Enraf-Nonius aXionXelect Zimmer User manual

Type
User manual
Instructions for Use
aXionXelect
EN
Contents
1
Introduction
2
Symbols
3
Package contents
4
Installation
5
Intended use and intended user
6
Precautionary instructions
7
Operation
8
Safety instructions
9
Cleaning, maintenance and troubleshooting
10
Specifications
11
Contact
12
Product liability
Introduction
1
EN109-1550752-41
Page 1
We are pleased that you have purchased an aXionXelect treatment couch.
This examination and treatment couch is specially designed for the demands of
the physical and manual therapist. The anatomically segmented top part
combined with the wide array of section positions enable you to position your
patient virtually however you wish.
During development and production of this aXionXelect a great deal of care
was put into the reliability, safety, ease of use and durability.
Symbols
2
EN109-1550752-41
Page 2
Follow the instructions in the Instructions for Use.
It is important that you read, understand and observe the precautionary and
operating instructions
Warning or Caution:
Indicates a hazardous situation which, if not avoided, could result in:
a. Death or serious injury to the patient (or)
b. Minor to moderate injury to the patient (or)
c. Damage to the equipment
Temperature Range.
Indicates acceptable temperature range
Humidity Limits.
Indicates acceptable relative humidity
Atmospheric Pressure.
Indicates the range of atmospheric pressure to which the medical device can
be safely exposed.
Waste electrical items that can be recycled.
Indicates the electrical and electronic components of the device can be
recycled and has to be disposed separately.
Manufacturer name, address and date of manufacture.
Model Number
Reference Number or Part Number
Serial Number
Indicates the serial number so that a specific medical device can be
identified.
Indicates that the item is a medical device.
Indicates a carrier that contains Unique Device Identifier information.
CC
Configuration code
IP
Ingress Protection Marking, classifies and rates the degree of protection
provided against intrusion (body parts such as hands and fingers), dust,
accidental contact, and water by mechanical casings and electrical enclosures.
Double Insulation
Symbols
2
EN109-1550752-41
Page 3
Alternating current
Product is in conformity with the European Union Medical Device Regulation
(EU) 2017/745
Risk of crushing hands
Package contents
3
EN109-1550752-41
Page 4
aXionXelect
Instructions for Use
Cable Clip
Information sheet disinfection of upholstery
Ordering data
For the ordering data of the aXionXelect standard accessories and optional
accessories we refer to www.zimmer.de
Installation
4
EN109-1550752-41
Page 5
Unpacking the
device
The device is heavy, it must be unpacked by at least 2 persons.
Do not lift the couch by the table top parts or the all-around control.
Inspection
Immediately after unpacking the couch, check whether the package contains
all the items listed above, and check the couch for any damage.
In case of damage, contact your local representative. Do NOT use the
equipment!
Installation
Read this Instructions for Use carefully before using the equipment.
Avoid exposure to direct sunlight, UV radiation, rain, excessive dust, moisture,
mechanical vibrations, and shocks.
The aXionXelect couches are not intended for use in so-called “wet rooms”
(hydrotherapy rooms).
The device may not be used in close proximity (i.e. less than 2 metres) to
shortwave or device.
The aXionXelect should be placed on a flat surface.
Turn the adjustable foot until it touches the ground and the couch is standing
stable.
For correct assembly and safe use of possible optional accessories one must
consult the separately provided documents included with the optional
accessories.
Connection to mains
supply
Mains supply connections must comply with the national requirements
regarding medical rooms.
Prior to connection of this device to the mains supply, check that the voltage
and frequency stated on the type plate correspond with the available mains
supply.
Connect the couch to the mains using the mains cable.
For couches with a digital lock, an audible beep will sound.
Installation
4
EN109-1550752-41
Page 6
Optionally, the couch can be equipped with a digital lock.
Enabling the digital lock will ensure the couch's motors cannot be operated.
