Thermo Fisher Scientific GeneChip HT Pico Kit Assay Automated Workflow User guide

Type
User guide
For Research Use Only. Not for use in diagnostic procedures.
GeneChip HT Pico Kit Assay Automated
Workflow
SITE PREPARATION GUIDE
for use with:
GeneChip Expression Arrays
GeneChip HT Pico Kit
Beckman Coulter Biomek FXP Target Prep Express Automated Workstation
Catalog Numbers 902790 and 902791
Publication Number MAN0018105
Revision B.0
Manufacturer:
Thermo Fisher Scientific Baltics UAB |
V.A. Graiciuno 8, LT-02241 |
Vilnius, Lithuania
Products:
GeneChip HT Pico Kit
Manufacturer:
Aymetrix Pte Ltd |
7 Gul Circle #2M-01 |
Keppel Logistics Building |
Singapore 629563
Products:
GeneChip expression arrays
For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE
LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR
ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Revision history: Pub. No. MAN0018105
Revision Date Description
B.0 09 October 2020 Added additional Bio-Rad Hard Shell PCR Plate catalog number.
A.0 27 December 2018 Initial release in Thermo Fisher Scientific document control system.
Supersedes legacy Aymetrix publication number 703399.
Updated to the current document template, with associated
updates to trademarks, logos, licensing, and warranty.
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these
products, you accept the terms and conditions of all applicable Limited Use Label Licenses.
TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.
©2020 Thermo Fisher Scientific Inc. All rights reserved.
Contents
CHAPTER 1 Biomek FXP Target Prep Express Automated
Workstation configuration .......................................................... 4
Equipment ..................................................................... 4
Instrument specifications and settings ............................................. 6
Other equipment required ....................................................... 10
Thermal cyclers ........................................................... 10
Spectrophotometer ........................................................ 10
Miscellaneous equipment required ........................................... 10
CHAPTER 2 Materials required and GeneChip HT Pico v1.0
method installation ................................................................ 11
Part 1: Equipment, reagents, consumables, and labware required .................... 11
Equipment ................................................................ 11
Reagents ................................................................. 13
Consumables and labware .................................................. 15
Labware templates ......................................................... 21
Part 2: Biomek HT Pico v1.0 method installation .................................. 23
Time requirements ......................................................... 23
Install the method files ..................................................... 24
Import the method into the Biomek FXP Target Prep Express
Automated Workstation .................................................. 26
Make required edits to Biomek Software v4.1.41 (Build 41) .................... 30
Import and configure the Biometra T-Robot thermal cycler program .............. 34
Set Microsoft Excel 2010 options ......................................... 40
Validate the method ........................................................ 40
Error recovery ............................................................. 41
Test the method with a water run ............................................ 42
Documentation and support ....................................................... 47
Related documentation ......................................................... 47
Customer and technical support ................................................. 48
Limited product warranty ........................................................ 48
GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP3
Biomek FXP Target Prep
Express Automated Workstation
configuration
Equipment ............................................................................ 4
Instrument specifications and settings ................................................... 6
Other equipment required ............................................................. 10
Equipment
The Beckman Coulter Biomek FXP Target Prep Express Automated Workstation is required to run
GeneChip High Throughput (HT) expression assays. This system includes the accessories, software,
and deck configuration listed in Table 1 and displayed in Figure 1. For information on upgrading a
Biomek liquid handler already at your site, contact Beckman Coulter Inc. Chapter 2, “Materials
required and GeneChip HT Pico v1.0 method installation” lists the components required to run the
GeneChip HT Pico Kit Assay Automated Workflow.
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Table 1 Biomek FXP Target Prep Express Automated Workstation components required for running
GeneChip HT methods for expression.
Item Quantity
Beckman
Coulter
Cat. No.
Biomek FXP Target Prep Express Automated Workstation
The part number for this workstation includes the components listed below:
1 A83103
Biomek FXP Dual Multichannel Span-8 1 A31844
Biomek Win 7 New Installation kit with Biomek Software 4.1 1 B40917
Packaged Tip Loader 1 719356
Span-8 Disposal ALP 1 719590
Low Volume Disposable Mandrel Kit, 8-pack 1 A39380
BFX 96-channel Disp, Tip Head, 200U 1 719368
1
4GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP
Table 1 Biomek FXP Target Prep Express Automated Workstation components required for running
GeneChip HT methods for expression. (continued)
Item Quantity
Beckman
Coulter
Cat. No.
