liquids coming into contact with any
internal parts of the table.
If you are unsure about whether the active ingredients are compatible, contact
the STERIS technical department.
Recommended daily maintenance:
-Check the table and hand control battery charge level (Optional if you are
using wireless remote control)
-Check the attachment of the extensions
-Check the table locking and its stability
-Check the operation of each movement from the main hand control
-Check the correct fastening of the pads on Velcro strips or snapLINX™
Preventative maintenance must be carried out at least once a year by a STERIS
technician, or another approved technician.
1- Safety points
-Attachment of the cylinders
-Clearance of the joints and protective covers
-Tightness of the side accessory rails
-Operation of the extension locking systems
-Operation of the table locking system
2- Inspection, maintenance, adjustment
-Slide condition
-Castor condition
-Battery condition (table + optional wireless hand control)
-Proper operation of charger / power supply
-Check oil level : The oil level contained in the tank should be between the
min and max levels indicated on the tank.
-Column play
-Condition of flexibles and connections
-Condition of cylinders
-Condition of cables and connectors and guides
-Lubrication of sliding rails
-Grease extension latches
-Condition and operation of the main/optional wireless hand control
-Condition and operation of the override system
-ALS operation
10. MAINTENANCE PROGRAMME
General Surgical Table - KERNA
[19]
11. KERNA™ SPARE PARTS LIST
In the event of a repeated or major malfunction, contact the STERIS technical department. Any modification or repair to this device by
personnel not approved by STERIS is prohibited.
FAULT DETECTED POSSIBLE CAUSE CORRECTIVE ACTION
The table will not switch on
The fuse is blown Replace the battery fuse or connect to Mains power
The hand control is defective Replace the hand control
The hand control cable is defective Replace the hand control
No movement occurs
The battery is completely discharged Connect the table to Mains power
The hand control is defective Replace the hand control
The hand control cable is defective Replace the hand control
The table remains in battery mode when the
Mains power cable is plugged in
Power cord bad connection Reconnect the Power cord
The Power cord is defective Replace the Power cord
A Mains fuse is blown Contact STERIS technical department
The extension will not lock or has a lot of
play
The extension is not correctly latched Re-lock the extension
The extension will not lock Contact STERIS technical department
12. TROUBLESHOOTING
The following table is provided for KERNA™ General Surgical Table operators faced with product fault scenarios.
Recycling the product: this equipment must be disposed by a legal body designated for the recycling of waste electrical and electronic equipment.
Hydraulic fluid: drain the hydraulic fluid off prior scrapping the table, and dispose this fluid properly.
Battery: take battery to an approved disposal facility. They are considered polluants.
13. DISPOSAL OF WASTE EQUIPMENT
DESCRIPTION REFERENCE
80mm mattress without URO cutout, welded seam, with velcro attachment S02CV3
80mm mattress with URO cutout, welded seam, with velcro attachment S01CV3
80mm mattress without URO cutout, welded seam, with snapLINX™ attachment S02CX3
80mm mattress with URO cutout, welded seam, with snapLINX™ attachment S01CX3
Optional wireless hand control (CMAX®) V543004910
HAND CONTROL CABLE (CMAX®) V542501208
HAND CONTROL (KERNA™) V543810995
VELCRO 1 METER LENGTH V654005006
SET OF 2 BATTERIES V342212555
CASTOR V606006408
General Surgical Table - KERNA
[20]
14. ELECTROMAGNETIC ENVIRONMENT
Essential performance of KERNA™ is supporting a patient without unwanted movement in a single fault condition.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
WARNING: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the KERNA™, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result. Before connecting any equipment, the user should make sure that they respect the applicable regulations and that their connections comply with the manufacturers’ recommendations. KERNA™ needs special precautions regarding
electromagnetic compatibility and needs to be installed and put into service according to EMC information provided in this instruction of use. The KERNA™ is compliant with standard IEC 60601-1-2 : 2014 (Ed.4).
Guidance and manufacturer’s declaration - Electromagnetic emissions
List of the cables & accessories in conformity with the emissions and
immunity testing
The KERNA™ are intended for use in an electromagnetic environment specified below. The customer or the user of KERNA™ should assure that it is used in such
environment. Hand control KERNA™
Remote control 542805998
Foot pedal TAB714
Mains cable 5m 689400010
Remote control cable 0,7m 542605104
Foot pedal cable 4m 542200409
Motorized legs LEG10M & LEG01M
Emission test Compliance Electromagnetic environment
guidance
RF emissions CISPR 11 Group 1
The KERNA™ use RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11 Class A The KERNA™ are suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that
supplies building used for domestic purposes.
NOTE The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services.
The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3 Conforms
Guidance and manufacturer’s declaration - Electromagnetic immunity
The KERNA™ are intended for use in an electromagnetic environment specified below. The customer or the user of KERNA™ should assure that it is used in such environment.
Immunity test IEC 60601-test level Compliance level Electromagnetic environment guidance
Electrostatic discharge (ESD))
IEC 61000-4-2 ±8 kV contact
±15 kV kV air
±8 kV contact
±15 kV kV air
Professional healthcare facility environment
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV
±1 kV Professional healthcare facility environment
Mains power quality should be that of a typical commercial or hospital environment
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV
±2 kV
Voltage dips, short interruptions and voltage variations
on power supply input lines
IEC 6100-4-11
<5 % UT (>95% dip in UT) for 0,5 cycle
40 % UT(60 % dip in UT) for 5 cycles
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
<5 % UT (>95% dip in UT) for 0,5 cycle
40 % UT(60 % dip in UT) for 5 cycles
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
Professional healthcare facility environment
Mains power quality should be that of a typical commercial or hospital environment. If the user of KERNA™
system requires continued operation during power mains interruptions, it is recommended that the
KERNA™ system be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Professional healthcare facility environment
Power frequency magnetic fields should beat at levels characteristic of a typical location in a typical
commercial or hospital environment.
RF Conducted IEC 61000-4-6 3V
150 kHz to 80MHz
6V in ISM bands between 0,15 MHz and 80 Mkz
0.8 MA at 1 kHz
3V
150 kHz to 80MHz
6V in ISM bands between 0,15 MHz and 80 Mkz
0.8 MA at 1 kHz
Professional healthcare facility environment
RF Radiated IEC 61000-4-3 3V/m
80 Mkz to 2,7 GHz
80% MA at 1 kHz
9V/m
710 MHz, 745 MHz
780 MHz, 5240 MHz
5550 MHz, 5785 MHz
27V/m
385 MHz
28V/m
450 MHz, 810 MHz
870 MHz, 930 MHz
1720 MHz, 1845 MHz
1970 MHz, 2450 MHz
3V/m
80 Mkz to 2,7 GHz
80% MA at 1 kHz
9V/m
710 MHz, 745 MHz
780 MHz, 5240 MHz
5550 MHz, 5785 MHz
27V/m
385 MHz
28V/m
450 MHz, 810 MHz
870 MHz, 930 MHz
1720 MHz, 1845 MHz
1970 MHz, 2450 MHz
Professional healthcare facility environment
NOTE UT is the a.c. mains voltage prior to application of the test level.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
(1) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which KERNA™ is used exceeds the applicable RF compliance level above, the KERNA™ operating table should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the KERNA™.
(2) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34

Steris Kerna Surgical Table Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI