Loctite Zeta 7216 Operating instructions

Type
Operating instructions

This manual is also suitable for

Loctite Zeta 7216 is a high intensity electrodeless curing chamber. The cure chamber can cure light sensitive products with an electrodeless bulb. In comparison to mercury vapor lamps, this leads to a number of advantages, e.g., longer bulb life, no warm-up time, no ozone generation, and lower energy consumption.

Loctite Zeta 7216 is a high intensity electrodeless curing chamber. The cure chamber can cure light sensitive products with an electrodeless bulb. In comparison to mercury vapor lamps, this leads to a number of advantages, e.g., longer bulb life, no warm-up time, no ozone generation, and lower energy consumption.

EQUIPMENT
OPERATION MANUAL
Loctite
®
Combined Manual
for
High Intensity
Cure Chamber
Zeta
®
7215
98007 Mercury Arc Lamp
Zeta
®
7216
98008 Electrodeless Lamp
®
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 1
Contents
Zeta® 7215 Mercury Arc Chamber Page No.
1 For Your Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.1 Ozone Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.2 Purpose of Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.3 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.4 Unpacking & Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.5 Items Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.6 Recommended Spare Parts / Optional Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
3 Installation & Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
3.1 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
3.2 Space & Utility Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
3.3 Mercury Lamp Installation & Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
4 Control & Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
4.1 Operating the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
4.2 Illustration of Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
4.3 Chamber Door Knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
5 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
5.1 Reflector Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
5.2 Lamp Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
6 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
7 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
8 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Zeta® 7216 Electrodeless Lamp Page No.
1 For Your Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
2.1 Ozone Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
2.2 Purpose of Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
2.3 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
2.4 Unpacking & Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
2.5 Items Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
2.6 Recommended Spare Parts / Optional Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
3 Installation & Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
3.1 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
3.2 Space & Utility Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
4 Control & Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
4.1 Operating the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
4.2 Illustration of Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
4.3 Chamber Door Knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
5 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
6 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
7 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
8 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Wiring Diagram for the Cure Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Wiring Diagram for the Mercury Arc Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 2
Zeta
®
7215
Mercury Arc Ultraviolet Curing Chamber
Instruction Manual 98007
Loctite
®
Zeta
®
7215
High Intensity Cure Mercury Arc Chamber
1
LAMP POWER
CONTROL POWER
OFF
ON
OFF
ON
BEFORE CLEANING, SERVICING OR RELAMPING,
OR LAMP SHIELDING PROVIDED WHEN LAMP IS IN USE.
SUITABLE EYE AND SKIN PROTECTION MUST BE WORN
NOT EXPOSED TO DIRECT OR REFLECTED RADIATION.
MUST BE TAKEN TO INSURE THAT PERSONNEL ARE
WHICH IS HARMFUL TO EYES AND SKIN. GREAT CARE
ULTRAVIOLET IRRADIATORS EMIT RADIATION
ALWAYS TURN THE POWER OFF.
CAUTION
P/N 984132
STANDBY MODE
MAIN POWER
HIGH MODE
ELAPSED TIME
7215
P/N 98007
READ OPERATOR'S MANUAL BEFORE USE
prevent user exposure to UV light.
This unit has been designed to
ULTRAVIOLET LIGHT
WARNING
CYCLE TIMER
CYCLE START
CYCLE STOP
ON
OFF
CHAMBER POWER
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 3
For Your Safety1
IMPORTANT DO NOT OPERATE UNIT BEFORE READING THIS SECTION
WARNING: Ultraviolet Light–Danger of severe eye or skin burns. Do not look at source or its
reflection without protective goggles. Do not defeat lamp shielding interlocks. Read operator's
manual before use or repair.
The Loctite
®
ZETA
®
7215 High Intensity Cure Chamber is a source of ultraviolet light used to
cure light sensitive products. The system incorporates ultraviolet shielding, an interlocked
chamber door, and a shutter to protect the user. It is safe when installed and used properly. The
following guidelines should be followed at all times to ensure protection from the ultraviolet light.
1. All personnel involved in the operation of the ZETA
®
7215 High Intensity Cure Chamber
should become familiar with the safety and operating procedures provided in the written material
provided with this equipment. Ultraviolet burns are not felt for several hours and short exposure
can cause severe burning of the skin. Never look directly at the output of the lamp or the output
of the lamp reflector. The ZETA
®
7215 is shielded and interlocked to prevent exposure to
ultraviolet radiation.
2. Do not defeat any of the system interlocks.
3. Do not attempt to open chamber door during the cure cycle. A cycle stop button is provided to
terminate curing and close the protective shutter. If the door is opened the unit turns down
power and closes the shutter, but the best procedure is to push cycle stop and wait for the
shutter to close prior to opening the chamber door.
4. Prior to any attempt at servicing, turn power off and remove power from all of the system
components.
5. When changing lamp, allow 15 minutes for the lamp to cool prior to removing.
Ultraviolet safety is the responsibility of the user. When in doubt about ultraviolet safety, consult
a Loctite technical specialist at 1-800-LOCTITE (562-8483).
2
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 4
Introduction2
2.1 OZONE WARNING
The ZETA
®
7215 Ultraviolet High Intensity Cure Chamber has been designed to operate safely
and reliably. However, if not used properly, it may become a potential hazard. All personnel
involved in the operation of the ZETA
®
7215 should be familiar with the operating safety
described in the Safety Considerations section of this manual. Ozone (O
3
) is a by-product of the
U.V. lamp. Ozone is a natural gas that is formed when Ultraviolet light from the sun is exposed
to the earth's atmosphere. Ozone has a pungent irritating odor. U.V. systems should not be
constantly operated in a confined room without proper ventilation. Exhausting the system to the
outside of the building will prevent ozone concentration in confined areas. Exhausting the ozone
is not dangerous to the atmosphere because the hot gas is unstable and breaks down to
oxygen usually before it leaves the ducting. Questions regarding installation, operation, or
safety should be addressed to Loctite Corporation Safety Health and Regulatory Affairs
Department in Rocky Hill, Connecticut, 1-800-LOCTITE (562-8483).
2.2 PURPOSE OF MANUAL
The purpose of this manual is to provide installation, operation and troubleshooting instructions
for the ZETA
®
7215 Ultraviolet High Intensity Cure Chamber. It is important the manual be
carefully read before any attempt to install or operate the equipment.
2.3 DESCRIPTION
The system is a bench top ultraviolet source designed for a wide range of Loctite
®
ultraviolet curing adhesives and coatings. The unit is suitable for laboratory use or
production requirements.
2.4 UNPACKING & INSPECTION
Visually inspect the shipping carton for physical damage. Damaged shipping cartons should be
reported to the carrier. Carefully open the shipping carton and remove the equipment, being
careful not to lose or damage any separately packed parts.
Check all parts against the packing list. Any damaged or missing parts should be reported to the
carrier and to your Loctite representative.
All packing materials and shipping carton should be saved in case the unit has to be returned at
some time in the future.
3
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 5
Introduction, cont.2
2.5 ITEMS SUPPLIED
1. Cure Chamber Base and Lamp Adapter Assembly LOCTITE PART # 984130
2. 6" Mercury Arc Lamp Power Supply LOCTITE PART # 984132
3. 6" Mercury Arc Lamp Irradiator Assembly LOCTITE PART # 984134
4. Equipment Manual LOCTITE PART # 984135
2.6 RECOMMENDED SPARE PARTS / OPTIONAL ITEMS
1. Reflector Liner Set LOCTITE PART # 984143
2. 6" Mercury Arc Lamp LOCTITE PART # 983091
3. Flow Through Air Panel LOCTITE PART # 984144
4. Capacitor, 18UF LOCTITE PART # 984145
5. Capacitor, 12UF LOCTITE PART # 984146
6. Capacitor, 6UF LOCTITE PART # 984147
7. Ballast, 9526 LOCTITE PART # 984148
8. Chamber Curing Shelf LOCTITE PART # 984142
9. Fuse, Chamber BUSS # ATQ15
10.Fuse, Power Supply BUSS # ATQ20
11. Fuse, Chamber Drive BUSS # AGC3
4
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 6
Installation & Start-Up3
3.1 INSTALLATION
Refer to the installation diagram on the next page. The diagram gives the exterior dimensions of
the High Intensity Cure Chamber and the Power Supply. Allow six inches of clearance around
the unit for air circulation and cable connections. Air is drawn from the bottom of the Chamber
and exhausted from the top out of the exhaust duct. Do not block air inlets.
