Preparing the Site
Biomek NXP System Specifications
Biomek NXP System Specifications
Table 1.1 presents system specifications and requirements for the Biomek NXP instrument.
Table 1.1 System Specifications
Item Description
Environment Indoor use only
Power Requirements 100–240 VAC, 50–60 Hz
Compressed Air Requirements
•NX-MC — Requires compressed air.
•NX-S8 — Only Span-8 instruments that
use a Microplate Shaking ALP require
compressed air.
Air Pressure: 60 - 100 psi of compressed air
414 - 690 kPa of compressed air
Air Flow: 2.25 cfm @ 60 psi
0.064 m3/min @ 414 kPa
System Fluid Requirements
•NX-S8 — Requires system fluid.
De-ionized or Distilled water
System fluid ambient operating temperature is
59°–86°F (15°–30°C).
System fluid should be degassed for 24 hours prior to use.
Dimensions base unit: •NX-MC — 36 in. (L) x 31 in. (W) x 34 in. (H); 91.44 cm (L)
x 78.74 cm (W) x 86.36 cm (H)
•NX-S8 — 36 in. (L) x 31 in. (W) x 44 in. (H); 91.44 cm (L)
x 78.74 cm (W) x 111.76 cm (H)
Weight:
Biomek NXP instrument (in packing crate)
Biomek NXP instrument
320 lbs. (145.2 kg)
190 lbs. (86.2 kg)
Ambient Operating Temperature 41°–86°F (5°–30°C)
Humidity Restrictions <85% (non-condensing) @ 86°F (30°C)
Altitude Restrictions up to 6562 ft (2000 m)
Installation Category Category II
Pollution Degree 2
Sound Pressure Level Maximum sound pressure: 82 dB
Maximum sound pressure at 1 meter away: 75 dB
Fuses US: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, UL recognized,
CSA certified
Europe: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, CENELEC
approved
Communications to Host RS-232 port
Communications to Tip Loader
•NX-MC — Uses a Tip Loader.
CAN
Communications to Active ALPs via the
Biomek NXP instrument
CAN