Short guide for the usage of rack holders
4
Affected systems
Systems
Link to latest version of
user documentation
Applicable
software version
Change
description
cobas p 512
(SN 64xxxxxx)
cobas p 612
(SN 63xxxxxx)
Operator's Manual cobas
p 512 (64x)
User Guide cobas p 612
(63x)
2.0.6 or higher
2.0.6 or higher
Instructions for the
usage of rack
holders
Table 1: Affected systems
Roche recommends that you familiarize yourself with the new and/or revised content
provided in this short guide.
If you print these pages, Roche recommends printing them single sided. In this way,
you can easily insert the new and/or
revised content in the appropriate locations in
Safety information
Pinching hazard
Closing the drawer may cause pinching
.
Close the drawer slowly and carefully to avoid pinching your fingers between the
drawers.
Do not bypass security installations.
Incorrect tube position
Incorrect positioning of the sample tubes causes problems set
ting them down during
contents may spill.
Manually check if the sample tubes are completely inserted in the rack. They should
rest against the bottom of the rack.
Use only racks with undamaged spring inserts.
• Check the position and the fill level of all trays and tubes.