Power requirements and the plug formats for the power cords are shown. Prior to applying
power, the Mercury Arc Lamp must be installed. Refer to page 7 Lamp Installation and
replacement. Install Lamp. The Lamp is a 1,500 watt Mercury Arc Lamp.
Connect all cables as shown in the Installation diagram.
A five inch exhaust connection is provided and should be connected as described in the
installation diagram.
After plugging in the Cure Chamber and The Mercury Arc Power Supply proceed to the
operating instructions.
5
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 7
Installation & Start-Up, cont.3
6
3.2 SPACE & UTILITY REQUIREMENTS
Remove power from all system
components prior to servicing.
Do not defeat lamp system interlocks.
WARNING
LB-11002
EXHAUST AIRFLOW
18 3/8
POWER SUPPLY ENCLOSURE
CURE CHAMBER BASE UNIT
DOOR SWINGS FROM RIGHT TO LEFT
INTERLOCK CABLE
HIGH VOLTAGE LAMP CABLE
POWER INPUT CORDPOWER INPUT CORD
21 1/2
7215
SHUTTER TIMER
CYCLE START
CYCLE STOP
ON
OFF
CHAMBER POWER
P/N 98007
READ OPERATOR'S MANUAL BEFORE USE
prevent user exposure to UV light.
This unit has been designed to
ULTRAVIOLET LIGHT
WARNING
LAMP POWER
CONTROL POWER
OFF
ON
OFF
ON
STANDBY MODE
P/N 984132
MAIN POWER
HIGH MODE
ELAPSED TIME
BEFORE CLEANING, SERVICING OR RELAMPING,
OR LAMP SHIELDING PROVIDED WHEN LAMP IS IN USE.
SUITABLE EYE AND SKIN PROTECTION MUST BE WORN
NOT EXPOSED TO DIRECT OR REFLECTED RADIATION.
MUST BE TAKEN TO INSURE THAT PERSONNEL ARE
WHICH IS HARMFUL TO EYES AND SKIN. GREAT CARE
ULTRAVIOLET IRRADIATORS EMIT RADIATION
ALWAYS TURN THE POWER OFF.
CAUTION
5 INCH EXHAUST VENT
HOSE, MAXIMUM LENGTH
40 FEET, NO INCOMING
POSITIVE PRESSURE
20.0014.25
10.25
Figure 1.Top View
Figure 2. Front View
Electrical Service Requirements
120 Volts 20 Amp Circuit
High Inrush Magnetic Breaker
120 Volts 15 Amp Circuit
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 8
Installation & Start-Up, cont.3
7
3.3 MERCURY LAMP INSTALLATION & REPLACEMENT
Disconnect power to chamber, lamp power supply and any other power source connected
to the system prior to start of Lamp installation.
3.3.1 REMOVING THE LAMP MODULE
1. Remove the power supply cable.
2. Remove the four Lamp module
retaining screws.
3. Lift module straight up using handles.
4. Place Lamp Module on bench inverted.
5. To make sure the Lamp is cool, allow
15 minutes from powering down.
3.3.2 INSTALLING THE LAMP
1. Loosen, do not remove, the four
screws on lamp holder. (2 on each side
of the lamp.)
2. Open the lamp retainers.
3. Wearing cotton gloves, remove the
lamp from packing, place in the holder.
4. Close the lamp retainers and tighten
the four screws. Lamp will move but
not come out of retainer.
5. Connect the wires on the lamp to
ceramic insulators.
6. Reinstall the lamp module.
7. Press the hourmeter reset on the
back of the power supply.
Lamp Loctite Part # 983091
POWER SUPPLY CABLE
SCREWSSCREWS
CERAMIC INSULATORS
SCREWS
SCREWS
Figure 3.Top of cure chamber
Figure 4. Lamp Module
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 9
Control & Operation4
8
4.1 OPERATING THE UNIT
All materials relating to safety should be reviewed prior to operating.The chamber and
lamp power supply should be properly installed. Assure that all cables are properly
installed, the lamp has been installed and the Chamber door is closed.
1. Refer to 7215 figures 5-7 on page 9. Push the Chamber Power On button located at the
lower right side of the Chamber. The amber light on the Chamber Power switch block
should light. See figure 5.
