  • Hello, I'm your chat assistant. I've reviewed the operation manual for the M400 M400/2XH Cond Ind transmitter. This document covers everything from installation and wiring to calibration, configuration, and troubleshooting. I'm here to help you navigate through the features and functions of this device, which includes a four-line LC display and support for various mounting options. Feel free to ask me any questions you have.
  • What type of sensors are compatible with the M400/2XH Cond Ind?
    Is the M400/2XH Cond Ind suitable for hazardous locations?
    What is the purpose of the 'H' symbol on the display?
Operation Manual
Transmitter M400/2XH Cond Ind
Transmitter M400/2XH Cond Ind 2
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
Transmitter M400/2XH Cond Ind 3
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
Operation Manual
Transmitter M400/2XH Cond Ind
Transmitter M400/2XH Cond Ind 4
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
Content
1 Introduction ___________________________________________________________________________________________ 7
1.1 Intended Use _____________________________________________________________________________________ 7
2 Safety instructions ______________________________________________________________________________________ 8
2.1 Denitionofequipmentanddocumentationsymbolsanddesignations ________________________________________ 8
2.2 Correctdisposaloftheunit _________________________________________________________________________ 9
2.3 ExinstructionsforM400seriesmulti-parametertransmitters _______________________________________________ 10
2.4 ExinstructionsforM400seriesmulti-parameterTransmitters–FMApproval ___________________________________ 12
2.4.1 InstructionsofusetobeconsideredunderFMapproval ___________________________________________ 12
2.4.1.1 General notes __________________________________________________________________ 14
2.4.1.2 Cautionarynotes,warningsandmarkings ____________________________________________ 15
2.4.1.3 ControlDrawings _______________________________________________________________ 17
3 Function and Design ____________________________________________________________________________________ 18
3.1 Function _______________________________________________________________________________________ 18
3.2 Design ________________________________________________________________________________________ 18
3.3 NavigationKeys _________________________________________________________________________________ 19
3.4 Displays _______________________________________________________________________________________ 20
3.5 MenuStructure __________________________________________________________________________________ 21
4 Mounting _____________________________________________________________________________________________ 22
4.1 UnpackingandinspectionofEquipment _______________________________________________________________ 22
4.2 Dimensions _____________________________________________________________________________________ 22
4.3 MountingProcedure ______________________________________________________________________________ 23
4.4 PanalMounting _________________________________________________________________________________ 24
4.5 WallMounting___________________________________________________________________________________ 24
4.6 PipeMounting ___________________________________________________________________________________ 25
5 Wiring _______________________________________________________________________________________________ 26
5.1 ElectricalConnectionoftheTransmitter ________________________________________________________________ 26
5.2 OverviewofTerminalBlocksinTransmitter _____________________________________________________________ 26
5.3 ConnectionofTerminalBlockTB2:SensorConnection ____________________________________________________ 28
5.3.1 AnalogInductiveConductivitySensors ________________________________________________________ 28
5.4 HARTCommunication _____________________________________________________________________________ 29
6 Placing Transmitter In, or Out of Service ___________________________________________________________________ 30
6.1 PlacingTransmitterinService _______________________________________________________________________ 30
6.2 PlacingTransmitterOutofService ____________________________________________________________________ 30
7 Quick Setup __________________________________________________________________________________________ 31
8 Sensor Calibration _____________________________________________________________________________________ 33
8.1 ProcessSensorCalibrationMode ____________________________________________________________________ 33
8.2 1 Point Sensor Calibration _________________________________________________________________________ 34
8.3 Zero Point Sensor Calibration _______________________________________________________________________ 35
9 Conguration _________________________________________________________________________________________ 36
9.1 EnterCongurationMode __________________________________________________________________________ 36
9.2 Measurement ___________________________________________________________________________________ 36
9.2.1 ChannelSetup ___________________________________________________________________________ 36
9.2.2 TemperatureSource _______________________________________________________________________ 37
9.2.3 Resistivity–ConductivityTemperatureCompensation _____________________________________________ 38
9.2.4 ConcentrationTable _______________________________________________________________________ 39
9.2.5 SetAveraging ____________________________________________________________________________ 41
9.3 AnalogOutputs __________________________________________________________________________________ 41
9.4 Set Points ______________________________________________________________________________________ 42
9.5 Alarm/Clean ___________________________________________________________________________________ 44
9.5.1 Alarm __________________________________________________________________________________ 44
9.5.2 Clean __________________________________________________________________________________ 45
9.6 Display ________________________________________________________________________________________ 45
9.6.1 Measurement ____________________________________________________________________________ 46
9.6.2 Resolution ______________________________________________________________________________ 46
9.6.3 Backlight _______________________________________________________________________________ 46
9.6.4 Name __________________________________________________________________________________ 47
9.7 HoldAnalogOutputs ______________________________________________________________________________ 47
10 System ______________________________________________________________________________________________ 48
10.1 SetLanguage ___________________________________________________________________________________ 48
10.2 Passwords _____________________________________________________________________________________ 49
