Mindray BeneLink Module User manual

Type
User manual
BeneLink Module
Operators Manual
BeneLink Module Operators Manual I
©
Copyright 2018 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights
reserved.
Rel
ease time: December 2018
Rev
ision 3.0.
WARNING
Federal Law (USA) restricts this device to sale by or on the order of a
physician or other practitioner licensed by U.S. state law to use or order
the use of this device.
II BeneLink Module Operators Manual
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this
manual. This manual may refer to information protected by copyrights or patents and
does not convey any license under the patent rights or copyright of Mindray, or of
others.
Mindray intends to maintain the contents of this manual as confidential information.
Disclosure of the information in this manual in any manner whatsoever without the
written permission of Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaption and translation or
any other derivative work of this manual in any manner whatsoever without the
written permission of Mindray is strictly forbidden.
is the trademark, registered or otherwise, of Mindray in China and
other countries. All other trademarks that appear in this manual are used only for
information or editorial purposes. They are the property of their respective owners.
This posting serves as notice under 35 U.S.C.§287(a) for Mindray patents:
http://www.mindrayna.com/patents.
BeneLink Module Operators Manual III
Manufacturers Responsibility
Contents of this manual are subject to changes without prior notice.
Mindray is responsible for the effects on safety, reliability and performance of this
product, only if:
all installation operations, expansions, changes, modifications and repairs of
this product are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable
national and local requirements;
the product is used in accordance with the instructions for use.
WARNING
Only skilled/trained clinical professionals should operate this
equipment.
It is important for the hospital or organization that uses this
equipment to perform a recommended service/maintenance plan.
Neglect of this may result in machine breakdown or personal injury.
IV BeneLink Module Operators Manual
Service
Mindray maintains a network of service representatives and factory-trained
distributors. Prior to requesting service, perform a complete operational check of the
instrument to verify proper control settings. If operational problems continue to exist,
contact Mindray service.
In North America contact the Service Department at (800) 288-2121, ext: 8116 for
Technical Support or (201) 995-8000 for assistance in determining the nearest field
service location.
Please include the instrument model number, the serial number, and a description of
the problem with all requests for service.
Any questions regarding the warranty should be directed to your local sales or service
representative.
BeneLink Module Operators Manual V
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address:
Mindray Building, Keji 12th Road South, High-tech industrial
park, Nanshan,Shenzhen 518057,P.R.China
Website: www.mindray.com
E-mail Address: service@mindray.com
Tel: +86 755 81888998
Fax: +86 755 26582680
Distributor: Mindray DS USA, Inc.
Address: 800 MacArthur Boulevard, Mahwah, New Jersey, 07430 USA
Tel: 1.800.288.2121, 1.201.995.8000
Website: www.mindray.com
VI BeneLink Module Operators Manual
Preface
Manual Purpose
This manual provides detailed information about the assembly, disassembly, and
testing of the equipment to support effective troubleshooting and repair. It is not
intended to be a comprehensive, in-depth explanation of the product architecture or
technical implementation. Use of the manual is a necessary for proper equipment
maintenance and will help eliminate equipment damage and personnel injury.
Intended Audience
This manual is for biomedical engineers, authorized technicians or service
representatives responsible for troubleshooting, repairing and maintaining the
monitors
Contact your local Mindray Service Organization for information on product courses
which address service and support for this product.
BeneLink Module Operators Manual 1
Contents
1 Safety ...................................................................................................................... 1-1
1.1 Safety Information .................................................................................................................. 1-1
1.2 Warnings ................................................................................................................................... 1-2
1.3 Cautions .................................................................................................................................... 1-2
1.4 Notes.......................................................................................................................................... 1-3
2 Interfacing with External Devices ............................................................................ 2-1
2.1 BeneLink Introduction ........................................................................................................... 2-1
2.2 Physical Specifications ........................................................................................................... 2-1
2.3 Data transmission ................................................................................................................... 2-2
2.4 Supported Devices ................................................................................................................. 2-2
2.4.1 Supported Anesthesia Machines ............................................................................ 2-2
2.4.2 Supported Ventilators ............................................................................................... 2-3
2.4.3 Supported tcGas Monitoring Devices .................................................................... 2-5
2.4.4 Supported Infusion System (For BeneVision N Series Monitors only) ............. 2-5
2.4.5 Supported Other Devices ......................................................................................... 2-6
2.4.6 Connecting External Devices via DIAP Protocol (For BeneVision N Series
Monitors only)
....................................................................................................................... 2-6
2.5 Configuring the ID Adapter .................................................................................................. 2-7
2.6 Connecting External Devices................................................................................................ 2-8
2.7 Accessories .............................................................................................................................2-10
3 Integrating the Anesthesia Machine ........................................................................ 3-1
3.1 Draeger Apollo ........................................................................................................................ 3-1
3.1.1 Output SignalsParameters ................................................................................... 3-1