2. Push the Control Power On button located on the Lamp Power Supply. The amber light on
Control Power Switch Block should light. See figure 6.
3. Push the Lamp Power On button located on the Lamp Power Supply. The amber light on
Lamp Power Switch Block should light. See figure 6.
4. Wait for the Cycle Start Lamp/Button to illuminate. (Located at the center right side of the
Chamber.) See figure 5.
5. Set Cycle timer for desired cure time. (Cycle timer setting instructions in figure 7.)
6. Open Chamber door and adjust shelf for desired intensity. The maximum ultraviolet intensity
is available at the top of the chamber and the minimum at the bottom shelf position. If you are
not sure about the proper shelf setting utilize a Loctite
®
ZETA
®
7010 Puck Radiometer. Using
the ZETA
®
7010, a measurement of intensity can be obtained at any shelf position. This
measurement can be used to determine proper part placement and day to day process
repeatability. The ZETA
®
7010 can be placed in the oven, the Chamber door closed, then
when the cycle start button is pushed, the ZETA
®
7010 will detect and record the ultraviolet
light when the shutter opens. After the shutter closes the ZETA
®
7010 can be removed from
the chamber and intensity or total energy read on the digital display. Contact your Loctite
representative at 1-800-LOCTITE for the ZETA
®
7010 Puck Radiometer part number 98001.
Parts should be placed near the center of the shelf. If you have no means of measuring
intensity, it is recommended that test samples are placed near the bottom of the chamber
then moved up if the intensity is not sufficient. Do not place parts so high as to interfere with
the shutter.
7. Close the Chamber door. Refer to the illustration of Chamber Door Knob figure 8 on page 10.
Push the Cycle Start button. The part can be observed through the window located above the
Chamber door. If it becomes necessary to end cure cycle, push the CYCLE STOP BUTTON
and the shutter will close.
8. At the end of the cycle remove parts from Chamber.
9. To shut down the system, first push the Lamp Power Off, then the Control Power Off, wait
three minutes then turn Chamber Power off. The three minute time period allows the fans to
provide a lamp cooling period.
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 10
Control and Operation, cont.4
9
4.2 ILLUSTRATION OF CONTROLS
20.0
CYCLE TIMER
7215
P/N 98007
READ OPERATOR'S MANUAL BEFORE USE
prevent user exposure to UV light.
This unit has been designed to
ULTRAVIOLET LIGHT
WARNING
CYCLE TIMER
CYCLE START
CYCLE STOP
ON
OFF
CHAMBER POWER
LAMP POWER
CONTROL POWER
OFF
ON
OFF
ON
BEFORE CLEANING, SERVICING OR RELAMPING,
OR LAMP SHIELDING PROVIDED WHEN LAMP IS IN USE.
SUITABLE EYE AND SKIN PROTECTION MUST BE WORN
NOT EXPOSED TO DIRECT OR REFLECTED RADIATION.
MUST BE TAKEN TO INSURE THAT PERSONNEL ARE
WHICH IS HARMFUL TO EYES AND SKIN. GREAT CARE
ULTRAVIOLET IRRADIATORS EMIT RADIATION
ALWAYS TURN THE POWER OFF.
CAUTION
P/N 984132
STANDBY MODE
MAIN POWER
HIGH MODE
ELAPSED TIME
Timer Set
Programs new time setting
after making changes.
Move to next digit on left,
wrap around at far left.
20.0
20.0
20.0
20.1
Increment digit up by 1
Not Used
Figure 7. Cycle Timer
Figure 6. Power Supply
Figure 5. Cure Chamber
If you step more than six digits to the left of the decimal
point, a minus appears. Continue to step (Do not hit the red
program button) and the selection will go back to the digit
two places to the left of the decimal point. If someone does
select the minus, it can be corrected by going back to the
minus and hitting the program (red) button to toggle the
minus off. You then can program the desired time normally.
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 11
Control & Operation, cont.4
10
4.3 CHAMBER DOOR KNOB
7215
P/N 98007
READ OPERATOR'S MANUAL BEFORE USE
prevent user exposure to UV light.