Pkg Static ALP Platform 4 719357
4 × 3 ALP Kit 1 719948
Static Peltier ALP 1 A93938
Shaking Peltier ALP 1 A93942
Deep well plate adapter (required for Axiom assays) 1 A83050
Thin PCR adapter (required for expression assays) 1 A19709
Frame for Modular Reservoirs 3 372795
24-Position Tube Rack Holder 1 373661
11-mm Diameter Tube Rack Inserts (case of 25) 1 373696
Blue Heater/Chiller 24-well Block
(referred to as Reagent Cold Block in this document)
1 A83054
Biometra T-Robot Integration - v4.1 1 969125
Chapter 1 Biomek FXP Target Prep Express Automated Workstation configuration
Equipment 1
GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP5
Instrument specifications and settings
To perform GeneChip HT Pico expression assays successfully on the Biomek FXP Target Prep
Express Automated Workstation instrument, the system must meet the following specifications and be
properly configured.
Biomek FXP Target Prep Express system:
Hybrid dual armed Biomek FXP
96MC head left pod (Pod 1) and Span-8 head right pod (Pod 2)
1-mL syringes and low-volume tubing on the Span-8 system
Thin PCR adapter mounted on the Shaking Peltier Device
Shaking Peltier ALP Labware Guides properly adjusted
ALPs secured to deck. See Figure 1.
Degassed water as system fluid
Clean compressed air at 40–80 psi
Computer workstation specifications:
Microsoft Windows 7 operating system
Internet Explorer v8.0
Microsoft Excel 2010 (must be 32bit version)
Biomek Software v4.1.31 (Build 31) or v4.1.41 (Build 41)
Biomek Software configuration:
Ensure that the Aymetrix deck is the default deck.
Ensure that deck location SPelt96_1 has the device ShakingPeltier1 associated with it.
Ensure that deck location Pelt_1 has the device StaticPeltier1 associated with it.
Ensure that deck location TR1 is ALP type Span8TrashRight (not Span8TipTrashRight).
Check for a properly framed deck for the Biomek 200-µL 96-Channel Disposable Pipetting Head
and the Biomek Span-8 Waste ALP pods.
Framing tolerance is 0.03 cm. Reframe if this is exceeded.
If the deck needs to be reframed, save a current copy of the instrument file for backup.
Exit and restart the Biomek Software after installing the Aymetrix deck and after any
changes in deck framing. This saves the new values in the default instrument (.bif) file.
Additional pod settings:
Navigate to Instrument4Hardware Setup4Pod24Additional Pod Settings.
Pod 2 (Span-8) settings (Figure 2)
Set System Trailing Airgap to 20 µL.
Navigate to Instrument4Hardware Setup4Pod14Additional Pod Settings.
Pod 1 (MC96) settings (Figure 3)
Check Always move to Z-Max when roving box.
Set the Additional Roving Height field to 0.45 cm or less (default is 0.75 cm).
Chapter 1 Biomek FXP Target Prep Express Automated Workstation configuration
Instrument specifications and settings
1
6GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP
Set the adjustable labware corner guides on the Shaking Peltier ALP:
Ensure that the thin adapter plate is installed on the peltier, if not, install the thin adapter plate.
To adjust the newer (tan) adjustable corner guides the Shaking Peltier ALP must be removed
from the deck to allow access to the rear screws; to adjust the older (blue) adjustable corner
guides the shaking peltier can remain on the deck.
The tan adjustable corner guides have 2 flathead screws present on each of the outer sides
of the four guides to perform adjustments to hold labware; the blue adjustable corner guides
have 2 hex screws on the top of each corner guide to perform adjustments to hold labware
Place a Costar Polystyrene 96-Well U-bottom Plate onto the Shaking Peltier ALP
Tan adjustable corner guides: tighten the flat head screws on the sides of each labware guide
(8 total screws) until a slight resistance is felt. Then tighten each screw an additional 1/2 turn.
Blue adjustable corner guides: adjust the corner guide screws until labware is held snugly.
Ensure that the Costar Polystyrene 96-Well U-bottom Plate is held snugly but can be removed
from the Shaking Peltier ALP.
Ensure that a Bio-Rad Hard-Shell 96-Well PCR Plate plate can be placed into and removed
from the Shaking Peltier ALP.
Reinstall the Shaking Peltier ALP on the deck and reframe the position for Pod 1 if necessary.
Chapter 1 Biomek FXP Target Prep Express Automated Workstation configuration
Instrument specifications and settings 1
GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP7
Figure 1 Biomek FXP Target Prep Express Automated Workstation system (top) and deck configuration
(bottom).
Chapter 1 Biomek FXP Target Prep Express Automated Workstation configuration
Instrument specifications and settings
1
8GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP
Figure 2 Hardware setup changes for the span-8 ALP on Pod 2.
Figure 3 Hardware setup changes for the 96-channel head on Pod1.
Chapter 1 Biomek FXP Target Prep Express Automated Workstation configuration
Instrument specifications and settings 1
GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP9
Other equipment required
Thermal cyclers
The GeneChip HT Pico v1.0 method has 2 thermal cycler options for reaction incubations: an on-deck
Biometra TRobot 96 Thermocycler or an o-deck thermal cycler.
For on-deck incubation using a Biometra TRobot 96 Thermocycler (Biometra Cat. No. 050-991), the
thermal cycler should be purchased directly from a local Biometra distributor. For information, see
analytikjena.com. The Biometra TRobot 96 Thermocycler integration solution must be purchased
separately from Beckman Coulter.
To use an o-deck thermal cycler, ensure that the thermal cycler can accept the full-skirted Bio-Rad
Hard-Shell 96 PCR plate. If the hybridization-ready target will be prepared for cartridge arrays, ensure
that the thermal cycler can accept FrameStar Break-A-Way PCR Plates. See Table 9.
Spectrophotometer
The GeneChip HT Pico v1.0 method has 2 options for measuring nucleic acid concentration: a UV/Vis
plate reader or a NanoDrop spectrophotometer (or equivalent instrument). NanoDrop instruments can
be purchased from Thermo Fisher Scientific. There are no unique plate reader requirements except that
the plate reader must accommodate Greiner UV-Star plates. See Table 9.
Miscellaneous equipment required
"MLS" indicates that the material is available from fisherscientific.com or another major laboratory
supplier.
Table 2 Other equipment required
Item Source
Benchtop cooler, –20°C MLS
Centrifuge for 96-well plates and 1.5-mL and 2.0-mL tubes MLS
Freezer, –20°C, not frost-free MLS
Ice bucket MLS
Pipette with pipette tips MLS
Refrigerator, 2–8°C MLS
Vortex mixer MLS
Chapter 1 Biomek FXP Target Prep Express Automated Workstation configuration
Other equipment required
1
10 GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP
Materials required and GeneChip
HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required .......................... 11
Part 2: Biomek HT Pico v1.0 method installation ........................................ 23
“Part 1: Equipment, reagents, consumables, and labware required” on page 11 of this chapter lists
the part numbers and suppliers for the equipment, reagents, consumables, and labware that have been
verified for use with the GeneChip HT Pico Kit on the Beckman Coulter Biomek FXP Target Prep
Express Automated Workstation. “Part 2: Biomek HT Pico v1.0 method installation” on page 23 of
this chapter provides detailed instructions for the installation and validation of the HT Pico v1.0 method
for the Biomek FXP Target Prep Express Automated Workstation.
Part 1: Equipment, reagents, consumables, and labware
required
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Equipment
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Table 3 Equipment required for target preparation.
Quantity Item Source
1 Biomek FXP Target Prep Express Automated Workstation Beckman Coulter
A83103
1 24-Position Tube Rack Holder Beckman Coulter
373661
11-mm Diameter Tube Rack Inserts (11 mm, white, case of 25) Beckman Coulter
373696
1 Reagent Cold Block, 24 Well Block Beckman Coulter
A83054
1 Thin PCR adapter (installed on the Shaking Peltier ALP) Beckman Coulter
A19709
2
GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP11
Table 3 Equipment required for target preparation. (continued)
Quantity Item Source
1 Centrifuge (for tube and plate) MLS
1 Pipettes MLS
1 Spectrophotometer (plate, single channel, or multichannel) MLS
1 Thermal cycler (for optional o-deck incubation) MLS
As
needed
Adhesive PCR plate seals (for optional o-deck incubation) MLS
1 Vortex mixer MLS
Table 4 Expression installation kits required for the HT Pico method.
Quantity Item Source
1 GeneChip HT 3' IVT PLUS and WT PLUS Core Installation Kit ,
components:
Alpaqua Magnetic Plate
Alpillo Plate Cushion
Arched auto-sealing lids with wide tabs (package of 4)
902453
1 GeneChip HT Pico Installation Kit, components:
HT Pico reagent cold block template
HT Pico reservoir frame stickers
PCR tube holder
902901
HT Pico v1.0 Biomek method file and support files 600757
(Installed by Thermo
Fisher Scientific FAS)
Table 5 Equipment required for array processing.
Item Source
GeneChip System for Cartridge Arrays
GeneChip Hybridization Oven 645
GeneChip Fluidics Station 450
GeneChip Scanner 3000 7G System
00-0331 (110/220V)
00-0079
00-0213
GeneChip AutoLoader with External Barcode Reader 00-0090 (GCS 3000 7G S/N 501)
00-0129 (GCS 3000 7G S/N 502)
GeneTitan Multi-Channel (MC) Instrument Contact Thermo Fisher Scientific
Chapter 2 Materials required and GeneChip HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required
2
12 GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP
Reagents
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Table 6 Contents and storage: GeneChip 3 IVT Pico Kit (Cat. No. 902790, 30 reactions) and
GeneChip HT Pico Kit (Cat. No. 902791, 96 reactions).
Component Storage
Pico Amplification Kit, Module 1: cRNA Step
Pico First-Strand Enzyme
–20°C
Pico First-Strand Buer
Pico Cleanup Reagent
Pico 3’ Adaptor Enzyme
Pico 3’ Adaptor Buer
Pico PCR Enzyme
Pico PCR Buer
Pico IVT Enzyme
Pico IVT Buer
Poly-A Control Stock
Nuclease-free Water Any temperature[1]
Pico Amplification Kit, Module 2: cDNA Step
Pico 2nd-Cycle Primers
–20°C
Pico 2nd-Cycle ds-cDNA Enzyme
Pico 2nd-Cycle ds-cDNA Buer
Pico Frag. & Label Enzyme
Pico Frag. & Label Buer
Pico Amplification Kit, Module 3
Purification Beads
4°C[2]
Pico Hydrolysis Buer
Pico Neutralization Buer
Control HeLa RNA
Control RNA (100 ng/μL HeLa total RNA) –20°C
Chapter 2 Materials required and GeneChip HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required 2
GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP13
Table 6 Contents and storage: GeneChip 3 IVT Pico Kit (Cat. No. 902790, 30 reactions) and
GeneChip HT Pico Kit (Cat. No. 902791, 96 reactions). (continued)
Component Storage
GeneChip Hybridization Control Kit
20X Hybridization Controls
–20°C
3 nM Control Oligo B2
Tubes Organizer: Plastic vinyl template for organization and storage of components in 9 x 9 array, 81-
places square wells, 5 1/4 in. x 5 1/4 in (for example, Nalgene Polycarbonate 9 x 9 CryoBox, Cat. No.
5026-0909 ).
[1] Store the Nuclease-free Water at20°C,C, or room temperature.
[2] Do not freeze.
Table 7 Additional reagents required.
Item Source
GeneChip Hybridization, Wash, and Stain Kit (30 reactions, for cartridge arrays) 900720
GeneTitan Hybridization, Wash, and Stain Kit for WT Array Plates (96 reactions) 901622
GeneTitan Hybridization, Wash, and Stain Kit for 3’ IVT Arrays (96 reactions) 901530
DNAZap PCR DNA Degradation Solutions AM9890
RNaseZap RNase Decontamination Solution AM9780
Nuclease-free Water (for elution, dilution and preparing 80% Ethanol Wash Solution):
~45 mL for 1 × 24 reactions, and
~120 mL for 1 × 96 reactions
MLS
100% Ethanol (molecular biology grade or equivalent) for preparing 80% Ethanol
Wash Solution:
~20 mL for 1 × 24 reactions, and
~50 mL for 1 × 96 reactions
MLS
Chapter 2 Materials required and GeneChip HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required
2
14 GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP
Consumables and labware
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Thermo Fisher Scientific does not oer a consumables kit specifically for use with the HT Pico v1.0
method. The HT WT PLUS Consumables for Biomek FXP Target Prep Express (Cat. No. 902508),
when supplemented with 2 additional Bio-Rad 'P' pads (Cat. No. MSL2032 ), is sucient to perform 4
complete HT Pico assays for cartridge arrays or 3 IVT array plates.
When supplemented with 2 additionalBio-Rad 'P' pads and 4 additional Bio-Rad Hard-Shell PCR
plate, clear well (Cat. No. HSP9631 or HSP9601), the kit will be sucient to perform 4 complete HT
Pico assays for WT or Clariom S array plates.
Table 8 Beckman Coulter tips and labware required for 1 full run of the GeneChip HT Pico Kit on the
Biomek FXP Target Prep Express Automated Workstation.
Item Packaging Total 24X Total 96X Source
Biomek P50 Sterile Filtered Tips (magenta
box)[1]
96/box
10 boxes/case
12 boxes 14 boxes Beckman
Coulter
A21586
Biomek P250 Sterile Filtered Tips (aqua
box), for AP96 multichannel pipettor
Plates
Cartridges
96/box
10 boxes/case
6 boxes
5 boxes
6 boxes
5 boxes
Beckman
Coulter
717253
Biomek P250 Span-8 Sterile Filtered
Tips(green box), for span-8 pipettor
96/box
10 boxes/case
1 box 2 boxes Beckman
Coulter
379503
Biomek P1000 Span-8 Conductive Sterile
Filtered Tips (yellow box) or Biomek P1000
Span-8 Nonconductive Sterile Filtered Tips
(orange box), for span-8 pipettor[2]
96/box
5 boxes/case
2 boxes 2 boxes Beckman
Coulter
987925 or
B01124
Quarter Reservoir, Divided by Width, 19 mL
per section
Plates
Cartridges
48 units/case
6 units
4 units
6 units
4 units
Beckman
Coulter
372792
Quarter Reservoir, 40 mL
Plates
Cartridges
48 units/case
5 units
4 units
5 units
4 units
Beckman
Coulter
372790
Frame for Modular Reservoirs 1 3 reusable
frames
3 reusable
frames
Beckman
Coulter
372795
[1] Two fewer boxes of this tip required for NanoDrop quantification option for method parts 2 and 4.
[2] The HT Pico v1.0 method can use conductive and nonconductive P1000 Span-8 tips interchangeably.
Chapter 2 Materials required and GeneChip HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required 2
GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP15
Table 9 Consumables and labware used on the Biomek FXP Target Prep Express Automated Workstation
deck.
Item Source Image
Biomek P50 Sterile Filtered Tips
(magenta box)
Beckman Coulter A21586
Biomek P250 Sterile Filtered Tips
(aqua box)
Beckman Coulter 717253
Biomek P250 Span-8 Sterile
Filtered Tips(green box)
Beckman Coulter 379503
Biomek P1000 Span-8 Conductive
Sterile Filtered Tips (yellow
box) or Biomek P1000 Span-8
Nonconductive Sterile Filtered Tips
(orange box)
Beckman Coulter
987925 (yellow) or
B01124 (orange)
Reservoir, Quarter
Quarter Reservoir, 40 mL
Quarter Reservoir, Divided by
Width, 19 mL per section
Beckman Coulter
372790
372792
Chapter 2 Materials required and GeneChip HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required
2
16 GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP
Table 9 Consumables and labware used on the Biomek FXP Target Prep Express Automated Workstation
deck. (continued)
Item Source Image
Frame for Modular Reservoirs
(3 total needed for HT Pico Method,
reusable for subsequent runs)
Beckman Coulter 372795
96-well deep well microplate,
polypropylene, 2 mL, non-sterile
Included in GeneChip HT WT
PLUS Consumables for Biomek
FXP Target Prep Express Automated
Workstation, 902508
(or Fisher Scientific AB0932 )
Greiner U-bottom (round) plate,
polypropylene
Used as a distribution plate for
cartridge hybridization mix and is not
interchangeable with the polystyrene
U-bottom plate.
Included in GeneChip HT WT
PLUS Consumables for Biomek
FXP Target Prep Express Automated
Workstation, 902508
(or Fisher Scientific 07-000-146)
Bio-Rad Hard-Shell PCR plate, clear
well
Included in GeneChip HT WT
PLUS Consumables for Biomek
FXP Target Prep Express Automated
Workstation, 902508
Note: One additional plate per run
required for WT or Clariom S array
plate option.
(or Bio-RadHSP9631 or HSP9601 )
A
B
C
D
E
F
G
H
1
2
3
4
5
6
7
8
9
10
11
12
A
B
C
D
E
F
G
H
1
2
3
4
5
6
7
8
9
10
11
12
Greiner UV-Star plate, nonsterile,
clear, flat bottom
Included in GeneChip HT WT
PLUS Consumables for Biomek
FXP Target Prep Express Automated
Workstation, 902508
(or Fisher Scientific 07-000-407)
Costar 96-Well U-bottom plate,
polystyrene
Included in GeneChip HT WT
PLUS Consumables for Biomek
FXP Target Prep Express Automated
Workstation, 902508
Chapter 2 Materials required and GeneChip HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required 2
GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP17
Table 9 Consumables and labware used on the Biomek FXP Target Prep Express Automated Workstation
deck. (continued)
Item Source Image
24-well plate, 10 mL, square well,
pyramid bottom, polypropylene
Included in GeneChip HT WT
PLUS Consumables for Biomek
FXP Target Prep Express Automated
Workstation, 902508
(or Seahorse Bioscience S30024)
'P' Pad for lid, autosealing
microplates, arched with Microseal
Included in GeneChip HT WT
PLUS Consumables for Biomek
FXP Target Prep Express Automated
Workstation, 902508
Note: One new 'P' Pad is required
per run for HT Pico.
(or Bio-Rad MSP1003, MSL2032
with lid)
FrameStar Break-A-Way plate Included in GeneChip HT WT
PLUS Consumables for Biomek
FXP Target Prep Express Automated
Workstation, 902508
(or Thomas Scientific 4ti-1000/B)
Nalgene Polycarbonate 9 x 9
CryoBox
5026-0909
Alpaqua Magnetic Plate Included in GeneChip HT 3' IVT
PLUS and WT PLUS Core Installation
Kit , 902453
Chapter 2 Materials required and GeneChip HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required
2
18 GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP
Table 9 Consumables and labware used on the Biomek FXP Target Prep Express Automated Workstation
deck. (continued)
Item Source Image
Alpillo Plate Cushion Included in GeneChip HT 3' IVT
PLUS and WT PLUS Core Installation
Kit , 902453
Arched auto-sealing lids with wide
tabs
Included in GeneChip HT 3' IVT
PLUS and WT PLUS Core Installation
Kit , 902453
Blue Heater/Chiller 24-well Block
Block must be chilled to 4°C before
use.
Beckman Coulter A83054
HT Pico Cold Block Template
The template is designed for the
GeneChip HT Pico Kit and fits
precisely onto the reagent cold
block. The cold block template helps
ensure proper placement of reagent
tubes into the block for each part of
the assay.
Included in the GeneChip HT Pico
Installation Kit, 902901.
Chapter 2 Materials required and GeneChip HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required 2
GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP19
Table 9 Consumables and labware used on the Biomek FXP Target Prep Express Automated Workstation
deck. (continued)
Item Source Image
HT Pico Reservoir Frame Stickers
Place modular reservoir template
stickers on the front of the modular
reservoir frames to ensure proper
placement of modular reservoirs and
reagents.
Included in the GeneChip HT Pico
Installation Kit, 902901.
PCR Tube Holder
For use with FrameStar Break-A-
Way PCR Plate in cartridge array
hybridization setup.
Included in the GeneChip HT Pico
Installation Kit, 902901.
(or Bio-Rad TRC-9601)
24-Position Tube Rack Holder with
11-mm Diameter Tube Rack Inserts
Beckman Coulter
373661 (rack)
373696 (inserts)
Thin PCR Adapter
This adapter is typically installed
on the Shaking Peltier ALP by
a Beckman Coulter field service
technician during new system
installation or a system upgrade.
Ensure that you have the thin PCR
adapter installed on the Shaking
Peltier ALP prior to running this
assay.
Beckman Coulter A19709
GeneTitan Hybridization Tray Part. No. 500867
(GeneTitan trays and lids are
provided with the array plate when
ordered. The part number provided
here is for identification purposes
only.)
Chapter 2 Materials required and GeneChip HT Pico v1.0 method installation
Part 1: Equipment, reagents, consumables, and labware required
2
20 GeneChip HT Pico Kit Assay Automated Workflow Site Preparation Guide—Biomek FXP
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50

Thermo Fisher Scientific GeneChip HT Pico Kit Assay Automated Workflow User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI