Roche cobas 8100 AliquoterModule User manual

Type
User manual
cobas® 8100 automated
workflow series
Operator’s Manual
Software Version 01-02
cobas® 8100 automated workflow series
Roche Diagnostics
2 Operator’s Manual · Version 1.0
Document information
Edition notice This Operator's Manual is for users of the cobas® 8100 automated workflow series.
The cobas® 8100 automated workflow series consists of a control unit PC and
workflow modules that combine to create an automated processing system.
Every effort has been made to ensure that all the information contained in this
document is correct at the time of printing. However, Roche Diagnostics GmbH
reserves the right to make any changes necessary without notice as part of ongoing
product development.
Intended use The cobas® 8100 automated workflow series is a collection of interlinked modules for
processing and transporting patient samples.
The system is designed to centrifuge patient samples, remove and insert sample tube
caps as required, apply barcode labels to secondary tubes, and prepare aliquots from
primary samples. It can sort samples for online and offline analysis, and temporarily
store samples. The system transports individual samples between modules and can
transport samples to connected analyzers using CLAS 2 components.
The cobas® 8100 automated workflow series is intended for use with analyzers that
perform tests in the areas of immunology, clinical chemistry, coagulation, urinalysis,
hematology, and serology.
It is essential that all users read this Operator's Manual thoroughly before using the
system.
Any user modification of the instrument will render the warranty or service
agreement null and void.
Copyright © 2013, Roche Diagnostics GmbH. All rights reserved.
Trademarks The following trademarks are acknowledged.
COBAS, COBAS C, COBAS INTEGRA, ELECSYS, and LIFE NEEDS ANSWERS are
trademarks of Roche.
All other trademarks are the property of their respective owners.
System approvals The cobas® 8100 automated workflow series meets the protection requirements
specified in IVD Directive 98/79/EC.
Compliance is demonstrated by the mark below.
Manual version Software version Revision date Changes
1.0 01-02 – –
Table 1 Revision history
Online help version Software version Revision date Changes
1.0 01-02 – –
Table 2 Revision history
The cobas® 8100 automated workflow series instrument
complies with IVD Directive 98/79/EC.
Roche Diagnostics
Operator’s Manual · Version 1.0 3
cobas® 8100 automated workflow series
Instrument approvals
Contact addresses
Manufacturer of cobas® 8100
automated workflow series
instrument
Authorized representative
Issued by Intertek for Canada and the US.
Hitachi High-Technologies Corporation
1-24-14 Nishi-Shinbashi
Minato-ku, Tokyo 105-8717, Japan
Roche Diagnostics GmbH
Sandhofer Strasse 116
D-68305 Mannheim, Germany
cobas® 8100 automated workflow series
Roche Diagnostics
4 Operator’s Manual · Version 1.0
Roche Diagnostics
Operator’s Manual · Version 1.0 5
cobas® 8100 automated workflow series
Table of contents
Document information 2
Contact addresses 3
Table of contents 5
About the cobas 8100 automated workflow
series 7
Where to find information 7
User information 7
Conventions used in this document 8
System Description
1 General safety information
Safety classifications 15
Safety precautions 16
Safety summary 18
Safety labels on the instrument 25
Disposal 28
2 Overview of system components
Overview 31
System configuration 35
General specifications 36
Instrument components 37
Sample and holder flow 44
Sample tubes 46
Sample and holder identification 47
Operating modes 50
Overview of alarms 51
3 Modules
Overview of sample stations 55
Input station 56
Aliquot station 71
Output station 80
RFM module 91
4 Overview of the control unit software
Introduction 97
The control unit PC 98
Overview of the control unit software 99
Overview of key windows 102
Online help 108
Operation
5 Safety information for operation
General information 119
Safety messages of the overall system 120
Safety messages for specific modules 122
6 Daily operation
Daily workflow 125
Prepare for system start-up 126
System start-up 128
Handling samples, sample trays, and tip trays 134
Loading samples 135
Monitoring the processing of samples 139
Monitoring and controlling the system 142
Searching for samples 144
Retrieving samples 146
Error samples 148
Unloading samples 152
Post-operation checks and inspections 156
Full system shutdown 157
System restart 160
Log off, shut down, and close
the control unit software 167
7 Extended operation
Overview of extended operation 171
Backing up and restoring parameters 172
Adding new users and modifying existing
users 173
Adding new tests and modifying existing
tests 174
Adding new processing groups and modifying
existing processing groups 175
Adding new sample types and modifying
existing sample types 176
Switching between ACU profiles 177
Setting alarm sounds and lights 178
8 Reports
Reports overview 181
Generating result reports 182
Generating sample summary reports 183
Generating process result reports 184
Generating sample count reports 185
Generating data log reports 186
Maintenance
9 Safety information for maintenance
General information 191
Safety messages of the overall system 192
Roche Diagnostics
6 Operator’s Manual · Version 1.0
cobas® 8100 automated workflow series
10 Maintenance actions
Overview 199
Maintenance schedule 202
Maintenance of the input station 205
Maintenance of the aliquot station 236
Maintenance of the output station 268
Maintenance of the RFM module 281
General maintenance actions 288
Troubleshooting
11 Troubleshooting
General information about troubleshooting 295
Troubleshooting alarms 296
Troubleshooting problems with the system 299
Troubleshooting problems with modules 309
12 Limited Operation
Overview of limited operation 351
Module masking 351
Masking processing groups and tests 356
Glossary
13 Glossary
Index
Index 365
Revisions
Roche Diagnostics
Operator’s Manual · Version 1.0 7
cobas® 8100 automated workflow series
About the cobas 8100 automated workflow series
The cobas 8100 automated workflow series is a fully automated system for sample
workflow management. The system provides a modular and flexible approach to the
processing of samples before and after online or offline analysis. The system is
optimized for high-throughput workloads, allowing a range of sample types to be
processed simultaneously and automatically.
The cobas 8100 automated workflow series allows a range of configurable
connections to online analyzers, including the cobas®
6000 analyzer series, the
cobas®
8000 modular analyzer series, the STAGO STA-R Evolution®, and post-
analytical units including the cobas®
p 501 and cobas® p 701.
Where to find information
The following documents can assist you in finding useful information quickly.
Operator's Manual This manual describes operation, maintenance and troubleshooting of the system.
An index at the end of the manual helps you to find information quickly. A glossary
offers detailed explanations for important terms.
Q
Observe the Operator's Manual for safe operation of the instrument
o If the instrument is used in a manner not specified in the Operator's Manual, the
protection provided by the instrument can be impaired.
o Keep this manual in a safe place to ensure that it remains easily accessible at all times.
Online help The software of the cobas 8100 automated workflow series offers online help
features. When the online help system is displayed it shows information related to the
current window.
U For more information, see: Online help (p. 108).
User information
User training Contact your local Roche sales representative for any questions or information about
training on the system.
Service agreement Contact your local Roche sales representative for further information about a service
agreement for the system.
Ordering information Spare parts and consumables for the cobas 8100 automated workflow series,
including caps, barcode labels, and aliquot tips, must be ordered from Roche
Diagnostics only. When ordering, use the Roche Diagnostics catalogue number and
reference name for each item. Contact Roche technical support for the detailed
ordering list.
cobas® 8100 automated workflow series
Roche Diagnostics
8 Operator’s Manual · Version 1.0
Conventions used in this document
Visual cues are used to help you locate and interpret information in this document
quickly. This section explains the formatting conventions used in this document.
Symbols The symbols below are used in this Operator's Manual.
Abbreviations The abbreviations below are used in this Operator's Manual.