10.2.1 ChangingPasswords ______________________________________________________________________ 49
10.2.2 ConguringMenuAccessforOperator _________________________________________________________ 49
10.3 Set/ClearLockout ________________________________________________________________________________ 50
Transmitter M400/2XH Cond Ind 5
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
10.4 Reset __________________________________________________________________________________________ 50
10.4.1 ResetSystem ____________________________________________________________________________ 50
10.4.2 ResetMeterCalibration ____________________________________________________________________ 50
10.4.3 ResetAnalogCalibration ___________________________________________________________________ 51
10.5 Set Date & Time _________________________________________________________________________________ 51
11 PID Setup ____________________________________________________________________________________________ 52
11.1 EnterPIDSetup __________________________________________________________________________________ 53
11.2 PIDAuto/Manual ________________________________________________________________________________ 53
11.3 Mode __________________________________________________________________________________________ 53
11.3.1 PID Mode _______________________________________________________________________________ 54
11.4 TuneParameters _________________________________________________________________________________ 54
11.4.1 PIDAssignment&Tuning __________________________________________________________________ 55
11.4.2 Setpoint&Deadband ______________________________________________________________________ 55
11.4.3 ProportionalLimits ________________________________________________________________________ 55
11.4.4 Corner Points ____________________________________________________________________________ 55
11.5 PIDDisplay _____________________________________________________________________________________ 56
12 Service ______________________________________________________________________________________________ 57
12.1 Diagnostics _____________________________________________________________________________________ 57
12.1.1 Model/Softwarerevision ___________________________________________________________________ 57
12.1.2 Digitalinput _____________________________________________________________________________ 58
12.1.3 Display ________________________________________________________________________________ 58
12.1.4 Keypad ________________________________________________________________________________ 58
12.1.5 Memory ________________________________________________________________________________ 58
12.1.6 Set OC _________________________________________________________________________________ 59
12.1.7 ReadOC _______________________________________________________________________________ 59
12.1.8 SetAnalogOutputs _______________________________________________________________________ 59
12.1.9 ReadAnalogOutputs ______________________________________________________________________ 60
12.2 Calibrate _______________________________________________________________________________________ 60
12.2.1 Calibrate Meter ___________________________________________________________________________ 60
12.2.1.1 Temperature ___________________________________________________________________ 60
12.2.2 CalibrateUnlock __________________________________________________________________________ 61
12.3 TechService ____________________________________________________________________________________ 61
13 Info _________________________________________________________________________________________________ 62
13.1 Messages ______________________________________________________________________________________ 62
13.2 Calibration Data _________________________________________________________________________________ 62
13.3 Model/SoftwareRevision __________________________________________________________________________ 63
14 Maintenance __________________________________________________________________________________________ 64
14.1 Frontpanelcleaning ______________________________________________________________________________ 64
15 Troubleshooting _______________________________________________________________________________________ 65
15.1 ErrormessagesforInductiveConductivitySensors–WarningandAlarmList __________________________________ 65
15.2 Warning-andAlarmindicationonthedisplay__________________________________________________________ 66
15.2.1 Warningindication________________________________________________________________________ 66
15.2.2 Alarmindication __________________________________________________________________________ 66
16 Accessories and Spare Parts _____________________________________________________________________________ 67
17 Specications _________________________________________________________________________________________ 68
17.1 Generalspecications _____________________________________________________________________________ 68
17.2 Electricalspecications ____________________________________________________________________________ 69
17.2.1 Generalelectricalspecications ______________________________________________________________ 69
17.2.2 4to20mA(withHART®) __________________________________________________________________ 69
17.3 Mechanicalspecications __________________________________________________________________________ 69
17.4 Environmentalspecications _______________________________________________________________________ 70
17.5 ControlDrawings ________________________________________________________________________________ 71
17.5.1 Installation,maintenanceandinspection ______________________________________________________ 71
17.5.2 ControlInstallationDrawingGeneralInstallation _________________________________________________ 72
17.5.3 Notes __________________________________________________________________________________ 76
18 Default Table _________________________________________________________________________________________ 77
18.1 Commonparameters _____________________________________________________________________________ 77
18.2 PIDDefaulValue _________________________________________________________________________________ 78
18.3 InductiveConductivityParameters ___________________________________________________________________ 78
19 Warranty _____________________________________________________________________________________________ 79
Transmitter M400/2XH Cond Ind 6
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
Transmitter M400/2XH Cond Ind 7
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
1 Introduction
1.1 Intended Use
TheM400/2XHCondIndisa2-wiretransmitterforanalyticalmeasurementswithHART®com-
municationcapability.Itisavailableas1-channelversionandcompatiblewithanaloginductive
conductivitysensors.
TheM400/2XHCondIndtransmitterisdesignedforuseintheprocessindustries.Additionally,
theM400/2XHCondIndtransmitterisintrinsicallysafecertifiedandsuitableforinstallationsin
hazardousareas.
METTLERTOLEDOacceptsnoliabilityfordamagesresultingfromincorrectuseoruseotherthan
thatintended.
M400/2XH Cond Ind parameter fit guide
Parameter M400/2XH Cond Ind
Analog
CondInd(InductiveConductivity)1)
1) InPro7250ST,InPro7250PFA.InPro7250HT
Table1: M400/2XHCondIndparameterfitguide
Transmitter M400/2XH Cond Ind 8
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
2 Safety instructions
Thismanualincludessafetyinformationwiththefollowingdesignationsandformats.
2.1 Denitionofequipmentanddocumentationsymbols
and designations
a WARNING:POTENTIALFORPERSONALINJURY.
a CAUTION:possibleinstrumentdamageormalfunction.
h NOTE:Importantoperatinginformation.
a Onthetransmitterorinthismanualtextindicates:Cautionand/orotherpossiblehazardinclud-
ingriskofelectricshock(refertoaccompanyingdocuments)
Thefollowingisalistofgeneralsafetyinstructionsandwarnings.Failuretoadheretothese
instructionscanresultindamagetotheequipmentand/orpersonalinjurytotheoperator.
– TheM400Transmittershouldbeinstalledandoperatedonlybypersonnelfamiliarwiththe
transmitterandwhoarequalifiedforsuchwork.
– TheM400Transmittermustonlybeoperatedunderthespecifiedoperatingconditions.Refer
tosection17"Specifications"onPage68.
– RepairoftheM400Transmittermustbeperformedbyauthorized,trainedpersonnelonly.
– Withtheexceptionofroutinemaintenance,cleaningproceduresorfusereplacement,asde-
scribedinthismanual,theM400Transmittermustnotbetamperedwithoralteredinany
manner.
– Mettler-Toledoacceptsnoresponsibilityfordamagecausedbyunauthorizedmodificationsto
thetransmitter.
– Followallwarnings,cautions,andinstructionsindicatedonandsuppliedwiththisproduct.
– Installequipmentasspecifiedinthisinstructionmanual.Followappropriatelocalandna-
tionalcodes.
– Protectivecoversmustbeinplaceatalltimesduringnormaloperation.
– Ifthisequipmentisusedinamannernotspecifiedbythemanufacturer,theprotectionpro-
videdbyitagainsthazardsmaybeimpaired.
WARNINGS:
Installationofcableconnectionsandservicingofthisproductrequireaccesstoshockhazard
voltagelevels.
MainpowerandOCcontactswiredtoseparatepowersourcemustbedisconnectedbeforeser-
vicing.
Switchorcircuitbreakershallbeincloseproximitytotheequipmentandwithineasyreachof
theOPERATOR;itshallbemarkedasthedisconnectingdevicefortheequipment.Mainpower
mustemployaswitchorcircuitbreakerasthedisconnectingdevicefortheequipment.
ElectricalinstallationmustbeinaccordancewiththeNationalElectricalCodeand/oranyother
applicablenationalorlocalcodes.
Transmitter M400/2XH Cond Ind 9
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
h NOTE: PROCESS UPSETS
Becauseprocessandsafetyconditionsmaydependonconsistentoperationofthistrans-
mitter,provideappropriatemeanstomaintainoperationduringsensorcleaning,replace-
mentorsensororinstrumentcalibration.
h NOTE:Thisisa2-wire-productwithtwoactive4–20mAanalogoutput.
2.2 Correct disposal of the unit
Whenthetransmitterisfinallyremovedfromservice,observealllocalenvironmentalregulations
forproperdisposal.
Transmitter M400/2XH Cond Ind 10
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
2.3 Ex instructions for M400 series multi-parameter
transmitters
M400seriesmulti-parametertransmittersareproducedbyMettler-ToledoGmbH.
IthaspassedtheinspectionofIECExandconformstofollowingstandards:
IEC 60079-0 : 2011
Edition: 6.0 Explosive atmospheres –
Part0:Generalrequirements
IEC 60079-11 : 2011
Edition: 6.0 Explosive atmospheres –
Part11:Equipmentprotectionbyintrinsicsafety"i"
IEC 60079-26 : 2006
Edition: 2 Explosive atmospheres –
Part26:Equipmentwithequipmentprotectionlevel(EPL)Ga
Ex Marking:
Ex ib [ia Ga] IIC T4 Gb
Ex ib [ia Da] IIIC T80°C Db IP66
Certificate No.:
IECEx CQM 12.0021X
SEV 12 ATEX 0132 X
1.SpecialConditionsofuse(X-markingintheCertificateNumber):
1. Avoidignitionhazardduetoimpactorfriction,preventmechanicalsparks.
2. Avoidelectrostaticdischargeonenclosuresurface,usewetclothonlyforcleaning.
3. Inhazardousarea,IP66cableglands(assupplied)mustbemounted.
2. Attention of use:
1. Ratedambienttemperaturerange:
–forgasatmosphere: –20~+60°C
–fordustatmosphere:–20~+57°C
2. Nooperationontheupgradeinterfaceinhazardousarea.
3. Usersshallnotarbitrarilyreplacetheinternalelectricalcomponents.
4. Wheninstallation,useandmaintenance,IEC60079-14shouldbeobserved.
5. Wheninstallationinexplosivedustatmosphere
 5.1CableglandorblankingplugtoIEC60079-0:2011andIEC60079-11:2011with
markingExiaIIICIP66shouldbeadopted.
