Thermo Fisher Scientific Orbitrap Tribrid Series Owner's manual

Type
Owner's manual
Orbitrap Tribrid Series
Hardware Manual
80000-97027 Revision A June 2018
© 2018 Thermo Fisher Scientific Inc. All rights reserved.
Automatic Gain Control, EASY-ETD, EASY-IC, EASY-Max NG, Fusion, ID-X, Lumos, OptaMax NG, and
Orbitrap ID-X are trademarks; Unity is a registered service mark; and Hypersil GOLD AQ, Orbitrap, Orbitrap
Fusion, Orbitrap Fusion Lumos, Pierce, Thermo Scientific, Tribrid, and Xcalibur are registered trademarks of
Thermo Fisher Scientific Inc. in the United States. Fisher Scientific is a registered trademark of Fisher Scientific
Co. in the United States.
The following are registered trademarks in the United States and other countries: COMBICON is a registered
trademark of Phoenix Contact GmbH & Co. Microsoft and Windows are registered trademarks of Microsoft
Corporation. Teflon is a registered trademark of E.I. du Pont de Nemours & Co.
The following are registered trademarks in the United States and possibly other countries: Liquinox is a
registered trademark of Alconox, Inc. MICRO-MESH is a registered trademark of Micro-Surface Finishing
Products, Inc. Nalgene is a registered trademark of Nalge Nunc International Corporation. Oerlikon Leybold
Vacuum is a registered trademark of OC Oerlikon Corporation AG. Rheodyne is a registered trademark of
IDEX Health & Science, LLC. SOGEVAC is a registered trademark of Oerlikon Leybold Vacuum. Swagelok is
a registered trademark of Swagelok Company. Tygon is a registered trademark of the division of Saint-Gobain
Performance Plastics Corporation. Upchurch Scientific is a registered trademark of IDEX Health & Science
LLC. Vespel is a registered trademark of E.I. du Pont de Nemours & Co. Viton is a registered trademark of
DuPont Performance Elastomers LLC.
Chemyx is a trademark of Chemyx Inc. MX Series II is a trademark of IDEX Health & Science, LLC.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Rev A, June 2018
Software version: (Thermo) Foundation 3.1 SP5 and later, Xcalibur 4.2 and later, Tune 3.1 and later
For Research Use Only. Not for use in diagnostic procedures.
Regulatory Compliance
Thermo Fisher Scientific performs complete testing and evaluation of its products to ensure full compliance with
applicable North American and European regulations. Your system meets the applicable requirements in the
electromagnetic compatibility (EMC) and product safety standards described in this section.
Unauthorized changes that you make to your system will void regulatory compliance and may defeat the built-in
protections for your instrument. Some examples of unauthorized changes include using replacement parts or adding
components, options, or peripherals that Thermo Fisher Scientific has not qualified and authorized. Unauthorized
changes can also result in bodily injury and/or damage to your system and laboratory.
Ensure continued compliance with regulatory standards:
Follow all installation instructions provided in the documentation that comes with your system.
Order replacement parts (as specified in the instrument manual) and additional components, options, and periph-
erals directly from Thermo Fisher Scientific or an authorized representative.
Regulatory compliance results for the following Thermo Scientific™ mass spectrometers:
Orbitrap Fusion Lumos
Orbitrap Fusion and Orbitrap ID-X
Orbitrap Fusion Lumos
Low Voltage Directive 2014/35/EU
This device complies with Low Voltage Directive 2014/35/EU and the harmonized safety standard IEC/EN/CSA/
UL 61010-1, 3rd Edition.
EMC Directive 2014/30/EU and other EMC test standards
This device was tested by TÜV Rheinland of North America and complies with the following EMC standards:
47 CFR 15, Subpart B, Class A: 2015 EN 61000-3-2: 2006 + A1 + A2 EN 61000-4-5: 2006
CISPR 11: 2009 + A1 EN 61000-3-3: 2008 EN 61000-4-6: 2009
ICES-003: 2014 EN 61000-4-2: 2009 EN 61000-4-8: 2010
EN 55011: 2009 + A1 EN 61000-4-3: 2006 + A1 + A2 EN 61000-4-11: 2004
EN 61326-1: 2013 EN 61000-4-4: 2004 + A1
Orbitrap Fusion and Orbitrap ID-X
FCC Compliance Statement
Notice on the Proper Use of
Thermo Scientific Instruments
In compliance with international regulations: This instrument must be used in the manner specified by Thermo Fisher
Scientific to ensure protections provided by the instrument are not impaired. Deviations from specified instructions on
the proper use of the instrument include changes to the system and part replacement. Accordingly, order replacement
parts from Thermo Fisher Scientific or one of its authorized representatives.
Low Voltage Directive 2014/35/EU
This device complies with Low Voltage Directive 2014/35/EU and the harmonized safety standard IEC/EN/CSA/
UL 61010-1, 3rd Edition.
EMC Directive 2014/30/EU and other EMC test standards
This device was tested by TÜV Rheinland of North America and complies with the following EMC standards:
47 CFR 15, Subpart B, Class A: 2012 EN 61326-1: 2013 EN 61000-4-4: 2004 + A1
CISPR 11: 2009 + A1 EN 61000-3-2: 2006 + A1 + A2 EN 61000-4-5: 2006
AS/NZS CISPR 22: 2009 + A1 EN 61000-3-3: 2008 EN 61000-4-6: 2009
ICES-003: 2012 EN 61000-4-2: 2009 EN 61000-4-8: 2010
EN 55011: 2009 + A1 EN 61000-4-3: 2006 + A1 + A2 EN 61000-4-11: 2004
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO THE
FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL INTERFERENCE,
AND (2) THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED, INCLUDING
INTERFERENCE THAT MAY CAUSE UNDESIRED OPERATION.
CAUTION Read and understand the various precautionary notes, signs, and symbols contained inside
this manual pertaining to the safe use and operation of this product before using the device.
WEEE Directive
2012/19/EU
Thermo Fisher Scientific is registered with B2B Compliance (B2Bcompliance.org.uk) in the UK and with the
European Recycling Platform (ERP-recycling.org) in all other countries of the European Union and in Norway.
If this product is located in Europe and you want to participate in the Thermo Fisher Scientific Business-to-Business
(B2B) Recycling Program, send an email request to weee.recycle@thermofisher.com with the following information:
WEEE product class
Name of the manufacturer or distributor (where you purchased the product)
Number of product pieces, and the estimated total weight and volume
Pick-up address and contact person (include contact information)
Appropriate pick-up time
Declaration of decontamination, stating that all hazardous fluids or material have been removed from the product
For additional information about the Restriction on Hazardous Substances (RoHS) Directive for the European Union,
search for RoHS on the Thermo Fisher Scientific European language websites.
IMPORTANT This recycling program is not for biological hazard products or for products that have been medically
contaminated. You must treat these types of products as biohazard waste and dispose of them in accordance with
your local regulations.
Thermo Scientific Orbitrap Tribrid Series Hardware Manual vii
C
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiii
Accessing Documentation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiii
Viewing the Product Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiv
Accessing the Help Menu Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiv
Viewing Online User Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiv
Providing Documentation Feedback. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
License for the 1M Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Special Notices, Symbols, and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Model Differences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xvii
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xviii
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Orbitrap Tribrid Series Mass Spectrometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Overview of an LC/MS Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
LC/MS Functional Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Electronic Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Controls and Indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
LEDs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Power Entry Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Communications Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Cooling Fans . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Chapter 2 Scan Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Scan Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
MS Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
MS2 Scan and MSn Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
SIM Scan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Scan Rates for the Ion Trap Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Scan Mass-To-Charge Ratio Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Resolutions for the Orbitrap Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Data Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Ion Polarity Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Contents
Contents
viii Orbitrap Tribrid Series Hardware Manual Thermo Scientific
Chapter 3 Vacuum System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Vacuum System Functional Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Schematic of the Internal Gas Supply Lines. . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Inlet Gases Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Helium Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Nitrogen Gas Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Vent Valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Vacuum Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Vacuum Gauges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Vacuum Pumps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Atmospheric Pressure Ionization Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
API Source Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Chapter 4 Ion Transmission and Mass Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Workflow for Mass Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Ion Optics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
MP00 Ion Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
MP0 Ion Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
MP1 Ion Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Curved Linear Trap. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Ion-Routing Multipole . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
MP3 Ion Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
DC Offset Voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Mass Analyzers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Quadrupole Mass Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Orbitrap Mass Analyzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Linear Ion Trap Mass Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Ion Detection Systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Chapter 5 Syringe Pump and Divert/Inject Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Syringe Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Divert/Inject Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Controlling the Divert/Inject Valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Chapter 6 System Shutdown, Startup, and Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Shutting Down the System in an Emergency. . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Placing the Mass Spectrometer in Standby Mode . . . . . . . . . . . . . . . . . . . . . . . 48
Turning On the Mass Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Shutting Down the Mass Spectrometer System Completely . . . . . . . . . . . . . . . 48
Starting the System after a Complete Shutdown . . . . . . . . . . . . . . . . . . . . . . . . 50
Starting the LC System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Starting the Data System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Starting the Mass Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Starting the Autosampler. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Contents
Thermo Scientific Orbitrap Tribrid Series Hardware Manual ix
Resetting the Mass Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Resetting Calibration Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Restarting the Data System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
On/Off Status for MS Components Under Varying Power Conditions. . . . . . . 54
Chapter 7 Daily Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Before Operating the Orbitrap Tribrid Series System . . . . . . . . . . . . . . . . . . . . 58
Checking the System Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Checking the Vacuum Pressure Levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Checking the Gas Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
After Operating the Orbitrap Tribrid Series System . . . . . . . . . . . . . . . . . . . . . 59
Flushing the Inlet Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Purging the Oil in the Forepump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Emptying the Solvent Waste Container. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Placing the System in Standby Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Chapter 8 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Maintenance Schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Tools and Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Maintaining the API Source Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Maintaining the API Source Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Cleaning the Ion Sweep Cone, Spray Cone, and Ion Transfer Tube . . . . . . . 68
Removing the API Source Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Cleaning the RF Lens, Exit Lens, MP00 RF Lens, and Lens L0. . . . . . . . . . . 74
Reinstalling the API Source Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Maintaining the Forepump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Maintaining the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Pumping Down the Mass Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Chapter 9 Replaceable Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Orbitrap Tribrid Series Chemicals Kit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Calibration Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
MS Setup Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Performance Specification Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Single Mechanical Pump Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Dual Mechanical Pumps Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
API Source Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Miscellaneous Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95
Contents
xOrbitrap Tribrid Series Hardware Manual Thermo Scientific
Thermo Scientific Orbitrap Tribrid Series Hardware Manual xi
F
Figure 1. Functional block diagram of the Orbitrap Tribrid Series system . . . . . . . . . . . . . 5
Figure 2. Orbitrap Tribrid Series front panel LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Figure 3. Power entry module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Figure 4. Communication connectors (left side of the MS) . . . . . . . . . . . . . . . . . . . . . . . . . 8
Figure 5. Functional block diagram of the vacuum system . . . . . . . . . . . . . . . . . . . . . . . . 16
Figure 6. Schematic of the internal gas supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Figure 7. Gas inlet ports (back of the MS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Figure 8. Placement of the three turbomolecular pumps . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Figure 9. Thermo Scientific API source (H-ESI mode) . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Figure 10. API source interface cross section (Orbitrap Fusion and Orbitrap ID-X
MSs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Figure 11. Exit lenses and RF lenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Figure 12. Workflow for mass analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Figure 13. Ion transmission path for the Orbitrap Fusion Lumos Tribrid MS (with
the ETD source and UVPD options) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Figure 14. MP00 RF lens (left, Orbitrap Fusion Lumos MS) and lens L0 (right) . . . . . . . . 30
Figure 15. Multipole MP0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Figure 16. TK lens (both sides) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Figure 17. Multipole MP1 (Orbitrap Fusion Lumos MS) . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Figure 18. C-trap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Figure 19. Ion-routing multipole (IRM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Figure 20. Multipole MP3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Figure 21. Quadrupole Q1, side and end views (Orbitrap Fusion and Orbitrap ID-X
MSs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Figure 22. Dual-split gate lens (both sides) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Figure 23. Polarity of the RF and DC voltages applied to the Q1 rods . . . . . . . . . . . . . . . . 35
Figure 24. Schematic of the Orbitrap cell and an example stable ion trajectory . . . . . . . . . . 36
Figure 25. Electrodynamic squeezing of ions in the Orbitrap analyzer (r) versus the
increased field strength (z) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Figure 26. Approximate shape of ion packets of different m/z values after the
voltages stabilize . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Figure 27. Linear ion trap (LIT) (Orbitrap Fusion and Orbitrap ID-X MSs) . . . . . . . . . . . 39
Figure 28. Assembly for the linear ion trap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Figure 29. Low pressure cell of the LIT mass analyzer in operation (representation) . . . . . . 42
Figure 30. Syringe pump setup (top view) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Figure 31. Divert/inject valve positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Figure 32. Divert/inject valve plumbed as a loop injector and as a divert valve . . . . . . . . . . 46
Figures
Figures
xii Orbitrap Tribrid Series Hardware Manual Thermo Scientific
Figure 33. Modular divert/inject valve (front view) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Figure 34. Ion sweep cone removed from the MS mount assembly . . . . . . . . . . . . . . . . . . 69
Figure 35. Ion transfer tube removal tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Figure 36. Ion transfer tube removal tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Figure 37. Spray cone, O-ring, ion transfer tube, and ion sweep cone . . . . . . . . . . . . . . . . 71
Figure 38. API source interface removed from the vacuum manifold (example). . . . . . . . . 73
Figure 39. Lens L0 removed from the back of the API source interface (example) . . . . . . . 74
Figure 40. MP00 RF lens assembly removed from the API source interface (example). . . . 75
Figure 41. RF lens removed from the API source interface cage (example). . . . . . . . . . . . . 75
Figure 42. Exit lens removed from the API source interface (example). . . . . . . . . . . . . . . . 76
Figure 43. Alignment slot and thumbscrew locations on the MP00 RF lens
(example) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Figure 44. Power entry module (right side of the MS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Thermo Scientific Orbitrap Tribrid Series Hardware Manual xiii
P
Preface
The Orbitrap Tribrid Series Hardware Manual describes the modes of operation and hardware
components for the Thermo Scientific OrbitrapTribrid™ Series mass spectrometer (MS)
system. It also provides the instruments cleaning and maintenance procedures.
This guide is intended for the following Thermo Scientific MSs:
Orbitrap Fusion™ (also known as Fusion™)
Orbitrap Fusion Lumos™ (also known as Lumos™)
Orbitrap ID-X™ (also known as ID-X™)
Contents
Accessing Documentation
Providing Documentation Feedback
License for the 1M Option
Special Notices, Symbols, and Cautions
Model Differences
•Contacting Us
Accessing Documentation
The Orbitrap Tribrid Series MS includes complete documentation.
Viewing the Product Manuals
Accessing the Help Menu Options
Viewing Online User Documentation
For system requirements, refer to the release notes on the software DVD.
Preface
xiv Orbitrap Tribrid Series Hardware Manual Thermo Scientific
Viewing the Product Manuals
The Thermo Fisher Scientific service engineer installs the instrument control applications and
the instrument manuals on the data system computer.
To view the product manuals
From the Microsoft™ Windows™ taskbar, choose Start > All Apps (Windows 10) or
All Programs (Windows 7) > Thermo Instruments > model x.x, and then open the
applicable PDF file.
Accessing the Help Menu Options
Follow this procedure to view the Help systems for the instrument-control applications.
To view the Help
Do the following as applicable:
Thermo Tune instrument-control application: Click the Options icon, , and
choose Tu ne He l p.
Thermo Xcalibur™ Method Editor application: Choose an option from the Help
menu (or press the F1 key).
Viewing Online User Documentation
Visit the Thermo Fisher Scientific website for product manuals and more.
To view user documentation from the Thermo Fisher Scientific website
1. Go to thermofisher.com.
2. Point to Services & Support and click Manuals on the left.
3. In the Refine Your Search box, search by the product name.
4. From the results list, click the title to open the document in your web browser, save it, or
print it.
To return to the document list, click the browser Back button.
Preface
Thermo Scientific Orbitrap Tribrid Series Hardware Manual xv
Providing Documentation Feedback
To suggest changes to the documentation or to the Help
Complete a brief survey about this document by clicking the button below.
Thank you in advance for your help.
License for the 1M Option
During the Orbitrap Fusion Lumos MS installation or upgrade process, the Thermo Fisher
Scientific field service engineer activates the optional 1M license (with purchase of the 1M
option).
Special Notices, Symbols, and Cautions
Make sure you understand the special notices, symbols, and caution labels in this guide. Most
of the special notices and cautions appear in boxes; those pertaining to safety also have
corresponding symbols. Some symbols are also marked on the instrument itself and can
appear in color or in black and white. For complete definitions, see Table 1.
Table 1. Notices, symbols, labels, and their meanings (Sheet 1 of 2)
Notice, symbol, or label Meaning
IMPORTANT Highlights information necessary to prevent damage to
software, loss of data, or invalid test results; or might contain
information that is critical for optimal performance of the
product.
Note Highlights information of general interest.
Tip Highlights helpful information that can make a task easier.
Caution: Read the cautionary information associated with this
task.
Chemical hazard: Observe safe laboratory practices and
procedures when handling chemicals. Only work with volatile
chemicals under a fume or exhaust hood. Wear gloves and other
protective equipment, as appropriate, when handling toxic,
carcinogenic, mutagenic, corrosive, or irritant chemicals. Use
approved containers and proper procedures to dispose of waste
oil and when handling wetted parts of the instrument.
Preface
xvi Orbitrap Tribrid Series Hardware Manual Thermo Scientific
Read and understand the following cautions that are specific to the shutdown of the mass
spectrometry system or to the removal of parts for cleaning.
Heavy object: The Orbitrap Tribrid Series MS, excluding its
workbench, weighs over 227 kg (500 lb). Never try to detach
and move the instrument from its workbench; you can suffer
personal injury or damage the instrument.
Hot surface: Before touching the API source assembly, allow
heated components to cool.
Pinch point: Keep hands away from the specified areas.
Risk of electric shock: This instrument uses voltages that can
cause electric shock and personal injury. Before servicing the
instrument, shut it down and disconnect it from line power.
While operating the instrument, keep covers on.
Risk of eye injury: Eye injury can occur from splattered
chemicals, airborne particles, or sharp objects. Wear safety
glasses when handling chemicals or servicing the instrument.
Risk of laser radiation—Orbitrap Fusion Lumos MS with the
optional ultraviolet photodissociation (UVPD) module only:
Failure to understand and comply with laser cautions and
operating instructions in the UVPD Module User Guide can
result in property damage or serious injuries to the user.
Sharp object: Avoid handling the tip of the syringe needle.
Trip obstacle: Be aware of cords, hoses, or other objects located
on the floor.
Table 1. Notices, symbols, labels, and their meanings (Sheet 2 of 2)
Notice, symbol, or label Meaning
CAUTION If you must turn off the MS in an emergency, turn off the main power switch
located on the right-side power panel. This switch turns off all power to the MS,
including the forepump, without harming components within the system. However, do
not use this method as part of the standard shutdown procedure. Instead, see Shutting
Down the Mass Spectrometer System Completely. To turn off the LC, autosampler, and
data system computer in an emergency, use their respective on/off switch or button.
Preface
Thermo Scientific Orbitrap Tribrid Series Hardware Manual xvii
Model Differences
This table lists the required number of forepumps and the available options for the
Orbitrap Tribrid Series MSs.
CAUTION To avoid an electrical shock, be sure to follow the instructions in Shutting
Down the Mass Spectrometer System Completely.
CAUTION Do not turn the instrument on if you suspect that it has incurred any kind of
electrical damage. Instead, disconnect the power cord and contact Thermo Fisher
Scientific technical support for a product evaluation. Do not attempt to use the
instrument until it has been evaluated. (Electrical damage might have occurred if the
system shows visible signs of damage, or has been transported under severe stress.)
CAUTION Do not disconnect the power cord at the MS while the other end is still
plugged into the electrical outlet.
CAUTION Do not place any objects (for example, the syringe pump or other containers
with liquids) on top of the instrument, unless instructed to in the documentation.
Leaking liquids might contact the electronic components and cause an electrical short
circuit.
CAUTION Hot surface. Allow heated components to cool to room temperature
(approximately 20 minutes) before servicing them.
Instrument Number of
forepumps
Available options
EASY-IC
ion source
EASY-ETD
ion source
Advanced
peak
determination
1M Orbitrap
resolution
UVPD
laser module
Orbitrap Fusion Lumos 2    
Orbitrap Fusion 1  
Orbitrap ID-X 1
Preface
xviii Orbitrap Tribrid Series Hardware Manual Thermo Scientific
Contacting Us
Contact Email Telephone QR Codea
aYou can use your smartphone to scan a QR Code, which opens your email application or browser.
U.S. Technical Support us.techsupport.analyze@thermofisher.com (U.S.) 1 (800) 532-4752
U.S. Customer Service
and Sales
us.customer-support.analyze@thermofisher.com (U.S.) 1 (800) 532-4752
Global Support To find global contact information or customize your request
1. Go to thermofisher.com.
2. Click Contact Us, select the country, and then select the type of support
you need.
3. At the prompt, type the product name.
4. Use the phone number or complete the online form.
To find product support, knowledge bases, and resources
Go to thermofisher.com/us/en/home/technical-resources.
To find product information
Go to thermofisher.com/us/en/home/brands/thermo-scientific.
Note To provide feedback for this document, go to surveymonkey.com/s/PQM6P62 or send an email message to
Technical Publications (techpubs-lcms@thermofisher.com).
Thermo Scientific Orbitrap Tribrid Series Hardware Manual 1
1
Introduction
The Orbitrap Tribrid Series MS is part of the Thermo Scientific family of MSs. The
Orbitrap Tribrid Series system consists of the MS, a syringe pump, a divert/inject valve, and
the Thermo Xcalibur data system.
Contents
Orbitrap Tribrid Series Mass Spectrometers
Overview of an LC/MS Analysis
LC/MS Functional Block Diagram
Electronic Assemblies
Controls and Indicators
Cooling Fans
Orbitrap Tribrid Series Mass Spectrometers
See these topics:
Orbitrap Fusion MS
Orbitrap Fusion Lumos MS
Orbitrap ID-X MS
For a summary of the significant differences, see Model Differences.
For descriptions of the various hardware components, see Chapter 3, “Vacuum System,
Chapter 4, “Ion Transmission and Mass Analysis,and Chapter 5, “Syringe Pump and
Divert/Inject Valve.
For an MS with the ETD option (see Model Differences), refer to the EASY-ETD and
EASY-IC Ion Sources User Guide. For an Orbitrap Fusion Lumos MS with the UVPD option,
refer to the UVPD Module User Guide.
Note The “Glossary defines some of the terms used in this manual.
1 Introduction
Orbitrap Tribrid Series Mass Spectrometers
2Orbitrap Tribrid Series Hardware Manual Thermo Scientific
Orbitrap Fusion MS
The Orbitrap Fusion is a mass spectrometer that contains three mass analyzers and includes
an external syringe pump, a divert/inject valve, and the Thermo Scientific EASY-Max NG™
API source. The instrument requires one forepump.
Orbitrap Fusion Lumos MS
The Orbitrap Fusion Lumos is a mass spectrometer that contains three mass analyzers and
includes an external syringe pump, a divert/inject valve, and the EASY-Max NG API source.
The instrument requires two forepumps.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118