Weinmann MEDUMAT Transport Instructions For Use Manual

Type
Instructions For Use Manual
MEDUMAT Transport
Ventilator
Instructions for Use for Devices from Software Version 6.7
2EN Contents
Contents
1. Overview . . . . . . . . . . . . . . . . . . . . . 4
2. Description . . . . . . . . . . . . . . . . . . . 21
2.1 Intended use. . . . . . . . . . . . . . . . . 21
2.2 Applications . . . . . . . . . . . . . . . . . 21
2.3 Operator and user qualification. . . 22
2.4 Function . . . . . . . . . . . . . . . . . . . . 22
3. Safety information . . . . . . . . . . . . . 24
4. Installation . . . . . . . . . . . . . . . . . . . 30
4.1 Connecting oxygen cylinder . . . . . 30
4.2 Connecting the hose system . . . . . 32
4.3 Connecting the inhalation
adapter. . . . . . . . . . . . . . . . . . . . . 34
4.4 Accessories from other
manufacturers . . . . . . . . . . . . . . . 35
4.5 Permanent installation of the unit. . 37
5. Operation . . . . . . . . . . . . . . . . . . . . 38
5.1 Controls . . . . . . . . . . . . . . . . . . . . 38
5.2 Switching the unit on/Self-test . . . 41
5.3 Navigating in menus . . . . . . . . . . . 44
5.4 Selecting emergency mode . . . . . . 46
5.5 Selecting a ventilation mode . . . . . 47
5.6 Changing the ventilation mode. . . 48
5.7 Selecting additional ventilation
functions . . . . . . . . . . . . . . . . . . . 49
5.8 Performing ventilation. . . . . . . . . . 50
5.9 Monitoring ventilation . . . . . . . . . 51
5.10 Performing inhalation . . . . . . . . . . 53
5.11 Alarm signals . . . . . . . . . . . . . . . . 54
5.12 Ventilation with breathing
system filters (not supplied with
the unit) . . . . . . . . . . . . . . . . . . . . 56
5.13 Ending ventilation. . . . . . . . . . . . . 56
5.14 Calculating the Oxygen level/
Operating time . . . . . . . . . . . . . . . 57
5.15 Alternative ventilation. . . . . . . . . . 58
5.16 Changing battery during use . . . . 58
5.17 Battery management . . . . . . . . . . 59
6. Ventilation modes . . . . . . . . . . . . . 63
6.1 Classification of the ventilation
modes . . . . . . . . . . . . . . . . . . . . . 63
6.2 Important ventilation parameters . . 65
6.3 Additional functions and safety
functions . . . . . . . . . . . . . . . . . . . 66
6.4 Pressure-controlled ventilation
modes . . . . . . . . . . . . . . . . . . . . . 68
6.5 Volume-controlled ventilation
modes . . . . . . . . . . . . . . . . . . . . . 78
7. Main menu . . . . . . . . . . . . . . . . . . . 84
7.1 Activating automatic alarm
limits. . . . . . . . . . . . . . . . . . . . . . . 84
7.2 Alarm Limits . . . . . . . . . . . . . . . . . 85
7.3 Curves . . . . . . . . . . . . . . . . . . . . . 86
7.4 Advanced ventilation
parameters . . . . . . . . . . . . . . . . . . 87
7.5 Apnea ventilation parameters . . . . 89
7.6 Audio/Video . . . . . . . . . . . . . . . . . 90
7.7 Options. . . . . . . . . . . . . . . . . . . . . 91
7.8 Night colors . . . . . . . . . . . . . . . . . 94
7.9 NVG (Night Vision Goggles) . . . . . 94
8. Operator menu . . . . . . . . . . . . . . . 96
8.1 Password Page . . . . . . . . . . . . . . . 99
8.2 Ventilation Modes. . . . . . . . . . . . . 99
8.3 File Export/Import . . . . . . . . . . . . 101
8.4 Software Update. . . . . . . . . . . . . 103
8.5 Options. . . . . . . . . . . . . . . . . . . . 105
8.6 User Settings. . . . . . . . . . . . . . . . 105
9. Hygienic preparation . . . . . . . . . . 109
9.1 MEDUMAT Transport . . . . . . . . . 109
9.2 Hose systems . . . . . . . . . . . . . . . 109
9.3 Parts and accessories. . . . . . . . . . 110
9.4 BiCheck flow sensor . . . . . . . . . . 110
9.5 Fittings . . . . . . . . . . . . . . . . . . . . 111
9.6 Hygiene input filter (optional) . . . 111
9.7 Cleaning, disinfection and
sterilization . . . . . . . . . . . . . . . . . 111
10. Function check . . . . . . . . . . . . . . . 115
10.1 Intervals . . . . . . . . . . . . . . . . . . . 116
10.2 Checking the system for leaks. . . 117
10.3 Checking the patient valve
(reusable hose system only). . . . . 118
10.4 Performing a function check . . . . 119
11. Troubleshooting . . . . . . . . . . . . . 123
Contents EN 3
11.1 Troubleshooting . . . . . . . . . . . . . 123
11.2 System alarms . . . . . . . . . . . . . . 125
11.3 Physiologic alarms . . . . . . . . . . . 127
12. Maintenance . . . . . . . . . . . . . . . . . 129
12.1 MEDUMAT Transport . . . . . . . . . 129
12.2 Sending in device . . . . . . . . . . . . 130
12.3 Batteries. . . . . . . . . . . . . . . . . . . 131
12.4 Accessories. . . . . . . . . . . . . . . . . 131
12.5 Changing the suction filter . . . . . 131
12.6 Changing the hygiene input
filter . . . . . . . . . . . . . . . . . . . . . . 132
12.7 Storage . . . . . . . . . . . . . . . . . . . 133
12.8 Disposal . . . . . . . . . . . . . . . . . . . 134
13. Product, accessories . . . . . . . . . . . 135
13.1 Standard scope of supply . . . . . . 135
13.2 Accessories. . . . . . . . . . . . . . . . . 136
13.3 Options . . . . . . . . . . . . . . . . . . . 139
13.4 Replacement parts . . . . . . . . . . . 140
14. Technical Data . . . . . . . . . . . . . . . 142
14.1 Specifications . . . . . . . . . . . . . . . 142
14.2 Battery specifications . . . . . . . . . 147
14.3 Block diagram . . . . . . . . . . . . . . 148
14.4 Separation distances. . . . . . . . . . 148
14.5 O2 consumption of the unit . . . . 149
14.6 Possible O2 concentration with
counterpressure . . . . . . . . . . . . . 149
14.7 Attainable tidal volume with
counterpressure . . . . . . . . . . . . . 151
14.8 Calculation of body weight on the
basis of body height . . . . . . . . . . 152
15. Glossary . . . . . . . . . . . . . . . . . . . . 153
16. Warranty . . . . . . . . . . . . . . . . . . . . 156
17. Declaration of Conformity . . . . . . 156
4EN Overview
1. Overview
1 Alarm LED
Glows to indicate alarms.
2 USB interface
Means of data transfer for servicing and
maintenance purposes.
3 O2/AIR inlet
Connection point, e.g., for an oxygen cylinder or
sterile compressed air.
4 O2/AIR inlet/outlet
This connection point enables oxygen to be ex-
tracted, e.g., using an inhalation device, or an ox-
ygen or sterile compressed air source to be
connected.
5 Filter compartment cover, air inlet
Covers the filter and ensures it is securely
positioned.
6 Ventilation connection terminal
The tube system is connected here.
7 Rechargeable battery
Provides mobile power supply to the unit.
8 DC connection
For DC power supply via an external power supply
unit or via the electrical circuit of an ambulance or
rescue vehicle.
9 External power supply unit
Provides power supply to the unit via a
100V - 240V grid.
Connections on MEDUMAT Transport
5Filter compartment
cover, air inlet
1Alarm LED
3O2/AIR inlet
2USB interface
7Rechargeable
battery
8DC connection
6Ventilation connection
terminal
9External power supply unit
4O2/AIR inlet/
outlet
Overview EN 5
1 CO2 measuring hose connection
The CO2 measuring hose of the patient hose sys-
tem is attached to this connection via the connec-
tion plug.
2 PEEP control hose connection
The PEEP control hose of the patient hose system
is attached to this connection via the connection
plug.
3 Pressure-measurement hose connection
The pressure-measurement hose of the patient
hose system is attached to this connection via the
connection plug.
4 Ventilation hose/inhalation adapter
connection
The ventilation hose of the patient hose system or
the inhalation adapter for an inhalation mask is
connected at this connection point.
5 BiCheck flow sensor connection line
connection
The BiCheck flow sensor connection line of the
patient hose system is attached to this connec-
tion.
Ventilation connection terminal
1CO2 measuring hose connection
2PEEP control hose connection
3Pressure-measurement hose
connection
4Ventilation hose/inhalation adapter
connection
5BiCheck flow sensor connection line connection
6EN Overview
1, 3, 5 Context-dependent function button
These buttons are used to set various ventilation
parameters, depending on the ventilation mode
selected.
2 Alarm mute button with LED
You can silence these acoustic alarms temporarily
(for 2 minutes) by pressing this button briefly
(< 1 s). If alarms are muted, the LED lights up.
Visual alarms are still displayed. The alarm menu
opens if this button is held depressed for longer
(> 2 s).
4 Function button for main menu
This button calls up the main menu.
6 Function button for 100% O2
This button calls up the 100% O2 function to ven-
tilate the patient briefly (for 2 minutes) with
100% O2 (FiO2 = 1.0).
7 Function button for inspiratory O2
concentration
This button calls up the O2 concentration menu.
The required inspiratory O2 concentration in the
respiratory gas can be set in this menu.
8 On/Standby/Off button
A short press switches the unit on and off. A long
press switches it off completely.
9 Navigation knob
For navigating in menus and confirming your set-
tings on the unit. During ventilation, this knob is
to set the I:E ratio.
10 Context-dependent control knobs
For setting various parameters, depending on
which ventilation mode is active. Settings made
here must be confirmed with the Navigation
knob.
11 Function buttons for emergency
ventilation
These buttons start emergency ventilation. By
pressing the buttons, preset parameters for in-
fants, children or adults are activated.
Controls of MEDUMAT Transport 1Context-dependent
function button
11Function buttons for
emergency ventilation
9Navigation knob
10 Context-dependent control knobs
4Function button for
main menu
3Context-dependent
function button
6Function button for
100% O2
7Function button for
inspiratory O2
concentration
5Context-dependent
function button
8On/Standby/Off
button
2Alarm mute button with
LED
Overview EN 7
1 Battery/Line operation indicators
Indicates whether the unit is being operated with
the external power supply unit (upper LED) or
with the internal battery (lower LED).
2 Function indicator for context-
dependent function buttons
The currently available function of the context-
dependent function buttons is indicated here.
3 Ventilation progress display
The ventilation procedure is shown here as a
curve or pressure gauge, depending on the
selected display mode.
4 Function indicator for context-
dependent control knobs
The currently available function of the context-
dependent control knobs is indicated here.
5 Function indicator for context-
dependent function buttons
The three directly selectable emergency ventila-
tion modes (Infant, child, adult) are indicated
here.
6 Numeric measurement display
The current measurements are shown here
numerically along with the corresponding
alarm limits.
7 Battery charge status
The battery charge status is indicated here.
8 Numeric measurement display
Displays the measured inspiratory O2 concentra-
tion (FiO2).
9 Info field
Information (error messages, visual alarms) about
the state of the patient and the ventilator is dis-
played here. The time of day is also displayed in
this field.
10 Mode indicator
The ventilation mode set by the user is indicated
here.
Display of MEDUMAT Transport
2Function
indicator for
context-
dependent
function
buttons
9Info field 10 Mode indicator
8Numeric measurement display
7Battery
charge
status
6Numeric mea-
surement display
3Ventilation progress display
1Battery/Line operation indicators
4Function indica-
tor for context-
dependent con-
trol knobs
5Function indica-
tor for context-
dependent func-
tion buttons
8EN Overview
1 Ventilation hose
The respiratory gas flows through the ventilation
hose to the patient valve.
2 Water filter for CO2 measurement
The water filter protects the measuring chamber
of the MEDUMAT Transport against moisture and
contamination from the patient's respiratory gas.
3 Connector
The measurement-tube system is connected to
MEDUMAT Transport by means of this connector.
4 PEEP control tube
With this tube, MEDUMAT Transport controls the
patient valve and the PEEP.
5 BiCheck flow sensor connection line
This electric lead transfers the measuring signals
from the BiCheck flow sensor to the MEDUMAT
Transport.
6 BiCheck flow sensor
This sensor supplies monitoring data on flow,
MVe, Vte and f.
7 Elbow
The mask/tube is connected here. The elbow is
removable, i.e., the mask/tube can also be
connected to the BiCheck flow sensor itself,
depending on the position of the patient.
8 Blanking plug
The blanking plug (Luer lock) is used to seal off
the CO2 outlet if your MEDUMAT Transport is not
equipped with CO2 measurement or this is not
activated.
9 Patient valve
Switchover between inspiration and expiration
happens here.
10 CO2 measuring hose
Test gas is removed via this tube if your unit is
equipped with the optional CO2 measurement
facility.
11 Pressure- measurement tube
For patient-side measurement of ventilation
pressure.
12 Velcro strap with clip
Used for fixing the patient hose system to the
patient's clothing.
13 Tube protection sleeve
Protects tubes and leads against soiling and
damage.
Notice:
Detailed information about the hose systems can be found in the "Patient Hose
System" instructions for use WM 66696.
Hose system (reusable and disposable versions available)
5BiCheck flow sensor
connection line
10 CO2 measuring hose
4PEEP control tube
11 Pressure-
measurement
tube
1Ventilation hose
7Elbow
6BiCheck flow sensor
9Patient valve
12 Velcro strap
with clip
2Water filter for CO2
measurement
3Connector
8Blanking plug
13 Tube
protection
sleeve
Overview EN 9
1 Filter grommet
Holds the suction filter in position.
2 Filter mount
For installing a hygiene input filter in the device.
3 Hygiene input filter
Protects the device from viral and bacterial
contamination.
Hygiene input filter (optional)
3Hygiene input filter
2Filter mount
1Filter grommet
10 EN Overview
1 Cover
Blocks the top two measuring ports on the device
during inhalation.
2 Inhalation mask
The patient inhales oxygen through the inhala-
tion mask.
3 Inhalation adapter
For connecting an inhalation mask to the device.
Inhalation adapter
2Inhalation mask
3Inhalation adapter
1Cover
Overview EN 11
MEDUMAT Transport Main menu
Alarm limits
AActivate automatic alarm limits
Curves
Audio/Video
Options
Night colors
PPressure, flow
Pressure, CO
2
Pressure, flow, CO
2
Pressure, flow,
measurements
Pressure, CO
2
,
measurements
Pressure gauge
Automatic alarm limits
MVe n
MVe p
f n
Apnea
etCO
2
n
etCO
2
p
Advanced ventilation parameters
Brightness/Day
Brightness/Night
Brightness/NVG
Volume
Alarm LED
Year
Month
Day
Hour
Minute
CO
2
configuration
Date, time
Device data
Pressure ramp
Flow ramp
Flow progress
decreasing
constant
Plateau time
Trigger thresholds
Inspiration
Expiration
Trigger time slot
Apnea ventilation parameters
AActivated
Apnea mode
BiLevel + ASB
SIMV + ASB
Settings
PEEP
pInsp
Vt
Freq
I:E
Suction activated
Unit
mmHG
Vol%
kPa
Med. O΍
Compressed air
Concentrator O΍
Bluetooth
Compressed gas supply
Hygiene input filter
NVG
MEDUMAT Transport Main menu
12 EN Overview
Symbols used on the display
Symbol Meaning
Emergency mode – Infant (up to approx. 1 year)
Emergency mode – Child (approx 1-12 years)
Emergency mode – Adult (approx 13 years and over)
Battery status indicator
Mode menu
NIV
BiLevel + ASB
aPCV
PCV
CPAP + ASB
PRVC + ASB
IPPV
S-IPPV
SIMV + ASB
Inhalation
Pre-oxygenation
Overview EN 13
Tick box: option activated
Radio button: function selected
Navigate upwards
Navigate downwards
Increase value
Decrease value
Confirm your selection
Navigation knob active
Bluetooth connection:
Symbol is gray when connection has been activated
Symbol is blue during communication
Acoustic alarm output activated
Acoustic alarm output deactivated
Alarm volume set to < 50%
Symbol Meaning
14 EN Overview
Acoustic alarm output permanently muted (NVG mode only)
Time
Trigger time slot
Symbol Meaning
Overview EN 15
Function of the controls during ventilation
Depending on the ventilation mode selected, you can set the following ventilation
parameters using the controls:
1 2345
6
7
8
Ventilation
mode
Control
knob
1
Control
knob 2
Control
knob 3
Control
knob 4
Navigation
knob 5
Function
button 6
Function
button 7
Function
button 8
BiLevel +
ASB
PEEP pInsp pMax Freq.
I:E and
Selection/
Confirmation
Δ
pASB Trigger Mode
aPCV
PEEP pInsp pMax Freq. I:E and
Selection/
Confirmation
Trigger
time slot Trigger Mode
PCV
PEEP pInsp pMax Freq.
I:E and
Selection/
Confirmation
--Mode
CPAP + ASB
PEEP - pMax - Only Selection/
Confirmation
Δ
pASB Trigger Mode
PRVC + ASB
PEEP Vt pMax Freq.
I:E and
Selection/
Confirmation
Δ
pASB Trigger Mode
IPPV
PEEP Vt pMax Freq.
I:E and
Selection/
Confirmation
--Mode
S-IPPV
PEEP Vt pMax Freq.
I:E and
Selection/
Confirmation
- Trigger Mode
16 EN Overview
SIMV + ASB
PEEP Vt pMax Freq.
I:E and
Selection/
Confirmation
Δ
pASB Trigger Mode
Inhalation
----Flow --Mode
Pre-
oxygenation
----Flow --Mode
Ventilation
mode
Control
knob
1
Control
knob 2
Control
knob 3
Control
knob 4
Navigation
knob 5
Function
button 6
Function
button 7
Function
button 8
Overview EN 17
Special markings
Symbol Meaning
MEDUMAT Transport type plate
1
Consult instructions for use
1
Date of manufacture
1
Degree of protection against electric shock: type BF device
1
Input
MEDUMAT Transport
Battery
1MEDUMAT Transport type plate
3STK and service label
9Cover of USB interface
8Filter compartment
cover
6O2/AIR inlet/outlet
2Voltage input
5Ventilation hose connection 4Follow the instructions for use
7O2/AIR inlet
10 Rechargeable battery type plate
Hygiene input filter (optional)
11 Hygiene input filter
18 EN Overview
1
DC voltage
1Imin
Imax
Minimum and maximum current
1
Do not dispose of the unit in the household waste.
1
Type of protection against electric shock: protection class II device
1IPX4 Protection against ingress of water
1
CE mark (confirms that the device complies with the applicable European
Directives).
1SN Serial number
Other marks
2
Input voltage
3
Service label: indicates when the next service is required.
3STK label: (only in the Federal Republic of Germany) indicates when the
next safety check in accordance with §6 Medical Device Operator
Ordinance (MPBetreibV) is required.
4
Follow the instructions for use
5
Maximum pressure 100 mbar
Symbol Meaning
Overview EN 19
6
Volume flow rate
7
Input 2.7 bar–6 bar O2 or sterile compressed air
8,9
Consult instructions for use
Rechargeable battery type plate
10
Do not dispose of the unit in the household waste.
10
Do not subject the unit to hard knocks or shocks.
10
Do not open the unit using force.
10
Protect the unit against heat.
10
Protect the unit against moisture.
Symbol Meaning
O
2
/ AIR
270 – 600 kPa
80 – 150 l/min
20 EN Overview
Labeling on the packaging
Safety information in these instructions for use
The safety instructions in these instructions for use are marked as follows:
Warning!
Warns of risk of injury and possible damage to the unit.
Caution!
Warns of material damage and possibly incorrect therapy results.
Notice:
Offers useful tips.
Hygiene input filter (optional)
11
Input
Consult instructions for use
Do not reuse
Symbol Meaning
MEDUMAT Transport:
SN Serial number of the unit
Permissible storage temperature: -30°C to +70°C
RH % 0-95 Permissible humidity for storage: up to 95% relative humidity
Symbol Meaning
3
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160