Performing a Self-Test
The AED Pro unit performs automatic or manual self-tests to verify its integrity and
readiness for emergency use. These tests verify the following:
• Battery energy — Verifies that the battery energy is sufficient for at least 2 hours of
continuous monitoring and ten shocks at maximum energy.
• Defibrillation electrodes connection — Verifies that defibrillation electrodes are properly
preconnected to the unit.
• ECG circuitry — Verifies that the ECG signal acquisition and processing electronics are
functional.
• Defibrillator charge and discharge circuitry — Verifies that the defibrillator electronics
are functional and can charge and discharge at 2 joules.
• Microprocessor hardware and software — Verifies the proper function of the
microprocessor electronics and the integrity of the software.
• CPR circuitry and sensor — Verifies that the CPR monitoring and compression depth
detection are functional (when CPR-D•padz are attached).
• Audio circuitry — Verifies that the audio output circuitry is functional.
• Display — Verifies that the visual indicators are functional.
Manual Self-Test
You can manually initiate a self-test by pressing and holding the On/Off button for 5
seconds.
The unit illuminates the Shock button and issues voice and text messages so that you
can verify the visual and auditory output functions. In addition, the screen shows
information about the unit’s hardware and software.
Preconnecting the Defibrillation Electrodes Cable
WARNING! To prepare for an emergency, keep the defibrillation electrodes cable
connected to the unit at all times.
WARNING! Do not reuse defibrillation electrodes.
The AED Pro unit supports both adult and paediatric electrode pads. The device adjusts
defibrillation energy to adult or paediatric levels depending on the type of electrodes
connected to it. Always use electrode pads that are appropriate for the patient.
WARNING! Do not use adult defibrillation electrode pads on patients under 8 years of
age.
The electrode packaging allows you to connect the cable to the unit while the pads
remain in a sealed envelope.
• To prepare for future emergencies, after each use connect a new package of electrodes
by plugging the electrode cable into patient cable connector.
• To ensure that the electrodes are fresh and ready to use in an emergency, regularly
check the electrode expiration date on the preconnected electrode pack.
• Replace expired electrodes.
• After completing its power-on self-test, the unit issues a voice and text message to
indicate the type of electrodes that are connected (ADULT PADS or PEDIATRIC PADS).
Verify that the connected electrodes are appropriate for the patient. If necessary, replace
the connected electrodes with appropriate ones.
If the electrode cable is not properly connected to the unit, the unit issues the voice and
text prompt PLUG IN CABLE.
If the electrodes are not properly attached to the patient, the unit issues the voice and text
prompt CHECK DEFIB PADS or ATTACH DEFIB PADS TO PATIENT’S BARE CHEST.