Thermo Fisher Scientific Partisol 2000i Air Sampler Quick start guide

Type
Quick start guide
Partisol 2000i Air Sampler
Quick Start Guide
PM2.5 Sampling Setup
Part Number 111160-00
31May2011
2 Partisol 2000i PM2.5 Sampler Quick Start Guide Thermo Fisher Scientific
Thermo Fisher Scientific Partisol 2000i PM2.5 Sampler Quick Start Guide 3
IMPORTANT NOTE This guide is meant to assist users in setting up the
Partisol 2000i Air Sampler for PM2.5 sampling. It describes the primary
steps involved in setting up the hardware.
Although this document serves as a general guide, it does not provide the
detailed information contained in the instruction manual. For installation
issues not covered in this guide, refer to the instruction manual supplied
with the instrument.
The Partisol 2000i Air Sampler ships in two boxes. The large rectangular
box contains the Sampler and the smaller box contains the PM10 inlet. The
long cylindrical carton houses the stand.
Unpack the box containing the instrument, and make sure that the package
contains the following components.
1 PM10 inlet
1 VSCC or other PM2.5 device
1 VSCC adapter
1 Filter transport container with 2 cassettes and carriers
3 rainhoods and associated hardware
1 Sample tube
1 Partisol stand
1 Ambient temperature sensor and cable
1 Flow audit adapter
1 Solid filter leak check/separator disk
1 9-to-9 pin null modem computer cable
1 Null modem adapter
1 Operating manual
1 Quick start guide
1 iPort package
1 RPComm package
Ensuring
Completeness of
Shipment
Ensuring Proper
Accessories
4 Partisol 2000i PM2.5 Sampler Quick Start Guide Thermo Fisher Scientific
Attach one gasket to each of the three rain hoods supplied with the unit.
Peel back the paper facing and apply the sticky side of the gasket to the
flanged part of the rain hoods.
The large rain hood goes on the right side of the Sampler. The smaller rain
hoods are installed on the back panel of the enclosure. The two smaller rain
hoods are interchangeable.
Using the provided screws, install the three rain hoods (with gaskets
attached) to the back, left and right sides of the enclosure.
Unscrew the Phillips screws on the left side of the instrument, and retain
the washers. Place the washers between the ambient temperature bracket
and the enclosure—not under the head of the screw, and secure the bracket
with the screws. Plug in the 3-pin connector on the back side of the
Sampler where “Temperature Sensor” is marked.
Attaching Gaskets
to Rainhoods
Installing Rain
Hoods on the
Enclosure
Installing the
Temperature Sensor
Thermo Fisher Scientific Partisol 2000i PM2.5 Sampler Quick Start Guide 5
Open the filter exchange mechanism by pulling forward on its handle an
allowing it to drop into its open position. Remove the filter cassette carrier
with the installed cassette.
Lift the rollers of the filter exchange mechanism through the slots in the left
and right guides, and allow to drop down.
Attach VSCC Adapter to the bottom of the cyclone. Install the VSCC with
adapter into the inlet adapter at the top of the enclosure pushing straight
upward until it hits a stop.
Insert the rollers of the filter mechanism into the slots on the left and right
guides.
Opening the Filter
Exchange
Mechanism
Installing VSCC
6 Partisol 2000i PM2.5 Sampler Quick Start Guide Thermo Fisher Scientific
Note If using a WINS impactor, prepare as directed in the Preventive
Maintencance section of the instruction manual before installing in the
Sampler.
Insert the 1-1/4-inch OD sample tube into the bulkhead of the Sampler.
Ensure that the tube is pushed past the upper and lower o-rings, which
provide some resistance, and hits the stop. Hand-tighten the dome
connector on the bulkhead to ensure a tight grip. Place the PM10 inlet over
the end of the sample tube. Ensure that the tube is pushed past the inlet’s
two O-rings, which provide some resistance, and hits a stop.
Snap apart the two halves of a filter cassette and insert a screen so that the
serial number is on the underside of the screen. Place a 47 mm diameter
filter over the screen. Snap the two halves of the filter cassette together.
Installing the PM10
Inlet
Assembling a Filter
Cassette
PM10 Inlet
VSCC
Sample Tube
VSCC Adapter
Thermo Fisher Scientific Partisol 2000i PM2.5 Sampler Quick Start Guide 7
Place the cassette in the rectangular filter cassette carrier.
Install the filter cassette carrier on the filter platform. The slot and hole line
up with the hardware on the platform. Push the handle of the filter
exchange mechanism backward to lock the filter cassette into its sampling
position.
Turn on the sampler.
Note Use the provided anti-static filter transport container when moving
the filter and carrier to and from the sampling site.
Installing a
Collection Filter
8 Partisol 2000i PM2.5 Sampler Quick Start Guide Thermo Fisher Scientific
The following items are included in the stand assembly:
Parts:
A (2) Top Front and Top Back 36-005479
B (4) Top and Bottom, Right and Left 36-005476
C (4) Legs 36-005478
D (2) Bottom Front and Bottom Back 36-005477
Hardware:
E (4) Hex Head bolt 1/4-20 x ½ inch 21-001291-0008
F (20) Hex Head bolt 1/4-20 x 3/4 inch 21-001291-0012
G (24) Flat washer, 1/4 inch 21-001275
H (24) Split ring washer, 1/4 inch 21-000848
Ensuring Proper
Stand Parts
Thermo Fisher Scientific Partisol 2000i PM2.5 Sampler Quick Start Guide 9
Note Always place the split ring washer between the head of the bolt and
the washer.
1. Assemble the bottom of the stand Items B and D by laying out the
front, back, and sides.
2. Place one leg on each bottom stand inside corner and fasten with
hardware F, G and H.
3. Attach the top front and back (Item A) to the outside of the legs with
hardware F, G and H. Do not tighten.
4. Fasten the right and left rails to the legs and top right and left rails with
hardware F, G and H.
5. Tighten all hardware.
Assembling the Stand
10 Partisol 2000i PM2.5 Sampler Quick Start Guide Thermo Fisher Scientific
Place the Sampler onto the stand, and secure using hardware E, G, and H.
© 2011 Thermo Fisher Scientific Inc. All rights reserved.
Specifications, terms and pricing are subject to change. Not all products are
available in all countries. Please consult your local sales representative for
details.
Thermo Fisher Scientific
Air Quality Instruments
27 Forge Parkway
Franklin, MA 02038
1-508-520-0430
www.thermo.com/aqi
Secure Sampler
to Stand
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10

Thermo Fisher Scientific Partisol 2000i Air Sampler Quick start guide

Type
Quick start guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI