TH30

Extech Instruments TH30, RHT30 User manual

  • Hello! I am an AI chatbot trained to assist you with the Extech Instruments TH30 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
  • What is the default logging interval of the Extech Instruments TH30?
    What is the maximum logging interval of the Extech Instruments TH30?
    What is the maximum delay before logging can begin on the Extech Instruments TH30?
    What temperature recording units can the Extech Instruments TH30 use?
    What display elements does the TH30 have?
    What data formats can be generated for the logging data?
USERMANUAL
USBMultiFunctionDataloggers
RHT30Humidity/TemperatureDatalogger
TH30DualTemperatureDatalogger
AdditionalUserManualTranslationsavailableatwww.extech.com
2RHT30_TH30enGB_V1.0(11/16)
Introduction
ThankyouforselectingtheExtechmultifunction,easytouse,portableUSB
datalogger.TheRHT30offerstemperatureandhumidityloggingandtheTH30offers
internal(air)andexternal(probe)temperaturelogging.
TheseUSBdataloggerscanbeconfiguredandcustomizedbytheuserbypluggingthem
intoaPCUSB
port,followingtheonscreenprompts,andmakingselectionsonthe
supplied‘PDFLoggerConfigurationTool’.
Thesedevicesareshippedfullytestedandcalibratedand,withproperuse,willprovide
yearsofreliableservice.Pleasevisitourwebsite(www.extech.com)tocheckforthe
latestversionofthisUserManual,ProductUpdates,andCustomerSupport.
Features
PlugandplaydataloggerwithUSBconnector.NoUSBdriverrequired
UserconfigurabledataloggingsettingsviasuppliedPCsoftwaretool
AdvancedDataloggerfunctionsincludingdelayedstart,high/lowalarms,
bookmarking,andpasswordprotection
AutomaticallygeneratePDFandMicrosoft
®
Excel
®
datalogreports
Highaccuracythermistortemperaturesensoronbothmodelswithcapacitive
humiditysensorontheRHT30
Compactlightweightdesign
LargeLCDforeasyviewing
Dualcolor(redandgreen)LEDstatusindicatorfordataloggerprogress,PC
connectionstatus,andhigh/lowalarms
Largememory
capacity:48,000readings
Descriptions
DataloggerDescription
1. Externaltemperaturesensor(TH30only)
2. Externaltemperaturesensorconnection
todatalogger(TH30only)
3. Protectivecoverandswivelstand
(removetoexposeUSBplug).Rotatethe
coverforbestdesktopviewingangle.
4. LCDdisplay.RefertotheDisplayDescriptionlaterinthissection.
1
2
34 567
8
9
3RHT30_TH30enGB_V1.0(11/16)
5. StatusLEDindicator(red/green).RefertotheStatusLEDIndicatorDescription
laterinthissection.
6. MAXMINbutton.RefertotheButton
Descriptionslaterinthissection.
7. Sensors(Thermistorinternalair
temperatureandcapacitiveRHsensor)
8. PowerandDatalogStart/Stopbutton.
RefertotheButtonDescriptionslater
inthissection.
9. MARK/Batterylifebutton.RefertotheButtonDescriptionslaterinthis
section.
10. USBplug
11. Batterycompartment(CR2032x2)on
rear
12. Wallhangingmount(rear).Mounting
screwsupplied.
ButtonDescriptions
POWERbutton.ShortpresstopowerONorOFF.Notethatthedatalogger
cannotbeswitchedOFFwhiledataloggingisinprogress.
START/STOPbutton(samephysicalbuttonasthePowerbuttonabove).Press
andholdfor5secondstostartorstopthedatalogger.Notethatthedataloggerwill
notstartunlessithasbeenconfiguredusingthe‘PDFLoggerConfigurationTool.exe’
programasdescribedlaterinthisusermanual.
MaximumMinimumReadingbutton.Atanytime,shortpresstostep
throughthehighest(MX)andlowest(MIN)readingsrecordedinthecurrent
dataloggingsession.
Bookmarkbutton.Manuallyplaceatimestamped‘bookmark’inthedatalog
reportbypressingandholdingthebuttonuntilthedisplayedmeasurementflashes3
times.Uptoeight(8)bookmarkscanbesaved.
BatteryStatusNote:ShortpresstheMARKbuttontodisplaytheapproximatenumber
ofdaysremaining
forthebatterylife.
10
11 12
4RHT30_TH30enGB_V1.0(11/16)
Display(LCD)Description
o
F
o
C Temperatureunitofmeasureasprogrammedbytheuser
RH% DisplayedwhenshowingRelativeHumidityreading(RHT30)
T1 Representstheinternalsensor(air)temperature(TH30)
T2Representstheexternalprobetemperature.Ifnoprobeisconnected,
thedisplaywillshowdashesfortheT2temperaturereading(TH30)
H HighAlarmtriggered
L LowAlarmtriggered
RECRECflashesoncepersecondwhenactivelydatalogging.Whenloggeris
programmedwitha‘StartDelay’,RECisdisplayedbutdoesnotflash
untilafterthedelaytimehaselapsedandtheloggingbegins
Batteryiconflashesoncepersecondwhenbatteryleveliscriticallylow
MXDisplayedwhenthemaximumrecordedreadingisselectedwiththe
MAXMINbutton
MN Displayedwhentherecordedminimum(MN)readingisselectedwith
theMAXMINbutton
LEDStatusIndicatorDescription
Thedualcolor(red/green)LEDstatusindicatorislocatedtotherightoftheLCD.
Red(Alarm):Blinksredevery10secondsinanalarmcondition
Green(Datalogging):Blinksgreenevery10secondswhileactivelydatalogging
Green(PCconnection):GlowssteadygreenwhileestablishingconnectiontoPC
5RHT30_TH30enGB_V1.0(11/16)
PDFLoggerConfigurationTool
ConnectthedataloggertoPCUSBport.ThePCwillprompttoviewfiles.Openthe‘PDF
LoggerConfigurationTool’fromthelistofavailablefiles.RefertotheOperation
Sectionforadditionaldetailsonthistool:

1. CONVERTTOEXCEL:Createaspreadsheetreportoftheloggeddata
2. CONVERTTOPDF:CreateaPDFreportandatrendgraphoftheloggeddata
3. CONFIGURATION:Clicktoconfigurethelogger
4. LANGUAGESELECTION
5. SAMPLINGRATE:Dataloggerrecordinginterval
6. STARTDELAY:Delayperiodfrom
whenSTARTispressedanddataloggingbegins
7. TEMP.UNIT:Temperatureunitselection(
o
C/
o
F)
8. PASSWORD:Selecta16characterpasswordfortamperprotection
9. COMPANYNAME:20Characteruserlabelforinclusioninloggingreport
10. SAVE:PressSAVEtoconfirmtheconfiguration
11. MANUAL:OpenthesuppliedPDFUserManual
12. CANCEL:Aborttheconfiguration
13. ALARMLIMIT:SelecttheRH%and
Temperaturelow/highalarmlimits.Notethat
fortheTH30,RH%isreplacedbyexternaltemperatureprobealarmlimit
14. ALARMDELAY:Dataloggerwillignorealarmsfortheprogrammedperiodoftime
15. ALARMTYPE:SelectAlarmmode
3
4
5
6
7
8
9
10 11
12
13
14
15
1
2
6RHT30_TH30enGB_V1.0(11/16)
Operation
NOTES:
1.AdobeReader®softwareisrequired.
2.EnsuretheloggerisatroomtemperaturebeforeinsertingitintoaPCport.
3.Theexecutablefileforprogrammingisnamed“PDFLoggerConfigurationTool”.
4.Adataloggingsessioncannotstartuntilthedataloggerisproperlyconfigured.
ConfiguringtheTH30andRHT30foraDataloggingsession
Atanytimebeforeloggingisstarted,theloggercanbeprogrammedandre
programmed.Thepreviousloggingsessionwillbedeletedonceanewsessionis
started.
1. ConnectthedataloggertoaPCviatheUSBport.TheloggerwillswitchON
automatically.Notethatashortpress
ofthepowerbutton willalsopower
thelogger.
2. TheLEDwillglowgreenwhilea
connectionwithcomputerisbeing
established.
3. AnAutoPlaypromptwillappearonthePC
whencommunicationisestablished.
4. Clickon‘Openfoldertoviewfiles’tosee
theavailablefiles.
Opentheprogram"PDFLogger
ConfigurationTool.exe".
5. ThedefaultlanguageisEnglish;selectGerman,French,Italian,Spanish,or
Portugueseasdesired.Thegeneratedloggingreportformatwillmatchthe
languageselected.
6. ToreviewtheUserManual,click‘Manual’toopentheUserManualPDFfile.
7. Theparametersneededtoconfigurethedataloggerfor
loggingaredefined
below.Refertotheearlierscreenshotofthe‘PDFLoggerConfigurationTool’
windowforreference:
DatalogSamplingRate
Selectthedataloggingsamplingrate(recordinginterval)from30seconds~
120minutes.Thedefaultsettingis30seconds.
DatalogStartDelay
Selectaperiodof
timethatmustelapsebeforerecordingcanbeginafterthe
Start/Stopbutton
ispressed(selectfrom0minto24hours).Thedefault
settingis0minutes.
TemperatureUnitofMeasurement
Selectthetemperatureunit°Cor°F.
7RHT30_TH30enGB_V1.0(11/16)
AlarmDelay
ThepresetalarmdelaytimefortheSingleEventAlarmmodeisalwayszero.
TheadjustabledelaytimerangefortheCumulativeAlarmis5~120minutes.
AlarmModes
SingleEventAlarmmode:TheLEDimmediatelystartsflashingredevery10
secondswhenthemeasuredvalueexceedsthealarmthreshold.
CumulativeAlarmmode:TheLEDbeginsflashingredonlyaftertheaverage
valueexceedsthealarmthresholdandalsoafteranyprogrammedAlarm
Delaytime.SeeAlarmDelayparameterabove.
AlarmDisable(default):Alarmfunctiondisabled.
AlarmNote:TheLEDwillcontinueflashingredevery10seconds(onceitis
triggeredbyanalarm)evenwhenthemeasurementlaterreturnstothe
normalrange.TostoptheLEDalarm,plugtheloggerintothePCUSBportto
generateareport,orpoweroffthelogger(shortpressofthepowerbutton
whileitisnotconnectedtothePC).
AlarmLimits
Selectthealarmthresholdrangesforairandexternalprobetemperature
(TH30)orairtemperatureandRH%(RHT30).Forexample,iftheexternal
probetemperaturerangeissetto2.0~8.0
o
C,whenthemeasurementfalls
below2
o
Corexceeds8
o
CtheLEDwillflashredevery10seconds.
Note:Ifanalarmoccursbetweentwologgingpoints,theloggeddatawillnot
showthealarmbuttheLEDmaystillflashredevery10seconds,especiallyin
SingleEventalarmmode.Ifdesired,usetheCumulativeAlarmmodeto
minimizethiseffect.
Password
A16alphanumericcharacter(max.)passwordmaybeusedtoprevent
unauthorizedreprogramming.
UserCompanyName
Inputacompanynametobeincludedonthelogreport(20charactersmax.).
ANoteOnTimeZones
Theloggerautomaticallysynchronizesthetimezonewith
theconnectedPC
when‘SAVE’isselected.Note:Timezonechangesoccurringduringloggingare
nottakenintoaccount.
Oncealltheprogrammingiscomplete,press‘SAVE’toconfirmthesettings.Closethe
setupwindow,removetheloggerfromthePCUSBport,andstartdataloggingas
describedinthenextsection.
8RHT30_TH30enGB_V1.0(11/16)
StartDatalogging
1. ConfiguretheloggeraspreviouslydescribedandthenholdtheStart/Stop
buttonfor5secondstostartlogging.
2. Measurementsupdateonthedisplayevery10seconds.FortheRHT30the
temperatureandRH%displaystoggleevery5seconds.
3. IftheloggerisprogrammedwithaStartDelay:Afterpressingandholdingthe
Start/Stopbuttonfor5secondstostart
logging,RECwillappearontheLCDbut
willnotflashuntilthedelaytimehaselapsedandlogginghasbegun.
4. Whenthedataloggerissuccessfullyrunning,theLEDflashesgreenevery10
seconds(aslongasthereisnoalarm).
MonitoringAlarmsWhileDatalogging
TheLEDwillflashRED
every10secondswhenthereisanalarm.Thedisplayalso
showsLforlowalarmandHforhighalarm.TostoptheLEDfromflashingred,
plugtheloggerintoaPC,orpoweroff(shortpressofthepowerbutton).
Placing‘BookmarksWhileDatalogging
To
placeabookmarkmanuallyduringdatalogging,pressandholdtheMARKbutton
untilthemeasuredreadingflashes3times.Uptoeight(8)bookmarkscanbe
placed;thebookmarkswillappearintheloggingreport.
MAXMINReadings
ShortpresstheMAXMINbuttontostepthroughthehighest(MX)and
lowest(MN)
readingsstoredinmemory.
CheckingBatteryStatus
Atanytime,shortpresstheMARKbuttontoviewtheapproximatenumberofdays
remainingforbatterylife.Ifbatterypoweriscriticallylow,thebatteryiconwill
blinkoncepersecondontheLCDdisplay.
DownloadingLoggedData(PDFandSpreadsheetDatalogReports)
1. PressandholdtheStart/Stop buttonfor5secondstostopthedatalogger.
2. PlugtheloggerintothePCUSBport(thisalsostopsthedatalogger).
3. Openthe‘PDFloggerConfigurationtool’fromthelistofavailablefiles.
4. Choose‘ConverttoPDF’or‘ConverttoExcel’togenerateareport.
5. Select
thePCfilelocationwherethereportistobesaved.
6. ThePDFreportincludestherecordeddata,bookmarks,andatrendgraph.The
spreadsheetreportcontainsthesamedataasthePDFreportminusthetrend
graph.
9RHT30_TH30enGB_V1.0(11/16)
Specifications
Display(LCD) 5digitLCDwithmultifunctionindicators
Statusindicator(LED) DualcolormultipurposeLED(red/green)foractivelogging,
alarms,andPCconnectionstatus
TemperatureSensors NTCThermistors(TH30/RHT30internaltemperature
sensorsandTH30externaltemp.probe)
Ext.Temperatureprobelength 2.1m(84”)(TH30)
Temperaturerange 30~70
o
C(22~158
o
F)internalandexternalprobes
Temperatureresolution 0.1
o
C/
o
F
TemperatureAccuracy ±0.5
o
C(0.9
o
F)forinternalandexternal(TH30)probe
HumiditySensor Capacitive(RelativeHumidity)RHT30only
Humidityrange 0.1~99.9%RH
Humidityresolution 0.1%RH
HumidityAccuracy ±3%at25
o
C(10~90%RH),±5%allotherranges
LoggerType Reusable
Dataloggercapacity TH30:24000airtemperature/24000externalprobetemp.readings
 RHT30:24000temperature/24000humidity(RH%)readings
Samplingintervals 30seconds and5,10,30,60,90,120minutes
Startdelaytime 0,5,30,45,60,90,120
minutes,and24hours
Alarmranges Temp:‐30~70
o
C(22~158
o
F);Humidity:1to99%RH
Alarmdelay(inhibit) 0,5,30,45,60,90,120minutes
Alarmtypes SingleEvent,Cumulative,andDisable
OperationButtons Power/StartStop,MARK,MAXMIN
Operatingtemperature 30~70
o
C(22~158
o
F)forlogging;Allowloggertoreturnto
roomtemperaturebeforeconnectingtoPC
OperatingRelativeHumidity <80%RH
Storagetemperature 40~85
o
C(40~185
o
F)
StorageHumidity<90%RH
Dimensions 80(L)x43(W)x2.5(D)mm(3.2x1.7x0.1”)
Weight 90g(3.2oz.)
Battery Two(2)3.0VCR2032batteries(preinstalled)
Batterylife Three(3)months
SafetyCompliance CE
Warranty OneYear

10RHT30_TH30enGB_V1.0(11/16)
Maintenance
BatteryReplacement
Thebatterycompartmentislocatedonthebackofthedataloggerasshowninthe
Descriptionssection.Whenthebatterysymbolflashesonthedisplay,replacethe
batteriespromptly.
1. Usingalargecoin,rotatethebatterycompartmentlidclockwiseapproximately
onequarterturntoopenthecompartment
2. Replacethe
two(2)CR2032batterieswithpositive(+)sidefacingup
3. Replacethebatterycompartmentlidandturncounterclockwisetosecureit
4. Disposeofbatteriesinaccordancewithcountry,state,andlocalregulations
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriatecollection
sites,theretailstorewherethebatterieswerepurchased,orwhereverbatteriesaresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuseris
obligatedtotakeendof
lifedevicestoadesignatedcollectionpointforthedisposalofelectricalandelectronicequipment.
BatteryStatusCheck
Tochecktheapproximatenumberofdaysofbatteryliferemaining,shortpressthe
MARKbutton.Thedisplaywillshowthenumberofdaysandthe‘d’fordays(XXd).
Cleaning
Wipethedataloggerwithadampcloth,donotusesolventsorabrasivesasthesemay
damagethedataloggerfinishanddisplaylens.
ErrorDisplayCodes
ERROR CAUSE SOLUTION
Dashes
‐‐‐‐‐.‐
Externalprobeunplugged(TH30only)
Checktheconnectionbetween
theexternalprobeandlogger
E02 Measuredvaluebelowspecifiedrange Operateinthespecifiedrange
E03 Measuredvalueabovespecifiedrange Operateinthespecifiedrange
E04 Temperatureerror(RHT30only) ContactExtechforservice
E11 Humiditycheckfailed(RHT30only) ContactExtechforservice
E31 Microprocessorfailed ContactExtechforservice
E33 Humiditymeasurementerror ContactExtechforservice
Copyright©2016FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
www.extech.com
/