Mettler Toledo M400 2 wire Wiring Diagram Installation guide

Type
Installation guide
Transmitter M400 2(X)H 18
© 03/2023 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland M400 2-Wire
Printed in Switzerland 30 748 687 A
4 Installation Instruction
4.1 Unpacking and Inspection of Equipment
Inspecttheshippingcontainer.Ifitisdamaged,contacttheshipperimmediatelyforinstructions.
Donotdiscardthebox.
Ifthereisnoapparentdamage,unpackthecontainer.Besureallitemsshownonthepacking
listarepresent.
Ifitemsaremissing,notifyMETTLERTOLEDOimmediately.
4.1.1 Panel Cutout Dimensional Information –
½DIN Models
½DINModeltransmittersaredesignedwithanintegralrearcoverforstand-alonewallmount
installation.
Theunitmayalsobewallmountedusingtheintegralrearcover.Seeinstallationinstructionsin
section4.1.2.
Belowarecut-outdimensionsrequiredbythe½DINmodelswhenmountedwithinaflatpanel
oronaflatenclosuredoor.Thissurfacemustbeflatandsmooth.Texturedorroughsurfacesare
notrecommendedandmaylimittheeffectivenessofthegasketsealprovided.
137 mm + 0.5 mm
0.0 mm
(5.39" + 0.02")
0.00"
PANEL CUT-OUT
137 mm + 0.5 mm
0.0 mm
(5.39" + 0.02")
0.00"
Optionalhardwareaccessoriesareavailablethatallowforpanel-orpipe-mount.
Refertosection15fororderinginformation.
Transmitter M400 2(X)H 19
© 03/2023 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland M400 2-Wire
Printed in Switzerland 30 748 687 A
4.1.2 Installation Procedure
General:
Orientthetransmittersothatthecablegripsfacedownward.
Wiringroutedthroughthecablegripsshallbesuitableforuseinwetlocations.
InorderprovideIP66enclosureratings,allcableglandsmustbeinplace.Eachcablegland
mustbefilledusingacable,orsuitableCableGlandHoleSeal.
For Wall Mount:
Removerearcoverfromfronthousing.
Startbyunscrewingthefourscrewslocatedonthefaceofthetransmitter,ineachcorner.This
allowsthefrontcovertoswingawayfromtherearhousing.
Removethehinge-pinbysqueezingthepinfromeachend.Thisallowsthefronthousingtobe
removedfromtherearhousing
Mountrearhousingtowall.SecuremountingkittotheM400accordingtothesupplied
instructions.Attachtowallusingappropriatemountinghardwareforwallsurface.Besureitis
levelandsecurelyfastenedandtheinstallationadherestoanyandallclearancedimensions
requiredfortransmitterserviceandmaintenance.Orientthetransmittersothatthecablegrips
arefacingdownward.
Replacethefronthousingtotherearhousing.Securelytightentherear-coverscrewstoensure
thatIP66/NEMA4Xenclosureenvironmentalratingismaintained.Theunitisreadytobewired.
For Pipe Mount:
Useonlymanufacturer-suppliedcomponentsforpipe-mountingtheM400transmitterand
installperthesuppliedinstructions.Seesection15fororderinginformation.
4.1.3 Assembly – ½DIN Version
1. 3M20X1.5cableglands
2. Plasticsplugs
3. 4screws
1
2
3
Transmitter M400 2(X)H 20
© 03/2023 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland M400 2-Wire
Printed in Switzerland 30 748 687 A
4.1.4 ½DIN Version – Dimension Drawings
4.1.5 ½DIN Version – Pipe Mounting
40 ...
60 mm
1.57...
2.36"
METTLER TOLEDO
M400
150 mm / 5.90"
90 mm / 3.54"
36 mm
1.41"
72 mm
2.83"
150 mm / 5.90"
136 mm / 5.35"
90 mm / 3.54"
75 mm / 2.95"
105 mm / 4,13"
31 mm
1.22"
22 mm
0.87"
137 mm
(+0,5mm/−0,0mm)
5.39"
(+0.02"/−0.0")
137 mm
(+0,5mm/−0,0mm)
5.39"
(+0.02"/−0.0")
Transmitter M400 2(X)H 21
© 03/2023 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland M400 2-Wire
Printed in Switzerland 30 748 687 A
4.2 Connection of Power Supply
Allconnectionstothetransmitteraremadeontherearpanelofallmodels.
a Besurepowertoallwiresisturnedoffbeforeproceedingwiththeinstallation.
Atwo-terminalconnectorontherearpanelofallM400modelsisprovidedforpowerconnection.
AllM400modelsaredesignedtooperatefroma14–30VDCpowersource.Refertospecifications
forpowerrequirementsandratingsandsizepowerwiringaccordingly(AWG16–24,wirecross-
section0.2mm2 to 1.5 mm2).