 5.2Theoverlayswitchofmulti-parametertransmittershallbeprotectedfromlight.
 5.3Avoidhighriskofmechanicaldangerontheoverlayswitch.
6. Observethewarning:potentialelectrostaticcharginghazard-seeinstructions,avoidignition
hazardduetoimpactorfrictionforGaapplication.
7. Forconnectiontointrinsicallysafecircuits,usethefollowingmaximumvalues
Transmitter M400/2XH Cond Ind 11
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
M400/2XH and M400G/2XH
Terminal Function Safety Parameters
10, 11 Aout1 Ui=30V Ii=100mA Pi = 0.8 W Li≈0 Ci=15nF
12, 13 Aout2 Ui=30V Ii=100mA Pi = 0.8 W Li≈0 Ci=15nF
1,2;3,4; DigitalInput Ui=30V Ii=100mA Pi = 0.8 W Li≈0 Ci≈0
6,7;8,9; OCOutput Ui=30V Ii=100mA Pi = 0.8 W Li≈0 Ci≈0
P, Q AnalogInput Ui=30V Ii=100mA Pi = 0.8 W Li≈0 Ci=15nF
N, O RS485Sensor Ui=30V
Uo=5.88V
Ii=100mA
Io=54mA
Pi = 0.8 W
Po = 80 mW
Li≈0
Lo = 1 mH
Ci=0.7μF
Co=1.9μF
A,E,G pHSensor Uo=5.88V Io=1.3mA Po = 1.9 mW Lo = 5 mH Co=2.1μF
B,A,E,G ConductivitySensor Uo=5.88V Io=29mA Po = 43 mW Lo = 1 mH Co=2.5μF
K,J,I TemperatureSensor Uo=5.88V Io=5.4mA Po = 8 mW Lo = 5 mH Co=2μF
H,B,D Dissolvedoxygensensor Uo=5.88V Io=29mA Po = 43 mW Lo = 1 mH Co=2.5μF
L One-wire Sensor Uo=5.88V Io=22mA Po = 32 mW Lo = 1 mH Co=2.8μF
M400/2XH Cond Ind
Terminal Function Safety Parameters
10, 11 Aout1 Ui=30V Ii=100mA Pi = 0.8 W Li≈0 Ci=15nF
12, 13 Aout2 Ui=30V Ii=100mA Pi = 0.8 W Li≈0 Ci=15nF
1,2;3,4; DigitalInput Ui=30V Ii=100mA Pi = 0.8 W Li≈0 Ci≈0
6,7;8,9; OCOutput Ui=30V Ii=100mA Pi = 0.8 W Li≈0 Ci≈0
D,E,F,G,H InductiveConductivity
Sensor Uo=5.36V Io=17.2mA Po = 23 mW Lo = 1 mH Co=3.2μF
K,J,I TemperatureSensor Uo=5.88V Io=5.4mA Po = 8 mW Lo = 5 mH Co=2μF
MODEL: M400/2XH
Made by METTLER TOLEDO in China
Electrical data see user manual
II2(1) G Ex ib[ia Ga] IIC T4 Gb
II2(1) D Ex ib[ia Da] IIIC T80 C Db IP66
–20 to 57°C (Dust)
Feeding Voltage: 14 to 30 VDC
METTLER TOLEDO
T[amb]: –20 to 60°C (Gas) 1258
IECEx
SEV 12 ATEX 0123 X
IECEx CQM 12.0021X
www.mt.com/pro
Mettler-Toledo GmbH
Im Hackacker 15
8902 Urdorf, Switzerland
P/N: S/N:
Fig.1: Label Model M400/2XH
MODEL: M400G/2XH
Made by METTLER TOLEDO in China
Electrical data see user manual
II2(1) G Ex ib[ia Ga] IIC T4 Gb
II2(1) D Ex ib[ia Da] IIIC T80 C Db IP66
–20 to 57°C (Dust)
Feeding Voltage: 14 to 30 VDC
METTLER TOLEDO
T[amb]: –20 to 60°C (Gas) 1258
IECEx
SEV 12 ATEX 0123 X
IECEx CQM 12.0021X
www.mt.com/pro
Mettler-Toledo GmbH
Im Hackacker 15
8902 Urdorf, Switzerland
P/N: S/N:
Fig.2: Label Model M400G/2XH
Transmitter M400/2XH Cond Ind 12
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
MODEL: M400/2XH Cond Ind
Made by METTLER TOLEDO in China
Electrical data see user manual
II2(1) G Ex ib[ia Ga] IIC T4 Gb
II2(1) D Ex ib[ia Da] IIIC T80 C Db IP66
–20 to 57°C (Dust)
Feeding Voltage: 14 to 30 VDC
METTLER TOLEDO
T[amb]: –20 to 60°C (Gas) 1258
IECEx
SEV 12 ATEX 0123 X
IECEx CQM 12.0021X
www.mt.com/pro
Mettler-Toledo GmbH
Im Hackacker 15
8902 Urdorf, Switzerland
P/N: S/N:
Fig.3: Label Model M400G/2XH Cond Ind
2.4 Ex instructions for M400 series multi-parameter
Transmitters – FM Approval
2.4.1 Instructions of use to be considered under FM
approval
M400seriesmulti-parametertransmittersareproducedbyMettler-ToledoGmbH.
