20
Quick Setup Guide TDL GPro™ 500
3 Installation
GeneralSetup(appliesforallparameters)
(PATH: Menu / Quick Setup)
While in Measurement mode press the [MENU] key to bring up the Menu selec-
tion. Select QUICK SETUP and press the [ENTER] key.
Display Convention:
1st line on display
→
a 3rd line on display
→
c
2nd line on display
→
b 4th line on display
→
d
Select the units of measurement for a and b. Only lines a and b can be cong-
ured in Quick Setup. Go to the Conguration Menu to congure lines c and d.
Channel Selection
Please select the type of Sensor:
Analog: For conventional analog sensors (will be displayed on channel “A”).
ISM: For ISM sensors (will be displayed on channel ”B”).
Please select the parameter requirement:
The choice of parameter depends on the level of transmitter. If an ISM sensor is
selected, the setting “Auto” means, all possible ISM sensors will be recognized
and accepted. If a special parameter is chosen, only this parameter will be
recognized and accepted on the transmitter.
Calibration
The GPro 500 is calibrated at the factory and does not require calibration at
installation & Startup.
TDLInstallation
(path: Quick Setup / TDL / Installation)
While in measurement mode press the key [MENU] . Press the
m
or
.
key to
select the TDL and then the Installation menu item.
In this mode, the current live value of the % transmission is displayed during
5 minutes until it automatically returns to the Measurement mode. Use this
value to rotate the blue sensor head attached with a loose clamp connection on
to the probe until the maximum transmission is found. In this position, and
secure the blue sensor head into position and tighten the clamp.
20.9 %V O
2
25.0 °C
MENU
Quick Setup
u
A
6.0 pH
A
25.0 °C
Channel Select=ISM
Parameter=Auto
u
B
20.9 %V O
2
B
20.9 %Trm
MENU
Quick Setup
u
B
20.9 %V O
2
B
20.9 %Trm
Transmission Signal
021 %
u