![](//vs1.manuzoid.com/store/data-gzf/0f5416b88e8974eeaba9f7c7c91818b3/2/002275864.htmlex.zip/bg1.jpg)
The Puritan Bennett™ 980 ventilator includes NIV+
software. It works with NeoMode 2.0 software to
provide a NIV+ calibration function on the Puritan
Bennett™ 980 ventilator’s more settings screen.
This NIV+ calibration is supplemental to but not
a substitute for the SST.
KEY FEATURES
The NIV+ software is designed to measure end
inspiratory and end expiratory pressures at
the patient interface. This provides valuable
information to the clinician and may reduce the
uncertainty around eective pressure delivery
to the patient. It also provides a robust means of
determining disconnect, especially when leaks are
present. The NIV+ software:
∙ Includes interface calibration for prongs
and mask
∙ Provides monitored data values that reect end
inspiratory (PI END IF) and end expiratory (PEEPIF)
pressures on the patient side of the
NIV interface
∙ Assesses eciency of neonatal ventilation with
varying interfaces
PERFORMING AN
INTERFACE CALIBRATION
1. Ensure the Puritan Bennett™ 980 ventilator is in
the “Waiting for patient connect,” “Stand-by,”
or “Disconnect”.
2. With the NIV ventilation type selected,
touch the More Settings tab.
3. Touch the Calibrate button for the NIV
Interface Cal.
4. Remove the proximal ow and ETCO2 sensors
from the patient connection port of the
breathing circuit. Make sure the port is open
to the air.
5. Connect the neonatal NIV interface to the
breathing circuit.
6. Position the NIV interface similarly to how it will
be positioned on the patient.
NIV+ SOFTWARE
QUICK REFERENCE GUIDE
Puritan Bennett™ 980 Ventilator
7. Ensure that the patient connection portion
of the interface is open to air.
8. Select the interface to be calibrated.
9. Touch Conrm.
10. Keep the breathing circuit and interface still
and in a position that simulates patient use
until the calibration is nished.
3
2
6
8
9