METTLER TOLEDO IND560 PLC Manual
1.4.1. Analog Output Setup Sub-Block
To configure the Analog Output Kit Option:
1. With power to the IND560 terminal removed, connect a volt or current meter to the appropriate
output. If the customer’s device is already connected, the meter is not necessary.
2. Apply power to the terminal and enter Setup. Navigate to PLC sub-block.
3. Select the Analog Output branch, then select the source. Choices are None, Displayed Weight
(the default),
Gross Weight, Rate. None disables the analog output. Displayed Weight outputs
an analog signal based on the displayed net or gross weight. When
Gross Weight is selected,
the analog signal is based on the gross weight regardless of what the net weight might be. In
order to be available as a source,
Rate must be configured at Scale > Rate.
4. Next, select the Channel. Options are Scale and None. Scale is the only option available now;
None is reserved for future use.
5. At the Zero Value prompt, enter the desired source value for which the analog output should be
zero. Typically this would be “0” in most applications; however, any valid value below the high
limit can be used.
6. At the Full Scale Value prompt, enter the desired source value at which the analog output
should be at its high limit. For sources of weight, this would typically be scale capacity, but it
could be lower. For rate, this should be the rate value that should provide a full analog output
signal.
7. After all these parameters have been entered, the analog output can be adjusted to meet the
customer’s requirements using the ZERO softkey and the SPAN softkey . To adjust
the zero reference analog signal, press the ZERO softkey .
8. Note that a display message is shown warning that during the adjustment, the analog output
will be set to zero and will not monitor changes in the source value. Press the ESCAPE softkey
to exit the zero adjustment process or press the OK softkey to continue the adjustment
process.
9. At the Analog Output - Cal Zero screen, use the softkeys to adjust the analog output signal to be
exactly zero on the customer’s device. The available softkeys are described in
Table 1-2.
Table 1-2: Softkey Descriptions
This adjusts the analog signal level down in large steps.
This adjusts the analog signal level down in small steps.
This adjusts the analog signal level up in small steps.
This adjusts the analog signal level up in large steps.
10. When the zero adjustment is complete, press the EXIT softkey to return to the Analog Output
screen.
11. Now, the full scale analog output value cab be adjusted by pressing the SPAN softkey . A
similar warning message will be shown indicating the analog output will be set to the high