This will prevent the couch from being adjusted unintentionally. The lock is now
enabled. Press both buttons ▲ and ▼ on the foot end of the couch for 2
seconds to disable the lock. (also see the chapter 'Operation')
Use the all-around control (or an optional other control element) to move the
couch up or down (also see the chapter 'Operation').
Connection to the
cable clip for mains
cable (optional)
Move the couch to the highest position.
Disconnect the mains cable.
The cable clip can be attached to the frame at either the head or foot side.
Make sure the desired surface is free of grease. Clean it by using a dry
cloth and cleaning spirit.
Remove the support layer.
Attach the cable clip firmly to the cleaned surface.
Put the mains cable in the cable clip. Make sure the mains cable is not
tightened, squeezed or hindered by moving parts.
Connect the mains cable.
Intended use and intended user
5
EN109-1550752-41
Page 7
Intended use
The aXionXelect examination and treatment couches are intended to support a
patient in the desired position for examination or treatment within the
physiotherapy, rehabilitation and adjacent disciplines.
Intended user
The axionXelect treatment couches are intended to be used, and shall only be
used, by or under the supervision of professional users in physiotherapy,
rehabilitation and adjacent areas.
Precautionary instructions
6
EN109-1550752-41
Page 8
If the use of this device may have caused or contributed to an undesirable
event such as death or serious injury to the user, the manufacturer AND the
competent authority of the Member State MUST be notified immediately!
Read this Instructions for Use carefully before using the equipment.
Do not lift the couch by the table top parts or the all-around control.
Immediately after unpacking the couch, check whether the package contains
all the items listed above, and check the couch for any damage.
In case of damage, contact your local representative. Do NOT use the
equipment!
Avoid exposure to direct sunlight, UV radiation, rain, excessive dust,
moisture, mechanical vibrations, and shocks.
The aXionXelect couches are not intended for use in so-called “wet rooms”
(hydrotherapy rooms).
The device may not be used in close proximity (i.e. less than 2 metres) to
shortwave or device.
The aXionXelect should be placed on a flat surface.
Turn the adjustable foot until it touches the ground and the couch is standing
stable.
For correct assembly and safe use of possible optional accessories one must
consult the separately provided documents included with the optional
accessories.
Mains supply connections must comply with the national requirements
regarding medical rooms.
Prior to connection of this device to the mains supply, check that the voltage
and frequency stated on the type plate correspond with the available mains
supply.
The aXionXelect is only meant to be used by competent personnel in
physiotherapy, rehabilitation or adjacent areas. Avoid any use of the
equipment by unauthorized personnel.
When used by children, supervision by a knowledgeable (para)medic is
obligatory. At all times keep children away from the equipment when no
supervision is at hand or use a safety switch.
Use both hands to change the couch settings: one hand for unlocking and
the other to adjust the section.
Precautionary instructions
6
EN109-1550752-41
Page 9
Never adjust a section when it supports the full weight of the patient!
Each section of this aXionXelect is supported by gas springs. As well as
easing operation, they offer full support of the sections. However, these gas
springs are not strong enough to lift a section while it supports the full weight
of the patient.
A maximum patient’s weight of 200 kg is allowed when the height of the
couch is being adjusted.
Try to limit the continuous use of the motors. For every 2 minutes that the
couch is operated in succession, it should not be used for 18 minutes in
order to cool down completely. If the motor is used significantly more than
this, it falls outside the normal use of the motor, and the performance of the
motor will decrease.
The couch is not anti-static.
For safety reasons, the wheels of the unit should be retracted before the
beginning of a treatment.
Never let your full weight rest on the switch bar!
Ensure that you will never rest your feet on the all-around-operation.
Especially when seated next to the couch it can lead to a dangerous situation
in which your leg can be jammed between the section and the all-around-
operation.
Do not sit on the head and leg section of the couch.
When adjusting this couch, keep hands and feet away from the couch for at
least 20 cm, especially on places that has been marked by these stickers.
When adjusting the sections, squeezing-danger can arise at contact points of
the section. Keep enough distance between you or your clients and these
points.
Head section: is not suited to sit upon and must be used to support the head
only.
Always use the original accessories that are meant for this product.