4.2.1 Housing (Wall Mount)
1
2
1: TB1–InputandOutputAnalogSignal
2: TB2–SensorSignal
Transmitter M400 2(X)H 22
© 03/2023 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland M400 2-Wire
Printed in Switzerland 30 748 687 A
4.3 Terminal Block (TB) Denitions
TB1
TB2
A Q
1 12 13 18
PowerconnectionsarelabeledA01+ / HART and A01– / HART
resp.A02+ and A02–for14to30VDC.
4.4 Terminal Block TB1
Terminal Designation Description
1V_EC
Easyclean
2GND_EC
3485A_EC
4485B_EC
5DI1+ Digitalinput1
6DI1−
7DI2+ Digitalinput2
8DI2−
9OC1+ Opencollectoroutput1(switch)
10 OC1−
11 OC2+ Opencollectoroutput2(switch)
12 OC2−
13 AO1+/HART Powerconnection14to30VDC
Analogoutputsignal1
HARTsignal
14 AO1−/HART
15 AO2+ Powerconnection14to30VDC
Analogoutputsignal2
16 AO2−
17 Notused –
18
Transmitter M400 2(X)H 23
© 03/2023 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland M400 2-Wire
Printed in Switzerland 30 748 687 A
4.5 Terminal Block TB2: Analog Sensors
4.5.1 Conductivity (2-e / 4-e) Analog Sensors
Terminal Function Color
A Cnd inner11) White
BCndouter11) White/blue
CCndouter1 –
DNotused –
E Cndouter2 –
F Cnd inner22 ) Blue
GCndouter2(GND)
2) Black
HNotused –
I RTDret/GND Bareshield
J RTDsense Red
KRTD Green
L Notused
MNotused –
N Notused
ONotused –
PNotused –
QNotused –
1) For third party Conductivity 2-e sensors a jumper between A and B may be required.
2) For third party Conductivity 2-e sensors a jumper between F and G may be required.
4.5.2 ph and Redox (ORP) Analog Sensors
pH Redox (ORP)
Terminal Function Color1) Function Color
A Glass Transparent Platinum Transparent
BNotused –
CNotused – – –
DNotused –
E Reference Red Reference Red
FReference
2) – Reference
2)
GSolutionGND
2) Blue
3) SolutionGND
2)
HNotused – – –
I RTDret/GND White
J RTDsense – – –
KRTD Green – –
L Notused – – –
MShield(GND) Green/yellow Shield(GND) Green/yellow
N Notused – – –
ONotused –
PNotused – – –
QNotused –
1) Grey wire not used.
2) Install jumper between F and G for ORP sensors and pH electrodes without SG.
3) Blue wire for electrode with SG.
Transmitter M400 2(X)H 24
© 03/2023 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland M400 2-Wire
Printed in Switzerland 30 748 687 A
4.5.3 Amperometric Oxygen Analog Sensors
InPro 6800(G) InPro 6900 InPro 6950
Terminal Function Color Color Color
ANotused – – –
B Anode Red Red Red
C Anode 1) 1)
DReference –1) 1) Blue
E Notused – – –
FNotused – – –
GGuard – Grey Grey
HCathode Transparent Transparent Transparent
I NTCret(GND) White White White
J Notused – – –
K NTC Green Green Green
L Notused – – –
MShield(GND) Green/yellow Green/yellow Green/yellow
N Notused – – –
ONotused – – –
P+Ain
2) –––
Q–Ain
2) –––
1) Install jumper between C and D for InPro 6800(G) and InPro 6900.
2) 4 to 20 mA signal for pressure compensation
4.6 Terminal Block TB2: ISM Sensors
4.6.1 pH, Amperometric Oxygen, Conductiviy (4-e)
and Dissolved Carbon Dioxide ISM Sensors
Terminal Function Color
ANotused –
BNotused –
CNotused –
DNotused –
E Notused
FNotused –
GNotused –
HNotused –
I Notused
J Notused
K Notused
L1-wire Transparent(cablecore)
MGND Red(shield)
N RS485-B
ORS485-A –
P+Ain1)
Q–Ain1)
1) Only for Oxygen sensors: 4 to 20 mA signal for pressure compensation
Transmitter M400 2(X)H 25
© 03/2023 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland M400 2-Wire
Printed in Switzerland 30 748 687 A
4.6.2 Optical Oxygen ISM Sensors
Optical Oxygen with VP8 Cable1) Optical Oxygen with other Cables
2)
Terminal Function Color Function Color
ANotused – Notused –
BNotused – Notused –
CNotused – Notused –
DNotused – Notused –
E Notused – Notused –
FNotused – Notused –
GNotused – Notused –
HNotused – Notused –
I Notused D_GND(shield) Yellow
J Notused – Notused –
K Notused – Notused –
L Notused – Notused –
MD_GND(shield) Green/yellow D_GND(shield) Grey
N RS485-B Brown RS485-B Blue
ORS485-A Pink RS485-A White
P+Ain
3) – +Ain
3)
Q–Ain
3) – –Ain
3)
1) Connect the grey +24DC wire and the blue GND_24V wire of the sensor separately to an external power supply.
2) Connect the brown +24DC wire and the black GND_24V wire of the sensor separately.
3) 4 to 20 mA signal for pressure compensation