This unit has been designed to
ULTRAVIOLET LIGHT
WARNING
CYCLE TIMER
CYCLE START
CYCLE STOP
ON
OFF
CHAMBER POWER
If the Chamber Interlock will not close, reduce the tension by turning the Black Tension Knob
counter-clockwise.
If the Chamber door does not close tightly, adjust the Tension Knob clockwise.
Chamber Interlock
Adjusts Chamber
Interlock Tension
Figure 5. Cure Chamber
Figure 8. Chamber Door Knob
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 12
Maintenance5
ALWAYS REMOVE POWER FROM ALL UNITS CONNECTED TO THE SYSTEM PRIOR TO
THE START OF ANY SERVICING!
With proper care and maintenance, the U.V. lamp can be expected to have a useful life of
approximately 1,000 hours. To install or change the lamp refer to section 3.3 in this manual
entitled Mercury Lamp Installation. In addition to lamp maintenance the reflectors should be
cleaned to insure proper output. Cleaning is suggested once every 40 hours. If deposits do not
appear on reflectors, cleaning can be delayed.
Cotton Gloves
Lamps should never be touched with bare hands. Always use clean cotton gloves when
handling U.V. lamps. Oils from fingers can become permanently etched on the bulb when turned
on. This will reduce the lamp output and reduce the life of the lamp.
5.1 REFLECTOR CLEANING
To clean the reflectors, wearing white cotton gloves, remove the lamp. Using a soft cotton cloth
dampened with alcohol, wipe the reflector surfaces gently. After cleaning, wipe surfaces dry with
a soft clean cotton cloth. Replace the lamp, then the Lamp assembly.
5.2 LAMP CLEANING
Wait until lamp is cool prior to cleaning. If deposits are present on the outside of the lamp, the
lamp can also be wiped clean with a slightly alcohol-dampened cotton cloth.
11
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 13
Troubleshooting6
Specifications7
12
SYMPTOM POSSIBLE CAUSE ACTION
System does not Open Interlock switch Check chamber door lamp
turn on module four retainers
Open fuse Fuse in Chamber
U.V. Lamp does Open fuse Fuse in power supply
not light
Lamp failure Replace Lamp
Cure speed declines End of Lamp’s useful life Replace Lamp
Lamp dirty See maintenance
Reflector dirty See maintenance
Outdated adhesive Replace adhesive
Lamp 1,500 Watt Mercury Arc Lamp
Power Supply 120 VAC, 20A, 60 HZ, 1 Phase
Power Supply Fuse 20 A
Power Supply Consumption 2,400 Watts
Dimensions See Installation Drawing
System Shipping Weight 145 Lbs.
Power Supply Fuse 20A
Chamber Fuse 15A
Chamber Drive Fuse 3A
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 14
Loctite expressly warrants that all products referred to in this Instruction Manual for the ZETA
®
7215 Mercury Arc Cure Chamber (hereafter called "Products") shall be free from defects in
materials and workmanship. Liability for Loctite shall be limited, at its option, to replacing those
Products which are shown to be defective either in materials or workmanship or to credit to the
purchaser the amount of the purchase price thereof (plus freight and insurance charges paid
therefor by the user). The purchaser's sole and exclusive remedy for breach of warranty shall be
such replacement or credit.
A claim of defect in materials or workmanship in any Products shall be allowed only when it is
submitted to Loctite in writing within one month after discovery of the defect or after the time the
defect should have reasonably have been discovered and in any event, within twelve months
after the delivery of the Products to the purchaser excluding the Lamp. No such claim shall be
allowed in respect of Products which have been neglected or improperly stored, transported,
handled, installed, connected, operated, used or maintained or in the event of unauthorized
modification of the Products including; where products, parts or attachments for use in
connection with the Products are available from Loctite, the use of products, parts or
attachments which are not manufactured by Loctite.
No Products shall be returned to Loctite for any reason without prior written approval from Loctite.
Products shall be returned freight prepaid in accordance with instructions from Loctite.
NO WARRANTY IS EXTENDED TO ANY EQUIPMENT WHICH HAS BEEN ALTERED,
MISUSED, NEGLECTED, OR DAMAGED BY ACCIDENT, OR IF THE SYSTEM IS USED
TO DISPENSE OR CURE ANY LIQUID MATERIAL OTHER THAN LOCTITE
CORPORATION PRODUCTS.
EXCEPT FOR THE EXPRESS WARRANTY CONTAINED IN THIS SECTION, LOCTITE MAKES
NO WARRANTY OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, WITH RESPECT TO
THE PRODUCTS.
ALL WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, AND
OTHER WARRANTIES OF WHATEVER KIND (INCLUDING AGAINST PATENT OR
TRADEMARK INFRINGEMENT) ARE HEREBY DISCLAIMED BY LOCTITE AND WAIVED BY
THE PURCHASER.
THIS SECTION SETS FORTH EXCLUSIVELY ALL OF LIABILITY FOR LOCTITE TO
THE PURCHASER IN CONTRACT, IN TORT OR OTHERWISE IN THE EVENT OF
DEFECTIVE PRODUCTS.
WITHOUT LIMITATION OF THE FOREGOING, TO THE FULLEST EXTENT POSSIBLE UNDER
APPLICABLE LAWS, LOCTITE EXPRESSLY DISCLAIMS ANY LIABILITY WHATSOEVER FOR
ANY DAMAGES INCURRED DIRECTLY OR INDIRECTLY IN CONNECTION WITH THE SALE
OR USE OF, OR OTHERWISE IN CONNECTION WITH, THE PRODUCTS, INCLUDING,
WITHOUT LIMITATION, LOSS OF PROFITS AND SPECIAL, INDIRECT OR CONSEQUENTIAL
DAMAGES, WHETHER CAUSED BY NEGLIGENCE FROM LOCTITE OR OTHERWISE.
Warranty8
13
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 15
14
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 16
15
Zeta
®
7216
Electrodeless Lamp Ultraviolet Curing Chamber
Instruction Manual 98008
Loctite
®
Zeta
®
7216
High Intensity Cure
Electrodeless
Lamp Chamber
7216
+
-
+
-
SYSTEM
UNIT
LAMP CONTROL
INDICATOR TEST
STANDBY
/RESET
LAMP OFF
LAMP ON
LAMP OUT
PWR IMBAL
SYS BLWR
EXT INTLK
RF INTLK
PWR ERROR
BLWR PRESSURE
PS TEMP
PHOTOCELL
TEST POINTS
(0-12 VDC)
P300M POWER SUPPLY
TEST POINTS
MAGNETRON
(0-800 mVDC)
POWER
O
I
P/N 98008
WARNING
ULTRAVIOLET LIGHT
This unit has been designed to
prevent user exposure to UV light.
READ OPERATOR'S MANUAL BEFORE USE
CHAMBER POWER
OFF
ON
CYCLE STOP
CYCLE START
CYCLE TIMER
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 17
16
For Your Safety1
IMPORTANT DO NOT OPERATE UNIT BEFORE READING THIS SECTION
WARNING: Ultraviolet Light–Danger of severe eye or skin burns. Do not look at source or its
reflection without protective goggles. Do not defeat lamp shielding interlocks. Read operator's
manual before use or repair.
The Loctite
®
ZETA
®
7216 High Intensity Cure Chamber is a source of ultraviolet light used to
cure light sensitive products. The system incorporates ultraviolet shielding, an interlocked
chamber door, and a shutter to protect the user. It is safe when installed and used properly. The
following guidelines should be followed at all times to ensure protection from the ultraviolet light.
1. All personnel involved in the operation of the ZETA
®
7216 High Intensity Cure Chamber
should become familiar with the safety and operating procedures provided in the written material
provided with this equipment. Ultraviolet burns are not felt for several hours and short exposure
can cause severe burning of the skin. Never look directly at the output of the lamp or the output
of the lamp reflector. The ZETA
®
7216 is shielded and interlocked to prevent exposure to
ultraviolet radiation.
2. Do not defeat any of the system interlocks.
3. Do not attempt to open the chamber door during the cure cycle. A cycle stop button is
provided to terminate curing and close the protective shutter. If the door is opened, the unit turns
down power and closes the shutter, but the best procedure is to push the cycle stop and wait for
the shutter to close prior to opening the chamber door.
4. Prior to any attempt at servicing, turn the power off and remove power from all of the system
components.
5. When changing the lamp, allow 15 minutes for the lamp to cool prior to removing.
Ultraviolet safety is the responsibility of the user. When in doubt about ultraviolet safety, consult
a Loctite technical specialist at 1-800-LOCTITE (562-8483).
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 18
Introduction2
17
2.1 OZONE WARNING
The ZETA
®
7216 Ultraviolet High Intensity Cure Chamber has been designed to operate safely
and reliably. However, if not used properly, it may become a potential hazard. All personnel
involved in the operation of the ZETA
®
7216 should be familiar with the operating safety
described in the Safety Considerations section of this manual. Ozone (O
3
) is a by-product of the
U.V. lamp. Ozone is a natural gas that is formed when Ultraviolet light from the sun is exposed
to the earth's atmosphere. Ozone has a pungent irritating odor. U.V. systems should not be
constantly operated in a confined room without proper ventilation. Exhausting the system to the
outside of the building will prevent ozone concentration in confined areas. Exhausting the ozone
is not dangerous to the atmosphere because the hot gas is unstable and breaks down to
oxygen usually before it leaves the ducting. Questions regarding installation, operation, or safety
should be addressed to Loctite Corporation Safety Health and Regulatory Affairs Department in
Rocky Hill, Connecticut, 1-800-LOCTITE (562-8483).
2.2 PURPOSE OF MANUAL
The purpose of this manual is to provide installation, operation and troubleshooting instructions
for the ZETA
®
7216 Ultraviolet High Intensity Cure Chamber. It is important that this manual, plus
the electrodeless lamp manual (F300S/F300SQ), be carefully read before any attempt to install
or operate the equipment.
2.3 DESCRIPTION
The system is a bench top ultraviolet source designed for a wide range of Loctite
®
ultraviolet
curing adhesives and coatings. The unit is suitable for laboratory use or low production
requirements.
2.4 UNPACKING & INSPECTION
Visually inspect the shipping carton for physical damage. Damaged shipping cartons should be
reported to the carrier. Carefully open the shipping carton and remove the equipment, being
careful not to lose or damage any separately packed parts.
Check all parts against the packing list. Any damaged or missing parts should be reported to the
carrier and to your Loctite representative.
All packing materials and shipping carton should be saved in case the unit has to be returned at
some time in the future.
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 19
18
Introduction, cont.2
2.5 ITEMS SUPPLIED
1. Cure Chamber Base and Lamp Adapter Assembly LOCTITE PART # 984136
2. Electrodeless Lamp power Supply Assembly, 208 VOLT LOCTITE PART # 984138
3. Electrodeless Lamp Irradiator Assembly LOCTITE PART # 984139
4. RF Detector Cable LOCTITE PART # 984158
5. Power Supply to Chamber Interlock Cable LOCTITE PART # 984159
6. Power Supply To Lamp High Voltage Cable LOCTITE PART # 984160
7. Equipment Manual LOCTITE PART # 984135
2.6 RECOMMENDED SPARE PARTS / OPTIONAL ITEMS
1. Capacitor, .85/.43 MFD LOCTITE PART # 984192
2. Diode Block LOCTITE PART # 984193
3. Irradiator Photocell Assembly LOCTITE PART # 983501
4. Irradiator Magnetron Kit LOCTITE PART # 983502
5. Irradiator RF Screen Kit LOCTITE PART # 983503
6. Irradiator Ignitor Bulb LOCTITE PART # 983504
7. Blower Filter, Set of 20 LOCTITE PART # 983505
8. Irradiator Air Pressure Switch LOCTITE PART # 983506
9. Irradiator "D" Bulb LOCTITE PART # 983469
10.Irradiator "H" Bulb LOCTITE PART # 983470
11. Irradiator "V" Bulb LOCTITE PART # 983471
12.Flow Through Air Panel LOCTITE PART # 984144
13.Chamber Curing Shelf LOCTITE PART # 984142
984135 Zeta 7215 & 7216 manl 8/3/00 3:24 PM Page 20
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32

Loctite Zeta 7216 Operating instructions

Type
Operating instructions
This manual is also suitable for

Loctite Zeta 7216 is a high intensity electrodeless curing chamber. The cure chamber can cure light sensitive products with an electrodeless bulb. In comparison to mercury vapor lamps, this leads to a number of advantages, e.g., longer bulb life, no warm-up time, no ozone generation, and lower energy consumption.