Symbol Used for
Safety caution symbol
Q Tip
P Start of procedure
S End of procedure
o List item
U Cross-reference
f Path to the active menu or window
Electrical and electronic equipment marked with this symbol are
covered by the European directive WEEE.
The symbol denotes that the equipment must not be disposed of in
the municipal waste system.
Table 3 Symbols used in this document
Abbreviation Definition
A
ACU Automatic Centrifuge Unit module
ANSI American National Standards Institute
AOB Add-on/Output Buffer module
AQM Aliquot Module
ASCII American Standard Code for Information Interchange
B
BC Barcode
BCL Barcode Labeler/Tube Feeder module
C
CE European Conformity marking
CL8 CLAS 2 Connection Line between the instrument and the UCL
CLAS 2 Clinical Laboratory Automation System 2
CSA Canadian Standards Association
CU Control Unit
D
DSP Destopper module
E
EN European standard
H
HIS Hospital Information System
Table 4 Abbreviations used in this document
Roche Diagnostics
Operator’s Manual · Version 1.0 9
cobas® 8100 automated workflow series
I
IEC International Electrical Commission
IPB Input Buffer module
L
LED Light Emitting Diode
LIS Laboratory Information System
N
n/a not applicable
O
OBS Output Buffer/Sorter module
P
PC Personal Computer
R
RCL Reformatter Connection Line
RFID Radio Frequency Identification
RFM Reformatter Module
RSF Restopper module for Archiving Caps
S
STAT Short Turnaround Time
Originally an abbreviation of Statim, meaning immediately or urgent.
U
UCL Universal Connection Line
UL Underwriters Laboratories Inc.
W
WEEE European Waste Electrical and Electronic Equipment Directive
Abbreviation Definition
Table 4 Abbreviations used in this document
cobas® 8100 automated workflow series
Roche Diagnostics
10 Operator’s Manual · Version 1.0
System Description
1 General safety information ..............................................................................................................13
2 Overview of system components ......................................................................................................29
3 Modules.............................................................................................................................................53
4 Overview of the control unit software .............................................................................................95
Roche Diagnostics
Operator’s Manual · Version 1.0 13
cobas® 8100 automated workflow series 1 General safety information
Table of contents
General safety information 1
Before you start working with the system, it is essential that you understand the
warnings, cautions, and safety requirements contained in this manual, as well as the
explanations of the safety labels to be found on the system.
In this chapter
Chapter
1
Safety classifications.............................................................................................................15
Safety precautions ................................................................................................................16
Operator qualification...................................................................................................16
Safe and proper use of the system................................................................................16
Miscellaneous safety precautions.................................................................................17
Safety summary ....................................................................................................................18
Warning messages .........................................................................................................18
Electrical safety.........................................................................................................18
Biohazardous materials...........................................................................................19
Flammable materials ...............................................................................................20
Waste.........................................................................................................................20
Caution messages...........................................................................................................20
Mechanical safety.....................................................................................................20
Working solutions...................................................................................................21
Insoluble contaminants in samples.......................................................................21
Evaporation of samples...........................................................................................22
Fatigue due to long hours of operation ................................................................22
Malfunction due to interfering electromagnetic fields.......................................22
Spillage ......................................................................................................................22
Instrument................................................................................................................22
Notices.............................................................................................................................23
Moving parts ............................................................................................................23
Circuit breakers and fuses ......................................................................................23
Spillage ......................................................................................................................24
Software.....................................................................................................................24
Safety labels on the instrument ..........................................................................................25
Locations of safety labels on the instrument..............................................................26
Disposal .................................................................................................................................28
Roche Diagnostics
14 Operator’s Manual · Version 1.0
1 General safety information cobas® 8100 automated workflow series
Table of contents
Roche Diagnostics
Operator’s Manual · Version 1.0 15
cobas® 8100 automated workflow series 1 General safety information
Safety classifications
Safety classifications
This section explains how precautionary information is presented in this manual.
The safety precautions and important user notes are classified according to the
ANSI
Z535.6 Standard. Familiarize yourself with the meanings and icons below.
Safety alert symbol
The safety alert symbol alone promotes awareness of hazards that are generic or directs
the reader to safety information provided elsewhere in the document.
These symbols and signal words are used for specific hazards.
WARNING
Warning
r Indicates a hazardous situation that, if not avoided, could result in death or serious
injury.
CAUTION
Caution
r Indicates a hazardous situation that, if not avoided, could result in minor or moderate
injury.
NOTICE
Notice
Indicates a hazardous situation that, if not avoided, can result in damage to the system.
U For more information about the product safety labels, see Safety labels on the instrument
(p. 25).
Important information that is not relevant to safety is indicated with the icon below.
Q
Tip
Indicates additional information on correct use or useful tips.
Roche Diagnostics
16 Operator’s Manual · Version 1.0
1 General safety information cobas® 8100 automated workflow series
Safety precautions
Safety precautions
Pay attention to the following safety precautions!
If you ignore these precautions, you can suffer minor to moderate injury. Each precaution
is important.
Operator qualification
As a user, you are required to have sound knowledge of relevant safety precaution
guidelines and standards and of the information and procedures contained in the
Operator's Manual.
o Do not carry out operation and maintenance unless you are trained to do so by
Roche Diagnostics.
o Carefully follow the procedures specified in the Operator's Manual for operation
and maintenance.
o Leave maintenance, installation, or service that is not described in the Operator's
Manual to trained Roche Service representatives.
o Follow Good Laboratory Practices, especially when working with biohazardous
material.
Safe and proper use of the system
Personal protective equipment o Wear appropriate protective equipment including, but not limited to, eye
protection with side shields, fluid resistant lab coat, and disposable laboratory
gloves.
o Wear a face shield if there is a chance of splash or splatter.
Installation Only trained, certified Roche Service representatives can perform an installation. The
user is responsible for providing the necessary facilities.
Operating conditions o Operation outside of the specified ranges can lead to incorrect processing or
malfunction of the system.
U See Operating conditions (p. 36).
o Use the system indoors only and avoid heat and humidity.
o Make sure that the ventilation openings of the system remain unobstructed at all
times.
o Perform maintenance according to the specified intervals to maintain the
operating conditions of the system.
o Keep the Operator's Manual in a safe place to ensure that it is not damaged and
remains available for use. The manual must be easily accessible for users.
Approved parts Only use parts and devices approved by Roche Diagnostics. Use of non-approved
parts or devices can result in malfunction of the system and can render the warranty
null and void.
Third-party software Do not install any non-approved third-party software. Installation of any third-party
software that is not approved by Roche Diagnostics can result in incorrect behavior
of the system.
Roche Diagnostics
Operator’s Manual · Version 1.0 17
cobas® 8100 automated workflow series 1 General safety information
Safety precautions
Miscellaneous safety precautions
Power interruption A power failure or momentary drop in voltage can damage the system or lead to data
loss. The items below are preventative measures.
o Roche recommends that you only operate the system with an uninterruptible
power supply.
o Do not switch off power while the control unit PC accesses the hard disk or
attached external storage devices. To ensure reliable operation, follow the shut
down procedures provided in this Operator's Manual.
Electromagnetic fields Devices that emit electromagnetic waves can cause the system to malfunction. Do not
operate the devices below in the same room where the system is installed.
o Mobile phone
o Transceiver
o Cordless phone
o Other electrical devices that generate electromagnetic fields
System unused for an extended
period of time
If the system is not to be used for an extended period of time, switch the main circuit
breaker to OFF. Remove, close, and refrigerate any remaining samples.
For further information, call technical support.
Relocation and transportation Do not attempt to relocate or transport the system. Leave relocation and
transportation to trained and authorized Roche personnel.
U For information about disposal of the system, see Disposal (p. 28).