Use a clean couch cover / towel / terry cloth / paper cover for every
individual. This is required to guarantee a hygienic use.
Do not use cleaning agents that contain alcohol or other aggressive cleaners.
Precautionary instructions
6
EN109-1550752-41
Page 10
The aXionXelect should be checked once a year by authorized personnel of
Zimmer MedizinSysteme or by its authorized distributor.
The manufacturer will not be held responsible for the results of maintenance
or repairs by unauthorized persons!
Operation
7
EN109-1550752-41
Page 11
Use both hands to change the couch settings: one hand for unlocking and the
other to adjust the section.
Never adjust a section when it supports the full weight of the patient!
Each section of this aXionXelect is supported by gas springs. As well as
easing operation, they offer full support of the sections. However, these gas
springs are not strong enough to lift a section while it supports the full weight
of the patient.
A maximum patient’s weight of 200 kg is allowed when the height of the couch
is being adjusted.
Try to limit the continuous use of the motors. For every 2 minutes that the
couch is operated in succession, it should not be used for 18 minutes in order
to cool down completely. If the motor is used significantly more than this, it
falls outside the normal use of the motor, and the performance of the motor
will decrease.
The couch is not anti-static.
The operation of your aXionXelect treatment couch is simple and safe.
The illustrations show how the various adjustments are done.
Moving the couch
The couch can be moved by using the foot
pedal to lower the wheels (optional).
Note:
Never move the couch with a person or load on it. Move the couch only on a
smooth surface. The couch is not designed to move across thresholds.
For safety reasons, the wheels of the unit should be retracted before the
beginning of a treatment.
Activation with
digital lock
autolockbox
(optional)
A digital lock is optional on the couch.
We recommend you always enable this lock when the couch is unattended.
Enabling it will ensure the couch's motors cannot be operated, and the couch
cannot be used. This will prevent the couch from being adjusted
unintentionally.
Press both buttons ▲ and▼ on the foot end of the couch for 2 seconds to
enable or disable the lock. One beep means the couch is locked. Two beeps
means the couch is ready for use.
When trying to operate the couch while the lock is engaged, this will also be
indicated by two beeps.
If the couch is not operated for 30 minutes, it will lock automatically.
Operation
7
EN109-1550752-41
Page 12
The default settings of the digital lock can be changed as follows:
Press and hold both buttons for 30 seconds until you hear 3 beeps. When you
hear the beeps, release the buttons. Now press 1 of the lock buttons several
times to select the desired mode. One or more beeps will sound. The number
of beeps represents the selected setting:
Press once (1 beep sounds) = lock enabled time of 15 minutes
Press twice (2 beep sounds) = lock enabled time of 30 minutes (default value)
Press three times (3 beep sounds) = lock enabled time of 60 minutes
Press four times (4 beep sounds) = lock can only be enabled and disabled
manually
Press five times (5 beep sounds) = lock completely disabled
If no button is pressed after this for 15 seconds, the new setting will be stored
and the setup mode is exited.
Electric height
adjustment
By pressing the foot switch continuously,
the required height can be reached.
Height adjustment
with the all-around
operation (optional)
By pressing or pulling the all-around
operation - which is located underneath
the framework - with your foot, the
required height of the couch can be set. At
the head section of the couch, the height
can be adjusted by turning the handle to
the left or to the right.
Never let your full weight rest on the switch bar!
Adjusting the back
section
Use one hand to pull the handle
underneath the back section, while
adjusting the section in the desired
position. Release the handle and the
section will be fixed.
Adjusting the head
section
Use one hand to pull the handle
underneath the head section while
adjusting the section in the desired
position. Release the handle and the
section will be fixed.
Operation
7
EN109-1550752-41
Page 13
Releasing the rolling
section (traction)
Release the rolling section by pulling out
the stop pin. Turn the pin slightly if you
want to prevent it from clicking back into
the locking mechanism.
Locking the rolling
section (traction)
To lock the rolling section again, turn the
pin such that it can click back into the
starting position.
Safety instructions
8
EN109-1550752-41
Page 14
Danger of squeezing
the extremities
On this couch warning stickers are placed.
When adjusting this couch, keep hands
away from the couch for at least 20 cm,
especially on places that has been marked
by these stickers.
Height adjustment
(general)
Ensure that neither the patient nor the
operator can get any part of the body or
clothes near or in between the moving
parts of the mechanism when the couch is
being raised or lowered.
Be particularly careful for those situations
indicated beside (the couch is set in flexion
position combined with the lowest possible
position):
Underneath the head section, between the
head section and floor.
Height adjustment
Ensure that you will never rest your feet on
the all-around-operation.
Especially when seated next to the couch
it can lead to a dangerous situation in
which your leg can be jammed between
the section and the all-around-operation.
Danger of tilting
Do not sit on the head and leg section of
the couch.
Safety instructions
8
EN109-1550752-41
Page 15
Adjustment of the
sections
When adjusting the sections, squeezing-
danger can arise at contact points of the
section. Keep enough distance between
you or your clients and these points.
Load
Head section: is not suited to sit upon and
must be used to support the head only.
Cleaning, maintenance and
troubleshooting
9
EN109-1550752-41
Page 16
Cleaning
Use a clean couch cover / towel / terry cloth / paper cover for every individual.
This is required to guarantee a hygienic use.
Do not use agents that contain alcohol or other aggressive cleaners.
The couch must be cleaned before it is first used and before it is
disinfected.
Clean the couch with a damp cloth.
Use a mild detergent to clean the surface.
Remove soap residue with a damp cloth and clean water.
Disinfection
Clean the couch as described above before proceeding to disinfect the
couch.
Follow the disinfectant's instructions for use.
Never use a mixture of disinfectants or a disinfectant and soap mixture.
Preferably use a cloth to apply the agent to the imitation leather.
Avoid water accumulation in the hinge points of the metal frame. (In the
worst case, exposing the hinges to water in the long-term can result in rust
formation.)
The disinfectants listed on the information sheet, are tested and safe for
use on the imitation leather of the treatment table.
For the most current version of the list see www.zimmer.de
Remark:
Since we have no control over the production and raw materials used for the
disinfectants, we are unable to give a general guarantee that these agents will
not damage the upholstery.
Maintenance
If you encounter any problem with the aXionXelect, stop using the couch and
contact your supplier.
Check the operation of every section adjustment regularly (at least once
every month). For correct operation keep the bars of the mechanism clean
and free of grease.
Check at least once every 6 months the functioning of the height adjustment
if this has not been used for a long time.
Disconnect the mains cable from the electricity supply if the couch is not
being used for a long period.
The aXionXelect should be checked once a year by authorized personnel of
Zimmer MedizinSysteme or by its authorized distributor.
The manufacturer will not be held responsible for the results of maintenance or
repairs by unauthorized persons!
Cleaning, maintenance and
troubleshooting
9
EN109-1550752-41
Page 17
Expected service life
This device will remain suitable for its intended use as long as it is subjected to
yearly maintenance by a qualified service engineer as described in the service
manual and the service engineer deems the device fit for use according to
specifications.
End of life
Your aXionXelect contains materials which can be recycled and/or are noxious
for the environment.
Please ensure that you are well informed of the local rules and regulations
regarding the removal of equipment and accessories.
Specifications
10
EN109-1550752-41
Page 18
Technical data
Height adjustment
ca. 48 106 cm
Length
ca. 207 cm (ca. 240 cm incl. traction frame)
Width
ca. 67 cm
Lifting capacity
200 kg
Motor speed
ca. 1.5 cm/sec
Mains voltage
100 240 Volt, 50/60 Hz
Current consumption
250 VA max.
Applied part
Type B: tabletop
IP classification
IPX0 (treatment couch)
Upholstery
Type AKV Peri
Fire safety
DIN EN 1021 part 1/2
Free of phthalates
yes
Biological
DIN EN ISO 10993-5
DIN EN ISO 10993-10
Technical modifications reserved!
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15