Philips Smart Pad Cartridge
Please be carefully when using.
Pattern Figure 1Figure 2Figure 3
Picture
(Sample)
Event Separated gel that has
folded onto itself when
peeled.
Separated, folded gel may
also have a discolored and/or
melted appearance.
Gel almost completely
separated from backing.
Action
Apply pads to the patient. Do not hesitate.
※Please follow the direction from the audio
guidance of AED.
•Replace pads cartridge if
a spare is available.
•If no spare is available,
perform CPR until help
arrives.
•If you notice the gel beginning to separate from the foam backing as you peel, try to prevent the
gel from folding onto itself if possible.
•The foil seal on the pads cartridge should be opened only for patient use in an emergency
because the pads will quickly dry out if the foil seal is broken.
•If possible, please ensure that you store a spare pads cartridge with your HS1/OnSite/Home AED.
Please contact TEL: 0120-556-857 if there is any questions about this matter.
Note that we cannot support for inquiries during patient use in an emergency.
Please enclose this guidance in carrying case according to the instruction on the back of this document.
(For adults:M5071A, For infant/child:M5072A)
Guidance for Safety Use
※Per JRC guideline 2020, please use alternative reading as follows below.
“For adults” => “For from elementary students until adults”
“For infant/child” => “For pre-school children”
Surface area of gel peeling
Confirmation
required when using