IthaspassedtheinspectionofNRTLcFMusandtofollowingrequirements:
Theequipmentisprovidedwithaninternalbondwiringandaninternalflyingleadwirefor
groundingpurposes.
US marking
Operatingtemperaturerange –20°Cto+60°C(–4°Fto+140°F)
Environmentaldesignation Enclosuretype4X,IP66
Intrinsicallysafe ClassI,Division1,GroupsA,B,C,DT4
ClassII,Division1,GroupsE,F,G
Class III
Intrinsicallysafe ClassI,Zone0,AExiaIICT4Ga
Parameters Entity:
Controldrawing12112601and12112602(4)
FISCO:
Controldrawing12112603and12112602(4)
Nonincendive ClassI,Division2,GroupsA,B,C,DT4A
ClassI,Zone2,GroupsIICT4
Certificateno. FM16US0216X
Transmitter M400/2XH Cond Ind 13
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
US marking
Standards FM3810:2005
ApprovalStandardforElectricalEquipmentfor
Measuerement,ControlandLaoratoryUse
ANSI/IEC-60529:2004
DegreesofProtectionProvidedbyEnclosures(IPCodes)
ANSI/ISA-61010-1:2004
Edition:3.0SafetyRequirementsforElectricalEquipment
forMeasurement,Control,andLaboratoryUse–Part1:
GeneralRequirements
ANSI/NEMA250:1991
EnclosuresforElectricalEquipement
(1,000VoltsMaximum)
FM3600:2011
ApprovalStandardforElectricalEquipmentforUsein
Hazardous(Classified)Locations–GeneralRequirements
FM3610:2015
ApprovalStandardforIntrinsicallySafeApparatusandAs-
sociatedApparatusforUseinClassI,II&III,Division1,
Hazardous(Classified)Locations
FM3611:2004
ApprovalStandardforNonincendiveElectricalEquipment
forUseinClassI&II,Division2,andClassIII,
Division1&2,Hazardous(Classified)Locations
ANSI/ISA-60079-0:2013
Edition:6.0ExplosiveAtmospheres–
Part0:GeneralRequirements
ANSI/ISA-60079-11:2012
Edition:6.0ExplosiveAtmospheres–
Part11:EquipementProtectionbyIntrinsicSafety"i"
Canadian marking
Operatingtemperaturerange –20°Cto+60°C(–4°Fto+140°F)
Environmentaldesignation Enclosuretype4X,IP66
Intrinsicallysafe ClassI,Division1,GroupsA,B,C,DT4
ClassII,Division1,GroupsE,F,G
Class III
Intrinsicallysafe Class I, Zone 0, Ex ia IIC T4 Ga
Parameters Entity:
Controldrawing12112601and12112602(4)
FISCO:
Controldrawing12112603and12112602(4)
Nonincendive ClassI,Division2,GroupsA,B,C,DT4A
Certificateno. FM16CA0119X
Transmitter M400/2XH Cond Ind 14
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
Canadian marking
Standards CAN/CSA-C22.2No.60529:2010
DegreesofProtectionProvidedbyEnclosures(IPCodes)
CAN/CSA-C22.2No.61010-1:2004
Edition:3.0SafetyRequirementsforElectricalEquipment
forMeasurement,Control,andLaboratoryUse–Part1:
GeneralRequirements
CAN/CSA-C22.2No.94:1976
SpecialPurposeExclosures–IndustrialProducts
CAN/CSA-C22.2No.213-M1987:2013
Non-IncendiveEquipmentforUseinCalssI,Division2
HazardousLocations–IndustrialProducts
CAN/CSA-C22.2No.60079-0:2015
Edition:2.0ExplosiveAtmospheres–
Part0:GeneralRequirements
CAN/CSA-C22.2No.60079-11:2014
Edition:2.0ExplosiveAtmospheres–
Part11:EquipementProtectionbyIntrinsicSafety"i"
2.4.1.1 General notes
TheMulti-parameterTransmitterM400/2(X)H,M400G/2XH,M400FF,M400PAandthe
M400/2XHCondInd aresuitableforuseinhazardousatmospheresofallcombustiblematerials
ofexplosiongroupsA,B,C,D,E,FandGforapplicationsrequiringClassI,II,III,Division1in-
strumentsandgroupsA,B,CandDforapplicationsrequiringClassI,Division2instruments
(NationalElectricalCode®(ANSI/NFPA70(NEC®),Article500;orCanadianElectrical(CE)
Code®(CECPart1,CAN/CSA-C22.1),
AppendixFwheninstalledinCanada,orofexplosiongroupsIIC,IIBorIIAforapplications
requiringClassI,Zone0,AEx/ExiaIICT4,Gainstruments(NationalElectricalCode®(ANSI/
NFPA70(NEC®),Article500;orCanadianElectrical(CE)Code®(CECPart1,CAN/CSA-C22.1),
AppendixFwheninstalledinCanada).
IftheMulti-parameterTransmitterM400/2(X)H,M400G/2XH,M400FF,M400PA,
M400/2XHCondIndisinstalledandoperatedinhazardousareas,thegeneralExinstallation
regulationsaswellasthesesafetyinstructionsmustbeobserved.
Theoperatinginstructionsaswellastheinstallationregulationsandstandardsthatapplyforex-
plosionprotectionofelectricalsystemsmustalwaysbeobserved.
Theinstallationofexplosion-endangeredsystemsmustalwaysbecarriedoutbyqualified
personnel.
Formountinginstructionsonspecificvalvesrefertothemountinginstructionssuppliedwiththe
mountingkit.MountingdoesnotaffectthesuitabilityoftheSVIFFpositionerforuseinapoten-
tiallyhazardousenvironment.
Theequipmentisnotintendedtobeusedaspersonalprotectiveequipment.Topreventinjury,
readthemanualbeforeuse.
Forlanguagetranslationassistancecontactyourlocalrepresentativeoremail
Pourlalanguedetraductionaide,contactezvotrereprésentantlocalouenvoyezune-mail
Transmitter M400/2XH Cond Ind 15
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
2.4.1.2 Cautionarynotes,warningsandmarkings
Hazardous location notes:
1. ForguidanceonUSinstallations,seeANSI/ISA-RP12.06.01,InstallationofIntrinsically
SafeSystemsforHazardous(Classified)Locations.
2. InstallationsintheUSshallcomplywiththerelevantrequirementsoftheNationalElectrical
Code®(ANSI/NFPA70(NEC®)).
3. InstallationsinCanadashallcomplywiththerelevantrequirementsoftheCanadianElec-
trical(CE)Code®(CECPart1,CAN/CSA-C22.1).
4. Wiringmethodsmustconformtoalllocalandnationalcodesgoverningtheinstallation,
andwiringmustberatedforatleast+10ºCabovethehighestexpectedambienttempera-
ture.
5. Wheretheprotectiontypeallowsanddependsonwiringglands,theglandsmustbecerti-
fiedforthetypeofprotectionrequiredandareaclassificationidentifiedontheequipmentor
systemnameplate.
6. Theinternalgroundingterminalshallbeusedastheprimaryequipmentgroundingmeans
andtheexternalgroundingterminalisonlyforasupplemental(secondary)bondingcon-
nectionwherelocalauthoritiespermitorrequiresuchaconnection.
7. Adust-tightconduitsealshallbeusedwheninstalledinClassIIconductiveandnon-con-
ductivedustenvironmentsandClassIIIcombustibleflyingsenvironments.
8. ApprovedsealsagainstingressofwaterordustarerequiredandtheNPTormetricthread
fittingsmustbesealedwithtapeorthreadsealantinordertomeetthehighestlevelofin-
gressprotection.
9. Whentheequipmentissuppliedwithplasticdustplugsintheconduit/cableglandentries;
itistheend-user’sresponsibilitytoprovidecableglands,adaptorsand/orblankingplugs
suitablefortheenvironmentinwhichtheequipmentisinstalled.Wheninstalledinahaz-
ardous(classified)location,thecableglands,adaptorsand/orblankingplugsshalladdi-
tionallybesuitableforthehazardous(classified)location,theproductcertification,and
acceptabletothelocalauthorityhavingjurisdictionfortheinstallation.
10. Theend-usermustconsultthemanufacturerforrepairdisclaimers,andonlycertifiedparts,
suchasentryplugs,mountingandcoverlockscrewsandgaskets,suppliedbythemanu-
facturerarepermitted.Nosubstitutionswithnon-manufacturersuppliedpartsarepermit-
ted.
11. Tightencoverscrewsto1.8Nm(15.8lb·in.).Overtorquingmaycauseenclosurebreak-
age.
12. TheminimumtighteningtorqueforM4(No.6)bindingscrewprotectiveconductortermi-
nalsis1.2Nm(10.6lb·in.)orgreater,asspecified.
13. Caremustbetakenduringinstallationtoavoidimpactsorfrictionthatcouldcreateanig-
nitionsource.
14. Usecopper,copper-cladaluminumoraluminumconductorsonly.
15. Therecommendedtighteningtorqueforfieldwiringterminalsis0.8Nm(7lb·in.)orgreat-
er,asspecified.
16. TheNonincendiveversionoftheMulti-parameterTransmitterM400/2(X)H,M400G/2XH
mustbeconnectedtolimitedoutputNECClass2circuits,asoutlinedintheNationalElec-
tricalCode®(ANSI/NFPA70(NEC®)),only.Ifthedevicesareconnectedtoaredundant
powersupply(twoseparatepowersupplies),bothmustmeetthisrequirement.
17. TheClassI,Zone2certificationsarebasedonDivisionevaluationsandthemarkingac-
ceptanceofArticle505oftheNationalElectricalCode®(ANSI/NFPA70(NEC®)).
Transmitter M400/2XH Cond Ind 16
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
18. TheMulti-parameterTransmitterM400/2(X)H,M400G/2XH,M400FF,M400PA,
M400/2XHCondIndassessedwerecertifiedbyFMApprovalsunderaType3Certification
SystemasidentifiedinISOGuide67.
19. Tamperingandreplacementwithnon-factorycomponentsmayadverselyaffectthesafe
useofthesystem.
20. Insertionorwithdrawalofremovableelectricalconnectorsistobeaccomplishedonly
whentheareaisknowntobefreeofflammablevapors.
21. TheMulti-parameterTransmitterM400/2(X)H,M400G/2XH,M400FF,M400PA,
M400/2XHCondIndisnotintendedforservicingormaintenanceoperation.Malfunction-
ingunitsoperatingoutofmanufacturer’sspecificationshouldbediscardedandreplaced
withanewoperationalunit.
22. Substitutionofcomponentsmayimpairintrinsicsafety.
23. Donotopenwhenanexplosiveatmosphereispresent.
24. Explosionhazard,donotdisconnectwhilecircuitisliveunlessareaisknowntobenon-
hazardous.
25. Explosionhazard,substitutionofcomponentsmayimpairsuitabilityforClassI,
Division2.
TheMulti-parameterTransmitterM400/2XH,M400G/2XHintrinsicallysafeapparatus,entityver-
sion,bearsthefollowinglabelmarking:
WARNING - EXPLOSION HAZARD. DO NOT REMOVE OR REPLACE WHILE CIRCUIT IS LIVE WHEN A FLAMMABLE OR COMBUSTIBLE ATMOSPHERE IS PRESENT.
WARNING - POTENTIAL ELECTROSTATIC CHARGE HAZARD. USE ONLY DAMP CLOTH WHEN CLEANING OR WIPING.DO NOT USE SOLVENT.
AVERTISSEMENT - RISQUE D'EXPLOSION. NE PAS DÉBRANCHER TANT QUE LE CIRCUIT EST
SOUS TENSION, À MOINS QU'lL NE S'AGISSE D'UN EMPLACEMENT NON DANGEREUX
Entry thread: Metric, 5xM20;
NOTE:
1. Conduit Hubs/Fittings Entry Thread;
2. Must use minimum Class I,Division2,Groups A,B,C,D,Type 4X and IP66
suitable Hubs/Fittings & Cable Glands to fulfill the complete FM certification.
Operation Manual No. 30031683
Enclosure Type 4X IP66
–20 ºC ≤ Ta ≤ +60 ºC
Control Drawing No. 12112601
FM16US0216X, FM16CA0119X
IS/I,II,III/1/ABCDEFG/T4
I/0/AEx ia/IIC/T4
I/0/Ex ia/IIC/T4
Intrinsically Safe Version
SÉCURITÉ INTRINSÈQUE, Exia
C/US
US
C
Entity
Fig.4: Label Model M400/2XH, M400G/2XH
TheTransmitterM400/2XHCondIndintrinsicallysafeapparatus,entityversion,bearsthefol-
lowinglabelmarking:
WARNING - EXPLOSION HAZARD. DO NOT REMOVE OR REPLACE WHILE CIRCUIT IS LIVE WHEN A FLAMMABLE OR COMBUSTIBLE ATMOSPHERE IS PRESENT.
WARNING - POTENTIAL ELECTROSTATIC CHARGE HAZARD. USE ONLY DAMP CLOTH WHEN CLEANING OR WIPING.DO NOT USE SOLVENT.
AVERTISSEMENT - RISQUE D'EXPLOSION. NE PAS DÉBRANCHER TANT QUE LE CIRCUIT EST
SOUS TENSION, À MOINS QU'lL NE S'AGISSE D'UN EMPLACEMENT NON DANGEREUX
Entry thread: Metric, 5xM20;
NOTE:
1. Conduit Hubs/Fittings Entry Thread;
2. Must use minimum Class I,Division2,Groups A,B,C,D,Type 4X and IP66
suitable Hubs/Fittings & Cable Glands to fulfill the complete FM certification.
Operation Manual No. 30396563
Enclosure Type 4X IP66
–20 ºC ≤ Ta ≤ +60 ºC
Control Drawing No. 12112601
FM16US0216X, FM16CA0119X
IS/I,II,III/1/ABCDEFG/T4
I/0/AEx ia/IIC/T4
I/0/Ex ia/IIC/T4
Intrinsically Safe Version
SÉCURITÉ INTRINSÈQUE, Exia
C/US
US
C
Entity
Fig.5: Label Model M400/2XH Cond Ind
Transmitter M400/2XH Cond Ind 17
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
TheMulti-parameterTransmitterM400/2Hnonincendiveversion,bearsthefollowinglabelmark-
ing:
WARNING - EXPLOSION HAZARD. DO NOT REMOVE OR REPLACE WHILE CIRCUIT IS LIVE WHEN A FLAMMABLE OR COMBUSTIBLE ATMOSPHERE IS PRESENT.
WARNING - POTENTIAL ELECTROSTATIC CHARGE HAZARD. USE ONLY DAMP CLOTH WHEN CLEANING OR WIPING.DO NOT USE SOLVENT.
AVERTISSEMENT - RISQUE D'EXPLOSION. NE PAS DÉBRANCHER TANT QUE LE CIRCUIT EST
SOUS TENSION, À MOINS QU'lL NE S'AGISSE D'UN EMPLACEMENT NON DANGEREUX
Entry thread: Metric, 5xM20;
NOTE:
1. Conduit Hubs/Fittings Entry Thread;
2. Must use minimum Class I,Division2,Groups A,B,C,D,Type 4X and IP66
suitable Hubs/Fittings & Cable Glands to fulfill the complete FM certification.
Operation Manual No. 30031683
Enclosure Type 4X IP66
–20 ºC ≤ Ta ≤ +60 ºC
FM16US0216X, FM16CA0119X
NI/1/2/ABCDEFG/T4A
NI/1/2/IIC/T4
Nonincendive Version
C/US
US
30 VDC, 100 mA, 0.8 mA
Fig.6: Label Model M400/2H
2.4.1.3 ControlDrawings
Refertosection17.5"ControlDrawings"onPage71.
Transmitter M400/2XH Cond Ind 18
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
3 Function and Design
3.1 Function
TheM400/2XHCondIndisa2-wiretransmitterforanalyticalmeasurementswithHART®com-
municationcapability.TheM400/2XHCondIndItisavailableas1-channelversionandcom-
patiblewithanaloginductiveconductivitysensors.
M400/2XH Cond Ind parameter fit guide
Parameter M400/2XH Cond Ind
Analog
CondInd(InductiveConductivity)1)
1) InPro7250ST,InPro7250PFA.InPro7250HT
Table2: M400/2XHCondIndparameterfitguide
3.2 Design
1
150 mm (5.90")
150 mm (5.90")
M400
Menu Cal Info Enter
ESC
2
3
4
5
Fig.7: Design(Overview)M400transmitter
1 Hardpolycarbonatehousing1/2DIN
2 Four-lineLCdisplay
3 Fivetactile-feedbacknavigationkeys
4 TB1–Inputandoutputanalogsignal
5 TB2–Sensorsignal
Transmitter M400/2XH Cond Ind 19
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
3.3 Navigation Keys
Navigation key Description
Menu
EnterMenumode.
Navigatebackwardswithinachangeabledataentryfield.
Cal
Enter Calibration mode.
Navigateforwardwithinachangeabledataentryfield.
Menu Cal
ESC
ReturntoMeasurementmode.
Pressthe and ckeysimultaneously(escape).
NOTE:Inordertobackuponlyonemenupage,movethecursorundertheUP
Arrowcharacter(u)atthebottomrightofthedisplayscreenandpress[ENTER].
Increaseadigit.
Navigatewithinaselectionofvaluesoroptionsofadataentryfield.
Info
Enter Info mode.
Decreaseadigit.
Navigatewithinaselectionofvaluesoroptionsofadataentryfield.
Enter
Confirmactionorselection.
Table3: Navigationkey
Transmitter M400/2XH Cond Ind 20
© 08/2017 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter M400/2XH Cond Ind
Printed in Switzerland 30 396 563
3.4 Displays
16.73
25.0
mS/cm
°C
A
A
16.73
25.0
mS/cm
°C
A
A
Channel Select = Analog
Parameter = Cond Ind u
A
A
2.8
9.5
%H3P04
ppkTDS
1
2
3
4
5
1
2
3
4
56
Fig.8: Left:Measurementmode(example),Right:Editmode(example)
1 Channelinformation
A:Analogsensorisconnected.
H:TransmitterisinHoldmode.
2 1stline(a),standardconfiguration
3 2ndline(b),standardconfiguration
4 3rdline(c),Measurementmode:Thescreendependsonconfiguration.
Editmode:Navigationthroughthemenuoreditingparameters
5 4thline(d):Measurementmode:Thescreendependsonconfiguration.
Editmode:Navigationthroughthemenuoreditingparameters
6 If a uisdisplayed,youcanusethecorthekeytonavigatetoit.Ifyouclick[ENTER]youwillnav-
igatebackwardsthroughthemenu(gobackonescreen).
Youcanconfiguretheinformationshownonthedisplayforeachline.Bydefaultinthemeasure-
mentmodenovaluesareshowninthethirdandinthefourthlineofthedisplay.RefertoChap-
ter9.2.1"ChannelSetup"onPage36.
Display Description
Hold mode Duringcalibration,clean,DigitalInwithAnalogOutput/OCthetransmitter
changesinHoldmode.Aftercalibrationorcleaniscompletedtheflashing"H"
stillremainsforfurther20seconds.The"H"isnotdisplayedanylongerwhen
DigitalInisdeactivated.
aIdicatesthatanalarmorerrorisoccurred.Thesymbolisdisplayedaslongas
thecauseforthealarmorerrorpersists.
Table4: Holdmodeandalarmsymbol
"Savechanges"
dialog
Description
Yes&Exit SavechangesandexittoMeasurementmode.
Yes&uSavechangesandgobackonescreen.
Usethisoptionifyouwanttocontinueconfiguringwithoutre-enteringtheEdit
mode.
No & Exit DonotsavechangesandexittoMeasurementmode.
Table5: "Savechanges"dialog
/