3
be discarded into a ‘safe’ container. (If circumstance
dictate that main filters are to be reused, they must be
stored in a manner that no contaminant is released into
the atmosphere or reach the clean side of the filter. The
useful life of filters must be determined by a competent
person taking site conditions into account).
c. The filter retainer may be cleaned in warm soapy water,
rinsed in cold water and allowed to dry naturally. DO
NOT use heat.
d. The outside of the Filter Unit should be cleaned with a
soft clean cloth, and warm soapy water. Do not use
excessive amounts of water or allow water to get inside
of Filter Unit. Wipe the Filter Unit with a clean cloth
moistened with cold water. Wipe off excess water with
a clean dry cloth and allow the Filter Unit to dry naturally.
DO NOT use heat.
CAUTION: Do not allow water to get into the Filter Unit. It will
cause irreparable damage to either the motor or
battery. Do not allow water to get into the battery
charging socket.
During cleaning it may be necessary to take
precautions against any harmful dusts and/or
vapors which may be on the Filter Unit or filters.
Do not use materials that are likely to damage
plastics, e.g.; gasoline, organic solvents, or abra-
sive cleaning agents to clean the Filter Unit.
1. CHANGING FILTERS (Refer to Fig. 2.1)
The Filter Unit comes with a prefilter and main filter.
a. Prefilter: The prefilter is fitted between the inside of the
louvre cover and the wire clip. It is replaced by unlatch-
ing the wire clip, removing used filler, placing the filter
against the inside of the louvre cover and pushing the
clip back into position until clicks.
TROUBLESHOOTING
NOTE: All troubleshooting procedures described below MUST be carried out in a clean air environment. There are NO user
serviceable parts inside the Respirator unit. Air flow should be checked according to “Airflow Test” described in Pre-
operational Inspection Section.
The following operational faults may be investigated by the user by observing these procedures:
Fault Possible Cause Actions
No air flow 1. Unit not switched on 1.Operate switch several times while observing air flow indicator
2. Battery pack disconnected 2.Check battery pack connection and perform air flow test again
3. Battery discharged 3.Recharge battery pack and repeat flow test
4. Clogged filter(s) 4.Check filter or filters for clogging or other obstruction; replace as
necessary
5. Blown fuse (2amp Littlefuse) 5.Located inside the filter unit. Remove filters and four(4) Phillip
screws to remove the inside plate.
Insufficient air flow 1. Battery discharged 1.Recharge battery pack and repeat air flow test.
2. Clogged filter(s) 2.Check filter or filters for clogging or other obstruction; replace as
necessary
Unit works with 1. Blown fuse (2amp Littlefuse) 1.Located inside the filter unit. Remove filters and four(4) Phillip
charger but not screws to remove the inside plate.
with battery pack
If the actions described above fail to produce a satisfactory result, then the Respirator unit must be returned to the supplier for
examination and/or repair.
b. Main Filter: Place louver filter retainer on top of main
filter.Rotate both red turn buckles to unlock position
(see sketch) and open over-center catch.
Place the main filter into the Filter Unit body, so the filter
seal is in contact with the body all around its periphery.
If the filter is placed “back-to-front” into the body it will be
obvious that the two pillars on the Filter Unit body, are
preventing the filter from being correctly positioned.
With turn buckles in “lock” position, close over-center catch
into red “I”bars. Rotate 90 degree to click into “locked” position.
2. REPLACEMENT OF BATTERY PACK
Remove belt from Filter Unit. The Battery pack is located under
the smaller of the two backplates of the Filter Unit.
Unlatch over-center catch plates. Unlatch and remove smaller
cover by releasing the two catches from under the belt loop
(see sketch). When replacing battery, ensure the plug is
correctly inserted into the socket and that the cable cannot be
trapped and damaged.
3. STORAGE
The Respirator should be stored in a clean, dry atmosphere
within the temperature range -4° F to 158° F. It should be
protected from direct sunlight and any material known to
damage plastics, e.g. gasoline and solvent vapors.
GENERAL
a. Respiratory Protection. The Filter Unit is designed to be
used with its own range of headgear. It is essential that
only genuine manufacturer’s spares and replacements
are used. Unauthorized alternatives may impair the
performance. Renew prefilters each shift and main
filters as determined by a competent person.
b. The Filter Unit is not recommended for use outside the
temperature range of 23° F to 122° F.