Roche Diagnostics
18 Operator’s Manual · Version 1.0
1 General safety information cobas® 8100 automated workflow series
Safety summary
Safety summary
This safety summary contains the most necessary general warning and caution
messages. Additionally, you can find specific safety information at the beginning of
the Operation Part and the Maintenance Part.
Warning messages
WARNING
List of warning messages
Failure to observe warning messages can result in death or serious injury.
r Before operating the system, read the warning messages contained in this summary
carefully.
Electrical safety
WARNING
Electrical shock by electronic equipment
Removing the covers of electronic equipment can cause electric shock as there are high
voltage parts inside.
r Do not attempt to work in any electronic compartment.
r Do not remove any cover of the system other than those specified in the Operator's
Manual.
r Only authorized and qualified Roche personnel are to perform installation, service, or
repair of the system.
r Observe the Safety labels on the instrument (p. 25), particularly Electrical safety (p. 26).
WARNING
Power interruption
On occurrence of momentary power voltage drop due to power interruption or lighting, the
control section of this instrument may become faulty, or the system software, application
software or data may be damaged.
r Take special care when recovering from power interruption.
WARNING
Circuit breakers and fuses
If any foreign material comes in contact with the circuit board, the ICs or other circuit
elements may be damaged.
r Should any of the instrument circuit breakers or fuses blow, DO NOT attempt to
operate the system before contacting your Roche Service representative.
Roche Diagnostics
Operator’s Manual · Version 1.0 19
cobas® 8100 automated workflow series 1 General safety information
Safety summary
Biohazardous materials
WARNING
Infection by samples and associated materials
Contact with samples containing material of human origin can result in infection. All
materials and mechanical components associated with samples of human origin are
potentially biohazardous.
r Follow Good Laboratory Practices, especially when working with biohazardous
material.
r Keep the top covers and the front doors and drawers closed while the system is
operating.
r Always switch the system to the correct operating mode, as given in the appropriate
procedure, before working with an opened cover, door, or drawer, for example,
cleaning or maintenance.
r Be sure to wear appropriate protective equipment, including, but not limited to, eye
protection with side shields, fluid resistant lab coat, and disposable laboratory gloves.
r Wear a face shield if there is a chance of splash or splatter.
r If any biohazardous material is spilled, wipe it up immediately and apply disinfectant.
r If any sample comes into contact with your skin, wash it off immediately with soap and
water and apply a disinfectant. Consult a physician.
WARNING
Biohazardous materials
All in vitro diagnostic equipment, pretreatment systems and patients samples used on this
system should be treated as potentially biohazardous materials.
r Use the personal protective equipment recommended by your facility when handling
any of these components.
r If any potentially biohazardous materials come into direct contact with your skin, wash
them off immediately and apply disinfectant. Consult a physician.
r It is possible to pierce through protective gloves. Therefore use extra caution when
working inside the instrument.
WARNING
Samples
r Avoid direct contact with samples that may be biohazardous.
r If sample spills on the instrument, wear protective clothing including, but not limited to,
a lab coat, disposable laboratory gloves, and eye protection, and wipe it up
immediately.
r Make sure that the samples do not contain clots, dust or other insoluble contaminants.
If insoluble contaminants are contained in a sample, correct measuring values may not
be obtained.
WARNING
Restopper caps
The following instructions apply to all restopper caps, including Archive caps.
r The caps are for protection from evaporation only.
r The caps are not guaranteed to be leakproof.
r Store capped tubes only in an upright position.
r If capped tubes fall, or are accidentally placed on their sides, remove the caps manually
before loading them onto the IPB module.
r Capped tubes should not be stored frozen.
r Do not push the caps into overfilled tubes.
r Do not push the cap into the sample liquid.
r Do not mix samples or re-insert used caps.
Roche Diagnostics
20 Operator’s Manual · Version 1.0
1 General safety information cobas® 8100 automated workflow series
Safety summary
Flammable materials
WARNING
Cleaning with alcohol
If equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
r When performing maintenance using isopropyl alcohol, do not bring a flame close to
these liquids as this may result in a person receiving burns.
WARNING
Flammable and explosive materials
Use of dangerous flammables on or around the instrument can result in fire or explosion.
r Do not work in any electric compartment.
r Service and repair can only be performed by authorized and qualified personnel.
Waste
WARNING
Contamination of the environment by liquid and solid waste
The waste of the system is potentially biohazardous.
r Dispose of waste according to the appropriate local regulations.
U For information about the disposal of the system, see Disposal (p. 28).
Caution messages
CAUTION
List of caution messages
r Before operating, read the caution messages contained in this summary carefully.
Failure to observe them can result in minor or moderate injury.
Mechanical safety
CAUTION
Personal injury due to contact with moving parts
r Do not touch any parts of the system other than those specified. Keep away from
moving parts during operation.
r Do not touch the printer of the BCL module during operation.
r Keep the top covers and front doors and drawers closed and in place while the system
is operating unless specifically directed otherwise by the Operator's Manual.
r Switch the system to Stop mode before working with an opened cover, drawer, or door,
unless otherwise directed by the Operator's Manual.
r On the IPB module, the OBS module, and the AOB module, wait for the white light of
the lock release button to stop flashing before you open a loading lane drawer.
r On the RFM module, wait for the white light of the loading lane button to start flashing
before you load empty 5-position racks into the loading lane.
r On the AQM module, do not replace tip trays or empty the solid waste disposal box
during operation.
r During operation and maintenance, proceed according to the instructions given in the
Operator's Manual.
r Do not attempt to disassemble the instrument.
r Observe the system safety labels illustrated from page p. 26 to p. 28.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284
  • Page 285 285
  • Page 286 286
  • Page 287 287
  • Page 288 288
  • Page 289 289
  • Page 290 290
  • Page 291 291
  • Page 292 292
  • Page 293 293
  • Page 294 294
  • Page 295 295
  • Page 296 296
  • Page 297 297
  • Page 298 298
  • Page 299 299
  • Page 300 300
  • Page 301 301
  • Page 302 302
  • Page 303 303
  • Page 304 304
  • Page 305 305
  • Page 306 306
  • Page 307 307
  • Page 308 308
  • Page 309 309
  • Page 310 310
  • Page 311 311
  • Page 312 312
  • Page 313 313
  • Page 314 314
  • Page 315 315
  • Page 316 316
  • Page 317 317
  • Page 318 318
  • Page 319 319
  • Page 320 320
  • Page 321 321
  • Page 322 322
  • Page 323 323
  • Page 324 324
  • Page 325 325
  • Page 326 326
  • Page 327 327
  • Page 328 328
  • Page 329 329
  • Page 330 330
  • Page 331 331
  • Page 332 332
  • Page 333 333
  • Page 334 334
  • Page 335 335
  • Page 336 336
  • Page 337 337
  • Page 338 338
  • Page 339 339
  • Page 340 340
  • Page 341 341
  • Page 342 342
  • Page 343 343
  • Page 344 344
  • Page 345 345
  • Page 346 346
  • Page 347 347
  • Page 348 348
  • Page 349 349
  • Page 350 350
  • Page 351 351
  • Page 352 352
  • Page 353 353
  • Page 354 354
  • Page 355 355
  • Page 356 356
  • Page 357 357
  • Page 358 358
  • Page 359 359
  • Page 360 360
  • Page 361 361
  • Page 362 362
  • Page 363 363
  • Page 364 364
  • Page 365 365
  • Page 366 366
  • Page 367 367
  • Page 368 368
  • Page 369 369
  • Page 370 370
  • Page 371 371
  • Page 372 372
  • Page 373 373
  • Page 374 374

Roche cobas 8100 AliquoterModule User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI