OPERATING INSTRUCTIONS
When turning on the device, the microprocessor, after a quick functional self test, automatically switches the ventilator to the default "controlled" mode and checks the
acoustic alarm inhibiting it for about 30 seconds, detailing any information on the front panel in the dedicated alarm section. Erogation of medical gas to the patient starts
in cycles at regular intervals, in correspondance with the relevant frequency selected by the command. The current volume of mix sent to each breathing act is detected in
the following way: read the set value on the "volume/minute" button and divide it by the set frequency.
At any time the operator can see the situation by observing the luminous writings on the front panel of the device. Particularly the functioning modes are visualised, the
breathing phases, spontaneous inspiration of the patient and any alarm situations. The real pressure of the patient circuit is always monitored by the luminous manovacu-
umeter on the front panel.
For each breathing act, the electronic controlling device sets off a luminous display that visualises, in sync with the real time passing, the various phases of the breathing
act. If the functioning modes set by the electronic controlling device are not adequate for the patient in question when the machine is turned on, then the operator can select
a different mode, using the select buttons on the front panel.
The different functioning modes are described in the next paragraph.
For the duration of functioning, the microprocessor regulates the cyclical erogation of medical gas and verifies the state of some parameters of breathing, moreover it sig-
nals the exceeding of safety limits activating, depending on the case, the acoustic and/or luminous signals.
The device detects the apnoea situation only if "assisted" or "assisted/controlled" mode is selected. Precisely, if in the "assisted/controlled" mode the patient does not show
any spontaneous breathing functions within the lapse of time between the theoretical term of the breathing act, it activates the luminous/acoustic signal of apnoea and com-
mutes automatically to "controlled" (erogation mix at fixed frequency), tuning on the luminous signal of apnoea. Should the patient start to breathe naturally the device will
automatically go back to "assisted/controlled" mode and cancel any alarm signals, which are still active.
5.2
8
It is absolutely necessary to regulate first the ventilator volume and then pressure, because a late volume regulation (lt/min) would bring to an immediate
change of the pressure limit. After choosing volume ventilation, keep closed with the palm of the hand the exit outlet of the patient valve (where the mask is
normally inserted), then regulate the pressure limit with the apposite knob.
Hyperextend the head and seek, after inserting an oral-pharynx cannula in the patient's mouth following the correct manoeuvre, in order to guarantee clear-
ness of the airways. Place the ventilation mask on the mouth and on the nose of the patient checking the adhesion of the soft part of the mask on the
patient's face for a good holding system. Selection method of the volume/minute knob (through trips) prevents from value changing during functioning, in
case it is accidentally knocked.
In case of intubation, take the ventilation mask off the patient valve, and insert in the free hose the endo-tracheal tube with its connector.
Resistance of the airways due to obstructions or external cardiac massaging does not cause a variation of the respiratory volume frequency. In case of reduction of the
compliance, the ventilator will react with a rise of respiratory pressure at a constant volume.
FUNCTIONING MODES
ASSISTED/CONTROLLED Breathing
Starting from halfway the exhaling phase, breathing extends for the duration of the whole exhaling phase. Any attempt of the patient to inhale sufficient amounts of prod-
uct in this interval determines the start of a new breathing act and the consequent erogation of medical gas flow.
Attempts to inspire effected outside this interval are not accepted: anticipated attempts because they could be caused by transitional situations in the patient circuit and
could simulate voluntary breathing; late attempts as they could not show, since the machine switches automatically to controlled mode if the maximum limiter of wait is
exceeded, thus inhibiting any attempt to inhale except for the final part of the exhaling phase.
In case of exceeded maximum time of wait, the APNOEA situation is signalled visually and acoustically.
The alarm remains active until acceptance on the part of the respirator of an attempted inhalation valid as from the patient or until the alarm cancelling button "RESET
ALARMS" is pressed.
Any time the respirator considers as valid the inhalation attempt of the patient, it is signalled by means of the luminous display "SPONTANEOUS" underneath the graph-
ics of the inhaling phase.
This functioning mode guarantees prompt intervention of the machine in case spontaneous breathing of the patient stops, and allows for an easy adaptation of the
patient in that the breathing frequency can vary up to 50% increasing or decreasing.
NOTE
It is important though that the breathing frequency selected is the closest to the natural breathing of the patient.
In order to make a precise selection, besides the physiological data and direct observation of the patient, it is useful to use the shape of the wave generated every time
the patient breathes in on the functions monitor.
The selection can be considered optimal when it will be able to illuminate completely by means of the wave shape and during a single breathing act, without setting of
the apnoea condition. A wave shape which cannot complete its cycle when turning on the machine means that the frequency value selected is too low. Consequently,
the patient is discomforted and the I/E ratio is altered significantly. In fact, the ratio is calculated at I/E=1/2 and the volume/minute of mix erogated increases proportional-
ly.
Should the apnoea alarm go off, it means that the value of frequency selected is too high and that the patient cannot keep up with the rhythm of the machine.
CONTROLLED Breathing (IPPV)
This functioning mode is selected automatically by the device when it is turned on. It consists in supplying to the patient packets of medical mix regularly at intervals of
standard duration as per the frequency selection.
When operating in this mode, the relationship between the duration of the inhalation phase and the duration of the exhalation phase is constantly 1 to 2.
ASSISTED Breathing (IPPV with TRIGGER)
This functioning mode can be selected by activating the "assisted" button.
This function ensures that flowing volumes are of fixed duration and they are erogated to the patient. These volumes can be set by means of the frequency knob and of
the volume/minute intervals, following the patient's natural breathing rhythm.
The aperture of the inhalation valve is regulated by pressure; it is necessary that the patient creates depression for the breathing flow erogation to start.
The inhaling and exhaling ratio depends on the characteristics of the patient's natural breathing, however can be corrected using the frequency knob, which determines
the duration of the inhaling phase only, in this circumstance.
The duration of the inhaling phase is calculated dividing by 3 the duration of the whole breathing act.
5.3
E.g.: frequency F = 10 acts/min.
inhalation INS = 60 / 3 * F = 2 sec.
F=5 INS=60/3*5=4 sec.
F=20 INS=60/3*20=1 sec.
TRIGGER
The trigger is a special device able to synchronise the insufflation with the beginning of spontaneous inhalation of the patient. The ventilation supporting technique is
normally employed on patients who are able to sustain inhalation work at brief intervals.
The patient's effort to produce spontaneous breathing generates inside the respirator a negative pressure which triggers rapidly the insufflation, thank to the trigger
mechanism.
The effort (or negative pressure) necessary to set off the trigger mechanism can be regulated and increased progressively in order to aid and prompt the patient towards
breathing spontaneously, re-educating and training the breathing muscles.
The range of regulation goes from 0 mbar to -12 mbar.