3.1.2 Output SignalsAlarms ........................................................................................... 3-4
3.1.3 Output Signals Waveforms (For BeneVision N Series Monitors only) ........... 3-8
2 BeneLink Module Operators Manual
3.2 Draeger Fabius GS /Fabius Tiro .............................................................................................3-9
3.2.1 Output Signals Parameters ....................................................................................3-9
3.2.2 Output Signals Alarms ......................................................................................... 3-11
3.3 Draeger Perseus A500 ......................................................................................................... 3-13
3.3.1 Output SignalsParameters................................................................................. 3-13
3.3.2 Output SignalsAlarms......................................................................................... 3-15
3.3.3 Output Signals - Waveforms (For BeneVision N Series Monitors only) ......... 3-19
3.4 GE Aespire 7900&7100/Aestiva 7900&7100 ................................................................... 3-20
3.4.1 Output Signals Parameters ................................................................................. 3-20
3.4.2 Output Signals Alarms ......................................................................................... 3-21
3.4.3 Output Signals - Waveforms (For BeneVision N Series Monitors only) ......... 3-24
3.5 GE Aespire View (for BeneVision N Series Monitors only) ............................................ 3-24
3.5.1 Output Signals Parameters ................................................................................. 3-24
3.5.2 Output Signals - Alarms .......................................................................................... 3-26
3.5.3 Output Signals Waveforms ................................................................................. 3-26
3.5.4 Output Signals Respiratory Loops..................................................................... 3-27
3.6 GE Aisys/ Aisys CS2/Avance ................................................................................................ 3-27
3.6.1 Output Signals Parameters ................................................................................. 3-27
3.6.2 Output Signals Alarms ......................................................................................... 3-29
3.6.3 Output Signals - Waveforms (For BeneVision N Series Monitors only) ......... 3-31
3.7 MAQUET FLOW-i ................................................................................................................... 3-31
3.7.1 Output Signals Parameters ................................................................................. 3-31
3.7.2 Output Signals Alarms ......................................................................................... 3-34
3.8 Mindray A7/A5/A4/A3 ......................................................................................................... 3-36
3.8.1 Output Signals Parameters ................................................................................. 3-36
3.8.2 Output Signals Alarms ......................................................................................... 3-39
3.8.3 Output Signals - Waveforms (For BeneVision N Series Monitors only) ......... 3-46
4 Integrating the Ventilator ........................................................................................ 4-1
4.1 Carefusion Avea (For BeneVision N Series Patient Monitor Only) .................................4-1
4.1.1 Output Signals - Parameters .....................................................................................4-1
BeneLink Module Operators Manual 3
4.1.2 Output Signals - Alarms ............................................................................................ 4-4
4.1.3 Output Signals Waveforms .................................................................................... 4-6
4.2 Carefusion Vela ........................................................................................................................ 4-6
4.2.1 Output Signals Parameters.................................................................................... 4-6
4.2.2 Output Signals Alarms ........................................................................................... 4-8
4.3 Draeger Babylog 8000 plus/Babylog 8000 ......................................................................4-10
4.3.1 Output SignalsParameters .................................................................................4-10
4.3.2 Output SignalsTechnical Alarms .......................................................................4-11
4.4 Draeger Babylog VN500 ......................................................................................................4-11
4.4.1 Output SignalsParameters .................................................................................4-11
4.4.2 Output Signals Alarms .........................................................................................4-13
4.5 Draeger Evita 2 ......................................................................................................................4-16
4.5.1 Output Signals Parameters..................................................................................4-16
4.5.2 Output Signals Alarms .........................................................................................4-18
4.6 Draeger Evita2 dura/Evita 4/Evita XL ................................................................................4-19
supports integration with ..............................................................................................4-19
4.6.1 Output Signals Parameters..................................................................................4-19
4.6.2 Output Signals Alarms .........................................................................................4-22
4.6.3 Output Signals - Waveforms (For BeneVision N Series Monitors only) ..........4-24
4.7 Draeger Infinity V500 ...........................................................................................................4-25
4.7.1 Output SignalsParameters .................................................................................4-25
4.7.2 Output SignalsAlarms .........................................................................................4-27
4.8 Draeger Savina 300 ...............................................................................................................4-30
4.8.1 Output SignalsParameters .................................................................................4-30
4.8.2 Output SignalsAlarms .........................................................................................4-32
4.9 GE Carescape R860 ...............................................................................................................4-34
4.9.1 Output SignalsParameters .................................................................................4-34
4.9.2 Output Signals WaveForms (For BeneVision N Series Patient Monitor Only)4-37
4.9.3 Output Signals Respiratory Loops (For BeneVision N Series Patient
Monitor Only)
......................................................................................................................4-37
4.9.4 Output SignalsAlarms .........................................................................................4-37
4 BeneLink Module Operators Manual
4.10 GE Engström Carestation.................................................................................................. 4-42
4.10.1 Output Signals – Parameters ............................................................................... 4-42
4.10.2 Output Signals Alarms....................................................................................... 4-45
4.11 Hamilton C1/C2/C3/T1 ...................................................................................................... 4-49
4.11.1 Output Signals Parameters ............................................................................... 4-49
4.11.2 Output Signals Alarms....................................................................................... 4-51
4.12 Hamilton G5 (Block Protocol) .......................................................................................... 4-52
4.12.1 Output Signals Parameters ............................................................................... 4-52
4.12.2 Output Signals Alarms....................................................................................... 4-54
4.12.3 Output Signals - Waveforms (For BeneVision N Series Monitors only) ....... 4-59
4.13 Hamilton G5 (Polling Protocol) ........................................................................................ 4-59
4.13.1 Output Signals Parameters ............................................................................... 4-59
4.13.2 Output Signals Alarms....................................................................................... 4-61
4.14 Hamilton Galileo (Polling Protocol) ................................................................................ 4-62
4.14.1 Output Signals Parameters ............................................................................... 4-62
4.14.2 Output Signals Alarms....................................................................................... 4-64
4.15 MAQUET SERVO-i/SERVO-s .............................................................................................. 4-65
4.15.1 Output Signals – Parameters ............................................................................... 4-65
4.15.2 Output Signals Alarms....................................................................................... 4-68
4.15.3 Output Signals Waveforms (For BeneVision N Series Only) ....................... 4-70
4.15.4 Output Signals - Loops (For BeneVision N Series Only) ................................. 4-70
4.16 MAQUET SERVO-U/SERVO-N ........................................................................................... 4-71
4.16.1 Output SignalsParameters .............................................................................. 4-71
4.16.2 Output SignalsAlarms ...................................................................................... 4-73
4.16.3 Output SignalsWaveforms (For BeneVision N Series Monitor Only) ....... 4-75
4.16.4 Output SignalsRespiratory Loops (For BeneVision N Series Monitor Only)
............................................................................................................................................... 4-75
4.17 Newport E360 ..................................................................................................................... 4-76
4.17.1 Output Signals Parameters ............................................................................... 4-76
4.17.2 Output Signals Alarms....................................................................................... 4-77
4.18 Philips Respironics V60 (SNDA Protocol) ....................................................................... 4-80
BeneLink Module Operators Manual 5
4.18.1 Output Signals Parameters ...............................................................................4-80
4.18.2 Output Signals Alarms .......................................................................................4-81
4.19 Philips Respironics V60 (VRPT Protocol) .........................................................................4-82
4.19.1 Output Signals Parameters ...............................................................................4-82
4.19.2 Output Signals Alarms .......................................................................................4-82
4.20 Puritan Bennett PB840/PB980 (SNDF Protocol)............................................................4-84
4.20.1 Output Signals Parameters ...............................................................................4-84
4.20.2 Output Signals Alarms .......................................................................................4-86
4.21 Puritan Bennett PB840/PB980 (SNDA Protocol) ...........................................................4-87
4.21.1 Output Signals Parameters ...............................................................................4-87
4.21.2 Output Signals Alarms .......................................................................................4-89
4.22 ResMed VSIII (for Passport M series only).......................................................................4-90
4.22.1 Output SignalsParameters ...............................................................................4-90
4.22.2 Output SignalsAlarms .......................................................................................4-90
5 Integrating the tcGas Monitoring Device ................................................................. 5-1
5.1 Biometer TCM CombiM/TCM Tosca/TCM40 ....................................................................... 5-1
5.1.1 Output Signals Parameters.................................................................................... 5-1
5.1.2 Output Signals Alarms ........................................................................................... 5-2
5.2 Biometer TCM4 (For BeneVision N Series Monitors only) ............................................... 5-2
5.2.1 Output Signals Parameters.................................................................................... 5-2
5.2.2 Output Signals Alarms ........................................................................................... 5-3
5.3 SenTec Digital Monitor ........................................................................................................... 5-3
5.3.1 Output Signals Parameters.................................................................................... 5-3
5.3.2 Output Signals Alarms ........................................................................................... 5-4
6 Integrating Infusion System (For BeneVision N Series Monitors only) ...................... 6-1
6.1 B.Braun Perfusor Space Pumps ............................................................................................ 6-1
6.1.1 Output Signals Parameters.................................................................................... 6-1
6.2 Fresenius Agilia Injectomat Infusion Pumps ..................................................................... 6-1
6.2.1 Output Signals - Parameters .................................................................................... 6-1
6 BeneLink Module Operators Manual
6.2.2 Output Signals Technical Alarms ..........................................................................6-2
6.3 Fresenius Agilia Volumat Infusion Pumps ..........................................................................6-3
6.3.1 Output Signals - Parameters .....................................................................................6-3
6.3.2 Output Signals Technical Alarms ..........................................................................6-3
7 Integrating Other Devices........................................................................................ 7-1
7.1 Masimo Radical 7 Pulse CO-Oximeter (For BeneVision N Series Monitors only) ........7-1
7.1.1 Output Signals Parameters ....................................................................................7-1
7.1.2 Output Signals Alarms ............................................................................................7-2
7.2 Organon TOF-Watch SX NMT Monitoring Device .............................................................7-4
7.2.1 Output Signals Parameters ....................................................................................7-4
7.2.2 Output Signals - Alarms .............................................................................................7-4
8 Connecting External Devices via DIAP Protocol (For BeneVision N Series Monitors
only)
........................................................................................................................... 8-1
8.1 Setting Parameter Output Properties .................................................................................8-1
8.2 Setting the ID of the ID Adapter ...........................................................................................8-1
8.3 Parsing the Monitor Data .......................................................................................................8-1
9 BeneLink Function Test ............................................................................................ 9-1
9.1 Preparing the Tools .................................................................................................................9-1
9.2 Checking the Function ...........................................................................................................9-1
9.3 Setting the Serial Ports ...........................................................................................................9-3
9.3.1 Setting the Serial Ports of the Anesthesia Machines ...........................................9-3
9.3.2 Setting the Serial Ports of the Ventilator ................................................................9-4
9.3.3 Setting the Serial Ports of the TcGas Monitoring Devices ..................................9-7
9.3.4 Setting the Serial Ports of the Pumps .....................................................................9-8
9.3.5 Setting the Serial Ports of Other External Devices ...............................................9-9
9.4 Troubleshooting Device Integration Failures ................................................................. 9-10
9.5 BeneLink Installation and Test Report .............................................................................. 9-11
10 Alarm ID of External Devices (For BeneVision N Series Monitors only) ................. 10-1
BeneLink Module Operators Manual 7
10.1 Alarm IDs of Anesthesia Machines ..................................................................................10-1
10.2 Alarm IDs of Ventilators .....................................................................................................10-5
10.3 Alarm IDs of the Pulse CO-Oximeter ...............................................................................10-8
10.4 Alarm IDs of tcGas monitoring devices ....................................................................... 10-13
10.5 Alarm IDs of NMT monitoring device .......................................................................... 10-13
10.6 Alarm IDs of Infusion Systems ....................................................................................... 10-14
FOR YOUR NOTES
BeneLink Module Operators Manual 1-1
1 Safety
1.1 Safety Information
WAR
NING
Indicates a potential hazard or unsafe practice that, if not avoided,
could result in death or serious injury.
CA
UTION
Indicates a potential hazard or unsafe practice that, if not avoided,
could result in minor personal injury or product/property damage.
NOT
E
Provides application tips or other useful information to ensure that you
get the most from your product.
1-2 BeneLink Module Operators Manual
1.2 Warnings
WARNING
The BeneLink module receives parameter, alarm, and waveform data
from external devices but will not modify this data. For example, the
external device may display "---", b u t send "0" to the patient monitor.
There can be differences between the alarm priorities displayed on
your patient monitors and the priorities displayed on external devices
interfaced through BeneLink. For details please
see "Output signals
Alarms" section corresponding with each external device.
1.3 Cautions
CAUTION
First installation and debugging should be executed by Mindray
service personnel or authorized technician.
Please check the compatibility of the external device and the ID
adapter before connection. Not doing so may may result in failure of
the two devices to work together as desired.
Ports on the BeneLink module are not conventional network
connectors. They are intended for connecting with the serial port of
designated devices only. Do not connect them to public network
interfaces.
The signal labels used on the patient monitor may be different from those
given on the external device. For details please see "Output signals
Parameters" section corresponding with each external device. The alarms
from external devices may be delayed before transmission to the patient
monitor
.
BeneLink Module Operators Manual 1-3
1.4 Notes
NOTE
Devices of the same category cannot be connected to the BeneLink
module simultaneously.
The BeneVision N Series monitor supports up to two BeneLink
modules, while the Passport M Series supports only one.
The alarm messages from external devices are derived from the open
protocol of corresponding external device. For more information about
these alarms, please see the operator’s manual of the corresponding
devices.
For the intended use and contraindication of the external devices, refer
to their operator’s manuals.
When displayed in the monitor main screen, parameter labels of
external devices are prefixed with the plus sign “+”. For example, if
SpO2 is from an external device, its label is displayed as +SpO2”, and
its waveform label is displayed as “+Pleth”.
A plus sign “+” is added before each alarm message from external
devices.
1-4 BeneLink Module Operators Manual
FO
R YOUR NOTES
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227

Mindray BeneLink Module User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI