Extech Instruments RH210 User manual

  • Hello! I am an AI chatbot trained to assist you with the Extech Instruments RH210 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
UserGuide
HygroThermometer
ModelRH210
RH210-en-GB_V1.0 12/14
2
Introduction
ThankyouforselectingtheExtechRH210HygroThermometerwhichsimultaneouslydisplays
RelativeHumidity,Temperature,andDewPoint.Measurementsaremadewitharemote
electrode.Thisdeviceisshippedfullytestedandcalibratedand,withproperuse,willprovide
yearsofreliableservice.Pleasevisitourwebsite(www.extech.com)forproductregistration,
support,andtocheckforthelatestversionofthisUserGuide.
Features
AirTemperature,RelativeHumidity,andDewpointmeasurements
Detachableprobeforeaseofuseinavarietyofmeasurementenvironments
Fasthumiditymeasuringresponsetime
Autoofffeaturetosavebatterylife
LCDdisplayshowsTemperatureand%RHorDewpointsimultaneously
DATAHOLDforholdingdisplayedreading
RecordandViewMin/Maxreadings
Programmingmodeforselecting
o
C/
o
Ftemperatureunitsandenabling/disablingtheauto
poweroffutility
Lowbatteryindicator
RH210-en-GB_V1.0 12/14
3
1
2
3
4
5
6
7
8
9
MeterDescription
1. HumidityorDewPointreading
2. Temperaturereading
3. Powerbutton
4. DataHoldbutton
5. Recordbutton
6. Functionbutton
7. Probeconnection
8. SETbutton(underflap)
9. Probe
NOTE:Batterycompartmentonbackofmeter
MeterPower
Themeterispoweredbyone9Vbattery;pressthePowerbuttontoswitchthemeterON.Press
andholdthePowerbuttonforatleast2secondstoswitchthemeterOFF.
WhenpoweringON,themeterperformsaselftestwherebytheLCDdisplaycountsdowntozero
beforestabilizingatthenormaloperationscreen.
IfthemeterdoesnotswitchONwhenprompted,orifthelowbatteryiconisshowingontheLCD,
pleasereplacethebattery.Refertothe
batteryinstallationsectionfordetailedinformation.
TheAutoPowerOFFfeatureswitchesthemeterOFFaftera10minuteperiodofinactivity.To
defeatthisfeatureplease refertotheSetupModesection.
RH210-en-GB_V1.0 12/14
4
Measurements
TakingRelativeHumidity/DewPointandAirTemperatureMeasurements
1. Connecttheelectrodetotheprobeinputsocketlocatedonthebottomofthemeter.
2. SwitchthemeterONbypressingthePowerbutton.
3. ThedisplaywillindicateAirtemperatureonthelowerleft(insmalldigitformat)and
either%RHorDewpointtemperatureontheupper
right.
4. PresstheFunctionbuttontoselecteither%RHorDewPointTemperature.
5. PresstheHoldbuttontofreezethecurrentvaluesonthedisplay.
6. PresstheRECbuttontoaccesstheMINMAXRecordingmode.Seedetailedinformation
intheMINMAXRecordingsection.
7. Press
andholdthePowerbuttonfor2secondstoturnthemeterOFF.
DataHold
1. PresstheHOLDbuttontofreezethedisplayedvalue.
2. TheLCDwilldisplayHOLDalongwiththeheldreading.
3. PresstheHOLDbuttonagaintoreleasethedataholdfunction.TheHOLDdisplayicon
willswitchOFF.
MINMAXRecording
Whenselected,theDataRecordingfunctionrecordstheMIN(lowest)andMAX(highest)
readings:
1. PresstheRECbuttononcetostartrecordingMINandMAXreadings.TheRECindicator
willappearonthedisplay.
2. PressRECagainandRECMAXwilldisplayonthescreenalongwiththe
maximumvalue.
3. PressRECagainandthedisplaywillshowRECMINalongwiththeminimumvalue.
4. ToviewtherealtimereadingswhilecontinuingtotracktheMINandMAXvalues,press
theHOLDbuttonwhileeithertheRECMAXorRECMINiconsarevisible;nowonly
the
RECiconwillbevisible.Themeterisnowcontinuingtomonitorthehighest(MAX)and
lowest(MIN)readingsbutisdisplayingrealtimemeasurements.ToviewtheMAXand
MINvaluesagainusetheRECbuttonasdescribedearlier.
5. ToexittheRecordmodeandresettheMINandMAXvalues,pressandholdtheREC
buttonforatleast2seconds.Thedisplaywillreturntothenormaloperatingmodeand
theREC/MAX/MINiconswillalldisappear.
RH210-en-GB_V1.0 12/14
5
Setupmode
Note:TheSETUPModecannotbeaccessediftheHOLDortheMINMAXfunctionisactive.
1. Opentheflaponthemeter’srightsidetoaccesstherecessedSETbutton.
2. UseapapercliporsmallscrewdrivertopressandholdtheSETbuttonforatleast3seconds
untilthemeterbeepstwice.
3. The‘Code1000’promptshouldappearonthedisplay.Proceed
withthestepsbelow:
ChangetheTemperatureUnitsofMeasure(F/C)
With‘Code1000’showingonthedisplay(asexplainedabove),followthestepsbelow:
1. PressRECandthelowerdisplaylinewillshowtwounitsofmeasureicons.Thesmaller
digitsindicatetheunitofmeasurethatwillbeadoptedwhentheseprogrammingsteps
arecomplete.
2. ToselectF,presstheFunctionbuttonsothatthenumber1appearsandthenpress
REC.
3. ToselectC,presstheFunctionbuttonsothatthenumber0appearsand
thenpress
REC.
4. PresstheHOLDbuttontosavethesettingandreturntothenormaloperatingmode.The
displayshouldnowbeindicatingtheprogrammedunitofmeasure.
EnableDisabletheAutoPowerOFFutility
With‘Code1000’showingonthedisplay(asexplainedabove),followthestepsbelow:
1. PresstheRECbuttonandthenpresstheSETbutton(briefly)toaccesstheAutoOff
mode.
2. ToenabletheAutoOFFmode,presstheFunctionbuttontoindicatea1onthedisplay,
andthenpresstheRECbutton.
3. TodisabletheAuto–OFFmode,presstheFunctionbuttontoindicatea0onthe
display,andthenpresstheRECbutton.
4. PresstheHOLDbuttontosavethesettingandreturntothenormaloperatingmode.

RH210-en-GB_V1.0 12/14
6
BatteryReplacement
The9Vbatterythatpowerstheinstrumentrequiresreplacingwhenthebatteryiconflasheson
thedisplay.Toreplacethebatteryfollowthestepsbelow:
1. SwitchthemeterOFF.
2. Disconnecttheprobefromtheinstrument.
3. OpentherearbatterycompartmentusingascrewdrivertoremovethetwoPhillipshead
screwsthatsecurethecompartment.
4. Carefullyremovetheoldbatteryfromthecompartmentandgentlyunsnapthebattery’s
terminalsfromthewired(red/black)connecto rs.
5. Installanew9Vbatterybysnappingthebatteryterminalsontothewiredconnectors
(observingcorrectpolarity)andinsertingthebatteryintothecompartment.
6. Installthebatterychamberintothemeterhousingandsecurewiththetwoscrews.
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriate
collectionsites,theretailstorewhere thebatterieswerepurchased,orwherever
batteries
aresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuserisobligatedtotake
endoflifedevicestoadesignatedcollectionpointforthedisposalofelectricalandelectronic
equipment.
RH210-en-GB_V1.0 12/14
7
Specifications
GeneralSpecifications
Measurements Temperature,RelativeHumidity,Dewpointtemperature
Datahold Dataholdfreezesdisplayedreading
Memoryrecall Min/Maxreadingsstoreandrecall
Autopoweroff Meterswitchesoffafter10minutes(canbedisabled)
Samplerate 0.8second(approx.)
Operatingconditions Temperature:0to50°C(32to122°F);Humidity:<85%RH
Probe:0to95%RH
Batterypower 9Valkalinebattery
PowerConsumption Approx.6.0mADC
Weight 295g(0.65lbs.)includingbatteries&probe
Dimensions Meter:135x60x33mm(5.3x2.4x1.3")
Probe:197mm(7.8”)length;15mm(0.6”)diameteroftip
ElectricalSpecificationsat23°±5°C(73°F±9°F)
Measurement Range Resolution Accuracy
Relativehumidity 10%to95%RH 0.1%RH
>70%RH:+/(3%rdg+1%RH)
<70%RH:+/3%RH
DewPoint
25.3°Cto48.9°C 0.1°C Calculatedfrom
Temperatureand%RH
13.5°Fto120.1°F 0.1°F
Temperature
0°Cto50°C 0.1°C ±0.8°C
32°Fto122°F 0.1°F ±1.5°F
Note:AbovespecificationtestsconductedunderanRFfieldstrengthlessthan3V/Mand
frequencylessthan30MHz