TF-NBase-T Instructions
935700-00 Rev A 2
Distortion Test
1. Connect the Master DUT to the Ethernet
port in PCA Section 1 using a CAT6 cable.
2. Attach the SMA-SMA cables to the A +/–
pair in Section 1, then connect each to a power
splitter SMA connector 1.
3. Attach an SMA-BNC cable to each power
splitter SMA connector 2, then connect them to
the oscilloscope inputs.
4. Attach an SMA-BNC cable to each power
splitter SMA connector S, then connect them to
the Disturbing Sine Wave Generator (Disturber).
5. Terminate unused inputs in Section 1.
6. As the test progresses, move the SMA-SMA cables to the next pair of inputs in Section 1, terminating unused pairs.
Other Tests
1. Connect the Master DUT to the Ethernet
port in PCA Section 1 using a CAT6 cable.
2. Attach SMA-BNC cables to the A +/- pair
in Section 1, then connect them to the
oscilloscope inputs.
3. Terminate unused inputs in Section 1.
4. As the test progresses, move the SMA-
BNC cables to the next pair of inputs in Section
1, terminating the unused pairs.
© 2023 Teledyne LeCroy, Inc. All rights reserved.
Teledyne LeCroy is a trademark of Teledyne LeCroy, Inc., Inc. Other product or brand names are trademarks or requested trademarks of their respective
holders. Information in this publication supersedes all earlier versions. Specifications are subject to change without notice.
Unless otherwise specified, all materials and processes are compliant with RoHS Directive 2011/65/EU in its entirety, inclusive of any further
amendments or modifications of said Directive.
May be subject to disposal and recycling regulations. For more information, visit teledynelecroy.com/recycle.
700 Chestnut Ridge Road
Chestnut Ridge, NY 10977–6499
Tel: (845) 578 6020, Fax: (845) 578 5985
teledynelecroy.com
tf-nbase-t-instructions_05oct23.pdf
October, 2023