  • Hello! I am an AI chatbot trained to assist you with the Extech Instruments DCP42 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
USERGUIDE
ModelDCP42
160WLaboratoryGradeConstantPower,
SwitchingModePowerSupply

DCP42enGB_V1.512/14
2
TableofContents
1. INTRODUCTIONANDFEATURES 4
2. SAFETY 5
3. POWERSUPPLYDESCRIPTION 6
4. CONTROLMODES 7
4.1NormalMode 7
4.2PresetMode 7
4.3SetMode 7
4.3.1SettingtheΔV/Δtvalue 7
4.3.2ActivatingtheRemoteControlMode 8
4.3.3DeactivatingtheRemoteControlMode 8
4.3.4FactoryDefaultReset 8
5. OPERATION 9
5.1PoweringtheSupply 9
5.2SwitchingthePowerSupplyONandSelfTestInformation 9
5.3UsingtheControlknobs 9
5.4ConnectingthePowerSupplyandrunningaTest 9
5.5KeypadLockoutfunction 9
5.6ManualOutputON/OFF 9
5.7Ammeter/WattmeterSelection 9
6.
SUPPLEMENTALFUNCTIONS 10
DCP42enGB_V1.512/14
3
6.1ΔV/ΔtFunction 10
6.1.2SettingtheDCgeneratorvoltagelevel 10
6.1.2SettingtheΔt 10
6.1.3GeneratingRampUporRampDownDCoutputcycles 10
6.2FunctionA/B/CWaveformGenerator 11
6.2.1SetFUNC(A/B/C): 11
6.2.2GeneratingWaveForm 11
6.3UpperVoltageandCurrentLimits 12
6.4RemoteSensing 12
6.4.1Connection 12
6.4.2Disconnection 12
6.5RemoteControlOperation 13
6.5.1REMOTECONTROLMETHOD1:UsingtwoexternalDCvoltagesources 13
6.5.2REMOTECONTROLMETHOD2:Usingtwo0to5KOhmvariableresistors13
6.5.3RemoteControlOutputON/OFF 14
7. PCSOFTWAREPROGRAM 14
8. AUTOMATICOUTPUTSTEPPROGRAMS 14
8.1ExternalTimedProgramWindow 15
8.2OutputON/OFFfunction 15
8.3AutomaticTimedOutputOperation 15
9. APPENDIXADDITIONALOPERATIONALEXAMPLES 16
10. SPECIFICATIONS 17
11. TROUBLESHOOTING 19
DCP42enGB_V1.512/14
4
1. IntroductionandFeatures
ThankyouforselectingtheExtechDCP42SwitchingModePowerSupply.Thisnewlyredesigned
laboratorygradepowersupplydiffersfromaconventionalpowersupplyinthatitcancalculate
andchangethevoltageandcurrentlimitpointsaccordingtotheavailablemaximumpower.
Thereforethemaximumlimitsofthevoltageandcurrentarechangeableaccordingtotherated
power.Aconventionalpowersupplyhasonlyasetmaximumvoltageandcurrentsuchthatallof
theoperatingV&Amustfallwithintheselimits.TheDCP42offerscostsavingandspacesaving
advantagesasoneconstantpowerpowersupply
candotheworkofseveralconventionalpower
supplies.
TheDCP42ishighlyefficient,incorporatesupgradedSMPScircuitrywithsmallformfactor,benefits
fromanautomaticcrossoverforCCandCV,hasthree(3)Voltage/Currentpresets,andcanbe
remotecontrolled.TheDCP42isperfectforsolvingavarietyofloadingconditionsandapplications.
Featuresincludethefollowing:
Dualaction(Course/Fine)tuningprovidessmooth,accurate,andquicksettingof
Voltage/Current
Setting,changing,andverifyingthecurrentlimitlevelisconvenientandcanbe
performedwithoutoutputpolespiking
TheRemoteControlfeaturepermitsoutputON/OFFandVoltage/Current
adjustments
withouttheneedtotouchthefrontpanelofthepowersupply
TheUSBportoffersPCconnectivityforprogrammingandrunningramp/soakcycleswith
20programmablesetsofVoltage/Currentandvaryingtestdurations(upto999cycles)
TheDCrampandwaveformgeneratorcaneitherbe
panelsetorsetbythesupplied
softwarewhichpreviewsthefinalwaveform.
ApplicationsandindustrieswheretheDCP42canbeemployedincludelaboratory,
telecommunications,production testing,fieldtesting,DCnetworkpowering,andmore
Three(3)userprogrammableVoltage/Currentsettingsofferquickrecalloffrequently
usedtestsettings.
Thisdeviceisshippedfullytestedandcalibratedand,withproperuse,willprovideyearsofreliable
service.Pleasevisitourwebsite(www.extech.com)tocheckforthelatestversionofthisUser
Guide,ProductUpdates,andCustomerSupport.
DCP42enGB_V1.512/14
5
2. Safety
Thismanualcontainsimportantsafetyandoperationinstructionsforcorrectuseofthepower
supply.Readthroughthemanualandpayspecialattentiontothemarkingsandlabelsofthisunit
andequipmenttobeconnected.
Payspecialattentiontothesetwotypesofnoticesusedinthismanual:
WARNING:
Failuretoobservethiswarningmaycauseinjurytopersonsanddamagetopower
supplyorconnectedequipment.
CAUTION:Failuretoobservethiswarningmayresultindamagetoequipmentand
improperfunctioningofthepowersupply.
WARNING
Donotusethispowersupplynearwater
Donotoperateortouchthispowersupplywithwethands
Donotopenthecasingofthepowersupplywhenitisconnectedtoacmains
Referallservicingtoqualifiedservicepersonnelonly
BeforereplacingtheACfuse,determinethecauseandclearupthecausefirst
ReplacetheACfusewiththesametypeandratingastheoriginalfuse
ThemaximumoutputvoltageoftheDCP42exceeds60VDC,avoidtouchingthemetal
contactpartsattheoutputterminals
CAUTION
Useagrounded3pinACsource
Thisunitisintendedfor
indooruseonly
Donotoperateorplacethisunit inahumid,dustylocat ion
Donotoperatethisunitin directsunlightornearanyheatsource
BeforepluggingintolocalACmains,checktheratinglabelatthebackoftheunit
Donotblock
anyventilationopeningsoftheunit
Thisunitmustbeusedwithinthespecifiedrating;regularexcessivecontinuousloadingmay
causedamagetothepowersupply
Thegaugesizeoftheinputpowercablemustbeatleast0.75mm
2
andthetotallengthof
powercablemustnotexceed3m
InputFuseRecommended:T3AL250V(3ATimeLag)
ENVIRONMENTALCONDITIONSFOROPERATION
1080%R.H.(RelativeHumidity)
Altitudeupto2000m(6561ft.)
Installationcategory:CAT2
Pollutiondegree:2
Mainssupplyvoltagefluctuation
upto±10%ofnormalvoltage
DCP42enGB_V1.512/14
6
3. PowerSupplyDescription
(1)LEDpanelmeterdisplaywithCC/CV/WIndictor
(2)RemoteControlIndicator(RemoteControl/USBmode)
(3)OutputVoltageControlKnob
(4)OutputCurrentControlKnob
(5)PowerON/OFFSwitch
(6)6buttonKeypad(withillumination):
‐PresetKeyd1,2&3
‐MenuKey(ShiftKey)
‐Lock/Unlock
Key(A/WmeterKey)
‐OutputON/OFF(ΔV/ΔtKey)
(7)OutputSafetyJacks(Rated10A)
(8)USBRemoteControlport(Foraccesstocomputertoruncyclicaloperationwithprogrammable
voltage,current,periodtimeandcycle)
(9)RemoteSensingTerminal
Warning!Shortingtheremotesensingterminalorconnectingthe
sensingterminalwiththe
polarityreversedwilldamagethepowersupply
(10)RemoteControlTerminal
(11)ACInletsocket(withfuseholder)
DCP42enGB_V1.512/14
7
4. ControlModes
Therearefourcontrolmodesforthepowersupply:NORMAL,PRESET,SET,andREMOTECONTROL
modes.ThepowersupplyisfactorypresettoNormalModewithmaximumcurrentlevelCC.
4.1NormalMode
Thisisthefactorydefaultmode;thepowersupplyoutputVoltageandCurrentarecontrolledby
thedualactiongainknobs.
1. Pushtheknobstotogglethecoarseandfinetuning,noticethesubtlechangesin
brightnessoftherelatedLED.
2. Adjusttheknobstothedesiredvaluesusingcoarseandthenfinetuning.
3. Tocheckthepresetcurrentlevel,turntheCurrentKnoblightlyinanydirection.
4. Thedisplaywillresumeitsnormalbrightnessafterafewsecondstoreaffirmthe
adjustment.
4.2PresetMode
1. TherearethreepresetoutputsP1/P2/P3available.
2. PressthePresetbutton(1,2,or3)toenterthepresetmode(thebutton’slightwill
switchON).
3. Thepresetvaluesarefactorysetasshowninthetablebelowandareretainedevenafter
theunithas
beenpowereddown.
4. Theusercanchangetheseoutputsettingsusingthevoltageandcurrentknobs.
RecallNo. OutputVoltage OutputCurrent
P1 5V Maximum
P2 13.8V Maximum
P3 40VMaximum
4.3SetMode
EntertheSetModebypushingandholdingtheMENUbuttonfor3seconds.IntheSetModethe
usercanadjusttheΔV/Δtvalue,accesstheremoteControlmode,orperformaFactoryReset
recallfunction.Note:Ifnoselectionismade after10secondsthepowersupplydefaults
toits
priorsettings.
4.3.1SettingtheΔV/Δtvalue
a.PressandholdtheMENUbuttonfor3secondstoaccesstheSetMode.
b.TurntheVoltagecontrolknobuntilthepanelmeterdisplays“dudtSet”asshownbelowand
thenpresstheVoltagecontrolknobtoenter
theΔV/Δtsettingmode.
c.AdjusttheVoltageandCurrentcontrolknobsasdesiredtosettheΔV/Δtvalue.Furtherdetails
areprovidedinthededicatedsectionforΔV/Δtmodelaterinthisguide.
d.Whenfinished,presstheVoltagecontrolknobtoconfirmandreturntoSetmenu
e.Pressthe
MenukeytoexittheSetmenu
f.TostarttheΔV/Δtfunction,presstheSHIFTbuttonandthenpresstheΔV/Δtbuttontorun,
g.TostoptheΔV/Δtfunction,presstheSHIFTbuttonandthenpresstheΔV/Δtbuttonagain.
DCP42enGB_V1.512/14
8
4.3.2ActivatingtheRemoteControlMode
a.PressandholdtheMENUbuttonfor3seconds.
b.TurntheVoltagecontrolknobuntilthepanelmeterdisplays“rCSet”asshownbelowand
thenpresstheVoltagecontrolknobtoconfirmentryintotheremotecontrolmode.
c.TurntheCurrentcontrolknobtoselectremoteYESorNO.
d.PresstheVoltagecontrolknobtoconfirmandtoreturntotheSetmenu.
e.PresstheMenubutton toexittheSetmenu.
f.DetailsoftheremotecontrolmodearelocatedinthededicateRemote
Controlsectionlaterin
thisguide.
Note:AfteractivatingtheRemoteControlMode,allfunctionkeyslocatedonthefrontpanelwill
belocked(theLock/UnlockbuttonlightswitchesON).
4.3.3DeactivatingtheRemoteControlMode
a. PresstheLock/Unlockbuttontounlockallofthefunctionkeys
b. Repeat
stepsatocinthe‘ActivatingtheRemoteControlMode’sectionaboveandset
RemoteControltoNOandthenpressthevoltagecontrolknob.
4.3.4FactoryDefaultReset
a.PressandholdtheMENUbuttonfor3seconds.
b.TurntheVoltagecontrolknobuntilthepanelmeter
displays“FACSSEt”asshownbelowand
thenpresstheVoltagecontrolknobtoconfirmentryintotheFactoryResetmode.
c.TurntheCurrentcontrolknobtoselectYesorNo.
d.PresstheVoltagecontrolknobtoconfirmandtoreturntotheSetmenu.
e.PresstheMenubutton toexittheSetmenu.
f.ThePowerSupplywillrebootandreverttoitsFactoryDefaultsettings.
DCP42enGB_V1.512/14
9
5. Operation
Note:TheDCP42TotalOutputVoltageRangeis42VDCandtheTotalRatedCurrentis0to10ADC
5.1PoweringtheSupply
ChecktheratinglabelofthepowersupplyandensurethatitcomplieswiththeACmainsvoltage
thatistobeused.ConnectthepowersupplytotheACMainsusingtheprovidedpowercord.
5.2SwitchingthePowerSupplyONandSelfTestInformation
ThepowersupplywillperformaseriesofselftestswhenitisswitchedON.TheLEDandother
indicatorsonthefrontpanelwillilluminateinsequence.Whenthecoolingfanisbeingchecked,
ahighspeedwindsoundcanbeheard.Aftertheselftests,theCV,V
andALEDindicatorsarelit
andtheunitdisplaysvoltageand0.0current.TofindtheCCcurrentlevel,turnthecurrent
controlknoboneclickineitherdirection.Thecurrentdisplayreturnsto0.0afterafewseconds.
5.3UsingtheControlknobs
Therotaryencodercontrolknobshavefineandcoarsetuningwith‘clicking’movement.Push
theknobstotogglebetweencoarseandfinetuning,noticethesubtlechangesinbrightnessof
therelatedLED.Adjusttheknobstothedesiredvaluesbyusingcoarseandthenfinetuning.
Thedisplaywillresumeitsnormalbrightnessafterfewsecondstoconfirmanadjustment.
5.4ConnectingthePowerSupplyandrunningaTest
1. Connecttheequipmentundertesttothepowersupply.Red(+)isconnectedtothepositive
polarityinputoftheequipmentandBlack()isconnectedtothenegativepolarityinputof
theequipment.
2. SwitchONthepowersupplyfirst;thepanelmeter&greenCVIndicatorshouldilluminate.
3. Next,switchONtheequipmentundertest;thepanelmeter&greenCVIndicatorshould
remainlitingreen.
4. Testingcannowbegin.
5. Whenfinished,switchOFFtheequipmentundertestfirstandthenswitchoffthepower
supply.
5.5KeypadLockoutfunction
1. PressLOCK/UNLOCKtolockorunlockallofthefunctionbuttonsandcontrolknob sonthe
frontpanel.
2. TheLOCK/UNLOCKbuttonlightswitchesONwhenthepushbuttonsarelocked.
5.6ManualOutputON/OFF
1. PressOUTPUTON/OFFbuttontomanuallysettheoutputONorOFF. 
2. TheOUTPUTON/OFFbuttonlightswitchesONwhentheoutputisON.
5.7Ammeter/WattmeterSelection
1. WiththeoutputONandallpresetsOFF,pressSHIFTandthenpressPRESET.
2. PresstheSHIFTbuttonagainandthenpresstheLOCK/UNLOCKbuttontochangefrom
AmmetermodetoWattmetermode.
3. Tochang e backtoAmmetermodefromWattmetermode,presstheSHIFTbuttonandthen
presstheLOCK/UNLOCKbutton.
DCP42enGB_V1.512/14
10
6. SupplementalFunctions
6.1ΔV/ΔtFunction
Thereare3DCvoltagegeneratorsettings(A,B,andC).
Preset1=A
Preset2=B
Preset3=C
ΔVab(fromvoltagelevelAtovoltagelevelB).
Δtab(timeinsecondsfromvoltagelevelAtolevelB,thistransittime
isadjustablefrom0to20
seconds)
6.1.2SettingtheDCgeneratorvoltagelevel
Thereare3generatorvoltagelevelspresets(A,B,andC).
PressthePresetbutton1,2and3andadjustthevoltagetothedesiredvalue.
6.1.2SettingtheΔt
Thereare6setsof
Δtavailable,turningtheVoltagecontrolknobcyclesthrough thefollowing
sequence:AB→BA→AC→CA→BC→CB
UsingtheABsettingasanexample,turntheVoltagecontrolknobuntilAbappearsonthe
voltmeterdisplayandthenturntheCurrentcontrolknobtosettodesiredtransittimeΔt(5
secondsforexample).
6.1.3GeneratingRampUporRampDownDCoutputcycles
Rampupgeneratorexample:SetA=5V;B=10V;Δtab=5seconds
1. ConnectasuitableloadtotheoutputandturnONtheoutputbypressing
the“OUTPUT
ON/OFF”button;notetheoutputLEDindicator.
2. Push“Preset1”andnotetheoutputvoltageof5V.Press“Preset2”andnotethatthe
outputvoltagerisesto10Vin5secondsshowingarampupvoltage(theoutputwill
remainat10V).
3. Pressing“Preset1”
willnowgeneratearampdownvoltagefrom10Vto5Vin5seconds
andwillremainatthe5Vlevel.
Note:Thefinaloutputvoltagelevelwillbethelastvoltagelevelgenerated.
DCP42enGB_V1.512/14
11
6.2FunctionA/B/CWaveformGenerator
Preset1=A
Preset2=B
Preset3=C
FUNC(A/B/C)setstheduration(from0to600seconds)thatthevoltagegeneratorremainsON
atthespecifiedoutputvoltagelevelbeforemovingtoanothervoltagelevel.
6.2.1SetFUNC(A/B/C):
1. EntertheΔV/Δtsettingfirst(RefertotheSetModesection4.3.)
2. Turnthe VoltagecontrolknobtoselectthedesiredfunctionandthenturntheCurrent
controlknobtosetthetimeperiod(0to600seconds)forthechosenvoltagelevel.
3.
PressingtheCurrentcontrolknobtogglescoarseandfinetuningofthetimeperiod.
4. Togeneratewaveforms,theFuncA/B/CbuttonisusedinconjunctionwithΔV/Δtbutton(the
timeperiodfromonevoltageleveltoanothervoltagelevel)
5. Usingtheabovefunctionssimultaneously,3reversiblepairs(AB&BA,BC&CB,AC&CA)can
beusedtogeneratewaveforms.
Example1:Pulsewaveformwith3secat10Vand1secat5V.
SetA(Preset1)=5V,setB(Preset2)=10V
SetFuncA=1secondbyturningtheCurrentcontrol
knob
FuncB=3secondsusingthesameprocedureasabove
SetΔtab=0,SetΔtba=0
6.2.2GeneratingWaveForm
TogeneratethewaveformpushSHIFTandthenpresstheΔV/Δtbutton
ThewaveformwillrepeatcyclicallyuntiltheSHIFTbuttonandthentheΔV/Δtbuttonispressed.
Example2:Trianglewaveform
SetA=5V,B=10V
SetΔtab=3seconds,SetΔtba=3seconds
SetFunc.A=3seconds,SetFunc.B=3seconds
Note:Thewaveformgeneratorcanbeoperatedviatheremoteprogrammingsoftware(with
waveformpreviewandgraphicaldataloggingoutputpresentationfeatures).Detailsareprovided
inthesection4titled‘PCSoftwareProgram’.
PleaseverifythatthePresetkeysarebothlitandthatoneisflashing.Theflashinglightwill
jump
fromonetotheother,AtoB,thenBtoA.ThewaveformwillrepeatcyclicallyuntilSHIFTand
thenΔV/Δtarepressedagain.Examples3,4,and5areprovidedinAppendixAlaterinthis
guide.
DCP42enGB_V1.512/14
12
6.3UpperVoltageandCurrentLimits
Whenthevoltageorcurrentattheoutputterminalexceedsthemaximumallowablelimits,the
outputwillautomaticallyswitchOFF.Thisadditionalprotectionfeatureisnecessaryforapower
supplywithsuchawiderangeofvoltageandcurrentcapabilities.
TosettheUVL:
1. TosettheUVL:PresstheSHIFTbuttonandthenpresstheVoltagecontrolknob.The
voltmeterwillshowthecurrentlimitsettingandtheAmmeterwillshowSuul.
2. Turnthe Voltagecontrolknobtothedesiredlimitvalue.PressSHIFTtoconfirmandto
returntonormaloperation.
3. TosettheUCL,pressSHIFTandthenpresstheCurrentcontrolknob.TheAmmeterwill
showthecurrentlimitsettingandtheVoltmeterwillshowSUCL.
4. Turnthe Currentcontrolknobtothedesiredlimitvalue.PressSHIFTtoconfirmandto
returntonormaloperation.
Note:
TheUVLandUCLsettingsdefaulttothemaximumoutputvalue.
6.4RemoteSensing
Note:ObservewarningsdisplayedbythePowerSupply,anincorrectdisconnectionsequence
willdamagethePowerSupply
Warnings:NevershorttheRemoteSensingTerminaland alwaysdisconnecttheRemoteSensing
Terminalfirst.
6.4.1Connection
1.Firstcompletethepowerconnectionsbetweenpowersupplyandequipment.
2.Checkandensurethatthe
powerconnectionsaresecure.
3.MakeconnectionsbetweenRemoteSensingandequipment.
Warning!:NevershorttheRemoteSensingTerminaland neverconnecttheRemoteSensing
Terminalinreversepolarity
ThediagrambelowshowstheconnectionsforRemoteSensing,Poweroutput,andEquipment.
TheremotesensingwireshouldbeATLEAST
22AWGwiresize.
6.4.2Disconnection
Warning!Incorrectdisconnectsequenceswilldamagethepowersupply
1.Firstdisconnecttheremotesensingconnections.
2.Thendisconnectthepowerconnectionsbetweenthepowersupplyandequipment.
DCP42enGB_V1.512/14
13
6.5RemoteControlOperation
Therearetwomethodsforremotelycontrollingcurrentandvoltageadjustments.Currentmust
becontrolledforbothmethodsotherwisetheunitwilldefaulttoCCmode.
6.5.1REMOTECONTROLMETHOD1:UsingtwoexternalDCvoltagesources
RemoteSocketPinAssignmentforexternalvariablevoltagesource
PIN FUNCTIONS REMARKS
1 InternalDC+5V Lowerthan50mA
2 Voltageadjust 0to5V
3 Currentadjust 0to5V
4 Ground
5 OutputOFF Shorttoground
6 N.A.
7 N.A.
8 N.A.
Checktheoutputvoltagerangeofthepowersupplybyvaryingtheexternalvoltagesource.
Shortcircuitthemainoutputwith14AWGwiretochecktheCCsettingonthe displaywhile
varyingtheexternalvoltagesource.
6.5.2REMOTECONTROLMETHOD2:Usingtwo0to5KOhm
variableresistors
RemoteSocketPinAssignmentforexternalvariableresistor
PIN FUNCTIONS REMARKS
1 InternalDC+5V Oneendofresistor
2 Voltageadjust Variablelegofresistor
3 Currentadjust Variablelegofresistor
4 Ground Otherendofresistor
5 OutputOFF Shorttoground
6 N.A.
7 N.A.
8 N.A.
Checktheoutputvoltagerangeofthepowersupplybyadjustingthe5Kohmvariableresistor.
Shortcircuitthemainoutputwith14AWGwiretochecktheCCsettingonthe displaywhile
adjustingthevariableresistor.
DCP42enGB_V1.512/14
14
6.5.3RemoteControlOutputON/OFF
RemoteoutputON/OFFcontrolcanbeactivatedinanyofthemodesNormal,Preset,Set,and
Remote.Reviewtheconditionsbelow:
Bydefault,Pin5isopenandtheoutputisON.
ShortingPin5toPin4(ground)switchestheoutputOFF.
WhentheoutputisOFF,theC.V.&C.C.LEDswillflash. Thepresentoutputvoltageand
currentsettingwillbeshownonthepanelmeter.
Theusercanadjusttheoutputusingthevoltage&currentcontrolknobtothedesired
valuewhentheoutputisOFF.
Remark:Useonlythe8pinremoteplugprovidedandconnectwith22AWGwires.See
diagrambelow.
Note: Pin numbers are marked on the black area of the socket
7. PCSoftwareProgram
ThismeterhasthecapabilitytoconnecttoandcommunicatewithaPC.Toinstallandusethe
software,pleaserefertotheinstructionsprovidedonthesuppliedCDROMand/ortheinstructions
providedintheHELPUtilitywithinthesoftwareprogram.
Checkthesoftwaredownloadpageofthewebsitewww.extech.comforthelatestversionofthePC
softwareanditsoperatingsystemcompatibility.
8. AutomaticOutputStepPrograms
TousetheAutomaticOutputSteputilitytheusermustfirstbecomefamiliarwiththePC
interfaceandthesuppliedsoftware.Toinstallandusethesoftware,pleaserefertothe
instructionsprovidedonthesuppliedCDROMand/ortheinstructionsprovidedintheHELP
Utilitywithinthesoftwareprogram.Oncefamiliaritywiththesoftwareisachievedrefertothe
instructionsbelow:
TheunitcanbeexternallyprogrammedviaaPCtorun20steps,eachstepwithapreset
voltage/currentlimitingvalueandapresettimeperiodof1secondto99minutes.Thetimed
programcanbesetto
runonecycleoruptoaninfinitenumberofcycles.
DCP42enGB_V1.512/14
15
8.1ExternalTimedProgramWindow
ClearTable:DeletealldataintheDisplayTableandpreparefornewdataentry.
Run(Stop):TorunandstoptheTimedProgram.
RunningCycle:Enterthenumberofdesiredrunningcycleshere.
Thenumericalrangeforcyclingis0999.Notethatthemaximumcyclescanbe setto‘infinite’
when0”cyclesisentered.TheExternalTimedProgramallowstheusertosettheoutputeither
ONorOFFbyselectingtheboxesinthelastcolumn.
8.2OutputON/OFFfunction
OutputONTicked:OutputisONforthatstep
OutputONUnticked:OutputisOFFforthatstep.
8.3AutomaticTimedOutputOperation
1.Clearolddatainthetablebyclicking[ClearTable].
2.Enterdatainthetableusing'UpDownLeftRight'keysonPCkeyboardfornewlocations.
3.Datathatexceedtheratedvoltageandcurrentwillnotbeaccepted.
4.VoltagesthatexceedtheUVL(UpperVoltageLimit)
willnotbeaccepted.
5.Ifretrievedorentered,datathatexceedthepresetUpperorLowerLimitsettingforvoltage/
current/timeperiodswillswitchtoredincolorandwillnotbeaccepted.
6.Whentherunningtimeperiodofanystepissettozero
minutesandzeroseconds,thisstep
becomestheterminatingstepandthecyclewillendatthisstep.Intheaboveexamplethereare
4stepseachwitha2secondtimeperiod,ifstep3issettozerominutes&seconds,theprogram
willonlycyclestep1and
2andwillnotgotostep4.
7.Enterthenumberofdesiredrunningcycles.
8.Click[Run]toruntheExternalTimeProgram.TheExternalTimedProgramfeatureallowsthe
usertosettheoutputeitherONorOFFbyselectingtheboxesinthelastcolumn.
DCP42enGB_V1.512/14
16
9. AppendixAdditionalOperationalExamples
ExamplesofFuncA/B/CandWaveFormGeneratorFeatures
TheseexamplesarecontinuedfromExamples1and2inSection6.2earlierinthisguide.
Example3‐Irregularwaveform
SetA=5V,B=10V
SetΔtab=1second,SetΔtba=2seconds
Set
Func.A=3seconds,SetFunc.B=3seconds
Example4‐Irregularwaveform
SetA=5V,B=10V
SetΔtab=2second,SetΔtba=2seconds
SetFunc.A=2seconds,SetFunc.B=6seconds
Example5‐Irregularwaveform
SetA=5V,B=10V
SetΔtab=2second,SetΔtba=0seconds
SetFunc.A=4seconds,SetFunc.B=3seconds
Example6‐Irregularwaveform
SetA=5V,B=10V
SetΔtab=4second,SetΔtba=4seconds
SetFunc.A=2seconds,SetFunc.B=2seconds
DCP42enGB_V1.512/14
17
10. Specifications
Output
VariableOutputVoltage(LimitedtoO/P160W)0‐42VDC
VariableOutputCurrent(LimitedtoO/P160W)0‐10A
NominalOutputPower(42V/3.8A)160W
PresetOutputs
Preset1(P1)Voltage:5V±0.2V,Current:10A±0.2A
Preset2(P2)Voltage:13.8V±0.2V,Current:10A±0.2A
Preset3(P3)Voltage:
20V±0.2V,Current:4.0A±0.2A
VoltageRegulation
Load(0100%RatedCurrent)≤50mV
Line(90264VACVariation)≤10mV
CurrentRegulation
Load(1090%RatedVoltage)≤50mA
Line(90264VACVariation)≤10mA
Ripple&Noise
Ripple&Noise(RMS)Voltage≤8mVrms
Ripple&Noise(peakpeak)Voltage≤80mVpp
CurrentRipple&
Noise(peakpeak)≤50mA
MeterType&Accuracy
VoltageMeter4DigitLEDDisplay±(0.1%+5counts)
CurrentMeter4DigitLEDDisplay±(0.1%+5counts)
Resolution
OutputVoltageResolution0.02V
OutputCurrentResolution0.01A
SettingAccuracy
OutputVoltage±(0.2%+0.05)
OutputCurrent±(0.2%+0.05)
OutputRise&FallTime
OutputTON(50%LoadCurrent)≤80ms
OutputTON(100%LoadCurrent)≤1200ms
OutputTOFF(50%LoadCurrent)≤90ms
OutputTOFF(100%LoadCurrent)≤50ms
Power
OperationInputVoltageRange90–264VAC
OperationInputVoltageFrequencyRange4565Hz~
NoLoadInputCurrent
(230Vac/100Vac):≤220mA
Full
LoadInputCurrent
(230Vac/100Vac):≤1.0A/2.3A
PowerConsumption
(230Vac/100Vac):≤190W/200W@42V/3.8A
Efficiency
(230Vac/100Vac):≥86%/84%@42V/3.8A
DCP42enGB_V1.512/14
18
SwitchingFrequency
4555KHz~
TrackingOverVoltageProtections
O/P010V:SetVoltage+(1.0±0.3V)
O/P1042V:105125%ofsetvoltage
PowerFactor
ControlPowerfactorcorrection:>0.91atoptimalload
CoolingMethod
NaturalConvection
LeakageCurrent
2mA
Protections
Adjustableuppervoltagelimit,Adjustableuppercurrentlimit,
ShortCircuit,Overload,
OverTemperature,TrackingOVP
InputFuse
T3.15AL250V
Safety&EMC
SafetyStandardCE:EN61010
WithstandingVoltageI/PO/P:3.0KVac,I/PF/G:1.5KVac,O/PF/G:0.5KVac
InsulationResistanceI/PO/P,I/PF/G,O/PF/G:100Mohm@500Vdc
EMCCE:EN55011
HarmonicsCurrentEmissionEN6100032
VoltageFluctuations&FlickerEN6100033
EMCImmunityEN6100061
Environmental
WorkingTemperature0to+40°C(32to+104°F)
WorkingHumidity1080%RHNoncondensing
StorageTemperatureandHumidity‐15to+70°C(5to158°F)1085%RHNoncondensing
PollutionDegreeClass2
General
CaseMaterialPolycarbonateFrontPanelandElectroGalvanizedSteelEnclosure
CaseProtectionIP20
ConnectionDia.4.0mmSafetyJackx3(OutputPositive,NegativeandFunctionGround)
Dimensions(WxHxD)200x90x250mm(7.9x3.6x10inch)
Weight2.5kg(5.5lb)
DCP42enGB_V1.512/14
19
11. Troubleshooting
OUP:OverVoltageProtection
Thisunithasabuiltintrackingovervoltageprotectionfeature.Intheeventofoutputvoltage
becominggreaterthanthesetvalue(seespecifiedrangefromspecificationstable),protection
willbetriggeredandtheoutputpowerwillbecutoffandOUPwarningappearsasbelow.
Toresetthewarning,switchofftheunitandremoveallloading.
Switchtheunitbackonagainanditshould resumenormaloperation.
Ifthisproblempersists,pleasecontactthecustomercaredepartmentorthepointofpurchase.
OTP:OverTemperatureProtection
Thereisathermosensorinsidetheunittomonitorandpreventtheunitfromundulyheating
internally.WhenanOTPerroroccurs,thereisnooutputandthefollowingwarningwillappear
ontheLEDdisplay.
Whenthiswarningappears,switchofftheunitandremoveallloading.
Checktheloadandoutputsetting.Allowtheunittocoolforatleast30minutes.
Checktoseeifanyventilationholesareblocked,alsocheckthatthereissufficientclearance
aroundthepowersupply.
Listencarefullyfortheshortwindsoundemanatingfromthecoolingfanwhentheunitis
switchedONagain.Ifthewindsoundisnotdetected,thefanmaybefaulty;donotusethe
powersupplyinthiscase,contactthecustomercaredepartmentortheagentfromthepointof
purchase.
OCP:OverCurrentProtection
NormallytheoverloadprotectionissustainedbytheCCconstantcurrentmode.WhentheCC
modefailsandgoesundetected,seriousdamagetothetestpieceorloadmayoccur.TheOCPis
usedtominimizetheextentofdamagetotheloadsintheeventofpowersupplyfailure.
Switchoffthepowersupplyassoonasthiswarningisseen(asshownbelow).
Toresetthiswarning,removeallloadingandpresstheOUTPUTON/OFFbutton.
Switchtheunitbackonagainanddoublecheckwithcaution.
Ifthisproblempersists,pleasecontactthecustomercaredepartmentorconsultwiththeagent
atthepointofpurchase.
Copyright©2014FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
www.extech.com
/