illumina iScan System Manual

Category
Software
Type
System Manual

This manual is also suitable for

Customize a short end-to-end workow guide with the Custom Protocol Selector
support.illumina.com/custom-protocol-selector.html
For Research Use Only. Not for use in diagnostic procedures.
August 2015
Document # 11313539 v01
Material # 20000396
Catalog # SY-950-1001DOC
ILLUMINA PROPRIETARY
iScan System Guide
ii Material # 20000396
Document # 11313539 v01
This document and its contents are proprietary to Illumina, Inc. and its affiliates ("Illumina"), and are intended solely for the
contractual use of its customer in connection with the use of the product(s) described herein and for no other purpose. This
document and its contents shall not be used or distributed for any other purpose and/or otherwise communicated, disclosed,
or reproduced in any way whatsoever without the prior written consent of Illumina. Illumina does not convey any license
under its patent, trademark, copyright, or common-law rights nor similar rights of any third parties by this document.
The instructions in this document must be strictly and explicitly followed by qualified and properly trained personnel in order
to ensure the proper and safe use of the product(s) described herein. All of the contents of this document must be fully read
and understood prior to using such product(s).
FAILURE TO COMPLETELY READ AND EXPLICITLY FOLLOW ALL OF THE INSTRUCTIONS CONTAINED HEREIN
MAY RESULT IN DAMAGE TO THE PRODUCT(S), INJURY TO PERSONS, INCLUDING TO USERS OR OTHERS, AND
DAMAGE TO OTHER PROPERTY.
ILLUMINA DOES NOT ASSUME ANY LIABILITY ARISING OUT OF THE IMPROPER USE OF THE PRODUCT(S)
DESCRIBED HEREIN (INCLUDING PARTS THEREOF OR SOFTWARE).
© 2015 Illumina, Inc. All rights reserved.
Illumina,24sure,BaseSpace,BeadArray,BlueFish,BlueFuse,BlueGnome,cBot,CSPro,CytoChip,DesignStudio,
Epicentre,ForenSeq,GAIIx,Genetic Energy,Genome Analyzer,GenomeStudio,GoldenGate,HiScan,HiSeq,HiSeq X,
Infinium,iScan,iSelect,, MiSeq,MiSeqDx,MiSeq FGx,NeoPrep,Nextera,NextBio,NextSeq,Powered by Illumina,
SeqMonitor,SureMDA,TruGenome,TruSeq,TruSight,Understand Your Genome,UYG,VeraCode,verifi,VeriSeq, the
pumpkin orange color, and the streaming bases design are trademarks of Illumina, Inc. and/or its affiliate(s) in the U.S. and/or
other countries. All other names, logos, and other trademarks are the property of their respective owners.
Revision History
Document Date Description of Change
Material # 20000396
Document # 11313539
v01
August
2015
Added troubleshooting information for malfunctioning status
lights in the iScan Reader Issues section.
Document # 11313539
Rev. B
February
2011
Updated to reflect iScan Control Software (ICS) version 3.2,
including automated genotype calling functionality and other
features.
Document # 11313539
Rev. A
April
2008
Part number changed to 11313539.
Document # 11308663
Rev. A
April
2008
Initial release.
iScan System Guide iii
iv Material # 20000396
Document # 11313539 v01
[This page intentionally left blank]
Table of Contents
Revision History iii
Table of Contents v
Chapter 1 Overview 1
Introduction 2
Additional Resources 3
Instrument Components 4
Consumables Required 9
iScan System Requirements 10
Scanning Process Overview 11
Chapter 2 Starting the iScan System 13
Procedure Summary 14
Powering up the iScan System 15
Starting the iScan Control Software 16
Using LIMS with the iScan System 17
Chapter 3 Loading BeadChips 19
Procedure Summary 20
Cleaning a BeadChip 21
Loading BeadChips onto a Carrier 22
Loading a Carrier into the iScan Reader 24
Rescanning Barcodes 27
Chapter 4 Configuring iScan Control Software 29
Introduction 30
Omitting BeadChips from a Scan 31
Changing Scan Settings 32
Chapter 5 Scanning 35
Introduction 36
Starting a Scan 37
Monitoring the Scan Progress 38
Pausing, Resuming, and Stopping a Scan 39
Completing a Scan 40
Chapter 6 Viewing Scan Results 41
Introduction 42
Log Files 43
Scan Metrics 44
Images 45
Generated Files 49
Chapter 7 Shutting Down, Maintenance, and Service 51
Introduction 52
Shutting Down the iScan System 53
Maintenance 54
Service 55
iScan System Guide v
vi Material # 20000396
Document # 11313539 v01
Appendix 8 Troubleshooting 57
Handling Errors 58
Issue Types 59
Registration Issues 60
Auto-Alignment Issues 62
iScan Reader Issues 64
Image Quality Issues 67
iScan Control Software Display Issues 69
Index 71
Technical Assistance 73
Chapter 1
iScan System Guide 1
Chapter 1 Overview
Overview
Introduction 2
Additional Resources 3
Instrument Components 4
Consumables Required 9
iScan System Requirements 10
Scanning Process Overview 11
Overview
2Material # 20000396
Document # 11313539 v01
Introduction
The iScan System is an easy-to-use, laser-based, high-resolution benchtop optical imaging
system that can rapidly scan and collect large volumes of data from Illumina DNA
analysis and RNA analysis high-density BeadChips.
With powerful scanning tools for gene expression and genotyping applications, the iScan
System supports the rapid, sensitive, and accurate imaging of Illumina BeadChip products.
Incorporating the highest performance optics and detection systems, the iScan System
delivers outstanding data quality and reproducibility with dramatically reduced scan time.
BeadChips are substrates used for multisample analysis in Illumina genotyping and gene
expression applications. Assay features are loaded into wells of a BeadChip to create an
organized array. The iScan System compiles a virtual representation of a BeadChip,
acquires images of the BeadChip features, records the information, and exports the data for
downstream analysis.
The iScan System can be integrated with Illumina LIMS (laboratory information
management software) and assay automation options, such as the AutoLoader
(AutoLoader2 or AutoLoader 2.x model), to maximize throughput to thousands of samples
per day. When used with Infinium® HD BeadChips and the AutoLoader, the iScan System
can report up to 225 million genotypes in a single day, offering the fastest path to
discovery.
Audience and Purpose
This guide describes the instrument components, software interface, consumables, and
operational procedures for scanning Illumina BeadChips on the iScan System.
This guide is for laboratory personnel and other individuals responsible for:
}Operating the iScan System
}Performing instrument and component maintenance
}Training personnel
WARNING
Using controls, making adjustments, or performing procedures other than the procedures
specified in this guide can result in hazardous laser light or radiation exposure.
Additional Resources
iScan System Guide 3
Additional Resources
The following documentation is available for download from the Illumina website.
Resource Description
iScan System Site Prep Guide
(document # 1000000000661)
Provides specifications for laboratory space, electrical
requirements, and environmental considerations.
iScan SystemQuickReference
Card (part # 11313555)
Provides instructions for starting up and shutting down the
iScan system.
iScan System Quick Reference
Guide (part #15020712)
Provides an overview of instrument components and
software, instructions for imaging data on BeadChips, and
instructions for starting up and shutting down the iScan
system.
iScan SystemSafety and
Compliance Guide (part #
15022905)
Provides safety information and product compliance
statements for the iScan system.
Visit the iScan System support page on the Illumina website for access to documentation,
software downloads, and online training.
Overview
4Material # 20000396
Document # 11313539 v01
Instrument Components
The iScan consists of the following components:
}iScan Reader
}Instrument control computer
}BeadChip carrier or carriers
}Power cords and other accessories
}[Optional] AutoLoader System
In addition to these components, purchase BeadChips developed for your application.
iScan Reader
The iScan Reader is a laser-based, high-resolution optical imaging system that includes red
and green lasers for detecting fluorescence information on BeadChips. BeadChips are
placed into carriers and the carriers are loaded 1 at a time into the adapter plate in the
iScan Reader tray.
NOTE
The adapter plate is calibrated to each individual iScan Reader. Do not switch adapter plates
between iScan Readers.
Figure 1 iScan Reader with Loaded BeadChips
ABeadChip Carrier
BAdapter Plate
CiScan Reader Tray
iScan Reader Barcode Scanner
The barcode scanner is internal to the iScan Reader, and allows you to identify each
BeadChip accurately.
Instrument Components
iScan System Guide 5
Status Lights
The iScan Reader status indicator lights and scan bar on the front panel show the status of
the instrument.
Figure 2 Status Lights and Scan Bar
Status Light Description
Power (blue) Steady blue indicates that the instrument is on.
Ready (green
check)
Steady green indicates that the instrument has been initialized and is ready
to scan. The Ready light flashes during initialization.
Warning
(amber
triangle)
Solid amber indicates that an instrument error has occurred. Try cycling the
power.
Scan Bar
(vertical blue
bar)
Steady indicates that the instrument is scanning. The Scan Bar is the blue
LED pipe to the left of the tray.
iScan Reader Tray
The iScan Reader tray accepts up to 4 BeadChips loaded in a BeadChip carrier.
Instrument Control Computer
The iScan Control Software (ICS) installed on instrument control computer allows you to
control the iScan Reader during BeadChip scanning.
Hard Drive Configuration
The iScan System contains 2 permanent drives (C: and D:) on the computer and 1
removable drive (H:) within the iScan Reader.
Drives C: and D: are physically separate drives; they are not part of 1 large drive
partitioned into 2 drives.
Overview
6Material # 20000396
Document # 11313539 v01
Drive Description
C: Drive C: is ~75GB in size and is installed with all the generic scanner software
required to run the iScan System. It has 2 partitions, 1 for the Windows operating
system and 1 for the BIOS. This drive contains at least 30GB in free disk space on
this drive to keep the scanner from slowing down or timing out during a scan.
D: The scanner uses drive D: to hold the temporary images and scan data before
*.idat files are created for a sample. Drive D: is about 500GB and is intended to
store scan data locally. It has a single partition for the entire drive.
H: The iScan Reader houses a removable drive H: that contains a reader-specific
configuration file. This file keeps reader-specific information with the iScan Reader,
in case the computer fails.
WARNING
Do not disconnect or format the H: drive. Doing so removes all reader-specific
information, requiring an Illumina Field Service Engineer to reinstall and reconfigure
the iScan System.
iScan Control Software
The iScan Control Software (ICS) allows you to control the iScan Reader when scanning
BeadChips. iScan Control Software includes a graphical user interface for:
}Loading BeadChips
}Acquiring images
}Automatically registering and extracting images
}Organizing and viewing the acquired images
Instrument Components
iScan System Guide 7
BeadChip Carriers
The iScan System comes with 1 BeadChip carrier. The carrier holds up to 4 BeadChips for
scanning at 1 time.
The BeadChip carriers consist of the following components.
Figure 3 BeadChip Carrier
ALatches (closed)
BRaised Stops
CBeadChip Slots
DLift Button
ERecessed Pin
Carrier and BeadChip Barcodes
BeadChip carrier barcodes are used to identify the individual BeadChip carriers and
determine when a BeadChip position is occupied or empty.
These barcodes are necessary when performing automated scanning. During automated
scanning, the barcodes enable problem BeadChips to be quickly located in the output or
error stacks. They also tell the scanner when to retry scanning a BeadChip position when it
does not successfully read a BeadChip barcode on the first attempt.
Overview
8Material # 20000396
Document # 11313539 v01
Figure 4 Top View of BeadChip Carrier Barcode
ACarrier Barcode
BEmpty Barcode
Figure 5 Side View of BeadChip Carrier Barcode
If the carrier does not have a barcode number, the carrier barcode number is listed as the
barcode number for the first BeadChip in the carrier.
The format for the number is _1stBeadChipBarcode.
Power Cords and Other Accessories
The iScan System comes with power and connection cords that are connected for you by
authorized Illumina personnel as part of the system installation. Do not unplug or
disconnect any cords unless instructed by Illumina Technical Support.
Consumables Required
iScan System Guide 9
Consumables Required
This section lists the Illumina kits and user-supplied consumables required to scan
BeadChips using the iScan System.
Illumina BeadChip Kits
Before you begin scanning, complete the appropriate assay for your BeadChip and
application.
The iScan System is compatible with the following types of BeadChips:
}Infinium HD
}DirectHyb
}Universal BeadChips
The iScan System is not compatible with the following products:
}16-1 BeadChips
}Universal Array Matrix (UAM)
User-Supplied Consumables
Make sure that you have the following user-supplied consumables before you begin
scanning BeadChips.
Consumable Supplier Purpose
Alcohol wipes, 70% Isopropyl,
Medium
VWR, catalog#15648-981 Cleaning the back side of the
BeadChip.
Ethanol, 99.5%, ACS, absolute
(optional)
Fisher Scientific, catalog #
AC61509-5000
Lab tissue, low-lint VWR, catalog#21905-026
Overview
10 Material # 20000396
Document # 11313539 v01
iScan System Requirements
Lab Equipment Requirements
If you are a current Illumina customer and already have the correct hardware package for
the assay you plan to run, you do not need to purchase any additional lab equipment
besides the iScan System itself.
For example, if you already have the Infinium upgrade package for a BeadArray Reader,
you do not need to purchase any additional equipment to scan Infinium BeadChips on a
new iScan Reader.
Air Table Specifications
The iScan Reader instrument is sensitive to vibrations. Therefore, an air table is required to
help isolate the reader from vibrations produced by the surrounding environment. The air
table is placed between the reader and the laboratory benchtop and is hooked up to a
regulator connected to the house air line.
Set the air table pressure to a nominal value of 25psi. Do not allow the air table pressure
exceed 40psi. The maximum pressure from the house line to the air regulator is 80psi.
Viewing iScan System Information
To view iScan System information, perform the following steps.
1Open the iScan Control Software and click the menu button in the upper left corner
of the screen.
2Select About. The iScan Control Software About screen appears, showing the iScan
Control Software release version, hardware information, and contact information for
Illumina Technical Support.
Scanning Process Overview
iScan System Guide 11
Scanning Process Overview
This section provides an overview of the steps involved in scanning BeadChips with the
iScan System. See the referenced sections for detailed instructions.
NOTE
This process does not describe how to use the AutoLoader with the iScan System to
automate BeadChip loading. For more information, see the AutoLoader2 User Guide or the
AutoLoader 2.x User Guide.
1Make sure that you have downloaded the *.dmap files (decode data) for your
BeadChips. Use the Illumina Decode File Client to download *.dmap files from
Illumina for your BeadChips. Download the Decode File Client from your MyIllumina
account.
2Start the iScan System.
3Load the BeadChips onto a carrier, load the carrier into the iScan Reader tray, and scan
the BeadChip carrier barcodes. For more information, see Loading BeadChips on page 19.
4If necessary, select a different image format, scan settings, data normalization settings,
and input/output path. Make sure that input and output paths are correct. For more
information, see Configuring iScan Control Software on page 29.
5Scan BeadChips. For more information, see Scanning on page 35.
6View BeadChip images. For more information, see Viewing Scan Results on page 41.
7Remove BeadChips and shut down the iScan System. For more information, see
Shutting Down the iScan System on page 53.
12 Material # 20000396
Document # 11313539 v01
Chapter 2
iScan System Guide 13
Chapter 2 Starting the iScan System
Starting the iScan System
Procedure Summary 14
Powering up the iScan System 15
Starting the iScan Control Software 16
Using LIMS with the iScan System 17
Starting the iScan System
14 Material # 20000396
Document # 11313539 v01
Procedure Summary
Perform the following steps to start the iScan System.
1Power up the iScan Reader.
2Power up the iScan System computer.
3Start the iScan ControlSoftware.
These steps are described in the following sections, along with information on using LIMS
with the iScan System.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80